SIGNUS SACRONAIL® Transsacral Stabilization System
Applicant
Signus Medizintechnik GmbH
Product Code
HWC · Orthopedic
Decision Date
Mar 10, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SIGNUS SACRONAIL® Transsacral Stabilization System is intended for fixation of fractures of the posterior pelvis, in areas of superior posterior inferior inferior iliac spine, for sacral fractures and fracture-dislocations of the sacro-iliac joint.
Device Story
The SACRONAIL Transsacral Stabilization System is an orthopedic implant designed for the surgical fixation of posterior pelvic and sacral fractures. The device consists of an intramedullary nail (135mm–194mm length, 8mm diameter) and two locking screws. During surgery, the nail is placed intraosseously to provide bi-iliac symmetric anchorage, with locking screws inserted at a 70-degree angle to the nail axis to secure the construct against the lateral cortex of the ilia. The system is used by orthopedic surgeons in a clinical/surgical setting to restore pelvic stability. By providing a stable, rigid fixation construct, the device facilitates the healing of fractures and fracture-dislocations, potentially reducing pain and restoring pelvic function for the patient.
Clinical Evidence
No clinical data (pre-marketing or post-market studies) or animal studies were performed. Substantial equivalence is supported by bench testing, including mechanical testing per ASTM F1264-16, ASTM F1717-15, and ASTM F543, and biocompatibility assessments per ISO 10993-1.
Technological Characteristics
Constructed from Titanium alloy (Ti6Al4V) per ASTM F136 / ISO 5832-3 with Type II anodization (AMS 2488). System includes an intramedullary nail and locking screws. Mechanical testing performed per ASTM F1264, F1717, and F543. Implants are provided sterile; instruments are provided non-sterile.
Indications for Use
Indicated for fixation of posterior pelvic fractures, specifically in the superior and posterior inferior iliac spine areas, sacral fractures, and sacro-iliac joint fracture-dislocations.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
M.U.S.T. Sacral Iliac and Pelvic Trauma System (K171595)
Submission Summary (Full Text)
{0}------------------------------------------------
March 10, 2022
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency name in a lighter blue. The text reads "FDA U.S. FOOD & DRUG ADMINISTRATION".
SIGNUS Medizintechnik GmbH % Carolyn Guthrie Director of Regulatory & Quality Kapstone Medical, LLC 520 Elliot St. Charlotte, North Carolina 28202
Re: K212755
Trade/Device Name: SACRONAIL Transsacral Stabilization System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC, JDS Dated: February 7, 2022 Received: February 9, 2022
Dear Carolyn Guthrie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212755
Device Name
SIGNUS SACRONAIL® Transsacral Stabilization System
Indications for Use (Describe)
The SIGNUS SACRONAIL® Transsacral Stabilization System is intended for fixation of fractures of the posterior pelvis, in areas of superior posterior inferior inferior iliac spine, for sacral fractures and fracture-dislocations of the sacro-iliac joint.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for SIGNUS, a company that specializes in spine care. The logo features a stylized spine in gray and red, with the company name in bold, black letters. Below the company name is the tagline "The Spine Sign" in a smaller font, with a thin red line above and below it. The logo is simple and modern, and it effectively communicates the company's focus on spine care.
# Traditional 510(k) Submission SIGNUS SACRONAIL® Transsacral Stabilization System
# 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR 807.92, the following summary of information is provided.
### 1. Date of Preparation:
October 07, 2021
# 2. Applicant
SIGNUS Medizintechnik GmbH Industriestrasse 2 63755 Alzenau, Germany
### 3. Official Correspondent
Kapstone Medical LLC 520 Elliot St. Charlotte, NC 28202
# 4. Contact Person:
Carolyn Guthrie Email: cguthrie@kapstonemedical.com
# 5. Device Name
Trade Name: SIGNUS SACRONAIL® Transsacral Stabilization System Common Name: Pelvic Joint Fixation Regulation Description: Smooth or Threaded metallic bone fixation fastener Regulation Number: 21 CFR 888.3040, 21 CFR 888.3030 Product Code: HWC Secondary Product Code: JDS Classification: Class II Panel: Orthopedic
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for SIGNUS - The Spine Sign. The logo features a stylized graphic on the left, consisting of a gray pill shape with a red triangle on either side. To the right of the graphic is the word "SIGNUS" in large, bold, black letters, with the tagline "The Spine Sign" in smaller letters underneath, separated by a thin red line.
# Traditional 510(k) Submission SIGNUS SACRONAIL® Transsacral Stabilization System
#### 6. Predicate Devices
| Primary Predicate | | | | | | |
|-------------------|-----------|------------------------------|---------------|--|--|--|
| Trade Name | Clearance | Claim of<br>Equivalence for: | 510(k) holder | | | |
| Sacral Bar | K001720 | Predicate Device | Depuy/Synthes | | | |
| Additional Predicates | | | |
|------------------------------------------------------|-----------|--------------------------------|-----------------------------|
| Trade Name | Clearance | Claim of<br>Equivalence for: | 510(k) holder |
| Dynanail Mini<br>Hybrid | K203381 | Additional<br>Predicate Device | MedShape Inc. |
| 7.3 mm Cannulated<br>Screws | K161616 | Additional<br>Predicate Device | DePuy/Synthes |
| M.U.S.T. Sacral Iliac<br>and Pelvic Trauma<br>System | K171595 | Additional<br>Predicate Device | Medacta<br>International SA |
#### Table 5.1: Predicate devices
#### 7. Indications for Use
The SIGNUS SACRONAIL® Transsacral Stabilization System is intended for fixation of fractures of the posterior pelvis, in areas of superior iliac spine and posterior inferior illac spine, for sacral fractures and fracture-dislocations of the sacro-iliac joint.
