K212891 · Cynosure, LLC · GEI · Mar 18, 2022 · General, Plastic Surgery
Device Facts
Record ID
K212891
Device Name
TempSure System
Applicant
Cynosure, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 18, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Small 10mm, 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids. The Small 10mm, 18mm, 25mm, 30mm, and 60mm handpieces and the FlexSure Applicators provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The massage device is intended to provide a temporary reduction in the appearance of cellulite. The following surgical modes are applicable to the generator: Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed. Cutting: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, tumors, cosmetic repairs, development of skin flaps, skin tags and blepharoplasty. Blended Cutting and Coagulation: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids, keratosis, verrucae, basal cell carcinoma, nevi, fistulas, epithelidma, cosmetic repairs, cysts, abscesses, and development of skin flaps. Fulguration: basal cell carcinoma, papilloma, cyst destruction, tumors. verrucae, hemostasis. Bipolar: pinpoint precise coagulation, pinpoint hemostasis in any, field (wet or dry), snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment and turbinate shrinkage.
Device Story
TempSure System is a radiofrequency (RF) generator for aesthetic and surgical procedures. Inputs include user settings via GUI and activation via foot/hand-switch. Device delivers RF energy to tissue to induce heating or surgical cutting/coagulation. Used in clinical settings by healthcare professionals. New Small 10mm handpiece includes an integrated accelerometer to detect motion; RF output disables if no motion is detected to prevent overheating. Output allows clinicians to perform precise tissue heating or surgical interventions. Benefits include wrinkle reduction, pain relief, and surgical hemostasis.
Clinical Evidence
Bench testing only. Tissue heating bench test confirmed the Small 10mm handpiece maintains target temperature for 10 minutes under simulated clinical conditions. Electrical safety and electromagnetic compatibility testing performed per IEC 60601-1, 60601-1-2, and 60601-2-2.
Indicated for patients requiring non-ablative treatment of mild to moderate facial wrinkles/rhytids, temporary relief of pain, muscle spasms, increased local circulation, temporary reduction in cellulite appearance, and various general surgical procedures (coagulation, cutting, hemostasis).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is written in blue letters next to it.
March 18, 2022
Cynosure, LLC Michael King Regulatory Affairs Specialist III 5 Carlisle Road Westford, Massachusetts 01886
Re: K212891
Trade/Device Name: TempSure System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI, PBX Dated: February 17, 2022 Received: February 18, 2022
Dear Michael King:
We have reviewed your Section 510/k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) BBCDOX K212891
Device Name TempSure System
#### Indications for Use (Describe)
The Small 10mm, 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids.
The Small 10mm, 18mm, 25mm, 30mm, and 60mm handpieces and the FlexSure Applicators provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.
The massage device is intended to provide a temporary reduction in the appearance of cellulite.
The following surgical modes are applicable to the generator:
Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed.
Cutting: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, turnors, cosmetic repairs, development of skin flaps, skin tags and blepharoplasty.
Blended Cutting and Coagulation: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids, keratosis, verrucae, basal cell carcinoma, nevi, fistulas, epithelidma, cosmetic repairs, cysts, abscesses, and development of skin flaps.
Fulguration: basal cell carcinoma, papilloma, cyst destruction, tumors. verrucae, hemostasis.
Bipolar: pinpoint precise coagulation, pinpoint hemostasis in any, field (wet or dry), snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment and turbinate shrinkage.
| Type of Use ( <i>Select one or both, as applicable</i> ) |
|------------------------------------------------------------------------------------------------|
| <span style="padding-left: 20px;"><b>☑</b> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
| <span style="padding-left: 20px;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Cynosure, LLC
Section 5
510(k) Summary
### 510(k) Summary for Cynosure TempSure
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| 807.92(a)(1) Submitter Information | |
|--------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | Cynosure, LLC |
| Address | 5 Carlisle Road, Westford MA, 01886 |
| Phone Number | 781-993-2454 |
| Fax Number | 978-256-6556 |
| Establishment Registration<br>Number | 1222993 |
| Contact Person | Michael King |
| Preparation Date | September 9, 2021 |
| 807.92(a)(2) Name of Device | |
| Trade or Proprietary Name | TempSure System |
| Common or Usual Name | Surgical RF Generator |
| Classification Name | Electrosurgical, Cutting & Coagulation & Accessories;<br>Massager, Vacuum, Radio Frequency Induced Heat |
| Classification Panel | General & Plastic Surgery |
| Regulation | 21 CFR 878.4440 |
| Regulatory Class | II |
| Product Code(s) | GEI, PBX |
| 807.92 (a)(3) Legally marketed device(s) to which equivalence is claimed | |
