Gas-Chex Indicator

K212592 · Propper Manufacturing Co., Inc. · JOJ · Jan 20, 2022 · General Hospital

Device Facts

Record IDK212592
Device NameGas-Chex Indicator
ApplicantPropper Manufacturing Co., Inc.
Product CodeJOJ · General Hospital
Decision DateJan 20, 2022
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.2800
Device ClassClass 2

Intended Use

The Gas-Chex Indicator is an integrating indicator designed to respond to all critical parameters over a specified range of Ethylene Oxide sterilization cycles. The Gas-Chex Indicator is intended to be placed in each pack, pouch, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles: 1. 37°C, 736 mg/l EO, ≥35% RH, 3 hours exposure 2. 37°C, 759 mg/l EO, ≥35% RH, 3 hours exposure 3. 38°C, 736 mg/l EO, 40-80% RH, 4.5 hours exposure 4. 38°C, 759 mg/l EO, 40-80% RH, 4.5 hours exposure 5. 55°C, 736 mg/l EO, ≥35% RH, 1 hour exposure 6. 55°C, 759 mg/l EO, ≥35% RH, 1 hour exposure 7. 55°C. 600 mg/l EO. 60% RH. 4 hours exposure

Device Story

Single-use chemical integrating indicator; consists of reactive EO-sensitive ink printed on paper substrate strip. Placed inside sterilization containment (pack, pouch, tray) by healthcare personnel. During EO sterilization, ink undergoes chemical reaction; color shifts from brown-red to green upon reaching critical parameters (time, temperature, humidity, gas concentration). If parameters are not met, indicator remains brown-red or brown-yellow. Provides visual confirmation of sterilization cycle adequacy; assists healthcare providers in verifying that individual items have been exposed to required sterilization conditions. Benefits patient by ensuring medical devices are properly processed before use.

Clinical Evidence

Bench testing only. Performance validated in EO BIER vessel against FDA guidance for chemical indicators. Testing confirmed color change response to critical parameters, stability of end-point color, and shelf-life (36 months). Biocompatibility and ink transfer testing performed per ISO 11140-1:2014. No clinical data.

Technological Characteristics

Single-use paper strip substrate with reactive chemical indicator ink. Dimensions: 4" x 9/16" or 8" x 9/16". Principle: Chemical reaction to EO gas (time, temperature, humidity, concentration). Shelf life: 36 months. No energy source, software, or connectivity.

