K212741 · Fci (France Chirurgie Instrumentation) Sas · HPZ · Jun 24, 2022 · Ophthalmic
Device Facts
Record ID
K212741
Device Name
EZYPOR
Applicant
Fci (France Chirurgie Instrumentation) Sas
Product Code
HPZ · Ophthalmic
Decision Date
Jun 24, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.3320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EZYPOR® orbital implants are designed to fill the orbital cavity following enucleation, evisceration or during secondary implantation.
Device Story
EZYPOR orbital implants are high-density porous polyethylene (UHMWPE) spheres used to fill the orbital cavity after enucleation, evisceration, or secondary implantation. The device is implanted surgically by an ophthalmologist or surgeon. The porous structure (40-60% porosity) allows for fibrovascular ingrowth, which helps stabilize the implant within the orbit. The device is provided sterile in various diameters (12-22 mm); some models include a suture platform for attachment to the extraocular muscles. By restoring orbital volume, the implant provides a foundation for a prosthetic eye, improving cosmetic outcomes and patient quality of life.
Clinical Evidence
Bench testing only. Testing included evaluation of diameter, tunnel dimensions, mass, sphericity, open porosity, compression resistance, and traction resistance. Biocompatibility was established per ISO 10993-1, ISO 10993-5, and ISO 10993-18. Shelf life and package integrity were verified via accelerated aging and ISTA 3A transport testing.
Technological Characteristics
Material: High-density porous polyethylene (UHMWPE). Porosity: 40-60%. Pore size: 350-600 µm. Form factor: Spherical, 12-22 mm diameters. Sterilization: Ethylene oxide (ISO 11135-1:2014). Connectivity: None (standalone implant).
Indications for Use
Indicated for patients requiring orbital volume replacement following enucleation, evisceration, or secondary implantation procedures.
Regulatory Classification
Identification
An eye sphere implant is a device intended to be implanted in the eyeball to occupy space following the removal of the contents of the eyeball with the sclera left intact.
Special Controls
*Classification.* Class II (special controls). The device, when it is an ocular peg which is supplied sterile only, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Predicate Devices
MEDPOR® Plus SST™ Sphere (Smooth Surface Tunnel Sphere) (K021357)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 24, 2022
FCI (France Chirurgie Instrumentation) SAS % Barbara Fant, PharmD Principal Regulatory Consultant Clinical Research Consultants. Inc. 3308 Jefferson Avenue, Upper Level Cincinnati. Ohio 45220
Re: K212741
Trade/Device Name: EZYPOR Regulation Number: 21 CFR 886.3320 Regulation Name: Eye Sphere Implant Regulatory Class: Class II Product Code: HPZ Dated: May 12, 2022 Received: May 13, 2022
Dear Dr. Barbara Fant:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anjana Jain, PhD Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212741
Device Name EZYPOR®
Indications for Use (Describe)
The EZYPOR® orbital implants are designed to fill the orbital cavity following enucleation, evisceration or during secondary implantation.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary -- K212741
| 510(k) Owner: | France Chirurgie Instrumentation SAS (FCI S.A.S.)<br>20-22 rue Louis Armand<br>75015 Paris, France |
|---------------|----------------------------------------------------------------------------------------------------|
| Telephone: | +33 1 53 98 98 98 |
| Facsimile: | +33 1 53 98 98 99 |
Barbara Fant, PharmD Contact Person: Clinical Research Consultants, Inc. 3308 Jefferson Avenue Upper Level Cincinnati, OH 45220 Phone: (513) 961-8200 Facsimile: (513) 961-2858
June 23, 2022 Date:
EZYPOR® Trade Name:
Common name: Orbital Implant
Classification Name: Implant, Eye, Sphere
Regulation Number: 886.3320
Product Code: HPZ
#### Identification of a Legally Marketed Predicate Device
EZYPOR® orbital implants are substantially equivalent to the MEDPOR® Plus SST™ Sphere (Smooth Surface Tunnel Sphere, 510(k) Premarket Notification Number K021357, FDA Product Code HPZ.
#### Identification of a Legally Marketed Reference Device
The MEDPOR® Sphere (K863943; Product Code HPZ) is a reference device for the technological characteristics of the 12 to 14 mm diameter porous, smooth spherical shape without a suture platform. The Silicone Orbital Implant (K911110) is a reference device for the technological characteristics of the 12 to 14 mm diameter spherical shape.
{4}------------------------------------------------
## General Description
EZYPOR® orbital implants are high density polyethylene (UHMWPE) implants designed to fill the orbital cavity following enucleation, evisceration or during secondary implantation procedures. The polyethylene material has an open porosity structure of 40 to 60%. The implants are available in six diameter sizes, 12, 14, 16, 18, 20 and 22 mm.
EZYPOR® orbital implants are supplied sterile and are sterilized by ethylene oxide. A green indicator on the label shows that the product followed a sterilization cycle validated by FCI. The EZYPOR® orbital implants are sterilized in their final double-blister packaging, to make them easier to handle in aseptic conditions.
### Indications for Use
The EZYPOR® orbital implants are designed to fill the orbital cavity following enucleation, evisceration or during secondary implantation.