| Predicate Devices | Cynosure TempSure (K200241) |
| 807.92(a)(4) Device Description | |
| | The Cynosure TempSure™ System is a radiofrequency generator with<br>a variety of applications both aesthetic and surgical procedures. The<br>intended action is achieved through application of radiofrequency<br>energy to the patient which results in minimization of heat dissipation<br>and cellular alteration. Output of energy is controlled via the guided<br>user interface (GUI) and the foot and/or hand-switch.<br><br>An additional Small 10mm Smart Handpiece is now included for the<br>purpose of tissue heating for selected medical conditions and the<br>treatment of mild to moderate facial wrinkles and rhytids. The<br>handpiece is capable of operating at maximum energy setting of 30<br>(28W) in the Smart Handpiece Mode. There have been no additional<br>changes to the previous existing handpieces (10, 15, 18, 20, 25, 30 or<br>60mm), FlexSure Applicators, or Surgical modes (COAG, Cut, Blend, |
| Section 5 | 510(k) Summary |
| | Bipolar, Fulgurate) of the system. The TempSure system is still used with the same existing electrosurgical accessories. |
| | The TempSure™ System includes: |
| | TempSure™ Generator Temperature Sensing Handpieces (Small 10, 10, 15, 18, 20, 25, 30, and 60mm) Massage Heads (25, 30, and 60mm) FlexSure™ Applicators (Large and Medium) IEC Power Cord Footswitch Disposable/Reusable Neutral Pads Surgical Fingerswitch/Foot Controlled Handpieces Monopolar Cables Disposable/Reusable Electrodes, Forceps |
| | 807.92(a)(5) Intended Use of the Device |
| | The Small 10mm, 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids.<br><br>The Small 10mm, 18mm, 25mm, 30mm, and 60mm handpieces and the FlexSure Applicators provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.<br><br>The massage device is intended to provide a temporary reduction in the appearance of cellulite<br><br>The following surgical modes are applicable to the generator:<br><br>Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed.<br><br>Cutting: snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, tumors, cosmetic repairs, development of skin flaps, skin tags, and blepharoplasty.<br><br>Blended Cutting and Coagulation: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids, keratosis, verrucae, basal cell carcinoma, |
| | nevi, fistulas, epithelidma, cosmetic repairs, cysts, abscesses, and<br>development of skin flaps. |
| | <b>Fulguration:</b> basal cell carcinoma, papilloma, cyst destruction,<br>tumors, verrucae, hemostasis. |
| | <b>Bipolar:</b> pinpoint precise coagulation, pinpoint hemostasis in any,<br>field (wet or dry), snoring, submucosal palatal shrinkage, traditional<br>uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage<br>control, epistaxis treatment, and turbinate shrinkage |
| | 807.92(a)(6) Summary of the Technological Characteristics of the Device Compared to |
{6}------------------------------------------------
{7}------------------------------------------------
Section 5
the Predicate - Please refer to the System Specification Comparison Table below
### 807.92(b)(1) Non-clinical tests submitted
The following non-clinical tests have been included in this 510k submission in determination of substantial equivalence between the test device and the referenced predicates.
#### Tissue Heating with Small 10mm Handpiece
Bench testing was conducted to show that the new Small 10mm Smart Handpiece, when used in a simulated clinical condition, was able to heat and maintain temperature of the treatment area for 10 minutes. Refer to Section 18 - Performance Testing, Bench for additional information.
### Electromagnetic Compatibility and Electrical Safety
Electrical safety testing for the Cynosure TempSure was also completed to prove the safe use of the device. These test reports are provided in accordance with FDA Guidance "Premarket Notification 510k Submissions for Electrosurgical Devices for General Surgery" - Section XII - Electrical Safety and Electromagnetic Compatibility". The following test reports are available in Section 17 -Electromagnetic Compatibility and Electrical Safety.
- IEC 60601-1, Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performances
- IEC 60601-1-2, Medical Electrical Equipment – Part 1 -2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances requirements and tests
- IEC 60601-2-2, Medical electrical equipment Part 2 -2: Particular requirements for the basic o safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
### 807.92(b)(2) Clinical tests submitted - N/A - No clinical tests submitted
807.92(b)(3) Conclusions drawn from clinical and non-clinical tests submitted
{8}------------------------------------------------
Page 4 of 7 Cynosure, LLC
Section 5
510(k) Summary
The non-clinical tests demonstrate that the TempSure generator with the Small 10mm handpiece is safe and effective and performs as well or better than the legally marketed predicate device. The Tissue Heating bench test demonstrated that the device is able to maintain temperature for at least the 10minute treatment time when used as intended under clinical conditions at various treatment areas and settings. In addition, the Electrosurgical Compatibility and Electrical Safety testing shows that the device is safe to use and meets required standards. These non-clinical tests show that the TempSure generator using the Small 10mm handpiece meets design specifications as well as performance requirements.
