SmartBand EMR Kit (SB-EMR-K, SB-EMR-K-12, SB-EMR-K-9.4), SmartBand EMR Pack (SB-EMR-P, SB-EMR-P-12, SB-EMR-P-9.4), SmartSnare EMR Hexagonal Snare (SS-230-1 or packaged with kit)
K212860 · STERIS Corporation · FDI · Oct 7, 2021 · Gastroenterology, Urology
Device Facts
| Record ID | K212860 |
| Device Name | SmartBand EMR Kit (SB-EMR-K, SB-EMR-K-12, SB-EMR-K-9.4), SmartBand EMR Pack (SB-EMR-P, SB-EMR-P-12, SB-EMR-P-9.4), SmartSnare EMR Hexagonal Snare (SS-230-1 or packaged with kit) |
| Applicant | STERIS Corporation |
| Product Code | FDI · Gastroenterology, Urology |
| Decision Date | Oct 7, 2021 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 876.4300 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The SmartBand® EMR Device is intended for endoscopic mucosal resection in the upper GI tract.
Device Story
Endoscopic mucosal resection (EMR) device for upper GI tract; kit/pack configuration. Components: ligation handle, loading device, connector tube, cylindrical barrel with 5 bands, deployment cord, hexagonal snare. Operation: manual ligation of tissue followed by diathermic resection using snare connected to electrosurgical generator (e.g., ERBE VIO 300-D, US Endoscopy gi4000). Used in clinical settings by physicians. Universal connector allows attachment to endoscopes lacking luer connections. Band lock secures bands during shipping. Black band indicates proximity to end of supply. Benefits: provides options for treatment area size; facilitates tissue resection; extended shelf life (24 months).
Clinical Evidence
Bench testing only. Includes verification testing (measurements, visual inspection, shipping, packaging), biocompatibility testing (ISO 10993-4, -5, -10, -11; USP38-NF33), and simulated use testing.
Technological Characteristics
Materials: stainless steel (snare loop, loading device), Teflon (sheath), Nitinol (core wire), polyurethane (strain relief, band lock), ABS (finger slide, handle, flush tube), non-latex synthetic polyisoprene (bands), polycarbonate (barrel), silicone (universal connector). Energy: diathermic (snare). Sterilization: Ethylene Oxide (SAL 10^-6). Connectivity: manual interface with electrosurgical generators.
Indications for Use
Indicated for patients undergoing endoscopic mucosal resection in the upper GI tract.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Predicate Devices
- SmartBand EMR Kit (K190512)
Related Devices
- K190512 — SmartBand EMR Kit, SmartBand EMR Pack, SmartSnare · Intelligent Endoscopy · Nov 22, 2019
- K140726 — CAPTIVATOR EMR DEVICE FOR STANDARD GASTROSCOPES - BOX 1, CAPTIVATOR EMR DEVICE FOR STANDARD GASTROSCOPES - BOX 5, CAPTIV · Boston Scientific Corp · Jul 21, 2014
- K050578 — COOK IRELAND DUETTE MULTI-BAND MUCOSECTOMY DEVICE · Cook Ireland, Ltd. · Apr 22, 2005
- K220846 — CORE-SNARE · Incore Co., Ltd. · Nov 21, 2022
- K955650 — OLYMPUS SD SERIES SNARES · Olympus America, Inc. · Mar 29, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 7, 2021
Steris Corporation Carroll Martin Regulatory Affairs Director 5976 Heislev Road Mentor, OH 44060
Re: K212860
> Trade/Device Name: SmartBand EMR Kit (SB-EMR-K, SB-EMR-K-12, SB-EMR-K-9.4), SmartBand EMR Pack (SB-EMR-P, SB-EMR-P-12, SB-EMR-P-9.4), SmartSnare EMR Hexagonal Snare (SS-230-1 or packaged with kit) Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical unit and accessories Regulatory Class: Class II Product Code: FDI Dated: September 7, 2021 Received: September 8, 2021
Dear Carroll Martin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212860
Device Name
SmartBand® EMR Kit (SB-EMR-K, SB-EMR-K-12, SB-EMR-K-9.4) SmartBand® EMR Pack (SB-EMR-P, SB-EMR-P-12, SB-EMR-P-9.4) SmartSnare™ EMR Hexagonal Snare (SS-230-1 or packaged with Kit)
Indications for Use (Describe)
The SmartBand® EMR Device is intended for endoscopic mucosal resection in the upper GI tract.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the first page of a document with the identifier K212860 and page number 1 of 6. The document features the STERIS logo in a large, bold font. Below the company name is a graphic of several horizontal, curved lines, resembling waves.