Indications for Use

Indicated for use as an independent monitor of critical parameters (time, temperature, relative humidity, and EO gas concentration) within packs, pouches, trays, or containment devices during Ethylene Oxide sterilization cycles.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. January 20, 2022 Propper Manufacturing Co., Inc. Andrew Sharavara Chief Technical Officer 36-04 Skillman Avenue Long Island City, New York 11101 Re: K212592 Trade/Device Name: Gas-Chex Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: December 20, 2021 Received: December 23, 2021 #### Dear Andrew Sharavara: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number: K212592 Device Name Gas-Chex Indicator Indications for Use (Describe) The Gas-Chex Indicator is an integrating indicator designed to respond to all critical parameters over a specified range of Ethylene Oxide sterilization cycles. The Gas-Chex Indicator is intended to be placed in each pack, pouch, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles: - 1. 37°C, 736 mg/l EO, ≥35% RH, 3 hours exposure 2. 37°C, 759 mg/l EO, ≥35% RH, 3 hours exposure 3. 38°C, 736 mg/l EO, 40-80% RH, 4.5 hours exposure 4. 38°C, 759 mg/l EO, 40-80% RH, 4.5 hours exposure 5. 55°C, 736 mg/l EO, ≥35% RH, 1 hour exposure 6. 55°C, 759 mg/l EO, ≥35% RH, 1 hour exposure 7. 55°C. 600 mg/l EO. 60% RH. 4 hours exposure The Gas-Chex Indicator for EO sterilization has the following minimum Stated Values determined in a Resistometer: 54℃, 42 min, EO 600 mg/l, RH 60±10% 37℃,75 min. EO 600 mg/l. RH 60±10% 55°C, 36 min, EO 736 mg/l, RH 50±10% 37°C, 75 min, EO 736 mg/l, RH 50±10% 55°C, 35 min, EO 759 mg/l, RH 50±10% 37°C, 75 min, EO 759 mg/l, RH 50±10% | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| | | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summarv K212592 | Submitted by: | Propper Manufacturing Company, Inc. | |-----------------|------------------------------------------------------------| | Address: | 36-04 Skillman Avenue,<br>Long Island City, New York 11101 | | Contact Name: | Andrew Sharavara, Ph.D., Chief Technical Officer | | Telephone: | (800) 832-4300 x149 | | Fax: | (718) 482-8909 | | E-mail: | as@proppermfg.com | | Date Submitted: | August 12, 2021 | #### Device information: | Device Trade Name: | Gas-Chex® Indicator | |----------------------|---------------------------------------------------| | Classification Name: | Physical/Chemical Sterilization Process Indicator | | Common Name: | Ethylene Oxide Gas Sterilization Indicator | | Product Code: | JOJ | | Classification: | Class II (21 C.F.R. 880.2800) | | 510k number | K212592 | ## Description of the Device The Gas-Chex® Indicator is a single use chemical integrating indicator designed for Ethylene Oxide (EO) sterilization monitoring. Each indicator consists of reactive EO indicator ink printed on a 4" x 9/16" substrate paper strip. It can be also printed on other substrate sizes, for example 8" x 9/16" paper. Gas-Chex® Indicators are sold in boxes of 250 strips. The indicator responds to all critical parameters of an EO sterilization cycle: exposure time, temperature, relative humidity and amount of Ethylene Oxide gas. During sterilization process the indicator ink chemicals react to EO gas forming a green compound. The degree of the reaction depends on the sterilization exposure. When the parameters achieve required level, the indicator ink chemistry changes color from brown-red to green. If the parameters do not achieve the required level, the indicator color will be brown-red, or brown-yellow. ## Indications for Use The Gas-Chex® Indicator is an integrating indicator designed to all critical parameters over a specified range of Ethylene Oxide sterilization cycles. The Gas-Chex® Indicator is intended to be placed in each pack, pouch, tray, or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles: {4}------------------------------------------------ 1. 37°C, 736 mg/l EO, ≥35% RH, 3 hours exposure 2. 37°C, 759 mg/l EO, ≥35% RH, 3 hours exposure 3. 38°C, 736 mg/1 EO, 40-80% RH, 4.5 hours exposure 4. 38°C, 759 mg/1 EO, 40-80% RH, 4.5 hours exposure 5. 55°C. 736 mg/l EO. >35% RH. 1 hour exposure 6. 