AliveCor QT Service

K212662 · AliveCor, Inc. · DQK · Apr 26, 2022 · Cardiovascular

Device Facts

Record IDK212662
Device NameAliveCor QT Service
ApplicantAliveCor, Inc.
Product CodeDQK · Cardiovascular
Decision DateApr 26, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1425
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
QT interval measurementDeep neural networks with adaptive filters and signal processingMean interval difference ≤ ± 20ms and standard deviation of interval difference ≤ 25ms—Training: >750K ECGs from >250K patients. Validation: Dataset 1 (>34K ECGs) and Dataset 2 (226 ECGs).———
Heart-rate corrected QT (QTc) interval measurementDeep neural networks with adaptive filters and signal processingMean interval difference ≤ ± 20ms and standard deviation of interval difference ≤ 25ms—Training: >750K ECGs from >250K patients. Validation: Dataset 1 (>34K ECGs) and Dataset 2 (226 ECGs).———

Indications for Use

AliveCor QT Service analyses 30 seconds of a previously acquired electrocardiogram (ECG) from AliveCor designed 6-Lead ambulatory ECG devices analyzed as normal sinus rhythm for QT interval measurements. AliveCor QT Service is intended for use in a professional medical facility, such as a hospital, clinic, or doctor's office by a qualified health care professional, including trained ECG technician. AliveCor QT Service is indicated for use on adult patients (older than 18 years). The device has not been tested for and is not intended for pediatric use. The service is not intended for use in life supporting, or sustaining systems, or continuous ECG monitors, or cardiac alarm devices, or OTC use only devices.

Device Story

Cloud-based SaMD; inputs 30-second ECG segments from AliveCor 6-Lead ambulatory devices (e.g., KardiaMobile 6L); integrates KardiaAI (K201985) to confirm Normal Sinus Rhythm; utilizes adaptive filters, signal processing, and deep neural networks (DNN) to compute QT and heart-rate corrected QTc intervals (Bazett/Fridericia formulas); accessed by healthcare professionals via REST-based API from MDDS or clinical software; output provides interval measurements to assist clinicians in cardiac assessment; intended for professional clinical settings; does not provide real-time monitoring or alarms.

Clinical Evidence

No clinical testing was required. Bench testing only: validated algorithm performance using >750K ECGs from >250K patients for training, and >34K ECGs for validation. Performance verified against expert-annotated 12-lead ECG reference intervals, achieving mean interval difference ≤ ±20ms and standard deviation ≤ 25ms.

Technological Characteristics

Cloud-based SaMD; utilizes adaptive filtering, signal processing, and deep neural networks (DNN). Computes QT and QTc (Bazett/Fridericia) intervals. Connectivity via REST-based API. Software-only form factor; no direct patient interface. Operates on previously recorded ECG data.

Indications for Use

Indicated for adult patients (older than 18 years) to analyze 30 seconds of previously acquired ECGs from AliveCor 6-Lead ambulatory devices, specifically those analyzed as normal sinus rhythm, to provide QT and QTc interval measurements. Not for pediatric use, life-supporting/sustaining systems, continuous monitoring, cardiac alarms, or OTC use.