# 510(k) Summary for the SmartBand EMR Kit (SB-EMR-K, SB-EMR-K-12, SB-EMR-K-9.4) SmartBand EMR Pack (SB-EMR-P, SB-EMR-P-12, SB-EMR-P-9.4) SmartSnare EMR Hexagonal Snare (SS-230-1 or packaged with Kit)
STERIS Corporation 5976 Heisley Road Mentor, OH 44060
Contact: Carroll Martin Regulatory Affairs Director Tel: 440-358-6259 Email: Carroll Martin@steris.com
Submission Date: September 7, 2021
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{4}------------------------------------------------
#### 1. Device Name
| Trade Name: | SmartBand EMR Kit (SB-EMR-K, SB-EMR-K-12, SB-EMR-K-9.4)<br>SmartBand EMR Pack (SB-EMR-P, SB-EMR-P-12, SB-EMR-P-9.4)<br>SmartSnare EMR Hexagonal Snare (SS-230-1 or packaged with Kit) |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Class: | Class II |
| Regulation Name: | Endoscopic Electrosurgical Unit and Accessories |
| Common/usual Name: | Ligation Device |
| Regulation Number: | 21 CFR 876.4300 |
| Product Code: | FDI |
| Review Panel: | Gastroenterology/Urology |
#### 2. Predicate Device
SmartBand EMR Kit (SB-EMR-K, SB-EMR-K-12) SmartBand EMR Pack (SB-EMR-P, SB-EMR-P-12) SmartSnare EMR Hexagonal Snare (SS-230-1 or packaged with Kit) All cleared under K190512
#### 3. Device Description
The SmartBand EMR Device is an endoscopic mucosal resection device that is provided as a kit or a pack that is designed for ligation assisted endoscopic mucosal resection (EMR) in the upper gastrointestinal (GI) tract. The SmartBand EMR Kit consists of the SmartBand Ligator Kit, supplied non-sterile and the SmartSnare EMR Hexagonal Snare, supplied sterile. The ligator kit contains a ligation handle, loading device, connector tube and a cylindrical barrel with 5 bands separated by a deployment cord. The SmartBand EMR Pack consists of a cylindrical barrel with 5 bands separated by a deployment cord. The SmartBand EMR Pack can be used when more bands are needed to complete a procedure. The SmartBand EMR Hexagonal Snare consists of a handle, finger slide and a hexagonal snare loop on the distal end.
#### 4. Indications for Use
The SmartBand® EMR Device is intended for endoscopic mucosal resection in the upper GI tract.
#### న్. Technological Characteristics Comparison Table
A comparison of technical characteristics between the proposed and predicate devices is summarized in Table 1.
{5}------------------------------------------------
| Features | SmartBand EMR Device<br>Predicate Device K190512 | Modified Device | Comparison |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The SmartBand® EMR Device is<br>intended for endoscopic mucosal<br>resection in the upper GI tract. | The SmartBand® EMR Device is<br>intended for endoscopic mucosal<br>resection in the upper GI tract. | Same |
| Device<br>Components | Ligation Device<br>5 non-latex bands (4 green and 1<br>blue)<br>Trigger cord<br>Loading Device<br>Flush Tube<br>Deployment Handle<br>Pentax Adaptor | Ligation Device<br>5 non-latex bands (4 green and 1<br>black)<br>Trigger cord<br>Loading Device<br>Flush Tube<br>Deployment Handle<br>Pentax Adaptor<br>Universal Connector<br>Band Lock | Similar. The color<br>change of one band<br>and the addition of<br>the Universal<br>Connector and<br>Band Lock do not<br>affect the intended<br>use or fundamental<br>technology of the<br>device. |
| | Snare<br>Stainless steel snare loop<br>Shaft<br>Finger Slide<br>Snare Handle Body | Snare<br>Stainless steel snare loop<br>Shaft<br>Finger Rings<br>Snare Handle Body | Same |
| Materials of<br>Construction | Snare<br>Snare loop - stainless steel<br>(patient contacting)<br>Sheath - Teflon (patient<br>contacting)<br>Core Wire - Nitinol (patient<br>contacting)<br>Stainless steel (snare loop) Strain<br>relief - polyurethane (non-patient<br>contacting)<br>Finger slide - ABS (non-patient<br>contacting)<br>Snare handle body - ABS (non-<br>patient contacting)<br>Ligation Device<br>Bands - Non-latex synthetic<br>polyisoprene (patient contacting)<br>Barrel - Polycarbonate (patient<br>contacting)<br>Trigger cord - Liquid crystal<br>polymer (patient contacting)<br>Loading device - Stainless steel<br>(non-patient contacting)<br>Flush tube - ABS (non-patient<br>contacting)<br>Handle - ABS (non-patient<br>contacting) | Snare<br>Snare loop - stainless steel<br>(patient contacting)<br>Sheath - Teflon (patient<br>contacting)<br>Core Wire - Nitinol (patient<br>contacting)<br>Stainless steel (snare loop) Strain<br>relief - polyurethane (non-<br>patient contacting)<br>Finger slide - ABS (non-patient<br>contacting)<br>Snare handle body - ABS (non-<br>patient contacting)<br>Ligation Device<br>Bands - Non-latex synthetic<br>polyisoprene (patient contacting)<br>Barrel - Polycarbonate (patient<br>contacting)<br>Trigger cord - Liquid crystal<br>polymer (patient contacting)<br>Loading device - Stainless steel<br>(non-patient contacting)<br>Flush tube - ABS (non-patient<br>contacting)<br>Handle - ABS (non-patient<br>contacting)<br>Universal Connector - Silicone<br>(non-patient contacting)<br>Band Lock - Polyurethane (non-<br>patient contacting) | Similar. The<br>materials of the<br>Universal<br>Connector and<br>Band Lock do not<br>affect the intended<br>use or fundamental<br>technology of the<br>device. |
