Laser Treatment System, Model: Hera, Armo

K212793 · Canadian Pioneer Medical Technology Corporation · GEX · Oct 25, 2021 · General, Plastic Surgery

Device Facts

Record IDK212793
Device NameLaser Treatment System, Model: Hera, Armo
ApplicantCanadian Pioneer Medical Technology Corporation
Product CodeGEX · General, Plastic Surgery
Decision DateOct 25, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Laser Treatment System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Device Story

Laser Treatment System (Hera/Armo models) uses 808 nm diode laser for hair removal via selective photothermolysis; melanin absorbs laser energy, causing rapid temperature increase to destroy hair follicles. System comprises main unit, handpiece(s), power supply, cooling system, and control system. Hera model features one handpiece connector; Armo model features two. Handpieces available in three spot sizes (12x12mm, 12x20mm, 15x27mm). Used in dermatological practice by clinicians. Microprocessor controls electronics; self-contained water cooling system manages thermal output. Healthcare providers use device to achieve long-term, stable hair reduction, benefiting patients through permanent hair removal across all skin types.

Clinical Evidence

Bench testing only. No clinical data provided. Performance supported by biocompatibility testing (ISO 10993-1), electrical safety (ANSI/AAMI ES60601-1), EMC (IEC 60601-1-2), and laser safety standards (IEC 60601-2-22, IEC 60825-1).

Technological Characteristics

Diode laser (808 nm); Class IV. Materials: Sapphire (handpiece), stainless steel (tip). Cooling: Self-contained water system. Power: 100-240V AC, 50/60Hz. Spot sizes: 12x12mm, 12x20mm, 15x27mm. Fluence: 1-120 J/cm². Pulse duration: 10-300ms. Frequency: 1-10Hz. Standards: ISO 10993-1, IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-2.

