K212758 · Cedars-Sinai Medical Center: Aim · LLZ · May 19, 2023 · Radiology
Device Facts
Record ID
K212758
Device Name
Autoplaque
Applicant
Cedars-Sinai Medical Center: Aim
Product Code
LLZ · Radiology
Decision Date
May 19, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K212758 · May 19, 2023
Autoplaque
Cedars-Sinai Medical Center: Aim
Clinically indicated coronary computed tomography angiography (CCTA) images from multiple U.S. and non-U.S. sites
Retrospective clinical CCTA images were used to validate the device's performance by comparing automated measurements against expert reader ground-truth annotations.
201 patients (53% men, 47% women; mean age 60 ± 12 years) with clinically indicated CCTA; Sample Size: 201 patients; 781 lesions; Number of Sites: Multiple U.S. sites
Total plaque volume, calcified plaque volume, non-calcified plaque volume, and diameter stenosis
Retrospective validation study
175 patients (53% men, 47% women; mean age 56 ± 9 years) with clinically indicated CCTA; Sample Size: 175 patients; 1081 lesions; Number of Sites: Multiple non-U.S. external sites
Two cardiologists and one radiologist (adjudication panel)
Total plaque volume, calcified plaque volume, non-calcified plaque volume, and diameter stenosis
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Coronary plaque and stenosis analysis
Deep learning-based vessel, plaque and lumen segmentation
—
—
—
—
Comparison with Expert Reader Measurements in a Multicenter US Patient Population: 201 patients (781 lesions). Comparison with Expert Reader Measurements in a Non-US Patient Population: 175 patients (1081 lesions). Comparative Study with the Predicate Device: 27 patients (30 lesions).
>1 (cardiologists) + 1 (radiologist)
Indications for Use
Autoplaque is intended to provide an optimized non-invasive application to analyze coronary anatomy and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images. Autoplaque is a workstation-based post processing application. It is a non-invasive diagnostic reading software intended for use by cardiologists and radiologists as an interactive tool for viewing and analyzing cardiac CT data for determining the presence and extent of coronary plaques and luminal stenoses. The software is not intended to replace the skill and judgment of a qualified medical practitioner and should only be used by people who have been appropriately trained in the software's functions, capabilities and limitations. Users should be aware that certain views make use of interpolated data. This is data that is created by the software based on the original data set. Interpolated data may give the appearance of healthy tissue in situations where pathology that is near or smaller than the scanning resolution may be present. Autoplaque must be installed on a suitable commercial computer platform. It is the user's responsibility to ensure the monitor quality and ambient light conditions are consistent with the clinical applications. Typical users of Autoplaque are trained medical professionals, including but not limited to radiologists, clinicians, technologists, and others.
Device Story
Autoplaque 3.0 is a standalone workstation-based software for post-processing coronary CT angiography (CTA) images. It ingests DICOM image data from local storage. The device utilizes deep learning-based algorithms to perform automated segmentation of coronary vessels, plaque, and lumen. It provides clinicians with visualization tools including multiplanar reformatting (MPR), curved MPR, straightened MPR, and 3D views. The software calculates quantitative metrics such as vessel/lumen diameter and area, plaque volume (calcified, non-calcified, low-density), plaque burden, and stenosis severity. Clinicians review and may manually edit the automated segmentations. The output aids cardiologists and radiologists in assessing coronary pathology and determining treatment paths. It is intended for use by trained medical professionals in clinical settings. The device offers faster analysis times compared to manual expert review and the predicate device, potentially improving clinical workflow efficiency.
Clinical Evidence
Clinical validation included three studies. 1) US multicenter study (n=201 patients, 781 lesions) showed excellent agreement between Autoplaque 3.0 and expert readers for plaque volumes and diameter stenosis. 2) Non-US study (n=175 patients, 1081 lesions) confirmed excellent agreement with expert readers and demonstrated a mean analysis time of 5.7 seconds per patient. 3) Comparative study (n=27 patients, 30 lesions) showed excellent correlation and intraclass correlation agreement between Autoplaque 3.0 and the predicate device, with Autoplaque 3.0 performing significantly faster (2 seconds/lesion vs 30 seconds/lesion).
Technological Characteristics
Standalone software for Windows and Mac OS. Uses deep learning-based segmentation for vessels, plaque, and lumen. Supports DICOM 3.1. Features include 2D/3D visualization, MPR, curved/straightened MPR, and quantitative analysis of plaque/lumen parameters. Software is user-editable. No specific hardware materials; runs on commercial computer platforms.
