Used to endoscopically ligate esophageal varices at or above the gastroesophageal junction or to ligate internal hemorrhoids.
Device Story
Multi-Band Ligator attaches to distal end of endoscope; facilitates serial ligation of esophageal varices or internal hemorrhoids. Device comprises control handle, loading catheter, irrigation adapter, barrel preloaded with 4-9 loops, and trigger cord. Physician operates device endoscopically; manipulates handle and trigger cord to deploy individual bands onto target tissue. Multi-band design enables multiple ligations without endoscope removal for reloading. Intended for single use; supplied non-sterile. Benefits include reduced procedure time and minimized need for device exchange during treatment.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via appearance, functional performance, chemical properties, mechanical loop properties, trigger cord tensile strength, irrigation adapter physical properties, and barrel leakage testing. Biocompatibility confirmed per ISO 10993.
Technological Characteristics
Materials: Polyurethane (irrigation adapter), Polycarbonate/silicone rubber (barrel), Polypropylene (trigger cord), Natural rubber (loops). Mechanical actuation via trigger cord/handle. Single-use, non-sterile. Shelf life 2 years (ASTM F 1980-16).
Indications for Use
Indicated for patients requiring endoscopic ligation of esophageal varices at or above the gastroesophageal junction or internal hemorrhoids.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 6, 2022
Beijing ZKSK Technology Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM. 1801, No.161, East Lujiazui Rd., Pudong Shanghai, Shanghai 200120 CHINA
Re: K213223
> Trade/Device Name: Multi-Band Ligator Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal ligator Regulatory Class: Class II Product Code: FHN, MND Dated: September 20, 2021 Received: September 29, 2021
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213223
Device Name Multi-Band Ligator
Indications for Use (Describe)
Used to endoscopically ligate esophageal varices at or above the gastroesophageal junction or to ligate internal hemorrhoids.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K213223
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
## 1.0 Submitter's Information
Name: Beijing ZKSK Technology Co., Ltd
Address: Building 9, 6 & No.6 Yuan Hengye North 7th Street, Yongle Economic Development Zone, Tongzhou District,Beijing 101105, China Tel: +86 -13811778090
Contact: Ma Li
### Designated Submission Correspondent
Contact: Mr. Boyle Wang Name: Shanghai Truthful Information Technology Co., Ltd. Address: Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn
Date of Preparation: Apr.26, 2022
# 2.0 Device Information
Trade name: Multi-Band Ligator Common name: Band Ligator Classification name: Hemorrhoidal ligator.
Production code: FHN, MND Regulation number: 21 CFR 876.4400 Classification: Class II Panel: Gastroenterology/Urology 3.0 Predicate Device Information
WILSON-COOK MEDICAL, INC. Manufacturer: Trade/Device Name: WILSON-COOK 10 SHOT MULTI-BAND LIGATOR 510(k) number: K974018
# 4.0 Device Description
{4}------------------------------------------------
Multi-band ligator inculding control handle (rotation handle, shaft and joint), loading catheter, irrigation adapter (irrigation pipe and base), barrel, trigger cord (cord and pellets) and loop.The barrel preloaded with 4~9 loops. This device is designed for attachment to the end of an endoscope for ligating esophageal varices or hemorrhoids. Once assembled and attached the endoscope is advanced to the desired banding site and individual bands are deployed via manipulation of the deployment handle and trigger cord. The multi-band feature allows for serial ligations, which reduces the need to remove the endoscope for reloading. This device is supplied non-sterile and is intended for single use only.
Models difference of the subject device depends on loops number and components, see Table 1.
| Models | Loop No. | Components |
|---------|----------|-----------------------------------------------------------------------------------------------------------------|
| BL18-4A | 4 bands | |
| BL18-5A | 5 bands | Control handle (rotation handle, shaft and joint), |
| BL18-6A | 6 bands | loading catheter, irragation adapter (irrigation pipe and base), barrel, trigger cord (cord and pellets), loop. |
| BL18-7A | 7 bands | |
| BL18-8A | 8 bands | |
| BL18-9A | 9 bands | |
| BL18-4M | 4 bands | Loading catheter, irrigation adapter (irrigation pipe and base), barrel, trigger cord (cord and pellets), loop. |
| BL18-5M | 5 bands | |
| BL18-6M | 6 bands | |
| BL18-7M | 7 bands | |
| BL18-8M | 8 bands | |
| BL18-9M | 9 bands | |
### Table 1 Multi-Band Ligator Models
### 5.0 Indication for Use Statement
Used to endoscopically ligate esophageal varices at or above the gastroesophageal junction or to ligate internal hemorrhoids.
# 6.0 Summary of Non-Clinical Testing
Summary of non-clinical and performance testing- Bench testing was performed to evaluate the performance and functionality of the subject device against requirement specification. The subject device has been subjected to compliance testing according
{5}------------------------------------------------
to Technical Requirements of Multi-Band Ligator. Results from testing performed confirms that the design requirement specification and user needs have been met. The subject device is confirmed to be safe and effective for the intended use.