Regulatory Classification

Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 11, 2022 AliveCor, INC. % Prabhu Raghavan Principal Consultant MDQR, LLC 189 N. Bernardo Ave. Suite 100 Mountain View, California 94043 Re: K212662 Trade/Device Name: AliveCor QT Service Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable diagnostic computer Regulatory Class: Class II Product Code: DQK, DPS Dear Prabhu Raghavan: The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated April 26, 2022. Specifically, FDA is updating this SE Letter (e.g., typo in correspondent address) as an administrative correction. Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Shih Kozen, OHT2: Office of Cardiovascular Devices, 301-796-5813, Jennifer.Shih(@fda.hhs.gov. Sincerely, # Jennifer W. Shih -S Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 26, 2022 AliveCor, Inc. % Prabhu Raghavan Principal Consultant MDQR, LLC 444 Castro Street, Suite 600 Mountain View, California 94041 Re: K212662 Trade/Device Name: AliveCor QT Service Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK, DPS Dated: March 25, 2022 Received: March 28, 2022 Dear Prabhu Raghavan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Jennifer W. Shih -S Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K212662 Device Name AliveCor QT Service Indications for Use (Describe) AliveCor QT Service analyses 30 seconds of a previously acquired electrocardiogram (ECG) from AliveCor designed 6-Lead ambulatory ECG devices analyzed as normal sinus rhythm for QT interval measurements. Alive or OT Service is intended for use in a professional medical facility, such as a hospital, clinic, or doctor's office by a qualified health care professional, including trained ECG technician. Alive Or OT Service is indicated for use on adult patients (older than 18 years). The device has not been tested for and is not intended for pediatric use. The service is not in life supporting, or sustaining systems, or continuous ECG monitors, or cardiac alarm devices, or OTC use only devices. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "AliveCor" in a teal color. The font is sans-serif and the letters are evenly spaced. There is a registered trademark symbol to the right of the word. ## 510(k) Summary for K212662 Prepared in accordance with the requirements of 21 CFR 807.92 #### Submitter Information [807.92(a)(1)] | Submitter/Applicant | AliveCor, Inc.<br>189 N. Bernardo Ave.<br>Mountain View, CA 94043<br>Phone: 650-396-8553<br>Fax: 650-282-7932 | |--------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Primary Contact Person | Prabhu Raghavan<br>Regulatory Consultant for AliveCor<br>MDQR, LLC<br>189 N. Bernardo Ave.<br>Mountain View, CA 94043<br>Phone: 408-316-5707<br>Fax: 650-282-7932<br>Email: prabhu@mdqr.solutions | | Submitter and<br>Secondary Contact Person | Susan Noriega<br>VP of Regulatory Affairs, Clinical Affairs and Quality<br>189 N. Bernardo Ave. Suite 100<br>Mountain View, CA 94043<br>Phone: 650-793-1966<br>Fax: 650-282-7932<br>Email: snoriega630@alivecor.com | | Date Prepared | April 25, 2022 | | Device Information [807.92(a)(2)] | | | Trade Name<br>Common Name<br>Classification<br>Device Class<br>Product Code<br>Subsequent Product Code | AliveCor QT Service<br>Programmable diagnostic computer<br>21 CFR§870.1425<br>II<br>DQK, Computer, Diagnostic, Programmable<br>DPS, Electrocardiograph | # Predicate Information [807.92(a)(3)] | Predicate | K170155, GE EK12 Algorithm | |------------------|----------------------------| | Reference Device | K201985, AliveCor KardiaAI | ## Device Description [807.92(a)(4)] AliveCor QT Service is a cloud-based Software as a Medical Device (SaMD) that is used to measure the QT and heart-rate corrected QT ("QTc") interval measurements from {6}------------------------------------------------ electrocardiograms (ECG) recorded from adult patients (older than 18 years) using AliveCordesigned 6-Lead ambulatory ECG devices. AliveCor QT Service provides QT and QTc interval measurements only on ECGs analyzed as Normal Sinus Rhythm by KardiaAI (K201985). AliveCor QT Service is a prescription (Rx) use only device intended for use by qualified healthcare professionals, including trained ECG technicians. Healthcare professionals can access AliveCor OT Service over the internet from patient management and ECG storage Medical Device Data Systems (MDDS), or from other prescription-use only medical data software devices. These devices provide a previously recorded