SIMPACT Sacroiliac Joint Fixation System

K212903 · Life Spine, Inc. · OUR · Dec 15, 2021 · Orthopedic

Device Facts

Record IDK212903
Device NameSIMPACT Sacroiliac Joint Fixation System
ApplicantLife Spine, Inc.
Product CodeOUR · Orthopedic
Decision DateDec 15, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Life Spine SIMPACT Sacroiliac Joint Fixation System is intended for sacroiliac fusion for the following conditions: · Sacroiliac joint dysfunction that is a direct result of sacroiliac joint disruption and degenerative sacroiliitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months. · To augment immobilization and stabilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion. · Acute, non-acute, and non-traumatic fractures involving the sacroiliac joint.

Device Story

SIMPACT Sacroiliac Joint Fixation System consists of fully and partially threaded cannulated titanium alloy screws; used for sacroiliac joint fusion and stabilization. Device implanted by surgeons in clinical settings to provide mechanical fixation of sacroiliac joint. System components are single-use; intended to facilitate fusion through immobilization. No software or electronic components involved.

Clinical Evidence

No clinical data. Performance data consists of bench testing/Design of Experiment to validate the Posterior Oblique approach.

Technological Characteristics

Fully and partially threaded cannulated screws; material: Titanium alloy 6AL-4V-ELI per ASTM F-136; single-use; mechanical fixation; no software or energy source.

Indications for Use

Indicated for sacroiliac fusion in skeletally mature patients with sacroiliac joint dysfunction due to disruption or degenerative sacroiliitis (including postpartum cases >6 months), patients requiring sacropelvic fixation during lumbar/thoracolumbar fusion, and patients with acute, non-acute, or non-traumatic sacroiliac joint fractures.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ December 15, 2021 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION". Life Spine, Inc. Angela Batker RA/QA Manager 13951 S. Quality Drive Huntley, Illinois 60142 Re: K212903 Trade/Device Name: SIMPACT Sacroiliac Joint Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: OUR, HWC Dated: October 14, 2021 Received: October 15, 2021 Dear Angela Batker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212903 Device Name Life Spine SIMPACT Sacroiliac Joint Fixation System #### Indications for Use (Describe) The Life Spine SIMPACT Sacroiliac Joint Fixation System is intended for sacroiliac fusion for the following conditions: · Sacroiliac joint dysfunction that is a direct result of sacroiliac joint disruption and degenerative sacroiliitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months. · To augment immobilization and stabilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion. · Acute, non-acute, and non-traumatic fractures involving the sacroiliac joint. | Type of Use (Select one or both, as applicable) | |-----------------------------------------------------------------------------| | <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary Life Spine SIMPACT Sacroiliac Joint Fixation System | Submitted By: | Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 | |------------------------|---------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Contact: | Angela Batker<br>Life Spine, Inc.<br>13951 S. Quality Drive<br>Huntley, IL 60142<br>Telephone: 847-884-6117<br>Fax: 847-884-6118 | | Date Prepared: | September 10th, 2021 | | Trade Name: | Life Spine SIMPACT Sacroiliac Joint Fixation System | | Common Name: | Spinal Interlaminal Fixation Orthosis | | Classification: | OUR, 21 CFR 888.3040, Class II<br>HWC, 21 CFR 888.3040, Class II | | | Smooth or Threaded Metallic Bone Fastener | | Primary Predicate: | Medtronic, Inc. Rialto Sacroiliac Fusion System (K161210) | | Additional Predicates: | Life Spine Sacroiliac Joint Fixation Screw System (K201538,K180749,K141246)<br>Medtronic, Inc. MSB Sacroiliac Joint Fusion Device (K110472) | # Device Description: The Life Spine SIMPACT Sacroiliac Joint Fixation System consists of fully threaded screws and partially threaded cannulated screws in various diameters and lengths to enhance sacroiliac joint fusion. All components are fabricated and manufactured from titanium alloy 6AL-4V-ELI per ASTM F-136. The purpose of this submission is to add the Posterior Oblique approach technique. All implants are intended for single use only and should not be reused under any circumstances. Do not use any of the Life Spine Simpact Sacroiliac Joint Fixation System components with components from any other system or manufacturer. The Life Spine Simpact Sacroiliac Joint Fixation System components should never be reused under any circumstances. {4}------------------------------------------------ ## Indications for Use of the Device: The Life Spine SIMPACT Sacroiliac Joint Fixation System is intended for sacroiliac fusion for the following conditions: · Sacroiliac joint dysfunction that is a direct result of sacroiliac joint disruption and degenerative sacroiliitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months. · To augment immobilization and stabilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion. · Acute, non-acute, and non-traumatic fractures involving the sacroiliac joint. ## Technological Characteristics: The Life Spine Simpact Sacroiliac Joint Fixation System is substantially equivalent to the predicate systems in terms of design, materials, indications for use and sizing. ## Material: This submission seeks clearance of a device made from implant grade (Ti-6AL-4V) Titanium alloy according to F136. This this is the same material used in the predicate devices. # Performance Data: There are no new devices in this submission, only Design of Experiment validating Posterior Oblique approach. # Substantial Equivalence: The Life Spine Simpact Sacroiliac Joint Fixation System was shown to be substantially equivalent to the predicate devices in indications for use, design, function, materials used and mechanical performance. # Conclusion: The information presented demonstrates the substantial equivalency of The Life Spine Simpact Sacroiliac Joint Fixation System.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...