The JFil® Ureteral Stents are used for temporary internal drainage of the ureteropelvic junction. The stents may be placed using endoscopic techniques. The stents are not intended as permanent indwelling devices, it is recommended that the indwelling time does not exceed 14 days. Target population: adults only (greater than 21 years of age).
Device Story
JFil Ureteral Stents are polyurethane tube-like devices designed for temporary internal drainage of the ureteropelvic junction. The device features a single proximal pigtail loop and a distal polypropylene-monofilament surgical thread for cystoscopic removal. Stents are 6.0 or 7.0 French in diameter and 8 or 16 cm in length. The kit includes a 63 cm pusher with a radio-opaque ring for placement; some kits include a Nitinol guidewire. Used in clinical or hospital environments by trained professionals, the device is placed endoscopically to treat ureteral injuries and obstructions. The output is the physical drainage of the ureter. The device benefits patients by providing temporary relief of obstructions. It is sterile, single-use, and intended for short-term (≤14 days) indwelling.
Clinical Evidence
Bench testing only. Testing included drainage capacity (ASTM F623-99), curl strength, tensile/yield strength, and shelf-life validation. Biocompatibility was confirmed per ISO 10993-1.
Technological Characteristics
Polyurethane stent; 6.0/7.0 Fr diameter; 8/16 cm length. Single pigtail loop (proximal); polypropylene-monofilament thread (distal). Radio-opaque pusher. Sterilized via Ethylene Oxide (EO). Shelf-life: 3 years. Guidewire compatibility: 0.035".
Indications for Use
Indicated for temporary internal drainage of the ureteropelvic junction in adults (>21 years). Not for permanent use; indwelling time ≤14 days.
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 10, 2022
Rocamed SAM Tahiana Rasolofoniaina Regulatory Affairs Specialist 9 Avenue Albert II Monaco, Monaco 98000 Monaco
Re: K212868
Trade/Device Name: JFil Ureteral Stents, JFil Ureteral Stents ECO KIT Regulation Number: 21 CFR 876.4620 Regulation Name: Ureteral stent Regulatory Class: Class II Product Code: FAD
Dear Tahiana Rasolofoniaina:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 26, 2022. Specifically, FDA is updating this SE Letter with corrected model numbers of your device as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jessica K. Nguyen, Ph.D., OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices, Jessica. Nguyen@fda.hhs.gov.
Sincerely,
Jessica K. Nguyen -S
Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA) along with the Department of Health & Human Services-USA. The FDA logo is in blue and includes the acronym "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION." To the left of the FDA logo is the Department of Health & Human Services logo, which features a stylized emblem. The logos are positioned horizontally, with the Department of Health & Human Services logo on the left and the FDA logo on the right.
January 26, 2022
Rocamed SAM Tahiana Rasolofoniaina Regulatory Affairs Specialist 9 Avenue Albert II Monaco, Monaco 98000 Monaco
Re: K212868
> Trade/Device Name: JFil® Ureteral Stents, JFil® ECO Kit Regulation Number: 21 CFR§ 876.4620 Regulation Name: Ureteral Stent Regulatory Class: II Product Code: FAD Dated: December 23, 2021 Received: December 29, 2021
Dear Tahiana Rasolofoniaina:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Angel A. Soler-garcia -S
For Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212868
Device Name JFil® Ureteral Stents, JFil® ECO KIT
Indications for Use (Describe) The JFil® Ureteral Stents are used for temporary internal drainage of the ureteropelvic junction.
The stents may be placed using endoscopic techniques.
The stents are not intended as permanent indwelling devices, it is recommended that the indwelling time does not exceed 14 days.
Target population: adults only (greater than 21 years of age).
Type of Use (Select one or both, as applicable)> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
K212868 Page 1 of 4
# 510 (k) Summary
# 1. Submitter Information
| 510 (k) submitter: | ROCAMED SAM |
|--------------------------|----------------------------------------------------------------|
| Address: | 9 Avenue Albert II<br>98000 Monaco<br>MONACO (Principality of) |
| Contact Person Name: | Tahiana Rasolofoniaina |
| Contact Title: | Regulatory Affairs Specialist |
| Contact Person's email: | tr@promepla.com |
| Contact Person's Number: | (+377) 979-842-44 |
| Preparation date: | January 25, 2022 |
## 2. Device Information
| Trade Name of the Device: | JFil® Ureteral Stents, JFil® ECO KIT |
|----------------------------|--------------------------------------|
| Common Name | Ureteral Stent |
| Classification Name: | Stent, Ureteral |
| Classification Regulation: | 21 CFR 876.4620 |
| Device Class: | II |
| Review Panel | Gastroenterology/Urology |
| Product Code | FAD |
### 3. Predicate Devices
| 510(k) number | K030503 | K173734 |
|---------------------------|------------------------------|--------------------------|
| Owner | Boston Scientific Corp. | Promepla SAM |
| Trade name of the Device | Polaris Loop Ureteral Stents | RocaJJ Soft Stents |
| Classification Name | Stent, Ureteral | Stent, Ureteral |
| Regulation Classification | 21 CFR 876.4620 | 21 CFR 876.4620 |
| Device Class | II | II |
| Panel | Gastroenterology/Urology | Gastroenterology/Urology |
| Product Code | FAD | FAD |
Both predicate devices have not been subjected to a design-related recall.