Indications for Use

Indicated for hair removal and permanent hair reduction in patients of all skin types (Fitzpatrick I-VI), including tanned skin.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font. October 25, 2021 Canadian Pioneer Medical Technology Corporation Rashid Reza Mir Sayah Managing Director unit 2, 210 Drumlin Circle, Concord Vaughan, Ontario L4k 3E3 Canada Re: K212793 Trade/Device Name: Laser Treatment System, Model: Hera, Armo Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: August 26, 2021 Received: September 1, 2021 Dear Rashid Reza Mir Sayah: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K212793 Device Name Laser Treatment System, Model: Hera, Armo Indications for Use (Describe) The Laser Treatment System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |-----------------------------------------------------------------------------------------------------------------------------------| | <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Section 5 510(k) summary #### I Submitter Canadian Pioneer Medical Technology Corporation Unit 2-210 Drumlin Circle, Vaughan, Ontario, L4K 3E3, Canada Contact person: Dr. Rashid Reza Mir Sayah Managing Director Phone: 4377727788 Email: Canadianpioneer@yahoo.com Canadianpioneermedical@gmail.com Date of preparation: October 7, 2021 #### II Subject Device Trade Name of Device: Laser Treatment System, Model: Hera, Armo Common name: Powered Laser Surgical Instrument Classification Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulation Number: 21 CFR 878.4810 Regulatory Class: II Product code: GEX Review Panel: General & Plastic Surgery #### III Predicate Devices Trade Name of Devices: Laser Treatment System, Model: NYX and Grace Common name: Powered Laser Surgical Instrument Regulation Number: 21 CFR 878.4810 Requlatory Class: II Product code: GEX Review Panel: General & Plastic Surgery 510(k) number: K210033 Manufacturer: Canadian Pioneer Medical Technology Corporation #### IV Device description The Laser Treatment System consists of the main unit and a hand piece. The system uses a diode laser as an active medium placed in an optical cavity to produce amplified beam at the {4}------------------------------------------------ wavelength of 808 nm. A microprocessor is used to control electronics for the front panel. A self-contained water cooling system is built into the power supply unit. The Diode Laser Therapy Devices is designed to be used in dermatological practice for stable, long term hair reduction. The principle of laser hair removal is selective photothermolysis. The wavelength of 808nm would be able to effectively penetrate deep into and absorbed by the target chromophore. The laser power is delivered to the treatment region via a delivery system. The proposed device includes power supply system, delivery system, control system, cooling system, laser system. The models object of this submission are Armo and Hera: Hera has one handpiece connector. Armo has two handpiece connectors. The 808 nm handpieces with different spot sizes (3 spot sizes : 12×12mm,12×20mm,15×27m) are available for both models. #### V Indications for use The Laser Treatment System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. #### VI Comparison of technological characteristics with the predicate devices The indication of proposed device is covered by the predicated devices. The proposed device is only intended to use for hair removal. The device includes three models of handpieces for clearance in this submission. The differences between models are their spot sizes and Density .These are covered by the predicated devices. | Device feature | Laser Treatment System<br>(subject device) | Laser treatment system<br>Nyx and Grace<br>(K210033) | |--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product code | GEX | GEX | | Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | | Indications for use | The Diode Laser Treatment System is intended for hair | The Diode Laser Treatment | | | removal, permanent hair<br>reduction on all skin types<br>(Fitzpatrick skin type I-VI),<br>including tanned skin.<br>Permanent hair reduction is<br>defined as the long-term,<br>stable reduction in the<br>number of hairs regrowing<br>when measured at 6, 9, and<br>12 months after the<br>completion of a treatment<br>regime. | System is<br>intended for hair<br>removal,<br>permanent hair<br>reduction on all<br>skin types<br>(Fitzpatrick skin<br>type I-VI),<br>including tanned<br>skin. Permanent<br>hair reduction is<br>defined as the<br>long-term, stable<br>reduction in the<br>number of hairs<br>regrowing when<br>measured at 6, 9,<br>and 12 months<br>after the<br>completion of a<br>treatment<br>regime. | | Operation<br>principle | Melanin could absorb the<br>energy from the laser, which<br>would result in temperature<br>rapid increase, to destroy<br>surrounding hair follicles, and<br>finally remove hair. | Melanin could<br>absorb the<br>energy from the<br>laser, which<br>would result in<br>temperature<br>rapid increase, to<br>destroy<br>surrounding hair<br>follicles, and<br>finally remove<br>hair. | | Laser<br>classification | Class IV | Class IV | | Light delivery<br>system | 808 nm | 808nm | | Spot Size | 12×12mm<br>12x20mm | 12x12mm | | | 15×27mm | | | Fluence | 1~120J/cm² at spot size 12×12mm<br>1~70 J/cm² at spot size12×20mm<br>1~40 J/cm² at spot size15×27mm | 1~120 J/cm² | | Frequency | 1-10Hz | 1-10 Hz | | Pulse Duration | 10-300ms | 10-300 ms | | Power supply | 100-240V AC, 50/60Hz | 100-240V AC, 50/60Hz | | Patient contact material | Sapphire in handpiece and handpiece tip (stainless steel) | Sapphire in handpiece and handpiece tip (stainless steel) | | Biocompatibility | Comply with ISO10993-1 | Comply with ISO10993-1 | | Electrical Safety | Comply with IEC60601-1, IEC60601-2-22 | Comply with IEC60601-1, IEC60601-2-22 | | EMC | Comply with IEC60601-1-2, | Comply with IEC60601-1-2, | | Laser safety | Comply with IEC60825-1, IEC60601-2-22 | Comply with IEC60825-1, IEC60601-2-22 | {5}------------------------------------------------ K212793 {6}------------------------------------------------ {7}------------------------------------------------ #### VII Performance data The following performance data were provided in support of the substantial equivalence determination. ### Biocompatibility testing Biocompatibility of the Laser Treatment System was evaluated in accordance with ISO 10993-1:2009 for the body contact category of "Surface –intact skin" with a contact duration of "Limited (< 24 hours)". The following tests were performed, as recommended: Cytotoxicity, Irritation and Sensitization. All evaluation acceptance criteria were met ## Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the Laser Treatment System. The system has been tested to comply with the following standards: - ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and -A2:2010/(R)2012" Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance; - IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests. ### Bench Testing - IEC 60601-2-22:2007 + A1:2012, Medical Electrical Equipment Part 2-22: Particular -Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment; - IEC 60825-1: 2007, Safety of laser products Part 1: Equipment classification and requirements. #### Software Verification and Validation Testing Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". #### VIII Conclusion The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...