Ultrasonic Surgical Aspirator System, Model: XD880B

K212750 · Smtp Technology Co., Ltd. · LFL · Sep 27, 2021 · SU

Device Facts

Record IDK212750
Device NameUltrasonic Surgical Aspirator System, Model: XD880B
ApplicantSmtp Technology Co., Ltd.
Product CodeLFL · SU
Decision DateSep 27, 2021
DecisionSESE
Submission TypeSpecial
Device ClassClass U
AttributesTherapeutic

Intended Use

The XD880B Ultrasonic Surgical Aspirator System is an ultrasonic surgical system consisting of handpieces and associated tips. The product is intended for the fragmentation, emulsification and aspiration of both soft and hard (i.e. bone and bone approximations) tissue, the intended use is as follows with different configurations: - Neurosurgery - Gastrointestinal and Affiliated Organ Surgery - Urological surgery - Plastic and Reconstructive surgery - General Surgery - Orthopedic Surgery - Gynecological Surgery - Thoracic Surgery - Wound Care - Laparoscopic Surgery - Thoracoscopic Surgery

Device Story

Ultrasonic surgical aspirator system (XD880B) for tissue fragmentation, emulsification, and aspiration. System comprises console, handpieces, tips, foot switch, and accessories (liquid-flow sleeve/tube, suction canister). Operates via ultrasonic vibration (39kHz/52kHz) to break down soft/hard tissue; irrigation system (peristaltic pump) and aspiration (vacuum) remove debris. Used in OR settings by surgeons. Console with touchscreen displays/controls parameters. Soft tissue protection feature included for bone cutting mode. Device aids in precise tissue debridement, cutting, and shaping. Benefits include controlled tissue removal with reduced collateral damage.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing only, including safety/performance testing for soft tissue protection, acoustic performance testing, and cutting efficiency/thermal testing for modified tips.

Technological Characteristics

Ultrasonic surgical system; 39kHz and 52kHz continuous wave vibration. Console-based control with touchscreen; peristaltic irrigation pump (5-120 ml/min); 600mmHg max vacuum. Power: 100-240 VAC, 50/60 Hz. Includes handpieces, tips, and accessories. Soft tissue protection feature in bone cutting mode. Dimensions: 18.0cm H x 31.0cm W x 53.5cm D; Weight: 9.3kg.

