← Product Code [ELW](/productcode/ELW) · K213175

# No Stress Impress (K213175)

_No Stress Impress, LLC · ELW · Jan 19, 2023 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K213175

## Device Facts

- **Applicant:** No Stress Impress, LLC
- **Product Code:** [ELW](/productcode/ELW.md)
- **Decision Date:** Jan 19, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

No Stress Impress is intended for use as a rebasing and relining impression material as well as for general impressions including but not limited to dentures, root surfaces, and posts.

## Device Story

No Stress Impress is a thermoplastic impression material provided in stick or wafer form. It is used by dental professionals to capture anatomical impressions of oral structures, including dentures, root surfaces, and posts. The material is softened, placed in the patient's mouth to record the impression, and then allowed to harden. The resulting physical impression is used by the clinician to facilitate dental procedures such as rebasing or relining. The device is a passive material and does not involve electronic processing, software, or automated analysis.

## Clinical Evidence

No clinical data. Bench testing performed per ISO 4823:2015 (Dentistry Elastomeric Material). Biocompatibility testing conducted per ISO 10993-1, including oral mucosal irritation (hamster), guinea pig maximization sensitization, and cytotoxicity (elution and agarose overlay methods), all yielding non-irritating, non-sensitizing, and non-reactive results.

## Technological Characteristics

Thermoplastic impression material supplied as sticks or wafers. Composition consists of ingredients similar to legally marketed impression materials. Performance testing conducted per ISO 4823:2015. Biocompatibility verified per ISO 10993-1.

## Regulatory Identification

Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.

## Special Controls

*Classification.* Class II (Special Controls).

## Predicate Devices

- Impression Compound ([K110378](/device/K110378.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym "FDA" in a blue square, followed by "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

January 19, 2023

No Stress Impress LLC % Cheryl Wagoner Consultant Wagoner Consulting LLG 5215 Crosswinds Drive Wilmington, North Carolina 28409

Re: K213175

Trade/Device Name: No Stress Impress Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: Class II Product Code: ELW Dated: December 6, 2022 Received: December 7, 2022

Dear Cheryl Wagoner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Bobak Shirmohammadi -S

For Michael E. Adjodha, M. ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

## Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K213175

Device Name No Stress Impress

#### Indications for Use (Describe)

No Stress Impress is intended for use as a rebasing and relining impression material as well as for general impressions including but not limited to dentures, root surfaces, and posts.

| Type of Use (Select one or both, as applicable)                                    |                                                                                 |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      X Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

(as required by 21 CFR 807.92)

| Submitter      | No Stress Impress                                                                                                             |
|----------------|-------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Misael Otero, DDS<br>President, No Stress Impress<br>2012 Scrimshaw Place<br>Wilmington, North Carolina 28405<br>917-304-7010 |
| Date Prepared  | April 25, 2022                                                                                                                |

| Trade Name     | No Stress Impress                  |
|----------------|------------------------------------|
| Common Name    | Impression material                |
| Classification | 21 CFR §872.3660, Product Code ELW |
| Class          | Class II                           |

| Product name        | Manufacturer                                           | 510(k) # | Basis for<br>Substantial<br>Equivalence                           |
|---------------------|--------------------------------------------------------|----------|-------------------------------------------------------------------|
| Primary Predicate   |                                                        |          |                                                                   |
| Impression Compound | Sybron Dental<br>Specialties (acquired<br>by KavoKerr) | K110378  | • Same intended use<br>• Same Indications<br>• Similar technology |

| Description | No Stress Impress is a thermoplastic impression material<br>available for full or partial impressions that comes as a stick or<br>wafer. |
|-------------|------------------------------------------------------------------------------------------------------------------------------------------|
|-------------|------------------------------------------------------------------------------------------------------------------------------------------|

| Intended<br>Use/Indications<br>for Use | No Stress Impress is intended for use as a rebasing and relining impression material as well as for general impressions including but not limited to dentures, root surfaces, and posts. |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics       | This product is made ingredients similar to ingredients found in other impression materials currently on the U.S. markets and substantially equivalent to the predicate devices.         |

{4}------------------------------------------------

| Performance Data | Material testing per ISO 4823:2015, 7.8 Dentistry Elastomeric Material was conducted with satisfactory results.                                               |                                      |
|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
|                  | Biocompatibility testing was performed in accordance with ISO 10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing” including: |                                      |
|                  | Test Performed                                                                                                                                                | Results                              |
|                  | ISO Oral Mucosal Irritation Study in Hamsters -<br>Collar Method - 24 hours                                                                                   | Non-irritating                       |
|                  | ISO Guinea Pig Maximization Sensitization Test                                                                                                                | Non-sensitizing                      |
|                  | Cytotoxicity Study Using the ISO Elution Method                                                                                                               | No lysis, no reactivity, no pH shift |
|                  | Cytotoxicity Study Using the ISO Agarose Overlay Method                                                                                                       | No lysis, no reactivity              |

| Conclusion |                                                                                                                                                                                                                                     |
|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|            | The Subject device has identical intended use and indications<br>as the primary predicate. The ingredients are similar to the<br>predicate. The labeling claims of the Subject device are the<br>similar to those of the predicate. |
|            | The labeling of the Subject device contains the similar warnings<br>and precautions as those in the labeling of the predicates.                                                                                                     |
|            | Any differences that exist between the Subject device and the<br>predicates have no significant effect on the safety or<br>effectiveness.                                                                                           |
|            | The Subject device is substantially equivalent to other<br>prescription impression materials cleared in the US in terms of<br>biocompatibility, technology, intended use, indications, and<br>suitability characteristics.          |

---

**Source:** [https://fda.innolitics.com/device/K213175](https://fda.innolitics.com/device/K213175)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