{9}------------------------------------------------
Cynosure TempSure 510(k)
KPending
| Characteristic | Cynosure TempSure System (KPending) | Cynosure TempSure System (K200241) |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Small 10mm, 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids.<br><br>The Small 10mm, 18mm, 25mm, 30mm, and 60mm handpieces and the FlexSure Applicators provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. | The 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids.<br><br>The 18mm, 25mm, 30mm, and 60mm handpieces and the FlexSure Applicators provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. |
| | The massage device is intended to provide a temporary reduction in the appearance of cellulite | The massage device is intended to provide a temporary reduction in the appearance of cellulite |
| | The following surgical modes are applicable to the generator: | The following surgical modes are applicable to the generator: |
| | Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed | Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed |
| | Cutting: snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage. skin incisions, biopsy, cysts, abscesses. tumors. cosmetic repairs, development of skin flaps. skin tags and blepharoplasty. | Cutting: snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage. skin incisions, biopsy, cysts, abscesses. tumors. cosmetic repairs, development of skin flaps. skin tags and blepharoplasty. |
| | Blended Cutting and Coagulation: snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment, | Blended Cutting and Coagulation: snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment, |
| Section 5 | 510(k) Summary | |
| Characteristic | Cynosure TempSure System (KPending) | Cynosure TempSure System (K200241) |
| | effective hemorrhage control. epistaxis treatment,<br>turbinate shrinkage, skin tags, papilloma keloids.<br>keratosis, verrucae, basal cell carcinoma. nevi. fistulas.<br>epithelidma. cosmetic repairs. cysts. abscesses. and<br>development of skin flaps. | turbinate shrinkage, skin tags, papilloma keloids.<br>keratosis, verrucae, basal cell carcinoma. nevi. fistulas.<br>epithelidma. cosmetic repairs. cysts. abscesses. and<br>development of skin flaps. |
| | Fulguration: basal cell carcinoma. papilloma. cyst<br>destruction, tumors. verrucae, hemostasis. | Fulguration: basal cell carcinoma. papilloma. cyst<br>destruction, tumors. verrucae, hemostasis. |
| | Bipolar: pinpoint precise coagulation. pinpoint<br>hemostasis in any, field (wet or dry). snoring.<br>submucosal palatal shrinkage, traditional<br>uvulopalatoplasty (RAUP). myringotomy with<br>effective hemorrhage control. epistaxis treatment and<br>turbinate shrinkage | Bipolar: pinpoint precise coagulation. pinpoint<br>hemostasis in any, field (wet or dry). snoring.<br>submucosal palatal shrinkage, traditional<br>uvulopalatoplasty (RAUP). myringotomy with<br>effective hemorrhage control. epistaxis treatment and<br>turbinate shrinkage |
| Rx/OTC | Prescription | Prescription |
| Energy Type | Radiofrequency | Radiofrequency |
| Modality | Monopolar, Bipolar | Monopolar, Bipolar |
| Temperature Sensing | Temperature-Sensitive Handpiece | Temperature-Sensitive Handpiece |
| Temperature Response Time | <1 second | <1 second |
| Handpiece Size | Small 10mm, 10mm, 15mm, 18mm, 20mm, 25 mm, 30<br>mm, 60mm, Large & Medium FlexSure Applicators | 10mm, 15mm, 18mm, 20mm, 25 mm, 30 mm, 60mm,<br>Large & Medium FlexSure Applicators |
| Massage Head | Yes (25mm, 30mm, 60mm) | Yes (25mm, 30mm, 60mm) |
| Surgical Accessories | The TempSure™ generator is compatible with<br>existing, previously cleared electrosurgical<br>accessories: neutral pads, electrodes, surgical<br>handpieces, cables, etc. | The TempSure™ generator is compatible with<br>existing, previously cleared electrosurgical accessories:<br>neutral pads, electrodes, surgical handpieces, cables,<br>etc. |
| Treatment Activation | Fingerswitch Footswitch | Fingerswitch Footswitch |
| Section 5 | | 510(k) Summary |
| Characteristic | Cynosure TempSure System (KPending) | Cynosure TempSure System (K200241) |
| esthetic Optimal<br>Temperature | 39-45°C | 39-45°C |
| Patient Contacting Material | Small 10mm Handpiece: Gold-plated 316 stainless<br>steel<br>Handpieces: PVDF coating, Gold-plated brass, Gold-<br>plated aluminum, Loctite (M-31CL), Polyetherimide<br>(Ultem 1000)<br>FlexSure: Polyimide, Hydrogel<br>Massage Heads: Polycarbonate (Sabic Lexan HP-1),<br>Delrin | Handpieces: PVDF coating, Gold-plated brass, Gold-<br>plated aluminum, Loctite (M-31CL), Polyetherimide<br>(Ultem 1000)<br>FlexSure: Polyimide, Hydrogel<br>Massage Heads: Polycarbonate (Sabic Lexan HP-1),<br>Delrin |
| Input Voltage | 110V | 110V |
| Output Waveform | 4.0 MHz Sin-wave CW, Fully Rectified, Partially<br>Rectified, and 1.7 MHz for Bipolar | 4.0 MHz Sin-wave CW, Fully Rectified, Partially<br>Rectified, and 1.7 MHz for Bipolar |
| Modes | Surgical (Coag, Cut, Blend, Fulgurate, Bipolar)<br>Smart Handpiece Mode…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.