MedSchenker Smart Transport Medium (STM15-A/STM20-A/STM30-A/SCS30-A) System
K212743 · Medschenker, Inc. · JSM · May 3, 2023 · Microbiology
Device Facts
Record ID
K212743
Device Name
MedSchenker Smart Transport Medium (STM15-A/STM20-A/STM30-A/SCS30-A) System
Applicant
Medschenker, Inc.
Product Code
JSM · Microbiology
Decision Date
May 3, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2390
Device Class
Class 1
Indications for Use
MedSchenker Smart Transport Medium (STM15-A/STM20-A/STM30-A/SCS30-A) System is intended for the collection and transport of upper respiratory clinical specimens, contaming respiratory viruses, from the collection site to the testing laboratory. MedSchenker Smart Transport Medium (STM15-A/STM30-A/SCS30-A) System is a culture-based medium that can be processed using standard clinical laboratory operating procedures for the recovery of infectious viral particles.
Device Story
MedSchenker STM is a non-propagating transport system for clinical viral specimens. System consists of screw-cap tubes containing Hank's Balanced Salt Solution (HBSS) enriched with protein, antimicrobial agents, and a pH buffer; some configurations include a sterile nasopharyngeal swab. Healthcare professionals collect specimens using the swab, place them into the medium, and transport to a laboratory. The medium maintains viral viability and prevents microbial overgrowth during transit. Laboratory personnel process the sample using standard culture procedures to recover infectious viral particles. Benefits include room-temperature stability and preservation of viral infectivity for up to 72 hours (HSV-1, HSV-2, Adenovirus) or 24 hours (Flu A, VZV) to facilitate accurate diagnostic testing.
Clinical Evidence
Bench testing only. Viral recovery studies performed using HSV-1, HSV-2, Adenovirus, VZV, and Influenza A. Samples were inoculated onto swabs, placed in STM, and stored at 4-8°C or 20-25°C for 0-72 hours. Viability assessed via cytopathic effects (CPE) using MTT assay. Results confirmed recovery of all tested strains within specified timeframes. Mycoplasma recovery was evaluated but data was insufficient to support a claim.
Technological Characteristics
Transport medium in plastic, conical-bottom screw-cap tubes. Formulation: Amphotericin B, Bovine Serum Albumin, Hank's Balanced Salt Solution, Vancomycin, Colistin, Gelatin, HEPES, L-cysteine, L-glutamic acid, Phenol Red, Sucrose. pH 7.3 ± 0.5. Shelf life 12 months. Storage 2-25°C. Single-use. Swab: Nylon tip with break point.
Indications for Use
Indicated for the collection and transport of upper respiratory clinical specimens containing respiratory viruses from the collection site to the testing laboratory for culture-based recovery of infectious viral particles.
Regulatory Classification
Identification
A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.
Predicate Devices
Copan Universal Transport Medium (utm-rt) System (K042970)
Submission Summary (Full Text)
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K212743
B Applicant
MedSchenker, Inc.
C Proprietary and Established Names
MedSchenker Smart Transport Medium (STM15-A/STM20-A/STM30-A/SCS30-A) System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JSM | Class I, reserved | 21 CFR 866.2390 - Transport Culture Medium | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain substantial equivalence determination for MedSchenker Smart Transport Medium (STM15-A/STM20-A/STM30-A/SCS30-A) System for the collection and transport of clinical specimens containing viruses from the collection site to the testing laboratory.
B Measurand:
Not Applicable
C Type of Test:
Non-propagating transport device with culture medium.
## III Intended Use/Indications for Use:
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
# A Intended Use(s):
See Indications for Use below.
# B Indication(s) for Use:
MedSchenker Smart Transport Medium (STM15-A/STM20-A/STM30-A/SCS30-A) System is intended for the collection and transport of upper respiratory clinical specimens, containing respiratory viruses, from the collection site to the testing laboratory.
MedSchenker Smart Transport Medium (STM15-A/STM20-A/STM30-A/SCS30-A) System is a culture-based medium that can be processed using standard clinical laboratory operating procedures for the recovery of infectious viral particles.
