Clinical validation study conducted in a NICU setting to evaluate the form, fit, and function of the OxySoftN sensor in the intended neonatal population.
NICU; Neonatal; Clinical validation; Form and fit
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
MDT20002VALNIC; Clinical validation study
Neonates (ages 1 to 9 days, weight 640 to 4710 grams); Sample Size: 27 patients for accuracy; 31 patients for form and fit
SaO2 values obtained from CO-oximeter analysis
SpO2 accuracy, form and fit
Indications for Use
The Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, model OxySoftN, is intended for use in non-invasive continuous or spot-check monitoring of functional oxygen saturation of arterial oxygen saturation (SpO2) and pulse rate for neonates and infants weighing less than 3 kg or adults and pediatrics weighing more than 40 kg.
Device Story
The Nellcor OxySoftN is a single-patient-use, non-invasive pulse oximetry sensor. It utilizes two light-emitting diodes (LEDs) and a photodiode to measure light absorbance across a pulsating arteriolar vascular bed (finger, foot, or hand). The sensor transforms the time-varying absorbance of tissue into a signal that is amplified and digitized by a compatible Nellcor pulse oximetry monitor to calculate functional arterial oxygen saturation (SpO2) and pulse rate. It is used by healthcare professionals or lay operators (in home settings) for continuous or spot-check monitoring. The device provides real-time SpO2 and pulse rate data to clinicians, facilitating clinical decision-making regarding patient oxygenation status. The sensor is provided sterile and is designed for single use, featuring an adhesive interface for secure attachment.
Clinical Evidence
Clinical validation included multiple studies: MDT20002VALNIC (NICU neonates, n=27, SpO2 accuracy 2.68%); MDT20037CYBOXY (elderly subjects, demonstrating gentler adhesive vs. predicate); MDT20047OXSNOF (sensor-off feature validation, 99.6% success rate); MDT20038CYBADH (repositioning adhesion study); MDT20028OXYLOV (healthy adults, SpO2 accuracy ±3% at 60-80% saturation, ±2% at 70-100%); and MDT20006OXYVMT (motion study, SpO2 accuracy ±3% at 70-100% saturation).
Technological Characteristics
Transmittance pulse oximetry sensor using dual-wavelength LEDs and photodiode. Materials include polyurethane, silicone, synthetic fabric, aluminum polypropylene, synthetic rubber, and PVC. Powered by compatible host monitors. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-1-12, and ISO 80601-2-61. Sterilized via EtO. Connectivity via Nellcor-compatible sensor cable.
Indications for Use
Indicated for neonatal, infant, pediatric, and adult patients, including those well or poorly perfused, during motion and no-motion conditions. Used in hospitals, hospital-type facilities, intra-hospital transport, mobile emergency medical applications (ground/air transport), and home environments. Contraindicated for patients with allergies to synthetic fabrics or silicone.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of the symbol, there is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in a larger font and "ADMINISTRATION" in a smaller font below it.
Covidien llc Greeshma Kayala Regulatory Affairs Manager 6135 Gunbarrel Avenue Boulder, Colorado 80301
### Re: K212555
Trade/Device Name: Nellcor OxySoft Neonatal-Adult SpO2 Sensor, OxySoftN, Nellcor OxySoft Neonatal-Adult SpO2 Sensor, OxySoftNHC Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: April 1, 2022 Received: April 4, 2022
### Dear Greeshma Kayala:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtnev Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K212555
#### Device Name
Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, OxySoftN/OxySoftNHC
Indications for Use (Describe)
The Nellcor™ OxySoftN single patient use sensor is indicated for use with neonatal, infant, pediatric and adult patients during both no motion and motion conditions and for patients who are either well or poorly perfused, in hospitals, hospital-type facilities, intra-hospit, mobile emergency medical applications including both ground and air transport and home environments. Transport environments include intra-hospital transport and air emergency transport: road ambulances and fixed-wing aircraft and helicopters. The Nellcor™ OxySoft SpO2 Sensor works in conjunction with monitoring systems that use OxiMax™ and Nellcor compatible pulse oximetry systems to facilitate spot checking or continuous monitoring of a patient's arterial hemoglobin (SpO2) and pulse rate via topical application of the sensor over a pulsating arteriolar vascular bed, such as a finger, foot or hand. This device is for prescription use only.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <b> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | <div> <span> <b> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </b> </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) Summary
### SUBMITTER INFORMATION
#### Submitted By:
Covidien, Ilc 6135 Gunbarrel Avenue Boulder, CO 80301
Date Prepared: May 2, 2022
#### Contact Person:
Official Correspondent: Elli Marrs, Regulatory Affairs Specialist
#### Secondary Correspondent: Greeshma Kayala, Sr. Regulatory Affairs Manager
Phone: (303) 579-1718 Email: Greeshma.kayala@medtronic.com
### DEVICE NAME
| Trade Name: | Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor OxySoftN<br>Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor OxySoftNHC |
|----------------------------|------------------------------------------------------------------------------------------------------------------|
| Common Name: | Oximeter |
| Classification Regulation: | 21 CFR 870.2700 |
| Classification Name: | Oximeter |
| Regulatory Class: | Class II |
| Product Code: | DQA |
| Review Panel: | Anesthesiology |
#### PREDICATE DEVICE
| Predicate Manufacturer: | Covidien |
|-------------------------|--------------------------------------------|
| Predicate Trade Name: | NellcorTM Neonatal-Adult SpO2 Sensor, MAXN |
| Predicate 510(k): | K012891 |
| Clearance Date: | March 7, 2002 |
No reference devices were used in this submission.