| Features | SmartBand EMR Device<br>Predicate Device K190512 | Modified Device | Comparison |
| # of bands/color | 5 non-latex bands (4 green and 1<br>blue) | 5 non-latex bands (4 green and 1<br>black) | Similar. This<br>change aids the<br>user in determining<br>when they are<br>getting close to<br>using all of the<br>bands, as the black<br>band is the second<br>to last band in line<br>on the barrel. The<br>change in color<br>does not affect the<br>intended use or<br>fundamental<br>technology of the<br>device. |
| Barrel<br>Dimensions | Inside diameter:<br>10.75mm and 12.0mm<br>Depth: 7.75mm | Inside diameter<br>9.4mm, 10.75mm, and12.0mm<br>Depth 7.75mm (10.75mm and<br>12.0mm sizes)<br>Depth 9.5mm (9.4mm size) | Similar. This<br>change gives the<br>user more options<br>based on the size<br>of the treatment<br>area. |
| Sterile/Non-<br>sterile | Sterile (snare) and Non-sterile<br>(ligation device) | Sterile (snare) and Non-sterile<br>(ligation device) | Same |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Same |
| Sterilization<br>Assurance Level | 10-6 | 10-6 | Same |
| Usage | Single used; snare is used with<br>diathermic energy | Single used; snare is used with<br>diathermic energy | Same |
| Target<br>Population | Patients undergoing an endoscopic<br>mucosal resection procedure | Patients undergoing an<br>endoscopic mucosal resection<br>procedure | Same |
| Energy<br>Used/Delivered | Diathermic energy (snare) | Diathermic energy (snare) | Same |
| Method of<br>Application | Manual | Manual | Same |
| Features | SmartBand EMR Device<br>Predicate Device K190512 | Modified Device | Comparison |
| Packaging<br>Change | No band lock | Band lock present to secure<br>the bands during shipping | Different. The band lock<br>prevents the bands from<br>dislodging from the barrel<br>during shipping. The<br>additional box is a<br>convenience for the user.<br>These changes have no effect<br>on how the device is used or<br>its fundamental technology. |
| | No stand-alone box for the<br>snare | Addition of stand-alone box<br>for snare | |
| Compatible<br>endoscopes | Pentax, Fujinon and Olympus | Pentax, Fujinon and Olympus | Same<br>Similar. The function of the<br>electrical snare of the<br>proposed device remains the<br>same in that it is attached to<br>the electrical surgical<br>generator to allow the flow of<br>electrical energy to the snare<br>loop for the purpose of<br>resecting tissue. The use of<br>the modified device with the<br>US Endoscopy gi4000<br>Electrosurgical Generator<br>does not change the intended<br>use or fundamental<br>technology of the modified<br>device. |
| Compatible<br>electrical surgical<br>generators | ERBE VIO 300-D | ERBE VIO 300-D<br>US Endoscopy gi4000<br>Electrosurgical Generator | |
| Addition of<br>universal<br>connector | Not present. | Present | Different. The universal<br>connector allows the user to<br>attach the ligation device<br>handle directly to the<br>accessory ports of endoscopes<br>that do not have a luer<br>connection on the accessory<br>port. This change has no<br>effect on how the device is<br>used or its fundamental<br>technology. |
| Expiration<br>Dating | 12 months | 24 months | Different. The extended shelf<br>life of the proposed device<br>provides the user greater<br>flexibility related to device<br>use. It has no effect on how<br>the device is used or its<br>fundamental technology. |
Table 1. Technological Characteristics Comparison Table
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### 6. Summarv of Non-Clinical Performance Testing
Verification testing consisting of measurements, visual inspection, ship testing and packaging testing.
Biocompatibility testing in accordance with ISO 10993-4, -5, -10 and -11 and USP38-NF33.
Simulated use testing.
#### 7. Conclusion
The minor changes that are the subject of this submission have no impact on the fundamental scientific technology of the device or how it is used. STERIS has determined that the modified device is as safe and effective as the legally marketed predicate device.