Diode Laser System GP900A8, Diode Laser System GP900Q8
K213225 · Shenzhen Gsd Tech Co., Ltd. · GEX · Dec 22, 2021 · General, Plastic Surgery
Device Facts
Record ID
K213225
Device Name
Diode Laser System GP900A8, Diode Laser System GP900Q8
Applicant
Shenzhen Gsd Tech Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Diode Laser System is intended for use in dermatologic and general surgical procedures. The Fast Mode is intended for hair removal of unwanted hair, and permanent hair reduction in hair regrowth. The Free Mode is intended for hair removal of unwanted hair, and permanent hair reduction in hair regrowth. The Diode Laser System is intended for use on all skin types (Fitzpatrick skin types I-VI), including tanned skin. The permanent reduction in hair regrowth is defined as long-term, stable reduction in the number of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Device Story
Diode Laser System (models GP900A8, GP900Q8) uses 810nm semiconductor diode laser to perform hair removal and permanent hair reduction. Device consists of main unit, laser handpiece, and footswitch. Operated by clinician in dermatologic/surgical settings. Laser energy delivered to treatment area via handpiece; emission activated by footswitch. Two operating modes: Fast and Free. System provides permanent hair reduction by targeting hair follicles. Safety features include cooling system and central control system. Device is floor-standing (GP900A8) or desktop (GP900Q8).
Clinical Evidence
Bench testing only. No clinical data provided. Safety and performance verified via electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), laser safety (IEC 60601-2-22, IEC 60825-1), and biocompatibility (ISO 10993-5, ISO 10993-10) testing.
Technological Characteristics
Class IV diode laser; 810nm ± 5nm wavelength; 2cm² spot size; 5-50J/cm² fluence; 1-10Hz repetition rate; 13-200ms pulse duration. Semiconductor diode source. Cooling system, central control system. Complies with IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-2. Biocompatible materials per ISO 10993-5/10.
Indications for Use
Indicated for hair removal and permanent hair reduction in patients of all skin types (Fitzpatrick I-VI), including tanned skin, for dermatologic and general surgical procedures.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services-USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
December 22, 2021
Shenzhen GSD Tech Co., Ltd. Huifang Yao Regulatory Engineer Building A, JUNSD Hi-Tech Park, West of Bao'An RD. Watch & Clock Base, Guangming District Shenzhen, Guangdong 518106 China
Re: K213225
Trade/Device Name: Diode Laser System GP900A8, Diode Laser System GP900Q8 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 24, 2021 Received: September 29, 2021
Dear Huifang Yao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213225
Device Name Diode Laser System
#### Indications for Use (Describe)
The Diode Laser System is intended for use in dermatologic and general surgical procedures. The Fast Mode is intended for hair removal of unwanted hair, and permanent hair regrowth. The Free Mode is intended for hair removal of unwanted hair, and permanent hair regrowth. The Diode Laser System is intended for use on all skin types (Fitzpatrick skin types I-VI), including tanned skin.
The permanent reduction in hair regrowth is defined as long-term, stable reduction in the number of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
### Date Prepared: Sept. 24, 2021
This 510(k) Summary is being submitted in accordance with requirements of Title 21 CFR Section 807.92.
### 1. Submitter information
Shenzhen GSD Tech Co., Ltd Add.: Building A, JUNSD Hi-Tech Park, West of Bao'An RD. Watch & Clock Base, Guangming District, Shenzhen, China 518106 Establishment Registration Number: 3006580954
Contact Person: Huifang Yao Position: Regulatory Engineer Phone: +86 15018526594 E-mail: zoe.yao190322@qq.com.com
### 2. Device information
Trade name: Diode Laser System Model number: GP900A8, GP900Q8 Regulation number: 21CFR 878.4810 Regulation name: Powered Laser Surgical Instrument Regulatory class: II Panel: General & Plastic Surgery Product code: GEX
#### 3. Predicate device information
510(k) Number: K142186 Device Name: Diode Laser Hair Removal System Manufacturer: Shenzhen GSD Tech Co., Ltd.
## 4. Device description
The Diode Laser System is a surgical device intended for use in dermatologic and general surgical procedure. It utilizes a semiconductor diode with invisible infrared radiation as a laser source (810 nm). The laser power is delivered to the treatment area via a laser handpiece. The emission laser is activated by a footswitch.
The proposed Diode Laser System has two models, GP900A8 and GP900Q8. Both GP900A8 and GP900Q8 have the same functions modules, such as the power supply system, central control system, cooling system, laser delivery system and safety feature, and same working mode, such as Fast mode and Free
{5}------------------------------------------------
mode.
The GP900A8 is a standard case with wheels to allow easy movement on the floor. The GP90008 is a desktop case.
## 5. Intended use
The Diode Laser System is intended for use in dermatologic and general surgical procedures. The Fast Mode is intended for hair removal of unwanted hair, and permanent hair reduction in hair regrowth. The Free Mode is intended for hair removal of unwanted hair, and permanent hair reduction in hair regrowth. The Diode Laser System is intended for use on all skin types (Fitzpatrick skin types I-VI), including tanned skin.
