The medical face masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.
Device Story
Single-use, flat-pleated, ear-loop medical face mask; non-sterile. Constructed with outer/inner spunbond polypropylene layers and middle meltblown polypropylene filter layer. Features adjustable nose clip (PE and galvanized iron wire) and nylon/spandex ear loops. Used in clinical settings by healthcare personnel and patients to provide barrier protection against microorganisms, body fluids, and particulates. Device performance verified against ASTM F2100-19 standards for fluid resistance, filtration efficiency, and differential pressure. Intended to reduce potential exposure to blood and body fluids during infection control practices.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM F1862/F1862M (fluid resistance), ASTM F2299/F2299M-03 (particulate filtration), ASTM F2101 (bacterial filtration), EN 14683 (differential pressure), and 16 CFR Part 1610 (flammability). Biocompatibility assessed per ISO 10993-5, ISO 10993-10, and ISO 10993-1.
Technological Characteristics
Flat-pleated, ear-loop mask. Materials: Spunbond polypropylene (outer/inner), meltblown polypropylene (filter), nylon/spandex (ear loops), PE/galvanized iron wire (nose clip). Performance meets ASTM F2100-19. Non-sterile. No software or electronic components.
Indications for Use
Indicated for use by patients and healthcare personnel to reduce exposure to microorganisms, body fluids, and particulate material in infection control settings.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Surgical Face Masks (Ear loops and Tie-on) (K160269)
Related Devices
K202628 — Disposable Medical Face Mask · Qinghai Zhong Dao Win-Win Medical Protective Equipment Co., · Mar 19, 2021
K210391 — Single-use medical face mask · Zhuhai Herald Datanetics Limited. · Jul 22, 2021
K221272 — Medical Face Mask · Shandong Xingyu Gloves Co., Ltd. · Jun 28, 2022
K202714 — Medical Face Mask · Chongqing Chaoke Industry Development Co., Ltd. · Jun 11, 2021
K201625 — Disposable Medical Surgical Mask · Tianjin Saiyuan Technology Co., Ltd. · Jan 12, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
January 17, 2022
Guangzhou Quantum Laser Intelligent Equipment Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai. 200120 China
Re: K212994
Trade/Device Name: Medical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: May 18, 2021 Received: September 20, 2021
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K212994
Device Name Medical face mask
Indications for Use (Describe)
The medical face masks are intended to be worn to protect both the patient and healthcare persomel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K212994
- Date of Preparation: 12/23/2021 1.
- 2. Sponsor Identification
## Guangzhou Quantum Laser Intelligent Equipment Co., Ltd.
Building B27 Huachuang Animation Industrial Park, Jinshan Village, Shiqi Town, Panyu District, Guangzhou, Guangdong, China
Establishment Registration Number: 3016792608.
Contact Person: Hao Zheng Position: Business Manager Tel: +86-20-39969520 Email: 2281976947@qq.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Jinlei Tang (Alternative Contact Person)
### Mid-Link Consulting Co., Ltd.
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Medical Face Mask Common Name: Mask
Regulatory Information Classification Name: Mask, Surgical Classification: II; Product Code: FXX; Regulation Number: 21CFR 878.4040 Review Panel: General Hospital
Intended Use/Indications for Use:
The medical face masks are intended to be worn to protect and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.
Device Description:
The Medical Face Masks are single use, flat-pleated, ear loop masks that are provided in blue. The medical face masks are available in three types, which are Level 1, Level 2 and Level 3 based on ASTM F2100-19, depending on marketing reasons. However, Level 2 and Level 3 masks are exactly the same. The outer and inner layers of the mask are made of spunbond polypropylene. The middle filter layer is made of meltblown polypropylene filter. The ear loops are made of nylon and spandex. The ear loops are held in place over the users' mouth and nose by two elastic ear straps. The nose of PE and galvanized iron wire. Users can adjust the nose clip according to the shape of the nose and fix the mask on the bridge of the nose to prevent the mask from falling off. The proposed devices are provided in non-sterile.
- Identification of Predicate Device ર.