# C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
# D Special Instrument Requirements:
Not Applicable
# IV Device/System Characteristics:
# A Device Description:
The MedSchenker Smart Transport Medium (STM15-A/STM20-A/STM30-A/SCS30-A) System ("MedSchenker STM") includes a universal transport medium that is room temperature stable and that can sustain infectivity of a plurality of clinically important viruses during transit to the testing laboratory. The formulation of the MedSchenker STM includes Hank's Balanced Salt Solution (HBSS) enriched with protein for stabilization, antimicrobial agents to minimize bacterial and fungal contamination, and a buffer system to maintain a neutral pH.
The MedSchenker STM is provided in labeled, screw-cap tubes designed for transport of the clinical sample. MedSchenker STM is also supplied as a sample collection kit that contains one screw-cap tube of STM-RT (room-temperature stable) medium and a peel pouch that contains one sterile nasopharyngeal specimen-collection swab. The MedSchenker STM is provided as follows in Table 1:
Table 1: Configurations to be marketed:
| SKU | STM Tube Description | Pack size |
| --- | --- | --- |
| STM15-A | 1.5 mL Screw cap with Tube | 50 Qty |
| STM20-A | 2.0 mL Screw cap with Tube | 50 Qty |
| STM30-A | 3.0 mL Screw cap with Tube | 50 Qty |
| SCS30-A | 3.0 mL Screw cap with Tube + Nasopharyngeal CavSwab Swab | 50 Qty |
K212743 - Page 2 of 8
{3}
K212743 - Page 3 of 8
B Principle of Operation:
The MedSchenker STM System is used to safely collect, transport, and maintain the viability of clinically important viruses. Intended for use by Health Care Professionals, the transport system allows for the collection of the specimen via the sterile swab, maintenance of viral particles via a buffered salt/serum solution, prevention of microbial overgrowth via antimicrobial agents, as well as a buffer system to maintain a neutral pH. Once the clinical sample is collected, it should be immediately placed in the MedSchenker STM where it is then transported to the laboratory for testing and other detection/identification methods for the presence of a target viruses.
V Substantial Equivalence Information:
A Predicate Device Name(s):
Copan Universal Transport Medium (utm-rt) System
B Predicate 510(k) Number(s):
K042970
C Comparison with Predicate(s):
| Device & Predicate Device(s): | Subject: K212743 | Predicate: K042970 |
| --- | --- | --- |
| Device Trade Name | MedSchenker Smart Transport Medium (STM15-A/STM20-A/STM30-A/SCS30-A) System | Copan Universal Transport Medium (UTM-RT) System |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | MedSchenker Smart Transport Medium (STM15-A/STM20-A/STM30-A/SCS30-A) System is intended for the collection and transport of upper respiratory clinical specimens, containing respiratory viruses, from the collection site to the testing laboratory. MedSchenker Smart Transport Medium (STM15-A/STM20-A/STM30-A/SCS30-A) System is a culture-based medium that can be processed using standard clinical laboratory operating | Copan Universal Transport Medium (UTM-RT) System is intended for the collection and transport of clinical specimens containing viruses, chlamydiae, mycoplasma or ureaplasma from the collection site to the testing laboratory. UTM-RT can be processed using standard clinical laboratory operating procedures for viral, chlamydial, mycoplasma and ureaplasma culture. |
{4}
K212743 - Page 4 of 8
| | procedures for the recovery
of infectious viral particles. | |
| --- | --- | --- |
| Media formulation | Amphotericin B
Bovine Serum Albumin
Hank’s Balanced Salt
Solution
Vancomycin
Colistin
Gelatin
HEPES
L-cysteine
L-glutamic acid
Phenol Red
Sucrose | Same |
| Container for medium | Plastic, conical bottom | Same |
| Product configuration | Medium Tubes; Kit with
Medium Tubes and
Swab Option | Same |
| Storage Temperature | 2 - 25°C (refrigerated and
room temperature) | Same |
| Shelf Life | 12 months | Same |
| Single Use | Yes | Same |
| **General Device**
**Characteristic Differences** | | |
| Supported claims to perform
culture, isolation, and
detection of: | Viruses:
Influenza A (H1N1)
Type 5 Adenovirus
Herpes Simplex 1
Herpes Simplex 2
Varicella-Zoster Virus | Viruses:
Adenovirus
Cytomegalovirus (CMV)
Echovirus Type 30 (Echo
30)
Herpes Simplex Virus Type
1 (HSV1)
HSV2
Influenza A
Parainfluenza Type 3
Respiratory Syncytial Virus
(RSV)
Varicella Zoster Virus
(VZV)
Chlamydiae:
*Chlamydia pneumoniae*
Strain CM-1
