Surgical Gown
K212812 · Xianning Marveldas Protective Articles Co., Ltd. · FYA · Nov 16, 2022 · General, Plastic Surgery
Device Facts
| Record ID | K212812 |
| Device Name | Surgical Gown |
| Applicant | Xianning Marveldas Protective Articles Co., Ltd. |
| Product Code | FYA · General, Plastic Surgery |
| Decision Date | Nov 16, 2022 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4040 |
| Device Class | Class 2 |
Intended Use
Surgical gowns are sterile, disposable, and single use devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. This device is sterilized using Ethylene Oxide (EO) following the Validation and routine control under ANSI/AAMI/ISO 11135 (2014).
Device Story
Surgical gown; sterile, disposable, single-use; worn by operating room personnel. Constructed of SMS nonwoven material with film covering, Velcro tape, and cuff material. Provides barrier protection against microorganisms, body fluids, and particulates. Available in sizes S, M, L, XL, XXL. EO sterilized per ANSI/AAMI/ISO 11135. Used in clinical settings to reduce cross-contamination risk between patient and staff.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via AATCC 127 (hydrostatic pressure >50 cm), AATCC 42 (impact penetration ≤1g), ASTM D5034 (tensile strength), ASTM D5733 (tearing strength), ASTM D1683 (seam strength), 16 CFR Part 1610 (flammability Class 1), and ASTM F1671 (bacteriophage penetration). Biocompatibility confirmed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).
Technological Characteristics
SMS nonwoven material with film covering, Velcro tape, and cuff material. Reinforced design. EO sterilization. Sizes S-XXL. Standards: AATCC 127, AATCC 42, ASTM D5034, ASTM D5733, ASTM D1683, 16 CFR 1610, ASTM F1671, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for operating room personnel during surgical procedures to protect patient and personnel from transfer of microorganisms, body fluids, and particulate material.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Related Devices
- K120192 — DISPOSABLE SURGICAL GOWNS · Weihai Hongyu Nonwoven Fabric Products Co., Ltd. · Oct 16, 2012
- K211809 — Surgical Gown · Wuhan Dymex Healthcare Co., Ltd. · Oct 18, 2021
- K223401 — Disposable Surgical Gowns (S,M,L,XL,XXL,XXXL) · Xiamen Probtain Medical Techology Co., Ltd. · Mar 7, 2023
- K252301 — Henan Yadu Level 4 Surgical Gowns (Reinforced) · Yadu Medical (Henan) Co., Ltd. · Feb 4, 2026
- K141467 — SURGICAL GOWN · Xuchang Zhengde Environstar Medical Products Co.,L · Mar 5, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 16, 2022
Xianning Marveldas Protective Articles CO.,Ltd. % Salon Chen OS Engineer IMD Medical & Drug Technology Service Institutions Room 308, building 11, No. 23, Jinqu Road, Wanjiang Street Dongguan, Guangdong 523069 China
Re: K212812
Trade/Device Name: Surgical Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYA Dated: October 12, 2022 Received: October 12, 2022
Dear Salon Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Bifeng Qian -S
Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212812
Device Name Surgical Gown
Indications for Use (Describe)
Surgical gowns are sterile, disposable, and single use devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. This device is sterilized using Ethylene Oxide (EO) following the Validation and routine control under ANSI/AAMI/ISO 11135 (2014).
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div></div> |
|----------------------------------------------|------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 20px;"> ☑ </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary (K212812)
#### 1. Submitter Information
- > Company Name: Xianning Marveldas Protective Articles CO., Ltd.
- Address: Guishan Road #30, Hi-New Tech Industrial, Xianning, Hubei P R C >
- > Phone: +86-1572066625
- > Contact Person (Title): Zhuxiaokang
- > E-mail: sale@marveldas.com
- A Date of Preparation: November 10, 2022
#### 2. Subject Device Information
- > Type of 510(k) submission: Traditional
- Trade Name: Surgical Gown A
- > Common Name: Surgical Gown
- > Classification Name: Surgical Apparel
- > Model: S、M、L、XL、XXL
- > Classification Product Code: FYA
- > Requlation Number: 21 CFR 878.4040
- > Classification: Class II
- > Review Panel: General Hospital
#### 3. Predicate Device Information:
- > 510(k) Number: K121152
- > Predicate Device Name: Surgical Gown
- > Manufacturer: Jiangsu Guangda Medical Material Co., Ltd.
