← Product Code [BTR](/productcode/BTR) · K213028

# Tenax Laser Resistant Endotracheal Tube (K213028)

_Bryan Medical, Inc. · BTR · Oct 21, 2021 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K213028

## Device Facts

- **Applicant:** Bryan Medical, Inc.
- **Product Code:** [BTR](/productcode/BTR.md)
- **Decision Date:** Oct 21, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 868.5730
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The Tenax® Laser Resistant Endotracheal Tube is intended for endotracheal intubation. It is indicated for use for all types of surgical procedures involving carbon dioxide (10.60 microns), KTP (532 nm) or Blue Laser (445mm) laser use (normal pulsed or continuous beam delivery in the non-contact mode), when endotracheal intubation is required to administer anesthetic gases or to overcome emergency obstruction of an airway.

## Device Story

Tenax® Laser Resistant Endotracheal Tube is a single-use, sterile, endotracheal tube with two inflatable cuffs; designed for use during surgical procedures involving CO2, KTP, or 445nm Blue lasers. Device features an inner aluminum scattering layer for laser resistance and an atraumatic outer layer to protect patient tissue; includes an internal dye as a warning indicator for puncture or damage. Used by clinicians in the trachea/upper GI tract to administer anesthetic gases or manage airway obstruction. Device is discarded after use. Benefits include protection against laser-induced airway fires or damage during laser surgery.

## Clinical Evidence

Bench testing only. Performance data provided for laser resistance verification and validation to support safety and equivalence to the predicate device.

## Technological Characteristics

Laser-resistant endotracheal tube; features inner aluminum scattering layer, atraumatic outer layer, and two inflatable cuffs. Includes internal dye indicator for puncture detection. Sterile (ethylene oxide), single-use. No software or electronic components.

## Regulatory Identification

A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.

## Predicate Devices

- Tenax® Laser Resistant Endotracheal Tube ([K200761](/device/K200761.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 21, 2021

Bryan Medical, Inc. % Martha Russell Director of Regulatory Affairs AlvaMed. Inc. 1116 Great Plain Avenue, Suite 1 Needham, Massachusetts 02492

Re: K213028

Trade/Device Name: Tenax Laser Resistant Endotracheal Tube Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal tube Regulatory Class: Class II Product Code: BTR, BSK Dated: September 17, 2021 Received: September 21, 2021

Dear Martha Russell:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

## K213028

Device Name Tenax ® Laser Resistant Endotracheal Tube

#### Indications for Use (Describe)

The Tenax® Laser Resistant Endotracheal Tube is intended for endotracheal intubation. It is indicated for use for all types of surgical procedures involving carbon dioxide (10.60 microns), KTP (532 nm) or Blue Laser (445mm) laser use (normal pulsed or continuous beam delivery in the non-contact mode), when endotracheal intubation is required to administer anesthetic gases or to overcome emergency obstruction of an airway.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for Bryan Medical, Inc. The logo consists of two blue, angled rectangles on the left side, followed by the word "bryan" in black, sans-serif font. Below "bryan" is the text "MEDICAL, INC." in a smaller, sans-serif font, also in black.

Special 510(k): Modification to Tenax® Laser Resistant Endotracheal Tube

Page 1

### 1.0 SPECIAL 510(K) SUMMARY FOR TENAX® LASER RESISTANT ENDOTRACHEAL TUBE

This Special 510(k) for the Tenax® Laser Resistant Endotracheal Tube is submitted based on the FDA Guidance document "The Special 510(k) Program: Guidance for Industry and Food and Drug Administration Staff' issued September 13, 2019. This Special 510(k) addresses labeling revisions to the Tenax® Instructions for Use to add the use of a 445 nm Blue Laser, and that the device is not made with natural rubber latex. In addition, the Indications for Use statement was modified to add the 445nm Blue Laser.

## 1.1. Submitter

Bryan Medical, Inc. 5725 Dragon Way, Suite 300 Cincinnati, OH 45227

Phone: (617) 517-4932

Fax: (617) 249-0955

Contact Person: Martha Kamrow-Russell, AlvaMed, Inc.

Date Prepared: September 17, 2021

#### 1.2. Device

Name of Device: Tenax® Laser Resistant Endotracheal Tube

Common or Usual Name: Endotracheal tube

Classification Name: Tracheal tube (21 CFR 868.5730)

Regulatory Class: II

Product Code: BTR, BSK

### 1.3. Predicate Device

Tenax® Laser Resistant Endotracheal Tube, K200761

### 1.4. Device Description

The Tenax® Laser Resistant Endotracheal Tube is an endotracheal tube with two inflatable cuffs. It is indicated for use in surgical procedures where intubation is required in the presence of KTP, 445nm Blue Laser, or CO2 lasers.

The device is provided sterile (EO). After use, the device is discarded and disposed of in accordance with local regulations. There are no associated device accessories.