Indications for Use
Indicated for cardiologists and radiologists to analyze coronary anatomy and pathology, determine presence/extent of coronary plaques and luminal stenoses, and aid in treatment planning using cardiac CT angiographic images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Cedars-Sinai Medical Center % Philip Won Hyman, Phelps & McNamara, P.C. 700 Thirteenth Street NW, Suite 1200 WASHINGTON DC 20005
Re: K212758
May 19, 2023
Trade/Device Name: Autoplaque 3.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH, LLZ Dated: April 24, 2023 Received: April 25, 2023
Dear Philip Won:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT 8B: Division of Radiological Imaging Devices and Electronic Products OHT 8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212758
Device Name Autoplaque 3.0
### Indications for Use (Describe)
Autoplaque is intended to provide an optimized non-invasive application to analyze coronary and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images.
Autoplaque is a workstation-based post processing application. It is a non-invasive diagnostic reading software intended for use by cardiologists and radiologists as an interactive tool for viewing and analyzing cardiac CT data for determining the presence and extent of coronary plaques and luminal stenoses.
The software is not intended to replace the skill and judgment of a qualified medical practitioner and should only be used by people who have been appropriately trained in the software's functions, capabilities and limitations. Users should be aware that certain views make use of interpolated data. This is data that is created by the software based on the original data set. Interpolated data may give the appearance of healthy tissue in situations where pathology that is near or smaller than the scanning resolution may be present.
Autoplaque must be installed on a suitable commercial computer platform. It is the user's responsibility to ensure the monitor quality and ambient light conditions are consistent with the clinical applications.
Typical users of Autoplaque are trained medical professionals, including but not limited to radiologists, clinicians, technologists, and others.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:10pt">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10pt">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
## Cedars-Sinai Medical Center's Autoplaque 3.0
In accordance with 21 CFR § 807.92 the following summary of information is provided:
#### I. Submitter
Cedars-Sinai Medical Center 8700 Beverly Blvd. Los Angeles, CA 90048
Contact Person: Damini Dey, Ph.D.
Title: Director, Quantitative Image Analysis Program, Biomedical Imaging Research Institute (BIRI), Department of Biomedical Sciences, Cedars Sinai Medical Center Cedars-Sinai Medical Center Phone: 310-423-1517
Fax: 310-423-8396
#### Date of 510(k) Summary Preparation: II.
May 2, 2023
#### III. Device
| Device Proprietary Name: | Autoplaque 3.0 |
|--------------------------|--------------------------------------------------------------------------------------------|
| Common or Usual Name: | Image Processing System, Radiological |
| Classification Name: | Automated Radiological Image Processing Software<br>Radiological Imaging Processing System |
| Regulation Number: | 21 CFR § 892.2050 |
| Product Code: | QIH, LLZ |
| Device Classification | II |
#### IV. Predicate Device
Primary Predicate Device:
- Autoplaque add-on ORS Visual, K122429, Object Research Systems (ORS) Inc. .
{4}------------------------------------------------
#### V. Device Description
Autoplaque 3.0. a stand-alone software, performs post-processing of coronary Computed Tomography Angiography (CTA) images and measurements of components of images using computerized algorithms.
Autoplaque 3.0 aids the physician with measurement of coronary artery stenosis and provides measurements for coronary plaque and coronary artery remodeling. Autoplaque 3.0 does not replace standard clinical practice or clinician decision making.
Autoplaque 3.0 allows for standardized characterization of plaque and stenosis from DICOM image data (loaded from the local computer hard drive) and includes the following features:
- Review of heart and coronary vessels in Multiplanar Reformatting (MPR), curved MPR, . and straightened MPR views:
- Measurement of vessel diameter and area; .
- Characterization and measurement of plaque parameters; and
- . Measurement of lumen diameter, area, and luminal stenosis.
Autoplaque 3.0 includes automated vessel, plaque and lumen segmentation, which is reviewed and can be edited, if necessary, by the clinician.
Autoplaque 3.0 can run on Windows or Mac computer platforms. The minimum hardware specifications are specified in the user manual.
#### VI. Indications for Use
Autoplaque is intended to provide an optimized non-invasive application to analyze coronary anatomy and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images.
Autoplaque is a workstation-based post processing application. It is a non-invasive diagnostic reading software intended for use by cardiologists and radiologists as an interactive tool for viewing and analyzing cardiac CT data for determining the presence and extent of coronary plaques and luminal stenoses.
The software is not intended to replace the skill and judgment of a qualified medical practitioner and should only be used by people who have been appropriately trained in the software's functions, capabilities and limitations. Users should be aware that certain views make use of interpolated data. This is data that is created by the software based on the original data set. Interpolated data may give the appearance of healthy tissue in situations where pathology that is near or smaller than the scanning resolution may be present.