55°C, 759 mg/l EO, ≥35% RH, 1 hour exposure 7. 55°C, 600 mg/l EO, 60% RH, 4 hours exposure The Gas-Chex® Indicator has the following minimum Stated Values determined in a Resistometer: 54℃, 42 min, EO 600 mg/1, RH 60±10% 37℃,75 min, EO 600 mg/1, RH 60±10% 55℃,36 min, EO 736 mg/1, RH 50±10% 37℃,73 min, EO 736 mg/1, RH 50±10% 55℃, 35 min, EO 759 mg/l, RH 50±10% 37℃, 73 min, EO 759 mg/1, RH 50±10% # Performance The performance of the Gas-Chex® Indicator complies with FDA guidance for industry and FDA Staff: Pre-market notification [510(k)] submissions for chemical indicators, 2003 for integrating indicators. ## Technological Characteristics Comparison to Legally Marketed Predicate Device Comparison of the subject device (Gas-Chex Indicator, Propper Manufacturing Co., Inc) to Predicate device (Integron IT12, k191021, by Terragene S.A.). | | Subject device | Predicate device | Comparison | design | indicator ink | | | |-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|---------------------------------------------------------------|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product name | Gas-Chex Indicator | Integron IT12 integrator,<br>k191021 | | Initial color | Brown-red | Purple | Similar | | Product<br>generic name | A physical/chemical sterilization<br>process indicator | A physical/chemical<br>sterilization process indicator | Identical | End point color | Green | Green | Identical | | Product code | JOJ | JOJ | Identical | Indicator category | Integrating indicator | ISO Integrator | Similar. Both indicators are designed to respond to all critical parameters over a specified range. | | Sterilization<br>method | Ethylene Oxide gas sterilization | Ethylene Oxide gas<br>sterilization | Identical | Single use | Yes | Yes | Identical | | Intended use | Sterilization process indicator | Sterilization process indicator | Identical | Shelf life | 36 months | 36 months | Identical | | Sterilization<br>cycles | 37°C,736 mg/l EO, ≥35% RH, 3hrs<br>37°C,759 mg/l EO, ≥35% RH, 3hrs<br>38°C,736 mg/l EO, 40-80% RH, 4.5 hrs<br>38°C,759 mg/l EO, 40-80% RH, 4.5 hrs<br>55°C,736 mg/l EO, ≥35% RH, 1 hr<br>55°C,759 mg/l EO, ≥35% RH, 1 hr<br>55°C,600 mg/l EO, 60% RH, 4 hrs | 55°C, 600 mg/l EO, 60% RH,<br>240min | Identical with<br>additional cycles for<br>Gas-Chex Indicator | Indications for use | The Gas-Chex Indicator is designed to respond to all critical parameters over a specified range of Ethylene Oxide sterilization cycles. The Gas-Chex Indicator is intended to be placed in each pack, pouch, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:<br>37°C,736 mg/l EO, ≥35% RH, 3hrs<br>37°C,759 mg/l EO, ≥35% RH, 3hrs<br>38°C,736 mg/l EO, 40-80% RH, 4.5 hrs<br>38°C,759 mg/l EO, 40-80% RH, 4.5 hrs<br>55°C,736 mg/l EO, ≥35% RH, 1 hr<br>55°C,759 mg/l EO, ≥35% RH, 1 hr<br>55°C,600 mg/l EO, 60% RH, 4 hrs<br><br>The Gas-Chex Indicator for EO sterilization has the following minimum Stated Values determined in a Resistometer:<br>54°C, 42 min, EO 600 mg/l, RH 60±10%<br>37°C, 75 min, EO 600 mg/l, RH 60±10%<br>55°C, 36 min, EO 736 mg/l, RH 50±10%<br>37°C, 73 min, EO 736 mg/l, RH 50±10%<br>55°C, 35 min, EO 759 mg/l, RH 50±10%<br>37°C, 73 min, EO 759 mg/l, RH 50±10% | The integrator Terragene Integron® IT12 is designed to chemically react over time with the critical parameters of ethylene oxide sterilization cycle within a specified tolerance. The integrating indicator strip is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:<br>55°C for 240 minutes, 600 mg/l, RH 60 %.<br><br>SV 37°C/75 min,<br>SV 54°C/30 min,<br>SV 55°C/28 min. | Similar. Gas-Chex indicator can be used in several additional cycles.<br><br>Stated value at 37°C is identical.