K212771 · Led Intellectual Properties, LLC · OHS · Jan 26, 2022 · General, Plastic Surgery
Device Facts
Record ID
K212771
Device Name
LightStim Elipsa
Applicant
Led Intellectual Properties, LLC
Product Code
OHS · General, Plastic Surgery
Decision Date
Jan 26, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LightStim Elipsa is a LED light therapy device which uses specific wavelengths of light emitting diodes (LEDs). It is intended to emit energy in the red and infrared region of the light spectrum to provide treatment for full face wrinkles. The blue and red light spectrum is intentent of mild to moderate inflammatory acne.
Device Story
LightStim Elipsa is an over-the-counter LED light therapy device for home use. It utilizes four LED wavelengths (612nm, 645nm, 655nm, 850nm) for wrinkle treatment and two wavelengths (410nm, 645nm) for acne treatment. The device is operated by the patient; no user settings or adjustments are required. The system includes the LED module, power supply, goggles, and travel case. By emitting specific light energy, the device targets dermatological conditions; patients apply the device to the skin for prescribed treatment durations. The output is intended to reduce wrinkles and treat inflammatory acne, providing a non-invasive therapeutic benefit.
Clinical Evidence
Bench testing only. The device underwent electrical safety testing (AAMI/ANSI ES60601-1), electromagnetic compatibility testing (IEC 60601-1-2), and performance testing per IEC 60601-1-6, IEC 60601-1-11, and IEC 60601-2-83. Usability and label comprehension studies were conducted to evaluate self-selection, training effectiveness, and device operation by intended users, confirming that instructions were effective for primary operating functions.
Technological Characteristics
LED-based light therapy device. Emits red/infrared (612nm, 645nm, 655nm, 850nm) and blue/red (410nm, 645nm) light. Powered by external supply. Tested to AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, and IEC 60601-2-83. No user-serviceable components. Over-the-counter use.
Indications for Use
Indicated for individuals with full-face wrinkles and individuals (men and women) with mild to moderate inflammatory acne.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 26, 2022
LED Intellectual Properties LLC Chase Marchese VP of Manufacturing 16552 Von Karman Ave Irvine. California 92606
Re: K212771
Trade/Device Name: LightStim Elipsa Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP, OHS Dated: December 1, 2021 Received: December 1, 2021
Dear Chase Marchese:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212771
Device Name LightStim Elipsa
#### Indications for Use (Describe)
The LightStim Elipsa is a LED light therapy device which uses specific wavelengths of light emitting diodes (LEDs). It is intended to emit energy in the red and infrared region of the light spectrum to provide treatment for full face wrinkles. The blue and red light spectrum is intentent of mild to moderate inflammatory acne.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K212771
510(k) Summary LED Intellectual Properties, LLC. Device: LightStim Elipsa K212771
## 1. General Information
Submitter: LED Intellectual Properties, LLC 16552 Von Karman Ave. Irvine, Ca. 92606
Contact Person: Steve Marchese & Chase Marchese Phone: Chase (714) 924-0492 Date Prepared: 2022/01/19 Emails: Chase@lightstim.com & Steve@lightstim.com
## 2. Device name and code
Device Proprietary Name: LightStim Elipsa
Product Code and name:
OHS - Light Based Over-the-Counter Wrinkle Reduction OLP - Over-the-Counter Powered Light Based Laser For Acne
Classification Name: Laser Surqical Instrument For Use in General And Plastic Surgery And in Dermatology. 21 CFR 878.4810
Common or Usual Name: Light-Based over-the-counter wrinkle reduction and acne treatment.
Requlatory Class: II
## 3. Predicate Devices
SHENZHEN KAIYAN MEDICAL CO LTD, DemarkQ (K203214) Medtek Skincare, LLC, Poly Clear (K183708) LED Intellectual Properties, LLC, LightStim Professional 2-Panel Light (K150098)
## 4. Device Description
The LightStim Elipsa is an over-the-counter light emitting diode (LED) device that emits energy for use in dermatology for the treatment of acne and wrinkles. The device uses four types of LEDs for Wrinkles: 612nm amber, 645nm red, 655nm red, and 850nm infrared and two types of LEDs for Acne: 410nm and 645nm. The treatment time is controlled by the user. There are no user settings or adjustments required.
{4}------------------------------------------------
#### 510(k) Summary LED Intellectual Properties, LLC. Device: LightStim Elipsa K212771
The LightStim Elipsa system components include the device containing the LED module, power supply, goggles, and travel case.
The LightStim Elipsa does not contain any user serviceable components. The device is sold as Over-the-counter (OTC)
#### 5. Indications for Use
The LightStim Elipsa is a LED light therapy device which uses specific wavelengths of light, produced by light emitting diodes (LEDs. It is intended to emit energy in the red and infrared region of the light spectrum to provide treatment for full face wrinkles. The blue and red light spectrum is intended for the treatment of mild to moderate inflammatory acne.