ECG from an AliveCor-designed 6-Lead ambulatory ECG device, such as the KardiaMobile 6L (K210753), in a compatible format using the AliveCor QT Service's REST-based application program interface (API). REST or REpresentational State Transfer is a software methodology that defines rules for creating web services to access resources over Hypertext Transfer Protocol (HTTP). AliveCor QT Service responds to the analysis request with the following measurements: - QT interval, measured from the first 30 seconds of the ECG - Heart-rate corrected QT (QTc) interval based on the Bazett's formula and the Fridericia's ● formula AliveCor QT Service utilizes various internal algorithms including deep neural networks (DNN) to analyze an ECG to compute the QT interval. AliveCor QT Service also includes algorithms to compute the RR-interval, which is used to provide both the Bazett's and Fridericia corrected OTc intervals. These algorithms were trained and validated on datasets with ECGs from patients representative of the device's intended use. The training included more than 750K ECGs from over 250K patients. The training dataset included more than 200K ECGs with approximately 49% data from females, approximately 70% from subjects who were reported as whites, 3% nonwhites, and 27% from those who did not report their race The validation was conducted on two datasets. The first dataset included more than 34K ECGs with approximately 51% data from females, approximately 80% from subjects who were reported as whites, 3.6% non-whites, and 1 7% from those who did not report their race. The second dataset included 226 ECGs with approximately 60% data from females, approximately 87% from subjects who were reported as whites, 3.2% non-whites, and 10% from those who did not report their race. ### Indications for use [807.92(a)(5)] AliveCor QT Service analyses 30 seconds of a previously acquired electrocardiogram (ECG) from AliveCor designed 6-Lead ambulatory ECG devices analyzed as normal sinus rhythm for QT interval measurements. AliveCor OT Service is intended for use in a professional medical facility, such as a hospital, clinic, or doctor's office by a qualified health care professional, including trained ECG technician. AliveCor QT Service is indicated for use on adult patients (older than 18 years). The device has not been tested for and is not intended for pediatric use. The service is not intended for use in life supporting, or sustaining systems, or continuous ECG monitors, or cardiac alarm devices, or OTC use only devices. {7}------------------------------------------------ 510(k) Summary for K212662, AliveCor QT Service AliveCor, Inc. #### Substantial Equivalence The Substantial Equivalence utilizes one predicate device and one reference device, namely: - 1. Predicate: K170155, EK12 Algorithm, that analyzes ECGs to provide rhythm and measurements including QT interval measurements. - 2. Reference Device: K201985, KardiaAI, which is an API library that analyzes ECGs from AliveCor-designed ECG devices and provides rhythm analysis. The reference device, like the subject device is used to analyze ECGs recorded from AliveCordesigned ambulatory ECGs including the K210753, KardiaMobile 6L, an AliveCordesigned 6-Lead ambulatory ECG device that the subject device is compatible with. The subject device also incorporates this reference device internally to perform rhythm analysis to determine if the provided ECG is Normal Sinus Rhythm. AliveCor QT Service has the same intended use as the predicate device, K170155, GE EK12 Algorithm. Both devices analyze prerecorded multi-lead ECG signals and measure the QT and heart-rate corrected OT intervals. Both devices are intended to be used by qualified health care professional, including trained ECG technicians. The subject device additionally shares key safety and technological characteristics with the predicate device: both devices are SaMDs that may not be used for active patient monitoring and do not provide any time sensitive information, continuous display of information, alarms, or alerts intended to alert a caregiver to take an immediate clinical action. Both devices do not directly interface with the patient, and both are provided digital ECG samples and output digital information containing the requested analysis. Both devices are provided in the form of a service. The predicate is additionally provided as a compiler linked library and as a command-line interface application. The subject