{5}------------------------------------------------
Autoplaque must be installed on a suitable commercial computer platform. It is the user's responsibility to ensure the monitor quality and ambient light conditions are consistent with the clinical applications.
Typical users of Autoplaque are trained medical professionals, including but not limited to radiologists, clinicians, technologists, and others.
#### VII. Comparison of Technological Characteristics
Autoplaque 3.0 and the predicate device are identical with respect to intended use and technological characteristics with the exception of the following:
- Compatibility of the subject device with Mac operating system (OS);
- Provision of the subject device as stand-alone software versus an add-on; and ●
- . Deep learning-based vessel, plaque and lumen segmentation, which is reviewed and can be edited, if necessary, by the clinician.
The table below compares the key technological features between the subject and predicate devices.
| Parameter | Subject Device: Autoplaque 3.0<br>(K212758) | Predicate Device: Autoplaque add-on<br>ORS Visual<br>(K122429) | | | | | | | | | | | | | | | |
|--------------------------------------------------------------|--------------------------------------------------------------------------------|----------------------------------------------------------------|--------------------------------------------|-----|------|---------------------------------------|-----|------|--------------------------------------------------------------|-----|------|---------------------|-----|------|---------------|-----|------|
| Computer operating system | Windows OS<br>Mac OS | Windows OS only | | | | | | | | | | | | | | | |
| Stand-alone software | Yes | No, ORS Visual add-on | | | | | | | | | | | | | | | |
| DICOM compliance | DICOM 3.1 | Same | | | | | | | | | | | | | | | |
| 2D Imaging | Review of coronary vessels in 2D<br>MPR, curved MPR, and straightened<br>view. | Same | | | | | | | | | | | | | | | |
| 2D Measurement | 2D measurement tools of vessel<br>diameter and contour | Same | | | | | | | | | | | | | | | |
| 3D Imaging | Review of structures in 3D | Same | | | | | | | | | | | | | | | |
| Maximum intensity<br>projection (MIP) | MIP with interactive control | Same | | | | | | | | | | | | | | | |
| Multiplanar reformatting<br>(MPR) | MPR with oblique slicing and<br>variable slab thickness | Same | | | | | | | | | | | | | | | |
| Quantitative Measurements | | | NCP volume: non-calcified<br>plaque volume | Yes | Same | CP volume: calcified plaque<br>volume | Yes | Same | LD-NCP volume: low-<br>density noncalcified plaque<br>volume | Yes | Same | Total plaque volume | Yes | Same | Vessel volume | Yes | Same |
| Quantitative Measurements | | | | | | | | | | | | | | | | | |
| NCP volume: non-calcified<br>plaque volume | Yes | Same | | | | | | | | | | | | | | | |
| CP volume: calcified plaque<br>volume | Yes | Same | | | | | | | | | | | | | | | |
| LD-NCP volume: low-<br>density noncalcified plaque<br>volume | Yes | Same | | | | | | | | | | | | | | | |
| Total plaque volume | Yes | Same | | | | | | | | | | | | | | | |
| Vessel volume | Yes | Same | | | | | | | | | | | | | | | |
{6}------------------------------------------------
| Parameter | Subject Device: Autoplaque 3.0<br>(K212758) | Predicate Device: Autoplaque add-on<br>ORS Visual<br>(K122429) |
|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| NCP burden: noncalcified plaque volume/analyzed vessel volume | Yes | Same |
| LD-NCP burden: low-density noncalcified plaque volume/analyzed vessel volume | Yes | Same |
| CP burden: calcified plaque volume/analyzed vessel volume | Yes | Same |
| Total plaque burden: total plaque volume/analyzed vessel volume | Yes | Same |
| Plaque composition NCP: Noncalcified plaque composition (NCP volume /total plaque volume) | Yes | Same |
| Plaque composition CP: Calcified plaque composition (CP volume/total plaque volume) | Yes | Same |
| Plaque composition LD-NCP: low-density noncalcified plaque composition (LD-NCP volume/ NCP volume) | Yes | Same |
| Diameter stenosis: maximal diameter stenosis, with respect to proximal and distal references | Yes | Same |
| QCAD: Maximal diameter stenosis, with respect to proximal and distal references | Yes | Same |
| Remodeling index: ratio of maximum vessel area/proximal and distal references | Yes | Same |
| Area stenosis: maximum area stenosis, with respect to proximal and distal references | Yes | Same |
| Plaque length: diseased vessel length | Yes | Same |
| Contrast density difference: maximum difference in contrast density over lesion with respect to proximal | Yes | Same |
| MLD: minimal luminal diameter over lesion | Yes | Same |
| Parameter | Subject Device: Autoplaque 3.0<br>(K212758) | Predicate Device: Autoplaque add-on<br>ORS Visual<br>(K122429) |
| MLA: minimum luminal<br>area over lesion | Yes | Same |
| Vessel profile: Area,<br>maximum diameter,<br>minimum diameter measured<br>from selected vessel cross<br>section | Yes | Same |
| Lumen profile: Area,<br>maximum diameter,<br>minimum diameter measured<br>from selected lumen cross<br>section | Yes | Same |
| Vessel, plaque, and lumen<br>segmentation | Deep learning based, validated with<br>ground truth by clinician expert<br>readers, with full option to edit. | Algorithm based, validated with ground<br>truth by clinician expert readers, with full<br>option to edit. |
{7}------------------------------------------------
Discussion: As shown above, there are slight technological differences between the subject and predicate devices with respect to computer operating system, provision of the product as standalone software, and vessel, plaque, and lumen segmentation method; however, these differences do not raise new questions of safety and effectiveness, and the differences are addressed through software verification and validation as well as clinical performance studies.