- > This predicate has not been subject to a design-related recall
- > No reference devices were used in this submission.
### 4. Application Correspondent
- A Company Name: IMD Medical & Drug technology service institutions
- > Phone: +86-18613190779
- > Fax: +86-755-62809168
- > Contact Person(Title): Salon Chen (System engineer)
- A E-mail: 33999439@qq.com
- > the address of application correspondent:308, Building 11, Zhongchuang Foreign Exchange Industrial Park, No. 23, Jinqu Road, Wanjiang Street, Dongguan, Guangdong, China
#### 5. Device Description
The Sterile Level 4 Surgical Gown is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. The proposed devices are single
{4}------------------------------------------------
use and EO sterilized. The Sterile Level 4 Surgical Gowns are available in five models including S、 M、L、XL and XXL.
The proposed surgical are constructed of a SMS nonwoven material (SMS+Film covering material+Velcro tape+Cuff material) and it is a kind of reinforced surgical gown and blue color.
#### 6. Intended Use /Indications for Use
Surgical gowns are sterile, disposable, and single use devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. This device is sterilized using Ethylene Oxide (EO) following the Validation and routine control under ANSI/AAMI/ISO 11135 (2014).
#### 7. Comparison of technological characteristics with the Predicate
| Elements of Comparison | Predicate Device | Subject Device | Comparison |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Company Name | Jiangsu Guangda Medical<br>Material Co., Ltd. | Xianning Marveldas Protective<br>Articles CO., Ltd. | / |
| 510(k) number | K121152 | K212812 | / |
| Device Name | Surgical Gown | Surgical Gown | / |
| Classification Product Code | FYA | FYA | same |
| Regulation | 21 CFR 878.4040 | 21 CFR 878.4040 | same |
| Classification Name | Surgical Apparel | Surgical Apparel | same |
| Class | 2 | 2 | same |
| Prescription or OTC | OTC | OTC | same |
| | | | |
| | Jiangsu Guangda's<br>Reinforced Surgical Gowns,<br>Model Number GD-SG-01,<br>are non-sterile, single use<br>surgical gowns intended to<br>protect surgical patients and<br>operating room personnel<br>from the transfer of<br>microorganisms, body fluids,<br>and particulate material.<br>This product may be<br>sterilized using Ethylene<br>Oxide (EO) following the<br>validation and routine control<br>under ANSI/AAMI/ISO | Surgical gowns are sterile,<br>disposable, and single<br>use devices that are intended<br>to be worn by operating roomprocedures to protect both the<br>surgical patient and the<br>operating room personnel from<br>transfer of microorganisms,<br>body fluids, and particulate<br>material. This device is<br>sterilized using Ethylene Oxide<br>(EO) following the Validation<br>and routine control under<br>ANSI/AAMI/ISO 11135 (2014). | same |
| Intended Use & Indications<br>for Use | | | |
| Materials | SMS and PE+PP TWO layer<br>compound | SMS+Film covering<br>material+Velcro tape+Cuff | similar |
| Weight Per square(g) | 45g/m² | 45g/m² | same |
| Durability | Disposable | Disposable | same |
| Size | XL | S、M、L、XL、XXL | different |
| Color | Blue | Blue | same |
| Reinforced area | PE+PP two layer compound<br>protective reinforcement<br>with 100% white Terylene | double barrier non-woven fabric | same |
| Style | Reinforced | Reinforced | similar |
| Hydrostatic pressure:<br>AATCC 127 | >20cm | >50 cm | different |
| Impact penetration:<br>AATCC 42 | ≤1 | ≤1 | same |
| Tensile strength: ASTM<br>D5034 | Passed | Length(lbf):37.6<br>Width(lbf):24.2 | similar |
| Tearing strength: ASTM<br>D5733 | Passed | Length yarns torn(lbf): 14.3<br>Width yarns torn(lbf): 8.1 | similar |