### 1.5. Indication For Use

The Tenax® Laser Resistant Endotracheal Tube is intended for endotracheal intubation. It is indicated for use for all types of surgical procedures involving carbon dioxide (10.60 microns), KTP (532 nm), or Blue (445 nm) laser use (normal pulsed or continuous beam delivery in the non-contact mode), when endotracheal intubation is required to administer anesthetic gases or to overcome emergency obstruction of an airway.

{4}------------------------------------------------

| Image: bryan MEDICAL, INC. logo | Special 510(k): Modification to<br>Tenax® Laser Resistant Endotracheal<br>Tube | Page 2 |
|---------------------------------|--------------------------------------------------------------------------------|--------|
|---------------------------------|--------------------------------------------------------------------------------|--------|

| Intended Population:         | Adults (≥22 years old) requiring endotracheal intubation during<br>laser surgery procedures in the upper gastro-respiratory tract.                                                                                                                                                                                                                                               |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Environment of Use: | Trachea/Upper GI Tract                                                                                                                                                                                                                                                                                                                                                           |
| Contraindications:           | The Tenax® Laser Resistant Endotracheal Tube should not be<br>used in patients with narrow airways, which could restrict<br>ventilation, resulting in excessive elevation of intratracheal<br>pressure. Do not use in any patient suffering from conditions that<br>may preclude endotracheal intubation. Do not use this device for<br>any purpose other than its intended use. |

{5}------------------------------------------------

Page 3

Image /page/5/Picture/2 description: The image shows the logo for Bryan Medical, Inc. The logo features a blue, stylized "B" shape on the left, followed by the word "bryan" in black, sans-serif font. Below "bryan" is the text "MEDICAL, INC." in a smaller, sans-serif font, also in black. The overall design is clean and professional.

# 1.6. Comparison of Technological Characteristics with the Predicate Device

|                           | Subject Device:<br>Tenax® Laser Resistant Endotracheal<br>Tube                                                                                                                                                                                                                                                                                                                                                                                                         | Predicate Device:<br>Tenax® Laser Resistant Endotracheal<br>Tube                                                                                                                                                                                                                                                                                                                                                                                       |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer              | Bryan Medical, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Bryan Medical, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| 510(k)<br>Number          | To be assigned.                                                                                                                                                                                                                                                                                                                                                                                                                                                        | K200761                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Product Code              | BTR, BSK                                                                                                                                                                                                                                                                                                                                                                                                                                                               | BTR, BSK                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Requlation<br>Number      | 868.5730                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 868.5730                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Regulation<br>Description | Tracheal tube                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Tracheal tube                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common<br>Name            | Endotracheal tube                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Endotracheal tube                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Indications<br>for Use    | The Tenax® Laser Resistant<br>Endotracheal Tube is intended for<br>endotracheal intubation. It is indicated for<br>use for all types of surgical procedures<br>involving carbon dioxide (10.60 microns),<br>Blue (445 nm), or KTP (532 nm) laser<br>use (normal pulsed or continuous beam<br>delivery in the non-contact mode), when<br>endotracheal intubation is required to<br>administer anesthetic gases or to<br>overcome emergency obstruction of an<br>airway. | The Tenax® Laser Resistant<br>Endotracheal Tube is intended for<br>endotracheal intubation. It is indicated for<br>use for all types of surgical procedures<br>involving carbon dioxide (10.60 microns)<br>or KTP (532 nm) laser use (normal<br>pulsed or continuous beam delivery in<br>the non-contact mode), when<br>endotracheal intubation is required to<br>administer anesthetic gases or to<br>overcome emergency obstruction of an<br>airway. |
| Sterility                 | Sterile (ethylene oxide)                                                                                                                                                                                                                                                                                                                                                                                                                                               | Sterile (ethylene oxide)                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Single Use                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

{6}------------------------------------------------

Page 4

Image /page/6/Picture/2 description: The image shows the logo for Bryan Medical, Inc. The logo consists of two blue rectangles stacked on top of each other on the left side. To the right of the rectangles is the word "bryan" in black, with the words "MEDICAL, INC." in smaller black letters underneath.

Special 510(k): Modification to Tenax® Laser Resistant Endotracheal Tube

Both devices are laser resistant endotracheal tubes. They share an identical indication for use (outside of the addition of the Blue 445 nm laser for the subject device) and are intended for use in the same manner. The subject and predicate devices also share the following technological characteristics:

- Laser resistance using an inner aluminum scattering layer, ●
- Atraumatic outer layer for protection of patient tissue, ●
- Inflatable cuffs,
- Use of internal dye as a warning indicator to alert practitioners of device puncture or ● damage.

#### 1.7. Performance Data

The following performance data were provided in support of the substantial equivalence determination:

- Laser resistance
#### 1.8. Conclusion

The non-clinical data support the safety of the device and the verification and validation testing demonstrate that the Tenax® Laser Resistant Endotracheal Tube shall perform as intended in the specified use conditions. The data demonstrate that the Tenax® Laser Resistant Endotracheal Tube performs comparably to the predicate device currently marketed for the same intended use.

---

**Source:** [https://fda.innolitics.com/device/K213028](https://fda.innolitics.com/device/K213028)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