# VIII. Performance Data
Software verification and validation were conducted on the subject device to demonstrate that the subject device meets specifications and works as intended.
In addition, clinical validation studies were undertaken to verify and validate that Autoplaque 3.0 performs as intended. A summary of these studies is provided below.
# 1. Comparison with Expert Reader Measurements in a Multicenter US Patient Population
Performance data was collected on a data set of clinically indicated coronary computed tomography angiography (CCTA) images from multiple U.S. sites. The patient population consisted of 201 patients (53% men, 47% women) with a mean age of 60 ± 12 years. A total of 781 lesions from 201 patients were analyzed. Ethnicity available in 80 patients was as follows: White alone (not Hispanic or Latino) 18%, Black 19%, Hispanic or Latino 61%, and American Indian and Alaska native alone 1%. The CT scanners to collect images included commercial CT scanners from Siemens, Philips and GE. Ground truthing was performed by a board-certified imaging cardiologist. This study established concordance and agreement between clinician expert reader and Autoplaque 3.0 measurements for total plaque volume, calcified plaque volume, non-calcified plaque volume, and diameter stenosis using the acceptance criteria of an intraclass correlation coefficient and correlation coefficient. An excellent agreement was
{8}------------------------------------------------
shown between the subject device and expert reader measurements for total plaque volume, noncalcified plaque volume, calcified plaque volume, and diameter stenosis.
### 2. Comparison with Expert Reader Measurements in a Non-US Patient Population
Performance data was collected on clinically indicated coronary computed tomography angiography (CCTA) images from multiple non-U.S. external sites. The patient population consisted of 175 patients (53% men, 47% women) with a mean age of age 56 ± 9 years. A total of 1081 lesions from 175 patients were analyzed. The CT scanners to collect images included commercial CT scanners from Siemens, Philips and Canon. Ground truthing was performed by two cardiologists with one additional highly experienced radiologist to resolve discrepancies. This study established concordance and agreement between clinician expert reader and Autoplaque 3.0 measurements for total plaque volume, calcified plaque volume, non-calcified plaque volume, and diameter stenosis using the acceptance criteria of an intraclass correlation coefficient and correlation coefficient. An excellent agreement was shown between the subject device and expert reader measurements for total plaque volume, noncalcified plaque volume, calcified plaque volume, and diameter stenosis. In addition, the average per-patient plaque analysis time was 5.7 seconds for the subject device versus 25-30 minutes taken by experts.
### 3. Comparative Study with the Predicate Device
To establish that the subject device provides clinical performance for total plaque volume, calcified plaque volume, non-calcified plaque volume, and diameter stenosis that is substantially equivalent to the predicate device, a total of 30 lesions (from 27 patients) throughout the coronary artery tree were analyzed using both the subject and predicate devices. The patient population were 60% men and 40% women with a mean of age 64.5 ± 10.1 years. The CT scanner used to collect the images was a commercial Siemens CT scanner. There was an excellent correlation between the subject and predicate devices for total plaque volume, noncalcified volume, calcified plaque volume, and diameter stenosis. These measurements also demonstrated excellent intraclass correlation agreement. It took less than 2 seconds per lesion for the subject device Autoplaque 3.0 to analyze, which was much faster than the predicate device (30 seconds per lesion).
#### IX. Conclusion
The information provided above supports that Autoplaque 3.0 is as safe and effective as the predicate device. Although there are minor differences between the subject and predicate devices, the differences are addressed through software verification and validation studies, and performance testing. Therefore, it is concluded that Autoplaque 3.0 is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.