| Seam strength: ASTM D1683 | Pass | Arm Opening sleeves<br>23.5 lbf (FTS)<br>24.2lbf (FR)<br>Shoulder<br>37.0 lbf (FR) | similar |
| Flammability:<br>16CFR Part 1610 | Class 1 | Class 1 | same |
| Sterilization method | Although sold non-sterile,<br>gown can be EO Sterilized | EO | same |
| Resistance to bacteriophage<br>penetration ASTM F1671 | Level 4 per AAMI PB70 | Level 4 per AAMI PB70 | same |
| Biocompatibility | Cytotoxicity, ISO<br>10993-5:2009 | Passed | ISO 10993-5; Under the<br>conditions of the study, the<br>proposed device extract was<br>determined to be non-cytotoxic |
| | Irritation, ISO<br>10993-10:2002 | Passed | ISO 10993-10; Under the<br>conditions of the study, the<br>proposed device extract was<br>determined to be non-irritating |
| | Sensitization, ISO<br>10993-10:2002 | Passed | ISO 10993-10; Under the<br>conditions of the study, the<br>proposed device extract was<br>determined to be non-<br>sensitizing |
| Test Method | Purpose | Acceptance<br>Criteria | Result |
| AATCC 127 | Hydrostatic pressure | >50 cm | Pass |
| AATCC 42 | Impact penetration | ≤1g | Pass |
| ASTM D 5034 | Tensile Strength | Length(lbf): ≥37.6<br>Width(lbf): ≥24.2 | Pass |
| ASTM D 5733 | Tearing strength | Length yarns<br>torn(lbf):14.3<br>Width yarns<br>torn(lbf):8.1 | Pass |
| ASTM D 1683 | Seam strength | Arm Opening<br>sleeves<br>23.5 lbf (FTS)<br>24.2lbf (FR)<br>Shoulder<br>37.0 lbf (FR) | Pass |
| 16 CFR Part 1610 | Flammability of Textiles | Class 1 | Class 1 |
| ASTM F1671 | Bacteriophage Phi-X174<br>penetration | No penetration | Pass |
| Biocompatibility | Cytotoxicity, ISO<br>10993-5:2009 | ISO 10993-5 | ISO 10993-5; Under<br>the conditions of the<br>study, the proposed<br>device extract was<br>determined to be<br>non-cytotoxic |
| | Irritation, ISO 10993-10:2002 | ISO 10993-10 | ISO 10993-10;<br>Under the conditions<br>of the study, the<br>proposed device<br>extract was<br>determined to be<br>non-irritating |
| | Sensitization, ISO<br>10993-10:2002 | SO 10993-10 | ISO 10993-10;<br>Under the conditions<br>of the study, the<br>proposed device<br>extract was<br>determined to be<br>non-sensitizing |
{5}------------------------------------------------
{6}------------------------------------------------
#### Analysis:
1) In term of key material composition, the surgical gowns are both constructed of a SMS nonwoven material. Other Secondary material is slightly different from those of the predicate device. The proposed device has been passed on the performance and biocompatibility. It is no effect on safety or efficacy.
2) Although the size is different from those of the predicate device. The proposed device has been passed on the performance testing. It is no effect on safety or efficacy.
3) In terms of general intended used, performance testing, material composition and configuration, hydrostatic pressure、tensile strength、tearing strength and seam strength are slightly different from those of the predicate device. The proposed device has been tested according to AATCC 127、ASTM D 5034、ASTM D 5733 and ASTM D 1683 respectively, and met the requirement of the standard.
#### 8. Discussion of Non-Clinical Performance Tests Performed
{7}------------------------------------------------
#### Sterile Level 4 Surgical Gown
Non clinical tests were conducted to verify that the proposed device met all design specifications as was similar to the predicate device. The test results demonstrated that the proposed subject device complies with the following standards:
{8}------------------------------------------------
# 9. Summary of Clinical Test
No clinical study is included in this submission.
## 10.Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(k), Surgical gowns is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K121152.