Safety butterfly type blood collection needle, Butterfly type blood collection needle, Pen type blood collection needle, Safety pen
Applicant
Jiangsu Rongye Technology Co., Ltd.
Product Code
JKA · Clinical Chemistry
Decision Date
Oct 27, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Intended Use
The Safety pen type blood Collection needle is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The Safety butterfly type blood Collection needle is intended to be used with vacuum blood collection tube for multiple collections of venous blood. The pen type blood Collection needle is intended to be used with vacuum blood collection of venous blood. The butterfly type blood Collection needle is intended to be used with vacuum blood collection tube for multiple collections of venous blood.
Device Story
Blood collection needles (safety butterfly, butterfly, pen type, safety pen type) facilitate venous blood collection into vacuum tubes. Devices consist of stainless steel needles, plastic hubs, and tubing (for butterfly variants). Safety versions include integrated needle-shielding mechanisms to prevent needlestick injuries. Used in clinical settings by phlebotomists or nurses. Operation involves inserting needle into patient vein and connecting to vacuum tube holder; vacuum pressure draws blood into tube. Safety mechanisms are activated post-draw to cover needle tip. Benefits include standardized blood sampling and reduced risk of sharps injury to healthcare personnel.
Clinical Evidence
Bench testing only.
Technological Characteristics
Stainless steel needles; plastic hubs/tubing; sterile; single-use; manual operation; no electronic components or software.
Indications for Use
Indicated for venous blood collection using vacuum tubes. Intended for use by healthcare professionals for single or multiple blood draws.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Related Devices
K172763 — Blood Collecting Needle, Safety Blood Collecting Needle, Blood Collecting Needle with Holder, Safety Blood Collecting Needle with Holder, Blood Collecting Set, Safety Blood Collecting Set, Blood Collecting Set with Holder, Safety Blood Collecting Set with Holder · Zhejiang Kindly Medical Devices Co., Ltd. · Jun 4, 2018
K182146 — Sol-M Blood Collection Needles, Sol-Care Safety Blood Collection Needles, Sol-Care Safety Blood Collection Needle with Holders, Sol-M Blood Collection Set · Sol-Millennium Medical, Inc. · Dec 3, 2018
K230080 — Blood Collection Set, Blood Collection Set with Holder, Safety Blood Collection Set, Safety Blood Collection Set with Holder, Blood · Yangzhou Medline Industry, Co., Ltd. · Jul 7, 2023
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, the acronym "FDA" is displayed in a blue box, followed by the words "U.S. FOOD & DRUG" in a larger, bold font. Below that, the word "ADMINISTRATION" is written in a smaller font.
October 27, 2023
Jiangsu Rongye Technology Co., Ltd. Tina Jiang Sales Manager Touqiao Town, Yangzhou City Jiangsu, 225109 China
Re: K213146
Trade/Device Name: Safety butterfly type blood collection needle, Butterfly type blood collection needle, Pen type blood collection needle, Safety pen type blood collection needle Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA, FMI Dated: September 25, 2023 Received: September 27, 2023
Dear Tina Jiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors
{2}------------------------------------------------
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213146
#### Device Name
Safety butterfly type blood collection needle, Butterfly type blood collection needle, Safety pen type blood collection needle
### Indications for Use (Describe)
The Safety pen type blood Collection needle is intended to be used with vacuum blood collection tube for multiple collections of venous blood.
The Safety butterfly type blood Collection needle is intended to be used with vacuum blood collection tube for multiple collections of venous blood.
The pen type blood Collection needle is intended to be used with vacuum blood collection of venous blood.
The butterfly type blood Collection needle is intended to be used with vacuum blood collection tube for multiple collections of venous blood.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.