## 6. Comparison of Technological Characteristics
| Company | SHENZHEN KAIYAN MEDICAL<br>CO LTD | Medtek Skincare, LLC | LED Intellectual Properties, LLC | LED Intellectual Properties, LLC |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | DemarkQ | Poly Clear | LightStim Professional 2-Panel Light | LightStim Elipsa |
| 510(k) Number | K203214 | K183708 | K150098 | Proposed Device |
| Product Code | OLP | OLP | OHS | OLP & OHS |
| FDA Device Classification | Class II | Class II | Class II | Class II |
| FDA Clearance | Over the Counter | Over the Counter | Over the Counter | Over the Counter |
| Indications for use | The DemarkQ is an Over-the-<br>Counter (OTC) device intended<br>for treatment of mild to<br>moderate inflammatory acne. | The Poly Clear combination of<br>Red (633nm±10nm) and Blue<br>(417nm±10nm) is intended to<br>emit energy in the red, blue<br>regions of the light spectrum to<br>treat dermatological conditions,<br>specifically indicated to treat<br>mild to moderate acne vulgaris. | Two (2) interchangeable 2-Panel LED Systems,<br>each mounted on a hands-free, fully articulating<br>arm:<br>System #1 = intended for the use in the<br>treatment of full-face wrinkles.<br>System #2 = Not applicable | The LightStim Elipsa is a LED light<br>therapy device which uses specific<br>wavelengths of light, produced by light<br>emitting diodes (LEDs). It is intended to<br>emit energy in the red and infrared<br>region of the light spectrum to provide<br>treatment for full face wrinkles. The blue<br>and red light spectrum is intended for<br>the treatment of mild to moderate<br>inflammatory acne. |
| Target Population | Women and Men with mild to<br>moderate acne. | Women and Men with mild to<br>moderate acne. | System #1 = People with full-face wrinkles.<br>System #2 =Not applicable | People with Full-face wrinkles. Women<br>and Men with mild to moderate<br>inflammatory acne. |
| Wavelengths (nm) | 630nm +/-10nm<br>415nm -/-10nm | 633nm +/-10nm<br>417nm -/-10nm | System #1 = 605nm, 630nm, 660nm, 855nm,<br>System #2 = Not applicable | RED: 612nm, 645nm, 655nm, 850nm<br>BLUE: 410nm & 645nm |
| Output in milliwatts | DemarkQ WOW:<br>Red: 5<br>Blue: 25<br>DemarkQ POP:<br>Red: 25<br>Blue: 25 +/- 5 | Combo Red/Blue head:<br>Red 30mW/cm^2 +/-10nm<br>Blue 20mW/cm^2 +/-10nm | System #1 = 65mW cm2<br>System #2 = Not applicable | RED: 11.8 mW/cm2<br>BLUE: 12.8 mW/cm2 |
| Treatment Time | 3 minutes per treatment | Blue: 13minutes.<br>Red: 15 minutes | System #1 = 3min<br>System #2 = Not applicable | RED: 16 minutes per area<br>BLUE: 12 minutes per area |
| Dose | RED: 0.9-4.5 J/cm2<br>Blue: 4.5 J/cm2 | Red: 52 J/cm2<br>Blue 26 J/cm2 | System #1 = 11.7 j/cm^2<br>System #2 = Not applicable | RED: 11.3 j/cm^2<br>BLUE: 9.2 j/cm^2 |
| Total dose/Treatment | 9 J/cm2 | 52 J/cm2 | 11.7 J/cm2 | Red 11.3 J/cm2<br>Blue 9.2 J/cm2 |
| Total dose/week | 63 J/cm2 | 78 J/cm2 | 58.5 J/cm2 | Red 56.5 J/cm2<br>Blue 46 J/cm2 |
{5}------------------------------------------------
### 7. Performance Testing
The LightStim Elipsa device results in patient contact with an electrically powered component: therefore, it was tested for conformance to AAMI/ANSI ES60601-1:2005(R) 2012 and A1:2012 4th Edition: Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance
The LightStim Elipsa device includes an electronic component, and it was tested for conformance to IEC 60601-1-2:2014 4th Edition Medical Electrical Equipment – Part 1-2: General Requirements for Safety and Essential Performance- Collateral standard: Electromagnetic Compatibility-Requirements and tests.
The LightStim Elipsa device was also tested for conformance to IEC 60601-1-6:2010. IEC 60601-1-11:2015, IEC 60601-2-83:2019 including gap analysis demonstrating conformance to IEC 60601-2-57:2011.
Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software in Medical Devices. The minor (A) level of concern was determined because malfunction of, or a latent design flaw in the software could not lead to a patient injurv
### 8. USABILITY TESTING:
The sponsor conducted a Usability/Label Comprehension study to acquire data in order to evaluate and measure labeling comprehension, device usability, self-selection component, and training effectiveness by "intended users" of the Elipsa device.
The results of this testing on the final version of the labeling and instruction plan showed that the instructions were effective and intended users were able to understand the labeling and apply this information to device use for the primary operating functions of the device.
#### 9. Conclusions
After analysis of safety, indications, intended uses, dose rates, performance, features, design materials, power output, technological properties, treatment areas, treatment reqimens and methods of operation, the manufacturer asserts that no significant differences exist between the subject device and predicate, and no different questions of safety and effectiveness arise. Therefore, the subject device is as safe, as effective, and performs as well as the referenced predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.