## Sterilization
The EZYPOR® Orbital Implants are distributed as a packaged, sterile device. The EZYPOR® Orbital Implants are sterilized by ethylene oxide (EO) in the sealed packaging. EO is an established Category A Method, in accordance with ISO 11135-1:2014 Sterilization of health care products -- Ethylene oxide -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices. Endotoxin testing was performed as one of the assessments to evaluate sterility for the EZYPOR® Orbital Implants.
# Brief Summary of Non-Clinical Tests and Results
Non-clinical bench testing for the EZYPOR® orbital implants included evaluation of diameter and tunnel dimensions, implant mass, sphericity and open porosity, implant resistance to compression, and implant resistance to traction. All nonclinical test results met the predetermined acceptance criteria for the device. In-process controls and final product quality controls, including finished product release testing and inspection, assure that EZYPOR® is manufactured within specifications.
Biocompatibility of the EZYPOR® orbital implants was established by a review of existing data and test results. The materials used in the implants and materials used in the manufacturing processes were reviewed for available toxicity and bioavailability data for each chemical component, and a justification for the tests conducted to evaluate all potential toxic end points was completed. Implant materials testing included a chemical characterization of materials (leachable test), cytotoxicity testing, sensitization and irritation. Biocompatibility testing was conducted in in accordance with:
{5}------------------------------------------------
ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity ISO 10993-18:2020 Biological evaluation of medical devices -- Part 18: Chemical Characterization of Medical Device Materials in a Risk Management Process
Shelf life of the EZYPOR orbital implants has been established at five years based on accelerated aging and package integrity. The functional performance of the device as well as the package integrity have been verified through the tests conducted before and after the aging; the device transport testing was conducted in accordance with the standard ISTA 3A, and package integrity (visual integrity, dye penetration and bubble leak test) was conducted.
### Basis of Substantial Equivalence
EZYPOR® orbital implants are substantially equivalent to the predicate MEDPOR® Plus SST™ Sphere (Smooth Surface Tunnel Sphere) cleared under K021357 as both devices have the same intended use and indications for use to fill the orbital cavity (void) following enucleation or evisceration of the eve. The EZYPOR® and MEDPOR® Plus SST™ Sphere orbital implants are both manufactured from porous polyethylene material that has the same pore size and porosity. The design of EZYPOR® and MEDPOR Plus SST™ predicate device are similar, with each having a spherical shaped body and available in diameters of 16 to 22 mm that have a suture platform which allows the device to be sutured to the muscularis. The EZYPOR® is also available in 12 mm and 14 mm diameter models that are designed without a suture platform. The EZYPOR® and MEDPOR devices are both sterilized by Ethylene Oxide.
The difference between the subject and predicate devices is that the EZYPOR® orbital implant is also available in 12 mm and 14 mm diameter sizes (without a suture platform), which is the same as the MEDPOR® Sphere; whereas the 16 mm is the smallable diameter of the predicate device. The 12 mm and 14 mm diameter sizes of the EZYPOR® are equivalent in size to Silicone Orbital Implants (K911110) manufactured and sold by FCI.
The similarities and differences between the EZYPOR® orbital implants and the predicate and reference devices are presented in the comparison table below. Based on this comparison, EZYPOR® orbital implants are substantially equivalent to the MEDPOR® Plus SST™ Sphere (Smooth Surface Tunnel Sphere, 510(k) predicate device (K021357).
{6}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Comparison of EZYPOR®+and the Predicate Device and Reference Devices | | | | |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| | EZYPOR® | MEDPOR® Plus<br>SST™ Sphere<br>(Smooth Surface<br>Tunnel Sphere)<br>K021357<br>(Predicate Device) | MEDPOR® Sphere<br>K863943<br>(Reference Device) | Silicone Orbital Implant<br>K911110<br>(Reference Device) |
| FDA<br>Product<br>Code | HPZ | HPZ | HPZ | HPZ |
| Mfg. | FCI S.A.S.<br>FCI SUD | Stryker | Stryker | FCI S.A.S.<br>FCI SUD |
| Intended<br>use | To fill the orbital<br>cavity following<br>enucleation or<br>evisceration | Orbital volume<br>replacement implant | Orbital volume<br>replacement implant | To fill the orbital cavity<br>following enucleation or<br>evisceration |
| Indications<br>for Use | The EZYPOR®<br>orbital implants are<br>designed to fill the<br>orbital cavity<br>following enucleation,<br>evisceration or during<br>secondary<br>implantation. | Void volume<br>replacement following<br>enucleation and/or<br>evisceration of the eye. | For orbital reconstruction<br>following enucleation and<br>evisceration procedures. | Fill the orbital cavity<br>following enucleation,<br>evisceration |
| Patient<br>Contact<br>Materials | Porous, high-density<br>polyethelene | Porous, high-density<br>polyethylene | Porous, high-density<br>polyethylene | Medical grade silicone |
| Pore Size | 350-600 um | 350-600 um | 350-600 um | Not applicable |
| Porosity | 40 - 60 % | 40 - 60 % | | |
| Mfg.<br>Process | Preformed shapes | Preformed shapes | Preformed shapes | Molded shape |
| Design | Image: white sphere | Image: white sphere | Image: multiple white spheres | Image: white sphere |
| Available<br>Diameters | 12, 14, 16, 18, 20, 22<br>mm | 16, 18, 20, 22 mm | 14 mm to 23 mm diameters | 12 mm to 22 mm<br>diameters |
| Single-Use<br>Only | Yes | Yes | Yes | Yes |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.