Indications for Use

Indicated for fragmentation, emulsification, and aspiration of soft and hard tissue (bone/bone approximations) in neurosurgery, GI, urological, plastic/reconstructive, general, orthopedic, gynecological, thoracic, wound care, laparoscopic, and thoracoscopic surgeries.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 27, 2021 SMTP Technology Co., Ltd. Songtao Zhan CTO 1F&4F, Building A, Emerging Industry Incubation Center Zhangjiagang Free Trade Zone, 215634 Jiangsu Province China Re: K212750 Trade/Device Name: Ultrasonic Surgical Aspirator System, Model: XD880B Regulatory Class: Unclassified Product Code: LFL Dated: August 26, 2021 Received: August 30, 2021 Dear Songtao Zhan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, {1}------------------------------------------------ including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and. if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations. please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Long H. Chen -S-5 Digitally signed by Long H. Chen Date: 2021.09.27 15:18:39-04'00' Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212750 #### Device Name Ultrasonic Surgical Aspirator System #### Indications for Use (Describe) The XD880B Ultrasonic Surgical Aspirator System is an ultrasonic surgical system consisting of handpieces and associated tips. The product is intended for the fragmentation, emulsification of both soft and hard (i.e. bone and bone approximations) tissue, the intended use is as follows with different configurations: - · Neurosurgery - · Gastrointestinal and Affiliated Organ Surgery - · Urological surgery - · Plastic and Reconstructive surgery - · General Surgery - · Orthopedic Surgery - · Gynecological Surgery - · Thoracic Surgery - · Wound Care - · Laparoscopic Surgery - · Thoracoscopic Surgery Type of Use (Select one or both, as applicable) | <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | |----------------------------------------------------------------------------------------------------------| | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary # Ultrasonic Surgical Aspirator System # 1. Date Prepared Sep. 27th, 2021 ### 2. Submitter's Information | Company Name: | SMTP Technology Co., Ltd. | |------------------|-----------------------------------------------------------------------------------------------------------------------| | Company Address: | 1F&4F, Building A, Emerging Industry Incubation Center, Zhangjiagang Free Trade Zone, 215634, Jiangsu Province, CHINA | | Fax: | +86 10 88572898 ext. 8005 | | Contact Person: | Songtao Zhan | | | CTO | | | Tel: +86 10 88572898 | | | zhansongtao@smtpmed.com | # 3. Trade Name, Common Name, Classification | Trade/Proprietary Name: | Ultrasonic Surgical Aspirator System | |-------------------------|--------------------------------------| | Common/Usual Name: | Ultrasonic Surgical Aspirator | | Classification Name: | instrument, ultrasonic surgical | | Regulation Number: | N/A, Pre-Amendment | | Product Code: | LFL | | Class: | Class II | | Classification Panel: | General & Plastic Surgery | ### 4. Identification of Predicate Devices(s) The identification of predicate device within this submission is as follow: | Manufacturer: | SMTP Technology Co., Ltd. | |---------------|--------------------------------------| | Trade Name: | Ultrasonic Surgical Aspirator System | {4}------------------------------------------------ Classification Name: Regulation Number: Product Code: Class: Classification Panel: FDA 510 (k) #: instrument, ultrasonic surgical N/A, Pre-Amendment LFL Class II General & Plastic Surgery K202299 # 5. Description of the Device The Ultrasonic Surgical Aspirator System (Model: XD880B) is an ultrasonic powered surgical tool. It can be used for precise breakage, aspiration and debridement of soft tissues, and can also be used for precise cutting and shaping of bone tissue, using with the liguid-flow sleeve and liquid-flow tube. The product can be applied to a number of departments according to the suitable parameters and accessories, including: - . Department of Neurosurgery, general surgery and other departments for soft tissue breakage and aspiration. - Department of Orthopedic surgery, Plastic and Reconstructive surgery, Department of Neurosurgery (only for bone) for bone cutting. - . As well as the Department of Dermatology, Department of Trauma Orthopedics and other departments used for debridement. The Ultrasonic Surgical Aspirator System consists of a console, application software, foot switch, and accessories. The accessories include handpieces, tips, wrenches, liquid-flow sleeves, liquid-flow tubes, sterilization tray, suction canister and mobile cart. The handpiece, tip, liquid-flow sleeve and liquid-flow tube assemble to ultrasonic tool part to complete the cutting and fragment of soft tissue and bone tissue. # 6. Indications for Use The XD880B Ultrasonic Surgical Aspirator System is an ultrasonic surgical system consisting of handpieces and associated tips. The product is intended for the fragmentation, emulsification and aspiration of both soft and hard (i.e. bone and bone approximations) tissue, the intended use is as follows with different configurations: - -Neurosurgery - -Gastrointestinal and Affiliated Organ Surgery - -Urological surgery - -Plastic and Reconstructive surgery - -General Surgery - Orthopedic Surgery - - -Gynecological Surgery - Thoracic Surgery - - Wound Care - {5}------------------------------------------------ - Laparoscopic Surgery - - Thoracoscopic Surgery - # 7. Substantial Equivalence Most the features and characteristics of the subject device Ultrasonic Surgical Aspirator System are identical to those of the predicate device under k202299, and where there are differences, such differences do not have an impact on the safety or effectiveness of the subject device. The following Table 7A compares the subject device Ultrasonic Surgical Aspirator System to the predicate device. | State | SUBJECT DEVICE | PREDICATE DEVICE<br>K202299 | |------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | SMTP Technology Co., Ltd. | SMTP Technology Co., Ltd. | | Classification | Unclassified | Unclassified | | Common/Usual<br>Name | Ultrasonic Surgical Aspirator<br>Ultrasonic Surgical System | Ultrasonic Surgical Aspirator<br>Ultrasonic Surgical System | | Regulation<br>Panel | General & Plastic Surgery | General & Plastic Surgery | | Product Code | LFL | LFL | | Indications for<br>Use | The XD880B Ultrasonic Surgical<br>Aspirator System is an ultrasonic<br>surgical system consisting of<br>handpieces and associated tips.<br>The product is intended for the<br>fragmentation, emulsification and<br>aspiration of both soft and hard<br>(i.e. bone and bone<br>approximations) tissue, the<br>intended use is as follows with<br>different configurations:<br>Neurosurgery<br>Gastrointestinal and Affiliated<br>Organ Surgery<br>Urological surgery<br>Plastic and Reconstructive surgery | The XD880B Ultrasonic Surgical<br>Aspirator System is an ultrasonic<br>surgical system consisting of<br>handpieces and associated tips.<br>The product is intended for the<br>fragmentation, emulsification and<br>aspiration of both soft and hard<br>(i.e. bone and bone<br>approximations) tissue, the<br>intended use is as follows with<br>different configurations:<br>Neurosurgery<br>Gastrointestinal and Affiliated<br>Organ Surgery<br>Urological surgery<br>Plastic and Reconstructive surgery | | | General Surgery | General Surgery | | | Orthopedic Surgery | Orthopedic Surgery | | | Gynecological Surgery | Gynecological Surgery | | | Thoracic Surgery | Thoracic Surgery | | | Wound Care | Wound Care | | | Laparoscopic Surgery | Laparoscopic Surgery | | | Thoracoscopic Surgery | Thoracoscopic Surgery | | Vibration<br>System | Continuous Wave<br>Frequency:<br>39kHz & 52kHz | Continuous Wave<br>Frequency:<br>39kHz & 52kHz | | Electrical Power<br>Supply | 100-240 VAC 50/60 Hz. | 100-240 VAC 50/60 Hz. | | Footswitch | Foot switch connected to the<br>device control unit by means of a<br>cord. | Foot switch connected to the<br>device control unit by means of a<br>cord. | | Console | The functional parameters are<br>displayed and controlled through a<br>console with touch screen. | The functional parameters are<br>displayed and controlled through a<br>console with touch screen. | | Irrigation<br>System | Peristaltic pump. | Peristaltic pump. | | Irrigation fluid | Adjustable between:<br>5 to 120 ml/min. | Adjustable between:<br>5 to 120 ml/min. | | Aspiration<br>Vacuum | 600mmHg (80kPa) maximum | 600mmHg (80kPa) maximum | | Soft Tissue<br>Protection in<br>Bone Cutting<br>Mode | Yes | No | | Blade with Slot<br>of Cutting Tips | With slot | Without slot. | | Dimensions | 18.0cm H ×31.0cm W × 53.5cm D | 18.0cm H ×31.0cm W × 53.5cm D | | Weight | 9.3kg | 9.3kg | # Table 7A – Comparison of Characteristics {6}------------------------------------------------ # Discussion of Substantial Equivalence: Comparison of General information, Product performance, Electrical characteristics, and Conditions between the predicate device and the subject device is identified minor differences; the differences were supported with safety and performance testing, as appropriate, and do not affect device safety and performance. The subject device has same indications for use with predicate device. The technological characteristics differences listed {7}------------------------------------------------ in the above table between the subject device and the predicate device do not raise any new harms of safety and effectiveness. #### Non-Clinical Performance Data: SMTP completed the following non-clinical tests to demonstrate safety and effectiveness of Ultrasonic Surgical Aspirator System to show substantial equivalence to the predicate device, the test results confirm that the performance specifications meets the modification inputs. Bench Testing - . Safety and Performance Testing for Soft Tissue Protection in Bone Cutting Mode. - Acoustic Performance Testing for modified tips. ● - Cutting Efficiency and Thermal testing Report for Modified Tips. - . Comparison of Substantial Equivalence for Modified Tips. #### Clinical Performance Data: Clinical evaluation is not applicable for the proposed device. ### 8. Conclusion The Ultrasonic Surgical Aspirator System has the same intended use as the predicate device, and the same or similar technological characteristics. The differences in technological characteristics do not raise new harms of safety and effectiveness. Safety and performance testing have demonstrated the Ultrasonic Surgical Aspirator System is as safe and effective as the predicate device. Therefore, the subject device Ultrasonic Surgical Aspirator System is substantially equivalent to the predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%