device utilizes a mix of algorithms that utilize adaptive filters, and features extracted using signal processing to compute a representative ECG beat, which is then used by a deep neural network to compute the QT interval measurement1. The predicate computes a representative beat and uses ECG signal processing, pattern recognition, regression methods, as well as modeling of cardiac action potentials to establish the key time points in the overall ECG and thus compute the QT interval2. Both devices account for poor T-waves in the ECG. Any differences between the technological characteristics of the interval measurement algorithms have been evaluated through performance testing and do not raise different questions of safety or effectiveness. Differences between the subject device and the predicate have been tested to ensure that the device meets its intended use. Therefore, AliveCor QT Service is substantially equivalent to the predicate device. <sup>1</sup> Giudicessi, Schram, et al. Artificial Intelligence-Enabled Assessment of the Heart Rate Corrected QT Interval Using a Mobile Electrocardiogram Device. Circulation. 2021 Mar 30;143(13):1274-1286. doi: 10.1161/CIRCULATIONAHA.120.050231. Epub 2021 Feb 1. <sup>2</sup> Xue J. Q. (2009). Robust QT interval estimation--from algorithm to validation. Annals of nominvasive electrocardiology : the official journal of the International Society for Holter and Noninvasive Electrocardiology, Inc, 14 Suppl 1(Suppl 1), S35-S41. https://doi.org/10.1111/j.1542-474X.2008.00264.x {8}------------------------------------------------ | Feature | AliveCor QT Service<br>(Subject Device) | K170155 EK12<br>Algorithm<br>(Predicate Device) | K201985 KardiaAI<br>(Reference Device) | |----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Code | DQK, Computer,<br>Diagnostic, Programmable<br>DPS, Electrocardiograph | MHX, Monitor,<br>Physiological, Patient<br>(With Arrhythmia<br>Detection or<br>Alarms) | DQK, Computer,<br>Diagnostic, Programmable<br>DPS, Electrocardiograph | | Regulation | 21 CFR§870.1425,<br>Programmable diagnostic<br>computer | 21 CFR 870.1025,<br>Arrhythmia detector and<br>alarm | 21 CFR§870.1425,<br>Programmable diagnostic<br>computer | | Classifi-cation | II | II | II | | Indications for<br>Use | AliveCor QT Service<br>analyses 30 seconds of a<br>previously acquired<br>electrocardiogram (ECG)<br>from AliveCor designed 6-<br>Lead ambulatory ECG<br>devices analyzed as normal<br>sinus rhythm for QT<br>interval measurements.<br>AliveCor QT Service is<br>intended for use in a<br>professional medical<br>facility, such as a hospital,<br>clinic, or doctor's office by<br>a qualified health care<br>professional, including<br>trained ECG technician.<br>AliveCor QT Service is<br>indicated for use on adult<br>patients (older than 18<br>years). The device has not<br>been tested for and is not<br>intended for pediatric use.<br>The service is not intended<br>for use in life supporting,<br>or sustaining systems, or | EK12 Algorithm analyzes<br>ten or more seconds of a<br>previously acquired<br>electrocardiogram (ECG)<br>from physiological ECG<br>signal recording devices<br>for rhythm and<br>measurements. EK12<br>Algorithm is used to create<br>reports intended for use by<br>a Qualified Medical<br>Professional, including a<br>trained ECG Technician<br>operating within<br>Independent Diagnostic<br>Testing Facility (IDTF)<br>requirements and<br>performance standards for<br>the review and assessment<br>of an ECG.<br>EK12 Algorithm is<br>indicated for use on adults<br>and pediatric patients older<br>than 2 years.<br>The device is intended for<br>use in an IDTF or a | KardiaAI is a software<br>analysis library intended to<br>assess ambulatory<br>electrocardiogram (ECG)<br>rhythms from adult<br>subjects (when prescribed<br>or used under the care of a<br>physician). The device<br>supports analyzing data<br>recorded in compatible<br>formats from any<br>ambulatory ECG devices<br>such as event recorders, or<br>other similar devices. The<br>library is intended to be<br>integrated into other device<br>software. The library is<br>not intended for use in life<br>supporting, or sustaining<br>systems, or ECG monitors,<br>or cardiac alarm devices,<br>or OTC use only devices.