Mariner Pedicle Screw System

K212692 · SeaSpine Orthopedics Corporation · NKB · Sep 21, 2021 · Orthopedic

Device Facts

Record IDK212692
Device NameMariner Pedicle Screw System
ApplicantSeaSpine Orthopedics Corporation
Product CodeNKB · Orthopedic
Decision DateSep 21, 2021
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use of the Mariner Pedicle Screw System in a posterior or anterolateral approach is to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. The indications for use are as follows: · degenerative disc disease (DDD) as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies, · spondylolisthesis, - · trauma (i.e., fracture or dislocation), · spinal stenosis, - · deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), - · spinal tumor, - pseudarthrosis, and/or - · failed previous fusion.

Device Story

Non-cervical spinal fixation system; provides immobilization/stabilization of spinal segments to facilitate biological fusion. System comprises single-use implants: fixed, uni-planar, polyaxial, and motion-limiting modular pedicle screws; connecting rods; hooks; connectors; crossbars; set screw locking elements. Implants placed via posterior or anterolateral approach by surgeons. Instruments facilitate placement, locking, and removal. System includes storage trays/caddies for steam sterilization. Output is a rigid construct providing temporary or permanent stabilization. Benefits include correction of spinal disorders and support for fusion process.

Clinical Evidence

No clinical data. Substantial equivalence based on non-clinical mechanical testing (static and dynamic) per ASTM F1717 and ASTM F1798, and risk analysis.

Technological Characteristics

Materials: Titanium alloy (ASTM F136) and/or cobalt chrome alloy (ASTM F1537 or ASTM F562). Components: Pedicle screws, rods, hooks, connectors, crossbars, set screws. Sterilization: Steam. Connectivity: None. Software: None.

Indications for Use

Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion for thoracic, lumbar, and sacral spine conditions including DDD, spondylolisthesis, trauma, spinal stenosis, deformities (scoliosis, kyphosis, lordosis), spinal tumor, pseudarthrosis, and failed previous fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ September 21, 2021 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below. SeaSpine Orthopedics Corporation Jesse Albright Senior Regulatory Affairs Specialist 5770 Armada Drive Carlsbad, California 92008 Re: K212692 Trade/Device Name: Mariner Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB, KWQ, KWP Dated: August 20, 2021 Received: August 25, 2021 Dear Jesse Albright: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for {2}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K212692 Device Name Mariner Pedicle Screw System Indications for Use (Describe) The intended use of the Mariner Pedicle Screw System in a posterior or anterolateral approach is to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. The indications for use are as follows: · degenerative disc disease (DDD) as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies, · spondylolisthesis, - · trauma (i.e., fracture or dislocation), · spinal stenosis, - · deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), - · spinal tumor, - pseudarthrosis, and/or - · failed previous fusion. Type of Use (Select one or both, as applicable) | <div> <span style="font-size: 10pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |----------------------------------------------------------------------------------------------------------| | <div> <span style="font-size: 10pt;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary #### Contact Details | Applicant Name: | SeaSpine Orthopedics Corporation | |------------------------------------------|---------------------------------------------------------------------------| | Address:<br>Phone number:<br>Fax number: | 5770 Armada Drive, Carlsbad, CA 92008<br>(760) 216-5176<br>(760) 683-6874 | | Contact Person: | Jesse Albright, Senior Regulatory Affairs Specialist | | Date Prepared: | August 20, 2021 | #### Device Name | Trade Name: | Mariner Pedicle Screw System | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | Thoracolumbosacral Pedicle Screw System<br>Spinal Intervertebral Body Fixation System<br>Spinal Interlaminal Fixation System | | Classification Name: | Thoracolumbosacral Pedicle Screw System (21 CFR 888.3070)<br>Appliance, Fixation, Spinal Intervertebral Body (21 CFR 888.3060)<br>Appliance, Fixation, Spinal Interlaminal (21 CFR 888.3050) | | Product Code(s): | NKB, KWQ, KWP | | Device Class: | 2 | #### Legally Marketed Predicate Devices | 510(k) Number | Product Code(s) | Trade Name | Manufacturer | |--------------------------------|-----------------|---------------------------------|-------------------------------------| | Primary Predicate Device | | | | | K173882 | NKB | Mariner Pedicle<br>Screw System | SeaSpine Orthopedics<br>Corporation | | Additional Predicate Device(s) | | | | | K211606 | NKB, KWP, KWQ | Mariner Deformity<br>System | SeaSpine Orthopedics<br>Corporation | #### Device Description The Mariner Pedicle Screw System is a non-cervical spinal fixation system used to build constructs within the body to act as temporary or permanent non-cervical spinal fixation devices and is intended for use as a posterior pedicle screw fixation system, a posterior non-pedicle fixation system, and/or an anterolateral fixation system to correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusion to occur. {5}------------------------------------------------ The Mariner Pedicle Screw System includes a variety of single-use implants manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136) and/or cobalt chrome alloy (Co-28Cr-6Mo per ASTM F1537 or Co-35Ni-20Cr-10Mo per ASTM F562) and is comprised of fixed, uni-planar, and polyaxial and motion-limiting modular pedicle screws, as well as connecting spinal rods, hooks, connectors, crossbars, and a separate set screw locking element. The instruments included in the Mariner Pedicle Screw System facilitate the placement, final locking, and removal, if necessary, of the system implants, and accessories to the system include trays and caddies for storage, protection, and organization prior to and during the steam sterilization process. The purpose of this submission is to add an additional set screw to the Mariner Pedicle Screw System. #### Intended Use/Indications for Use The intended use of the Mariner Pedicle Screw System in a posterior or anterolateral approach is to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. The indications for use are as follows: - . degenerative disc disease (DDD) as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies, - . spondylolisthesis, - trauma (i.e., fracture or dislocation), - spinal stenosis, - deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), - spinal tumor, - pseudarthrosis, and/or - failed previous fusion. #### Summary of Technological Characteristics The Mariner Pedicle Screw System is identical or similar to the cited predicate devices in regard to intended use/indications for use, device description, technological characteristics (e.g., operating principle, design, components, manufacturing, labeling, sterility, etc.), and non-clinical performance (i.e., mechanical testing). #### Non-Clinical Testing demonstrated The Mariner Pedicle Screw System substantially equivalent mechanical performance to the predicate devices through static and dynamic mechanical testing with reference to ASTM F1717 and ASTM F1798 and a risk analysis compared to the predicate devices. {6}------------------------------------------------ ## Clinical Testing Not applicable. The determination of substantial equivalence is not based on an assessment of clinical performance data. #### Conclusions The submitted data demonstrate that the Mariner Pedicle Screw System is substantially equivalent to the cited legally marketed predicates.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...