TrabEx Pro

K213173 · Microsurgical Technologies, Inc. · MRH · Jun 8, 2022 · General Hospital

Device Facts

Record IDK213173
Device NameTrabEx Pro
ApplicantMicrosurgical Technologies, Inc.
Product CodeMRH · General Hospital
Decision DateJun 8, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The TrabEx Pro Handpiece is a manual ophthalmic surgical instrument with irrigation and aspiration functions.

Device Story

TrabEx Pro is a single-use, sterile, manual ophthalmic knife featuring an incision sealing sleeve. It connects to an external ophthalmic irrigation/aspiration (I/A) system to provide fluid management during surgery. Operated by a surgeon in a clinical setting, the device facilitates irrigation and aspiration to maintain anterior chamber stability or clear debris. The device does not contain internal electronics or software; it functions as a mechanical conduit for fluid flow. Clinical benefit is derived from its ability to perform precise surgical maneuvers while managing intraocular pressure and fluid dynamics via the connected I/A system.

Clinical Evidence

Bench testing only. No clinical data provided. Performance was validated by comparing irrigation and aspiration flow rates against the Trabectome (K040584). Biocompatibility testing included cytotoxicity, guinea pig maximization, ocular irritation, acute systemic toxicity, and material-mediated pyrogenicity per ISO 10993-1. Sterility was validated to a SAL of 10^-6 using radiation sterilization. Shelf life of 2 years was supported by accelerated aging per ASTM F1980-16 and package integrity testing per ISO 11607-1/2.

Technological Characteristics

Single-use, sterile, manual ophthalmic knife with incision sealing sleeve. External communicating device with limited (<24 hours) tissue contact. Irrigation/aspiration performance: min 3 ml/min at 0.59 psi (40 cm bottle height); aspiration 4 ml/min at max 4.8 psi (250 mmHg). Sterilization: radiation (SAL 10^-6). Shelf life: 2 years (ASTM F1980-16). Packaging: ISO 11607-1/2 compliant.

Indications for Use

Indicated for use as a manual ophthalmic surgical instrument providing irrigation and aspiration during ophthalmic procedures.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. June 8, 2022 MicroSurgical Technologies Inc Ms. Angela Mallery Principle Product Development Strategist, Regulatory 8415 154th Ave NE Redmond, Washington 98052 Re: K213173 Trade/Device Name: TrabEx Pro Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: MRH Dated: Mav 3, 2022 Received: May 3, 2022 Dear Ms. Angela Mallery: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Anjana Jain, PhD Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K213173 Device Name TrabEx Pro Indications for Use (Describe) The TrabEx Pro Handpiece is a manual ophthalmic surgical instrument with irrigation and aspiration functions. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | 510(k) SUMMARY | | | | |----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------| | Submitted by: | MicroSurgical Technology, Inc.<br>8415 154th Ave NE Redmond, WA 98052<br>(425)861-4002<br>www.microsurgical.com | | | | Date Prepared | June 7, 2022 | | | | Contact | Dave Scott<br>COO<br>(425)861-4002<br>info@microsurgical.com | | | | Trade/Proprietary Name | TrabEx Pro | | | | Regulatory Name and Classification | Infusion Pump;<br>21 CFR 880.5725 (Product code: MRH) | | | | Predicate | K993039 Bausch & Lomb Storz Millennium Viscous Fluid Injector System | | | | Reference Devices | K040584 Trabectome | | | | Device Description | The MicroSurgical Technology TrabEx Pro is a single use, sterile, manual ophthalmic knife.<br>The device can be connected to an ophthalmic Irrigation/Aspiration (I/A) system to provide<br>irrigation and aspiration while the procedure is being performed.<br>TrabEx PRO has an incision sealing sleeve. | | | | Indications for Use | The TrabEx Pro Handpiece is a manual ophthalmic surgical instrument with irrigation and<br>aspiration functions. | | | | Comparative Technology Characteristics | The subject and predicate device both include an irrigation and aspiration feature.<br>This medical device product has functions subject to FDA premarket review as well as<br>functions that are not subject to FDA premarket review. For this application, if the product has<br>functions that are not subject to FDA premarket review, FDA assessed those functions only to<br>the extent that they either could adversely impact the safety and effectiveness of the functions<br>subject to FDA premarket review or they are included as a labeled positive impact that was<br>considered in the assessment of the functions subject to FDA premarket review. | | | | Technological Characteristics | Irrigation/aspiration performance of the TrabEx Pro was compared to the Trabectome<br>(K040584) to demonstrate substantial equivalence of the TrabEx Pro to the Trabectome.<br>The TrabEx Pro and the Trabectome were tested using the same test methods for establishing<br>irrigation and aspiration rates; the rates are identical. | | | | | Irrigation and Aspiration Rates | | | | | Subject device TrabEx Pro | Trabectome K040584 | | | | Intended Use | Ophthalmic surgical instrument | | | | | | K213177 | | | Ability for Irrigation<br>and Aspiration | Irrigation Rate Min 3 ml/min at<br>40 cm bottle height (0.59 psi) | Irrigation Rate Min 3 ml/min at<br>40 cm bottle height (0.59 psi) | | | | Aspiration Rate 4 ml/min at no<br>more than 250 mmHg (4.8psi) | Aspiration Rate 4 ml/min at no<br>more than 250 mmHg (4.8psi) | | | | | | | Non-Clinical tests<br>performed | Bench Testing | The Trabectome was identified as a reference device for<br>demonstration of acceptable performance for the<br>irrigation/aspiration function-under review | | | | Biocompatibility | TrabEx Pro is an external communicating device with limited (<24<br>hours) contact with tissue. The device components with direct and<br>indirect tissue contact were assessed for biocompatibility per FDA<br>Biocompatibility Guidance "Use of International Standard ISO<br>10993-1, 'Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management process,'" and<br>included the following tests:<br>• Cytotoxicity (test extract method)<br>• Guinea pig maximization<br>• Ocular Irritation<br>• Acute systemic toxicity<br>• Material mediated pyrogenicity<br>The primary packaging does not contact the device components with<br>direct tissue contact and was evaluated using chemical<br>characterization assessment." | | | | Sterility | The TrabEx Pro was adopted into an existing radiation sterilization<br>process. The radiation sterilization dose range provided a Sterility<br>Assurance Level (SAL) of 10-6 in accordance with FDA Guidance,<br>"Submission and Review of Sterility Information in Premarket<br>Notification (510(k)) Submissions for Devices Labeled as Sterile"<br>issued on January 21, 2016. Limulus Amebocyte Lysate (LAL)<br>testing was performed to support the direct and indirect contacting<br>components of the TrabEx Pro are non-pyrogenic. | | | | Shelf Life | The labeled 2-year shelf life was supported with device performance<br>testing and package integrity testing as per ISO 11607-1 and -<br>2:2019, "Requirements for materials, sterile barrier systems and<br>packaging systems." Samples were aged in compliance with ASTM<br>F1980-16, "Standard Guide for Accelerated Aging of Sterile Barrier<br>Systems for Medical Devices." | | | Conclusion | Conclusion(s) drawn from the nonclinical tests demonstrate the device is substantially<br>equivalent to the identified legally marketed device. | | | {4}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...