{6}------------------------------------------------
## 4. Device Description
The JFil® Ureteral Stents consist of a tube-like device that is inserted into the ureter to allow the passage of the urine in the treatment of ureteral injuries and ureteral obstructions. This polyurethane stent has a single loop on the proximal tip and a polypropylene-monofilament surgical thread on the distal tip. JFil stents are 6.0 or 7.0 French (Fr) in diameter and 8 or 16 cm in specified length. The stent is provided with a 63 cm length pusher and may include a Nitinol Guidewire in the kit. The pusher consists of a radio-opaque ring connected to a tubing, and is intended to allow pushing, placing, and releasing the stent inside the patient. This stents can only be used by trained professionals in a clinic or hospitals environment. The JFil Ureteral Stents are sterile and for single use only. The stents are also available in ECO-kits which do not include a guidewire. The stents can be removed cystoscopically by gently pulling on the surgical thread.
| Designation | Ø (Fr) | Length<br>(cm) | Presence<br>of guide | Designation of the product |
|-------------|--------|----------------|----------------------|---------------------------------------------|
| ROJV0608ST | 6 | 8 | Yes | JFil® - Ureteral Stent - 6Fr ; 8cm |
| ROJV0616ST | 6 | 16 | Yes | JFil® - Ureteral Stent - 6Fr ; 16cm |
| ROJV0708ST | 7 | 8 | Yes | JFil® - Ureteral Stent - 7Fr ; 8cm |
| ROJV0716ST | 7 | 16 | Yes | JFil® – Ureteral Stent – 7Fr ; 16cm |
| ROJV5708ST | 7 | 8 | No | JFil® ECO KIT - Ureteral Stent - 7Fr ; 8cm |
| ROJV5716ST | 7 | 16 | No | JFil® ECO KIT - Ureteral Stent - 7Fr ; 16cm |
The models are described in the table below:
### 5. Intended Use
The JFil® Ureteral Stents are used for temporary internal drainage of the ureteropelvic junction.
The stents may be placed using endoscopic techniques.
The stents are not intended as permanent indwelling devices, it is recommended that the indwelling time does not exceed 14 days.
Target population: adults only (greater than 21 years of age).
#### 6. Comparison of The Technological Characteristics with Predicate Devices:
| Device & Predicate<br>Trade name<br>(Applicant) | Device under<br>evaluation<br>JFil® Ureteral<br>Stents<br>(Rocamed) | Primary Predicate<br>- K030503<br>Polaris™ Loop<br>Ureteral Stents<br>(Boston<br>Scientific) | Secondary Predicate<br>- K173734 RocaJJ<br>Soft Ureteral Stents<br>(Promepla) |
|-------------------------------------------------|---------------------------------------------------------------------|----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Regulation Number | 876.4620 | 876.4620 | 876.4620 |
| Regulation Name | Stents, Ureteral | Stents, Ureteral | Stents, Ureteral |
| Product Code | FAD | FAD | FAD |
| Classification | Class II | Class II | Class II |
| Stent Type | Single J Stent | Single J Stent | Double J Stent |
{7}------------------------------------------------
| K212868 |
|-------------|
| Page 3 of 4 |
==============================================================================================================================================================================
| Intended Use<br>Conditions | -Surgical suite<br>in a hospital<br>environment<br>- Medical facilities<br>-Under aseptic<br>conditions<br>(clothing, sterile<br>surgical gloves,<br>and a controlled<br>atmosphere) | The use conditions for the<br>device are a surgical suite<br>in a hospital environment.<br>Ureteral stents are handled<br>by surgeons under aseptic<br>conditions (clothing, sterile<br>surgical gloves, and a<br>controlled atmosphere).<br>They will be placed for a<br>length of time specified by<br>the urologist surgeon. | - Surgical suite in a<br>hospital environment<br>- Medical facilities<br>- Under aseptic<br>conditions<br>(clothing, sterile<br>surgical gloves,<br>and a controlled<br>atmosphere) |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prescription of use | Sale by or on the<br>order of a physician | Sale by or on the order<br>of a physician | Sale by or on the<br>order of a physician |
| Reuse Status | Single Use | Single Use | Single Use |
| Sterilization | Ethylene oxide (EO) | Ethylene oxide (EO) | Ethylene oxide<br>(EO) |
| Size (Fr) | 6/7 | 5/6/7/8 | 4,8/6/7/8 |
| Length (cm) | 8<br>16 | 16<br>10/12/14/18 /20/22/<br>24/26/28/30 | 24/26/28/30 |
| Stent type | Simple J | Simple J | Double J |
| Renal loop | Pigtail loop | Pigtail loop | Pigtail loop |
| Bladder loop | Suture thread | Suture loop | Pigtail loop |
| Suture<br>material | Polypropylene | Polypropylene | Polypropylene |
| Guidewire<br>Compatibility | 0.035" | 0.038" | 0.035" |
| Ink Mark | No | Yes | Yes |
| Shelf-life | 3 Years | Unknown | 3 Years |
| Radiopacity | Yes | Yes | Yes |
| Placement | Placed using<br>endoscopic<br>techniques | placement endoscopically<br>or<br>fluoroscopically | Placed using<br>endoscopic<br>techniques |
| Indwelling time | 14 days | 14 days | 6 days |
{8}------------------------------------------------
As evidenced by the above table, both the subject and the predicate devices have the same intended use, but the subject and predicate devices have different technological characteristics. However, these differences do not raise different questions of safety or effectiveness and the testing mentioned below showed that the subject is substantially equivalent with the predicates.
### 7. Performance Data
#### Performance Data - Bench
In support of this premarket notification, Rocamed conducted the following bench performance testing -
- Drainage capacity, Per ASTM F623-99, "Standard Performance Specification for Foley ● Catheter"
- Curl strength ●
- Tensile and elongation/yield strength
- . Shelf life validation
All bench testing results confirmed that the products described in this submission met the necessary specification.
#### Biocompatibility Testing
Biocompatibility of the subject stents was confirmed per ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The device met the requirements.
#### 8. Conclusion
Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.