<br>The KardiaAI library<br>provides the following<br>capabilities:<br>●<br>Filtering ECG noise, | | Feature | AliveCor QT Service<br>(Subject Device) | K170155 EK12<br>Algorithm<br>(Predicate Device) | K201985 KardiaAI<br>(Reference Device) | | | continuous ECG monitors,<br>or cardiac alarm devices,<br>or OTC use only devices. | professional medical<br>facility, such as a hospital,<br>clinic, or doctor's office. | Reporting heart rate measurement from ECGs,Detecting noisy ECGs,Reporting ECG rhythm analysis for the presence of sinus rhythm, atrial fibrillation, bradycardia, and tachycardia; for ECGs detected as sinus rhythm, detecting normal sinus rhythm, sinus rhythm with wide QRS, sinus rhythm with premature ventricular contractions (PVC), and sinus rhythm with supraventricular ectopy,Detecting QRS complexes in an ECG,For ECGs detected as sinus rhythm, classifying individual beats as a PVC or non-PVC beat, andGenerating an average beat from an ECGThe device is not intended for use in patients who have pacemakers, ICDs, or other implanted electronic devices | | Target<br>population | Adults (over 18) | Adults and pediatric<br>patients older than 2 years | Adults (over 18) | | Feature | AliveCor QT Service<br>(Subject Device) | K170155 EK12<br>Algorithm<br>(Predicate Device) | K201985 KardiaAI<br>(Reference Device) | | Indicated Use<br>Method | Prescription Use | Prescription Use | Prescription Use | | Form Factor | Software only | Software only | Software only | | Compatible<br>ECG Leads | Lead-I and Lead-II from<br>AliveCor-designed six-lead<br>ambulatory ECG | 12-lead or any subset of<br>12-leads | Lead-I from AliveCor-<br>designed ambulatory ECG | | Technology | Software algorithm | Software algorithm | Software algorithm | | ECG Analysis | QT and heart-rate<br>corrected QT intervals | Rhythm and measurement<br>analysis including QT and<br>heart-rate corrected QTc<br>intervals | Rhythm analysis | | Technological<br>characteristics | Adaptive filtering, signal<br>processing, deep neural<br>networks | Adaptive filtering, signal<br>processing, regression<br>models | Adaptive filtering, signal<br>processing, deep neural<br>networks | ## Comparison of Technological Characteristics with the Predicate Device [807.92(a)(6)] {9}------------------------------------------------ # Comparison of Technological Characteristics with the Predicate Device [807.92(a)(6)] {10}------------------------------------------------ # Comparison of Technological Characteristics with the Predicate Device [807.92(a)(6)] ### Performance Data [807.92(b)] All necessary testing was conducted on AliveCor QT Service to support a determination of substantial equivalence to the predicate device. ### Nonclinical Testing Summary [807.92(b)(1)] AliveCor conducted bench performance testing of the subject device to validate that the device meets its performance specifications and to demonstrate that, in light of the technological differences between AliveCor OT Service and its predicate, the subject device is at least as safe and as effective as the predicate device for the same intended use. Nonclinical testing was conducted to assess algorithm performance and to verify that the AliveCor QT Service estimated QT and heart-rate corrected QT intervals have a mean interval difference ≤ ± 20ms and a standard deviation of interval difference ≤ 25ms when compared to an expert annotated reference interval from a supine. diagnostic bandwidth, 12-lead ECG. Algorithm performance was evaluated using standards-based and AliveCor proprietary ECG databases. The databases were selected or developed to demonstrate that the QT interval measurement is accurate and can be used to measure the interval on ECGs recorded from AliveCor designed six-lead ECG devices by adults. Additionally, the subject device and the predicate were comparatively evaluated against appropriate databases to demonstrate substantially equivalent performance. {11}------------------------------------------------ #### 510(k) Summary for K212662, AliveCor QT Service AliveCor, Inc. In summary, the results of the nonclinical testing demonstrate that AliveCor QT Service performs to its specifications, meets its intended use with substantially equivalent performance to that of the predicate device and does not raise different questions of safety or effectiveness. #### Clinical Testing Summary [807.92(b)(2)] No clinical testing was required to establish substantial equivalence. #### Conclusions [807.92(b)(3)] AliveCor QT Service has the same intended use as the predicate device, and the labeling differences do not raise different questions of safety or effectiveness. Therefore, AliveCor QT Service is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...