Chlamydia trachomatis Type
1 Strain UW-12/UR
Mycoplasma:
*Mycoplasma hominis* |
{5}
K212743 - Page 5 of 8
| | | Mycoplasma pneumoniae
Ureaplasma:
Ureaplasma urealyticum |
| --- | --- | --- |
| pH Stability | 7.3 ± 0.5 maintained up to 12 months | 7.3 ± 0.2 maintained up to 12 months |
| Swab material | Nylon tip with break point | Polyester |
VI Standards/Guidance Documents Referenced:
1. ISO 11137-2:2013 Sterilization of health care products-Radiation- Part 2: Establishing the radiation dose.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Not Applicable
2. Linearity:
Not Applicable
3. Analytical Specificity/Interference:
Not Applicable
4. Assay Reportable Range:
Not Applicable
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Shelf-life: The shelf-life of the MedSchenker STM System was determined to be 12 months from the date of manufacture, based on the data provided. The shelf-life of MedSchenker STM System was established using an endpoint real time stability study. Three random samples from each of three non-consecutive lots were assessed using a pH meter at day 0 and day 12 post manufacture and the average pH of all samples analyzed was $7.3 \pm 0.5$. The pH of the media was again assessed in media stored for 12 months post manufacture and showed a pH average of $7.8 \pm 0.3$. The media were also assessed for any color change (from pink/violet to any yellowing), or turbidity (any particulates) and no changes were observed at 12 months of storage. The data support a stability claim of 12 months.
Sterilization: The MedSchenker STM is not provided sterile to the end user. Sterile nasopharyngeal CavSwabs are provided with the MedSchenker STM. The manufacturing process of the swabs includes sterilization by electron beam irradiation set at 15 KiloGray
{6}
(kGy) to achieve an SAL of $10^{-6}$ as confirmed by a bioburden assay in accordance with ISO 11137-2:2013. The manufacturing process of the MedSchenker STM include cleaning by electron beam irradiation to achieve an SAL of $10^{-3}$ in accordance with ISO 11137-2:2013.
# 6. Detection Limit:
# Performance Testing- Recovery Studies:
# Culture-based Viral Recovery Studies:
Stock Preparation. Strains of Herpes Simplex Virus Type 1 and 2 (HSV-1 and 2), Varicella Zoster Virus (VZV), Influenza A 2009 virus (H1N1), and Type 5 Adenovirus were used for media validation. The following cell lines were used for viral recovery: BHk 21 (for HSV-1 and 2), Vero cells (for VZV), MDCK (for H1N1) and Hep-2 (for Type 5 Adenovirus). The cell lines were grown to near confluency, harvested, diluted in DMEM with $5 - 10\%$ FBS to obtain a stock solution of $2 - 3\times 10^{5}$ cells/mL and plated in microwell plates 24 hours prior to inoculation. The following viral titers were used as the initial stock for the viral recovery studies: HSV-1 $(1.32\times 10^{6}\mathrm{TCID}_{50} / \mathrm{mL})$ , HSV-2 $(3.47\times 10^{5}\mathrm{TCID}_{50} / \mathrm{mL})$ , VZV $(6.31\times 10^{6}\mathrm{TCID}_{50} / \mathrm{mL})$ , H1N1 $(1.02\times 10^{7}\mathrm{TCID}_{50} / \mathrm{mL})$ and Type 5 Adenovirus $(1.07\times 10^{5}\mathrm{TCID}_{50} / \mathrm{mL})$ .
Viral recovery studies. Strains of HSV-1, HSV-2, VZV, Flu A, and Adenovirus were used for transport media validation. Each viral strain was mixed with negative nasopharyngeal clinical matrix, followed by a 1:2 dilution with $0.85\%$ saline to obtain diluted viral samples for testing. CavSwabs were used to soak up $100~\mu \mathrm{L}$ of each concentration of viruses in three replicates. Three replicates of the inoculated swabs were then transferred into tubes with the MedSchenker STM, one swab per tube, and incubated at room temperature $(20 - 25^{\circ}\mathrm{C})$ or refrigerated $(4 - 8^{\circ}\mathrm{C})$ for 0, 24, 48, and 72 hours for HSV-1, HSV-2, and Adenovirus, and for 0 and 24 hours for Flu A and VZV. Each time point was assessed using 3 lots of media. After each time point, the swabs were centrifuged and then removed from the transport media tube and a $50~\mu \mathrm{L}$ aliquot of the test suspension was seeded onto the appropriate host cell monolayer and incubated at $37^{\circ}\mathrm{C}$ $(5\% \mathrm{CO}_{2})$ for 3-5 days. Viral viability was assessed via cytopathic effects (CPE) using the MTT assay. Table 2 below shows the summary and recovery of the 1:2 dilution of each viral strain at the indicated times and temperatures. Viral recovery is represented as $\mathrm{TCID}_{50} / \mathrm{mL}$ and percent change over time.