510(k) Number: K160269 Product Name: Surgical Face Masks (Ear loops and Tie-on)
{5}------------------------------------------------
#### Summary of Technological characteristics 6.
| ITEM | Proposed Device | Predicate Device K160269 | Remark |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product Code | FXX | FXX | Same |
| Regulation No. | 21 CFR 878.4040 | 21 CFR 878.4040 | Same |
| Class | II | II | Same |
| Indications<br>for<br>Use | The medical face masks are intended to be<br>worn to protect both the patient and<br>healthcare personnel from transfer of<br>microorganisms, body fluids and<br>particulate material. These face masks are<br>intended for use in infection control<br>practices to reduce the potential exposure<br>to blood and body fluids. This is a single<br>use, disposable device, provided non-<br>sterile. | The surgical face masks are intended to be<br>worn to protect both the patient and<br>healthcare personnel from transfer of<br>microorganisms, body fluids, and particulate<br>material. These face masks are intended for<br>use in infection control practices to reduce<br>the potential exposure to blood and body<br>fluids. This is a single-use, disposable<br>device, provided non-sterile. | Same |
| Mask style | Flat-pleated | Flat-pleated | Same |
| ASTM<br>F2100<br>Level | Level 1, Level 2, Level 3 | Level 1, Level 2, Level 3 | Same |
| Design feature | Ear loop | Ear loop and Tie-on | Different |
| Color | Blue | Blue, White | Different |
| Dimension | Mask body: 175mm×95mm<br>Nose Clip: 100±5mm<br>Ear loop: 170±5mm | 175×90mm<br>Nose Clip: 100±5mm<br>Ear loop: 180×90mm | Different |
| Sterility | Non-Sterile | Non-Sterile | Same |
| Use | Single Use, Disposable | Single Use, Disposable | Same |
| Particulate<br>filtration<br>efficiency | Pass at 99.2% | Level 1: Pass at 99.6%<br>Level 2: Pass at 99.6%<br>Level 3: Pass at 99.7% | Different |
| Bacterial<br>filtration<br>efficiency | Pass at 99.2% | Level 1: Pass at 98%<br>Level 2: Pass at 98%<br>Level 3: Pass at 99% | Different |
| Differential<br>pressure | Pass at 3.8 mm H2O/cm² | Level 1: Pass at 2.0 mmH2O/cm²<br>Level 2: Pass at 1.6 mmH2O/cm²<br>Level 3: Pass at 2.5 mmH2O/cm² | Different |
| Flammability | Class 1 | Class 1 | Same |
| Fluid resistance | Pass at 160 mmHg | Level 1: Pass at 80 mmHg<br>Level 2: Pass at 120 mmHg<br>Level 3: Pass at 160 mmHg | Different |
| Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same |
| Patient Contacting Material | | | |
| Outer facing layer | Spunbond Polypropylene | Polypropylene | Different |
| Middle layer | Meltblown Polypropylene Filter | 1. Polypropylene spunbond<br>2. Polypropylene meltblown | |
| Inner facing layer | Spunbond Polypropylene | Polypropylene | |
| Nose clip | PE and Galvanized iron wire | Polyethylene coated steel wire | Same |
| Ear loop | Nylon and Spandex | Polyester, polyurethane | |
| Biocompatibility | | | |
| Cytotoxicity | Under the conditions of the study, the proposed device was non-cytotoxic. | Under the conditions of the study, the subject device was non-cytotoxic. | |
| Sensitization | Under the conditions of the study, the proposed device was non-sensitizing. | Under the conditions of the study, the subject device was non-sensitizing. | |
| Irritation | Under the conditions of the study, the proposed device was non-irritating. | Under the conditions of the study, the subject device was non-irritating. | |
# Table 1 Comparison of Medical Face Mask
{6}------------------------------------------------
### Different - Design feature
The proposed masks are ear-loop masks. The design features of the proposed device can be covered by the design features of the predicate device. Thus, this difference will not affect the safety and effectiveness of the proposed device.
# Different - Color
The proposed device is blue and the predicate device is provided in two colors, the proposed device can be covered by the predicate device. In addition, the biocompatibility test has been conducted and test results did not show any adverse effects. Thus, this difference will not affect the safety and effectiveness of the proposed device.
## Different - Dimension
The dimension for the proposed device is different from the predicate device. However, the difference does not affect the indication for use and will not raise safety issues. Thus, this difference will not affect the safety and effectiveness of the proposed device.
## Different - Particulate filtration efficiency
The test result for particulate filtration efficiency for the proposed device is different from the predicate
{7}------------------------------------------------
device. The proposed devices are available in Level 1, Level 3 based on ASTM F2100-19, depending on marketing reasons. However, Level 1, Level 3 masks are exactly the same. Therefore, the particulate filtration efficiency test was conducted on Level 3 masks and the test result can meet the requirements based on ASTM F2100-19. Thus, this difference will not affect the safety and effectiveness of the proposed device.