The permanent reduction in hair regrowth is defined as long-term, stable reduction in the number of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
| Description | Subject device | Predicate device | Rem |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| | | | ark |
| Company | Shenzhen GSD Tech Co., Ltd. | Shenzhen GSD Tech Co., Ltd. | / |
| Device name and<br>model | Diode Laser System<br>Models: GP900A8, GP900Q8 | Diode Laser Hair Removal System<br>Models: GP900A, GP900Q | / |
| 510 (k) number | / | K142186 | / |
| Classification name | Powered Laser Surgical Instrument | Powered Laser Surgical Instrument | SE |
| Product code | GEX | GEX | SE |
| Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | SE |
| Class | II | II | SE |
| Indications for use/<br>Intended use | The Diode Laser System is intended<br>for use in dermatologic and general<br>surgical procedures. The Fast Mode is<br>intended for hair removal of unwanted<br>hair, and permanent hair reduction in<br>hair regrowth. The Free Mode is<br>intended for hair removal of unwanted<br>hair, and permanent hair reduction in<br>hair regrowth. The Diode Laser<br>System is intended for use on all skin<br>types (Fitzpatrick skin types I-VI),<br>including tanned skin.<br>The permanent reduction in hair<br>regrowth is defined as long-term,<br>stable reduction in the number of hair | The Diode Laser System is intended<br>for use in dermatologic and general<br>surgical procedures. The Fast Mode<br>is intended for hair removal of<br>unwanted hair, and permanent hair<br>reduction in hair regrowth. The Free<br>Setting Mode is intended for hair<br>removal of unwanted hair, and<br>permanent hair reduction in hair<br>regrowth. The Diode Laser System is<br>intended for use on all skin types<br>(Fitzpatrick skin types I-VI),<br>including tanned skin.<br>The permanent reduction in hair<br>regrowth is defined as long-term, | SE |
| | | | |
| | regrowing when measured at 6, 9 and<br>12 months after the completion of a<br>treatment regime. | stable reduction in the number of<br>hair regrowing when measured at 6,<br>9 and 12 months after the completion<br>of a treatment regime. | |
| Configuration | Main unit<br>Handpiece<br>Foot control | Main unit<br>Handpiece<br>Foot control | SE |
| Energy source | Diode laser | Diode laser | SE |
| Laser classification | Class IV | Class IV | SE |
| Wavelength | $810nm\pm5nm$ | 808nm | SE |
| Spot size | 2cm² (10mm*20mm) | 1.44cm²(12mm*12mm) | Note<br>1 |
| Working mode | Fast mode & Free mode | Fast mode & Free setting mode | SE |
| Fluency<br>(energy density) | 5~50J/cm² | 1~120J/cm² | Note<br>2 |
| Repetition rate | 1~10Hz | 1~10Hz | SE |
| Pulse duration | 13-200ms | 5-200ms | Similar |
| Product appearance<br>design | GP900A8: Standard case<br>GP900Q8: Desktop case | GP900A: Standard case<br>GP900Q: Desktop case | SE |
| Dimension | GP900A8: 450mmx540mmx1080mm<br>GP900Q8: 650mmx460mmx370mm | GP900A: 445mmx554mmx1228mm<br>GP900Q: 505mmx363mmx344mm | Note<br>3 |
| Weight | GP900A8: 54kg<br>GP900Q8: 30kg | GP900A: 52kg<br>GP900Q: 22kg | Note<br>4 |
| Electrical Safety | Comply with IEC 60601-1, IEC<br>60601-2-22 and IEC 60825-1 | Comply with IEC 60601-1, IEC<br>60601-2-22 and IEC 60825-1 | SE |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | SE |
| Biocompatibility | Comply with ISO 10993-5 and ISO<br>10993-10 | Comply with ISO 10993-5 and ISO<br>10993-10 | SE |
### 6. Technological characteristics and substantial equivalence:
{6}------------------------------------------------
Note 1: Spot size only affects the area of treatment, not the therapeutic effect. Therefore, this difference will not affect the safety and effectiveness.
Note 2: The fluence of the subject device is within the range of the predicate devices. And the subject device has passed the IEC60601-1 test, IEC60601-1-2 test, IEC60601-2-22 test and IEC60825-1 test, the safety and performance of the product can be ensured. So the subject device is determined to be as safe, as effective, and performs as well as the legally marketed predicate devices.
Note 3/4: The subject device is different in dimension and weight from the
{7}------------------------------------------------
predicate device. By complying with IEC 60601-1, the mechanical performance of the proposed device is determined to be accepted, therefore, this difference will not affect the safety and effectiveness.
### 7. Nonclinical tests submitted
#### ● Safety test
IEC 60601-1: 2005/A1: 2012
Medical electrical equipment Part1: General requirements for basic safety and essential performance
#### ● EMC test
IEC 60601-1-2: 2014
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances-Requirements and tests
#### ● Reliability test
IEC 60601-2-22: 2019
Medical Electrical Equipment - Part 2-22: Particular requirements for the basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.
IEC 60825-1: 2014
Safety of laser products - Part 1: Equipment classification and requirements.
#### ● Biocompatibility test
ISO 10993-5: 2009
Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity ISO 10993-10: 2010
Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
None of the tests demonstrated any design characteristics that violate the requirements of the Reviewer Guidance or show any safety hazards. It is our conclusion that the subject device tested met all relevant requirements of the aforementioned tests.
## 8. Conclusion
The subject devices have the same intended use and same technological characteristics as the predicate device. Moreover the differences between the subject and predicate don't raise new questions of safety or effectiveness. Thus, the subject device is substantially equivalent to the Predicate Device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.