#### 8. Device Description
The SACRONAIL system provides a sustained, stable restoration of dorsal pelvic fractures and sacral fractures through its intraosseous location and its direct, bi-iliac symmetric anchorage.
The implant consists of a nail of various lengths from135mm to 194mm with a constant diameter of 8mm and 2 locking screws in different lengths. This variation ensures adaptation to the respective anatomy of the patient. The intersections of the screw axes to the implant axis have a distance of 15mm to the respective end face of the nail. This ensures a solid support on the lateral cortex of the Ilia. The angle of the screw axes to
{5}------------------------------------------------
K212755 - Page 3 of 5
Image /page/5/Picture/1 description: The image contains the logo for SIGNUS. The logo consists of a stylized shape in red and gray on the left, followed by the word "SIGNUS" in bold, black letters. Below "SIGNUS" is the tagline "The Spine Sign" in a smaller font, underlined by a thin red line.
# Traditional 510(k) Submission SIGNUS SACRONAIL® Transsacral Stabilization System
the implant axis corresponds to the anatomical conditions and is 70°. The locking screw ensures a stable connection of the components. The tips of the locking screws are rounded to facilitate insertion into the nail while the dorsal ends of locking screws are a hexagon. The caps serve to close the instrument connection of the nail. The implants are made from the Titanium alloy (Ti6Al4V) according to ASTM F 136 / ISO 5832-3, with a Type II anodization according to AMS 2488.
There are Instruments which are specific to the implantation of the SACRONAIL device. These include, but are not limited to, the Guide Frame, the Guide Bracket, the Fixation Bracket, various Instrument Guides, various dilators and drills, screwdrivers as well as the specific tools for revision of the device.
#### 8.1 Product Name
SIGNUS SACRONAIL® Transsacral Stabilization System
#### Intended Performance 8.2
The SACRONAIL system is intended to a sustained, stable restoration of dorsal pelvic fractures and sacral fractures through its intraosseous location and its direct, bi-iliac symmetric anchorage.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image is a logo for SIGNUS - The Spine Sign. The logo consists of a stylized gray and red shape on the left, followed by the word "SIGNUS" in bold, black letters. Below "SIGNUS" is the phrase "The Spine Sign" in a smaller font, underlined with a thin red line.
# 9. Comparison of Technological Characteristics with the Predicate Devices
The subject device SACRONAIL® Transsacral Stabilization System was shown to be substantially equivalent and have the same technological characteristics to a previously FDA cleared predicate device through comparison in areas including, intended use, material composition, and function through the means of mechanical and biomechanical cadaveric comparison. Any technical differences, which were identified and discussed in this submission, do not result in new questions of safety or effectiveness.
### 10. Performance Data
There are no clinical data generated and held by the manufacturer, i.e., no pre-marketing or post-market clinical studies or animal studies have been performed. The following information is provided in support of substantial equivalence.
# 10.1 Biocompatibility
The SACRONAIL® Transsacral Stabilization System is classified as an implant device with tissue/bone contact and permanent contact. Therefore, according to ISO 10993-1 the biological evaluation was assessed for potential effects.
The SACRONAIL® Transsacral Stabilization System is manufactured from Titanium Alloy per ASTM F136. The evaluation was based on product-specific tests and on published literature data. This approach is considered feasible as Titanium Alloy is a wellestablished material in the medical field.
The available data from the raw material manufacturer, the testing performed on the finished device, and published literature form the basis of the biological evaluation for the SACRONAIL® Transsacral Stabilization System.
The presented results in this submission show that pedicle screw systems made from Titanium Alloy per ASTM F136 have a high demonstrable biological safety. No concerns arose that would preclude clinical use of SACRONAIL® Transsacral Stabilization System. The requirements of the ISO 10993 standard series are fulfilled.
#### 10.2 Mechanical Testing
Testing according to
- ASTM F1264-16 Standard Specification and Test Method for Intramedullary . Fixation Devices
- . ASTM F1717-15: Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model
- . ASTM F543 Standard Specification and Test Methods for Metallic Medical Bone Screws
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image is a logo for SIGNUS. The logo consists of a stylized pill shape in gray and red on the left, followed by the word "SIGNUS" in bold, black letters. Below the word "SIGNUS" are two horizontal red lines, and below the lines is the phrase "The Spine Sign" in a smaller font.
# Traditional 510(k) Submission SIGNUS SACRONAIL® Transsacral Stabilization System
was performed to demonstrate equivalence to predicates.
The detailed mechanical testing shows that in summary, the SACRONAIL® Transsacral System provides equivalent mechanical properties compared to Stabilization the predicate for this intended use.
### 10.3 MR Safety Testing
The SACRONAIL® Transsacral Stabilization System has not been evaluated for safety, heating, migration, or compatibility in the MR environment. The SACRONAIL® Transsacral Stabilization System has not been tested for heating, migration or image artefact in the MR environment. The safety of the SACRONAIL® Transsacral Stabilization System in the MR environment is unknown. Scanning a patient who has this device may result in patient injury.
#### 11 Conclusion
The SIGNUS SACRONAIL® Transsacral Stabilization System and the predicate device have the similar intended use in sacral and pelvic fractures and are available by prescription only. The SACRONAIL System implants are provided sterile (Instruments are provided non-Sterile). Any technical differences, which were identified, do not result in new questions of safety or effectiveness.
Through assessment of technological characteristics, intended use and the biomechanical performance data provided, it can be concluded that SIGNUS SACRONAIL® Transsacral Stabilization System is substantially equivalent to the predicate device for this intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.