### Different - Bacterial filtration efficiency
The test result for bacterial filtration efficiency for the proposed device is different from the predicate device. The proposed devices are available in Level 1, Level 3 based on ASTM F2100-19, depending on marketing reasons. However, Level 1, Level 2 and Level 3 masks are exactly the same. Therefore, the bacterial filtration efficiency test was conducted on Level 3 masks and the test result can meet the requirements based on ASTM F2100-19. Thus, this difference will not affect the safety and effectiveness of the proposed device.
#### Different - Differential pressure
The test result for differential pressure for the proposed device is different from the predicate device. The proposed devices are available in Level 1, Level 2 and Level 3 based on ASTM F2100-19, depending on marketing reasons. However, Level 1, Level 2 and Level 3 masks are exactly the same. Therefore, the differential pressure test was conducted on Level 3 masks and the test result can meet the requirements of the standard ASTM F2100-19 which was recognized by FDA. Thus, this difference will not affect the safety and effectiveness of the proposed device.
### Different - Fluid resistance
The test result for fluid resistance for the proposed device is different from the predicate device. The proposed devices are available in Level 1, Level 2 and Level 3 based on ASTM F2100-19, depending on marketing reasons. However, Level 1, Level 2 and Level 3 masks are exactly the same. Therefore, the fluid resistance test was conducted on Level 3 masks and the test result can meet the requirements based on ASTM F2100-19. Thus, this difference will not affect the safety and effectiveness of the proposed device.
#### Different - Patient Contacting Material
The patient contacting material for the proposed device is different from the predicate device. However, biocompatibility test has been conducted on the proposed device and the result does not show any adverse effect. Thus, this difference will not affect the safety and effectiveness of the proposed device.
{8}------------------------------------------------
#### 7. Summary of Non-Clinical Test
Non clinical tests were conducted to verify that the proposed device met all design specifications as was same/similar to the predicate devices. The test results demonstrated that the proposed device complies with the following standards:
- > 16 CFR Part 1610 Standard for the Flammability of Clothing Textiles;
- > ASTM F1862/F1862M: 2017 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
- > ASTM F2299/F2299M-03 (2017) Standard Test Method for Determining the Initial Efficiency of Material Used in medical Face Masks to Penetration by Particulates using Latex Spheres
- > ASTM F2101: 2019 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
- > ASTM F2100: 2019 Standard Specification for Performance of Materials Used in Medical Face Masks
- > EN 14683: 2019, Annex C, Medical face masks- Requirements and test methods
- > ISO 10993-5:2009 Biological evaluation of medical device- Part 5: Tests for in vitro cytotoxicity
- > ISO 10993-10:2010 Biological evaluation of medical device- Part 10: Tests for irritation and skin sensitization
- > ISO 10993-1:2018 Biological evaluation of medical devices - Part 1 Evaluation and testing within a risk management process.
| Performance<br>Characteristics | Test Method | Test result | Acceptance Criteria | Conclusion |
|--------------------------------------|-----------------------------------|---------------------------|-------------------------------|----------------------|
| Fluid Resistance<br>Performance | ASTM<br>F1862/F1862M:<br>2017 | Pass at 160<br>mmHg | No penetration at 160<br>mmHg | Meet the requirement |
| Particulate Filtration<br>Efficiency | ASTM<br>F2299/F2299M-03<br>(2017) | Pass at 99.2% | ≥98% | Meet the requirement |
| Bacterial Filtration<br>Efficiency | ASTM F2101:<br>2019 | Pass at 99.2% | ≥98% | Meet the requirement |
| Differential Pressure<br>(Delta-P) | EN 14683: 2019,<br>Annex C | Pass at 3.8 mm<br>H2O/cm2 | <6.0 mmH2O/cm2 | Meet the requirement |
| Flammability class | 16 CFR Part 1610 | Class 1 | Class 1 | Meet the requirement |
#### 8. Clinical Test Conclusion
No clinical study is included in this submission.
#### 9. Conclusion
The conclusion drawn from the non-clinical tests demonstrates that the medical face mask is as safe, as
{9}------------------------------------------------
effective, and performs as well as or better than the legally marketed predicate device cleared under K160269.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.