7.1 Shelf life of Multi-Band Ligator is have successfully been tested according to ASTM F 1980-16. The label shelf life is 2 years.
7.2 Biocompatibility testing of Multi-Band Ligator has successfully been tested for cytotoxicity, sensitization and intracutaneously irritation. The test results verify that the biocompatibility criteria given in ISO 10993 are fulfilled. Multi-Band Ligator is biocompatible.
7.3 Performance testing – Bench: The performance of Multi-Band Ligator have been verified. Tests as described in table 1 have been completed.
| Test Item | Description |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Appearance | To confirm that subject device is integrity and cleanliness<br>and hygiene. |
| Functional performance | To confirm that the functions of the subject device<br>complied with the Indication for Use. |
| Chemical properties | To confirm that pH, Total content of heavy metals,<br>Potassium permanganate reductive substance,<br>Evaporative residues and Ultraviolet Absorbance meet the<br>related requirements. |
| Mechanical performance of<br>the Loop | To confirm Hardness, Tensile strength, Tensile elongation<br>and Tensile deformation meet the company's technical<br>requirements. |
| Tensile strength | To conform the Tensile strength of the Trigger cord. |
| Physical properties of<br>Irrigation adapter | To conform connection firmness, Inner surface and Luer<br>connector |
| Leakage of the barrel | To check the tightness of the barrel, it shall have no<br>leakage |
| Table 1: Performance testing summary - Bench | | |
|----------------------------------------------|--|--|
| | | |
# 7.0 _Summary of Clinical Testing
No clinical study is included in this submission.
### 8.0 Technological Characteristic Comparison Table
| Item | Subject Device | Predicate Device |
|----------------|-----------------|------------------|
| 510(k) No. | K213223 | K974018 |
| Product Code | FHN, MND | FHN, MND |
| Regulation No. | 21 CFR 876.4400 | 21 CFR 876.4400 |
#### Table 2- Comparison of Technology Characteristics
{6}------------------------------------------------
| | I | II |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Class | | |
| Indication for Use | Used to endoscopically ligate esophageal varices at or above the gastroesophageal junction or to ligate internal hemorrhoids. | Used to endoscopically ligate esophageal varices at or above the gastroesophageal junction or to ligate internal hemorrhoids. |
| Components | 1) Non-patient Contacting<br>2) Loading Catheter;<br>3) Irrigation Adapter;<br>4) Barrel with preloaded loops<br>5) Trigger Cord<br>6) Multi-Band Ligator handle | 1) Non-patient Contacting<br>2) Loading Catheter;<br>3) Irrigation Adapter;<br>4) Barrel with 10 Ligation Bands<br>5) Trigger Cord<br>6) Deployment Handle |
| Patient<br>Contacting<br>Materials | Irrigation Adapter: Polyurethane<br>Barrel: Polycarbonate (PC), silicone rubber<br>Trigger Cord: Polypropylen (PE)<br>Loop: Natural rubber | Friction Fit Adapter:Polyurethane<br>Barrel:Polycarbonate<br>Trigger Cord: Vectran<br>Bands: Latex Rubber |
| Method of Use | Friction Fit adapter and barrel with ligation loops mounted to the distal tip of the endoscope. Band release from the barrel accomplished by trigger cord and Multi-Band Ligator handle. | Friction Fit adapter and barrel with ligation bands mounted to the distal tip of the endoscope. Band release from the barrel accomplished by trigger cord and deployment handle. |
| Loops(Bands) | Multiple loops (4~9 )can perform serial ligations without removal of the endoscope for reloading of bands | Multiple bands (10 )can perform serial ligations without removal of the endoscope for reloading of bands |
| Sterile | Non-sterile | Non-sterile |
| Shelf Life | 2 years | 3 years |
| Single Use | Single Use | Single Use |
| Performance<br>Comparison<br>testing | The test was conducted to the predicate device and current device to compare their functional performance. | |
| Biocompatibility | Conform with ISO10993-1 (ISO10993-5, ISO10993-10,) | Conform with ISO 10993 standards |
The technological characteristics of the subject device are identical to those of predicate device. The subject device has the same basic design as the predicate device. The comparison between the subject and predicate devices is based on the following:
• Same intended use
- Same indications for use
- Similar material types that meet ISO 10993 biocompatibility requirements
{7}------------------------------------------------
· Same fundamental technology/principal of operation/user interface
The quantity of the loop of the subject device is 4,5,6,7,8,9 pcs, which is different with those of predicate device which is 10 pcs. But this difference will not raise any risk in safety and effectiveness.
Performance comparison testing was conducted to the predicate device and current device to compare their physical performance.It can be found the subject device incorporates substantially equivalent device materials, design, configuration, fundamental technology, sterilization process and intended use as those featured in the predicate device.
## 9.0 Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device in K974018 and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.