<br><br>Gas-Chex indicator has longer stated value time in BIER vessel at 54°C which is within typical performance range and conforms to the requirement not to achieve end point before biological indicator spore death. | | Device | Paper strip printed with indicator ink | Paper strip printed with | Identical | | | | | {5}------------------------------------------------ Propper BI-OK® EO Gas Biological Test Pack, K972747, was used as a reference device for stability and shelf-life review because this reference device includes EO sterilization record card – a paper sheet printed with EO indicator ink very similar to the ink of Gas-Chex Indicator. Terragene ChemDye CD16 EO Multivariable process indicator, k191021, was used as a reference device for color change comparison. {6}------------------------------------------------ # Summary of Non-Clinical Testing Provided below is the summary of non-clinical testing that was performed to demonstrate that the subject device met the acceptance criteria for each standard or test method. | Test | Purpose | Acceptance Criteria | Result | |-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | Performance<br>testing in EO<br>BIER vessel | Demonstrate conformance of<br>Gas-Chex Indicator with the<br>FDA guidance for industry and<br>FDA Staff: Pre-market<br>notification [510(k)]<br>submissions for chemical<br>indicators, 2003 for integrating<br>indicators. | Color changes:<br>54°C, 42 min, EO 600 mg/l, RH 60%: green<br>49°C, 33.6min, EO 510 mg/l, RH 60%:<br>brown, brown-red, brown-yellow<br>37°C, 75 min, EO 600 mg/l, RH 60%: green<br>32°C, 60 min, EO 510 mg/l, RH 60%:<br>brown, brown-red, brown-yellow<br>55°C, 36 min, EO 736 mg/l, RH 50%: green<br>50°C, 28.8min, EO 625mg/l, RH 50%:<br>brown, brown-red, brown-yellow<br>37°C, 73 min, EO 736 mg/l, RH 50%: green<br>32°C, 58.4min, EO 625mg/l, RH 50%:<br>brown, brown-red, brown-yellow<br>55°C, 35 min, EO 759 mg/l, RH 50%: green<br>50°C, 28 min, EO 645 mg/l, RH 50%:<br>brown, brown-red, brown-yellow<br>37°C, 73 min, EO 759 mg/l, RH 50%: green<br>32°C, 58.4min, EO 645mg/l, RH 50%:<br>brown, brown-red, brown-yellow | Passed | | Testing in "No<br>EO Gas" cycles | Confirm that Gas-Chex<br>indicator does not change color<br>in absence of Ethylene Oxide<br>gas | When tested in cycle with temperature<br>60°C±2°C, RH≥85%, time 90 min±1min, the<br>indicator should not achieve end point color. | Passed | | Testing against<br>biological<br>indicator | Establish correlation between<br>performance of Gas-Chex<br>indicator and EO Biological<br>indicator | The integrator does not achieve end-point<br>color before the biological indicator is<br>inactivated and demonstrates parallel<br>performance to the biological indicator. | Passed | | Single parameter<br>variation testing | Confirm that Gas-Chex<br>indicator is sensitive to critical<br>sterilization parameters. | Variation of one parameter while other ones<br>are maintained steady. Gas-Chex Indicator<br>should not reach specified end-point green<br>color. | Passed | | Testing in cycles<br>with parameters<br>typical for<br>healthcare | Demonstrate Gas-Chex<br>indicator achieves specified end<br>color in typical cycles. | Color change from brown-red to green. | Passed | | Biocompatibility<br>study and ink<br>transfer test | Demonstrate that the indicator<br>does not create biocompatibility<br>issues to health care<br>professionals and patients. | Evaluation of individual components for<br>biocompatibility and review of<br>biocompatibility of indicators with similar<br>formulation with history on the market.<br><br>Testing according to ISO 11140-1:2014.<br>Requirement: 6.2.2. No ink transfer should be<br>observed on unprocessed and EO processed<br>samples. | Passed | | End-point<br>stability and<br>shelf- life study | Confirm that Gas-Chex<br>indicator has acceptable<br>stability after processing when<br>achieved and not achieved end<br>point color ("Pass" and "Fail"<br>conditions).<br><br>Demonstrate that Gas-Chex<br>indicator maintains its<br>performance when tested using<br>real-time shelf-life exposure<br>method. | Gas-Chex indicators processed in Pass and<br>Fail cycles at various time points after<br>production and at the end of shelf life should<br>demonstrate stable color for at least 3 months.<br><br>Meet specifications after real-time 36 months<br>shelf-life exposure. | Passed | {7}------------------------------------------------ # Conclusion The conclusion drawn from the nonclinical test demonstrate that the Gas-Chex Indicator is as safe, as effective, and performs as well as or better that the legally marketed predicate device, K191021.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...