Table 2: Summary of recovered viral viability at $4 - {25}^{ \circ }\mathrm{C}$
| Viral strains | | 4-8°C | | 20-25°C | |
| --- | --- | --- | --- | --- | --- |
| | Duration (hours) | 1:2 | | 1:2 | |
| | | TCID50/mL | Percent change (%) (-ve indicates reduction) | TCID50/mL | Percent change (%) (-ve indicates reduction) |
| HSV-1 | 0 | 4.77 x 105 | 0 | 4.28 x 105 | 0 |
| | 24 | 4.18 x 105 | -12.40 | 2.87 x 105 | -32.91 |
| | 48 | 3.20 x 105 | -32.89 | 4.09 x 105 | -4.47 |
| | 72 | 3.32 x 105 | -30.30 | 2.71 x 105 | -36.57 |
| HSV-2 | 0 | 1.49 x 105 | 0 | 1.52 x 105 | 0 |
| | 24 | 1.45 x 105 | -2.88 | 1.24 x 105 | -18.60 |
| | 48 | 1.54 x 105 | +3.78 | 1.26 x 105 | -14.35 |
| | 72 | 7.56 x 104 | -49.24 | 8.13 x 104 | -46.55 |
K212743 - Page 6 of 8
{7}
| Adenovirus | 0 | 1.11 x 10^{5} | 0 | 1.03 x 10^{3} | 0 |
| --- | --- | --- | --- | --- | --- |
| | 24 | 9.95 x 10^{4} | -10.72 | 1.00 x 10^{5} | -2.57 |
| | 48 | 9.26 x 10^{4} | -16.89 | 8.76 x 10^{4} | -14.62 |
| | 72 | 7.87 x 10^{4} | -29.33 | 5.97 x 10^{4} | -41.77 |
| VZV | 0 | 1.23 x 10^{6} | 0 | 1.08 x 10^{6} | 0 |
| | 24 | 4.33 x 10^{5} | -64.85 | 1.43 x 10^{6} | +32.13 |
| Influenza A | 0 | 7.11 x 10^{6} | 0 | 5.52 x 10^{6} | 0 |
| | 24 | 5.79 x 10^{6} | -18.57 | 2.21 x 10^{6} | -60.05 |
Conclusion of the culture-based viral recovery study: The MedSchenker STM demonstrated the recovery of HSV-1, HSV-2, and Adenovirus in all replicates at tested incubation times and storage conditions. These data support the transportation of HSV-1, HSV-2, and Adenovirus in MedSchenker STM at refrigerated (4-8°C) or room temperature (20-25°C) for up to 72 hours.
The MedSchenker STM also demonstrated the recovery of VZV and Flu A in all replicates at refrigerated (4-8°C) or room temperature (20-25°C), up to 24 hours.
Performance of the MedSchenker STM was also evaluated for mycoplasma viability at different incubation times and temperatures. Data was insufficient to support inclusion of mycoplasma in the intended use of the MedSchenker STM. This is reflected in the following labeling limitation:
The MedSchenker STM was also evaluated for mycoplasma viability at different incubation times and temperatures. Data was insufficient to support use of this medium for transport of specimens for mycoplasma testing.
7. Assay Cut-Off:
Not Applicable
B Comparison Studies:
1. Method Comparison with Predicate Device:
Not Applicable
2. Matrix Comparison:
Not Applicable
C Clinical Studies:
1. Clinical Sensitivity:
Not Applicable
2. Clinical Specificity:
K212743 - Page 7 of 8
{8}
Not Applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not Applicable
D Clinical Cut-Off:
Not Applicable
E Expected Values/Reference Range:
Not Applicable
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and does support a substantial equivalence decision.
K212743 - Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.