CLISIS SYSTEMS, Discovery RF180

K213081 · General Medical Merate S.P.A · IZF · Oct 26, 2021 · Radiology

Device Facts

Record IDK213081
Device NameCLISIS SYSTEMS, Discovery RF180
ApplicantGeneral Medical Merate S.P.A
Product CodeIZF · Radiology
Decision DateOct 26, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1740
Device ClassClass 2
AttributesPediatric

Indications for Use

The CLISIS SYSTEMS, Discovery RF180 is indicated for performing general radiography, fluoroscopy and angiography procedures. Applications and techniques: - Gastroenterology - -Skeleton - Thorax and lungs - - Paediatrics - - Urology and gynecology - - Emergency/traumatology - - Digital angiography - Linear tomography - Auto Image Paste (Stitching) - Tomosynthesis

Device Story

CLISIS SYSTEMS, Discovery RF180 is a tomographic X-ray system for general radiography, fluoroscopy, and angiography. It captures X-ray images for diagnostic visualization in clinical settings. The system supports specialized techniques including linear tomography, tomosynthesis, and Auto Image Paste (stitching). Operated by trained healthcare professionals (radiologists, radiographers) in clinical environments. Output consists of digital radiographic and fluoroscopic images used for clinical decision-making in gastroenterology, orthopedics, pulmonology, urology, gynecology, and emergency medicine. Benefits include diagnostic imaging capabilities for various patient populations, including pediatric patients.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and technological comparison to predicate devices.

Technological Characteristics

Tomographic X-ray system. Components include X-ray source and digital image acquisition hardware. Supports digital angiography, linear tomography, and tomosynthesis. Connectivity for digital image processing and display. Complies with 21 CFR 892.1740.

Indications for Use

Indicated for general radiography, fluoroscopy, and angiography procedures across diverse clinical applications including gastroenterology, skeletal, thoracic, pediatric, urological, gynecological, and emergency/traumatology imaging. Prescription use only.

Regulatory Classification

Identification

A tomographic x-ray system is an x-ray device intended to be used to produce radiologic images of a specific cross-sectional plane of the body by blurring or eliminating detail from other planes. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Submission Summary (Full Text)

{0}------------------------------------------------ October 26, 2021 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is in blue text. General Medical Merate S.P.A % Marisa Testa CEO & QA/RA Consultant Thema S.R.L. Via Saragat 5 Imola, BO 40026 ITALY Re: K213081 Trade/Device Name: CLISIS SYSTEMS, Discovery RF180 Regulation Number: 21 CFR 892.1740 Regulation Name: Tomographic x-ray system Regulatory Class: Class II Product Code: IZF, JAA, MQB, IZI, KXJ Dated: September 21, 2021 Received: September 23, 2021 Dear Marisa Testa: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, , for Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K213081 Device Name CLISIS SYSTEMS, Discovery RF180 ## Indications for Use (Describe) The CLISIS SYSTEMS, Discovery RF180 is indicated for performing general radiography, fluoroscopy and angiography procedures. Applications and techniques: - Gastroenterology - - -Skeleton - Thorax and lungs - - Paediatrics - - Urology and gynecology - - Emergency/traumatology - - Digital angiography - Linear tomography - Auto Image Paste (Stitching) - Tomosynthesis | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...