### DEVICE DESCRIPTION
The Nellcor™ OxySoft™ Neonatal/Adult SpO2 Sensor, OxySoftNHC monitors the amount of oxygen in the patient's blood (oxygen saturation) or SpO2 and pulse rate after applying the sensor on the finger, foot, or hand. The sensor utilizes a pair of light emitting diodes (LEDs) employing light at two wavelengths, and the time varying absorbance of the
Covidien IIc
{5}------------------------------------------------
tissue due to the pulsatile blood signal is obtained, amplified from a photodiode to determine functional arterial oxygen saturation.
A clinician will prescribe the sensor for use as spot check or continuous use in hospitals, hospital-type facilities, intra-hospital transport, mobile emergency medical applications including both ground and air transport and home environments. The OxySoftNHC sensor can be used with all Nellcor™ compatible systems.
The sensor consists of a sensor assembly and cable assembly and is provided sterile (EtO) and is meant for single use.
NOTE: The two model numbers represent only one sensor type, OxySoftN. The 'HC' suffix denotes the model number that includes an additional Home Use Guide for the lay user for home use.
### INTENDED USE AND INDICATIONS FOR USE
### Intended Use:
The Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, model OxySoftN, is intended for use in non-invasive continuous or spot-check monitoring of functional oxygen saturation of arterial oxygen saturation (SpO2) and pulse rate for neonates and infants weighing less than 3 kg or adults and pediatrics weighing more than 40 kg.
### Indications for Use:
The Nellcor™ OxySoftN single patient use sensor is indicated for use with neonatal, infant, pediatric and adult patients during both no motion conditions and for patients who are either well or poorly perfused, in hospital-type facilities, intra-hospital transport, mobile emergency medical applications including both ground and air transport and home environments. Transport environments include intra-hospital transport and both ground and air emergency transport: road ambulances and fixed-wing aircraft and helicopters. The Nellcor™ OxySoft™ SpO2 Sensor works in conjunction with monitoring systems that use OxiMax™ and Nellcor compatible pulse oximetry systems to facilitate spot checking or continuous monitoring of a patient's arterial hemoglobin (SpO2) and pulse rate via topical application of the sensor over a pulsating arteriolar vascular bed, such as a finger, foot or hand. This device is for prescription use only.
### CONTRAINDICATIONS
The OxySoftN sensor is contraindicated for use on patients who exhibit allergic reactions to synthetic fabrics or silicone.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, models OxySoftN and OxysoftNHC are designed for the same functionality as the predicate device sensors, Nellcor™ Neonatal-Adult SpO2 Sensor, MAXN, measuring functional oxygen saturation.
The following technological characteristics were compared between the subject device and predicate device to demonstrate substantial equivalence in Table 1 below.
{6}------------------------------------------------
### Table 1. Comparison of Technological Characteristics
| Characteristic | Subject Device | Predicate Device | Comments |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Nellcor™ OxySoft™<br>Neonatal-Adult SpO2<br>Sensor, OxySoftN and<br>OxySoftNHC | Nellcor™ Neonatal-Adult<br>SpO2 Sensor, MAXN<br>(K012891) | |
| Device | Oximeter | Oximeter | Same |
| Product Code | DQA | DQA | Same |
| Device Class | II | II | Same |
| Indications for<br>Use | The Nellcor™ OxySoftN<br>single patient use sensor is<br>indicated for use with<br>neonatal, infant, pediatric<br>and adult patients during<br>both no motion and motion<br>conditions and for patients<br>who are either well or poorly<br>perfused, in hospitals,<br>hospital-type facilities, intra-<br>hospital transport, mobile<br>emergency medical<br>applications including both<br>ground and air transport<br>and home environments.<br>Transport environments<br>include intra-hospital<br>transport and both ground<br>and air emergency<br>transport: road ambulances<br>and fixed-wing aircraft and<br>helicopters. The Nellcor™<br>OxySoft SpO2 Sensor<br>works in conjunction with<br>monitoring systems that use<br>OxiMax™ and Nellcor<br>compatible pulse oximetry<br>systems to facilitate spot<br>checking or continuous<br>monitoring of a patient's<br>arterial hemoglobin (SpO2)<br>and pulse rate via topical<br>application of the sensor<br>over a pulsating arteriolar<br>vascular bed, such as a<br>finger, foot or hand. This<br>device is for prescription<br>use only. | The OxiMax™ single<br>patient use sensors (which<br>includes the MaxN) is<br>indicated for use with<br>neonatal, infant, pediatric,<br>and adult patients during<br>both no motion and motion<br>conditions and for patients<br>who are either well or poorly<br>perfused, in hospitals,<br>hospital-type facilities, intra-<br>hospital transport, and<br>home environments. These<br>devices are for prescription<br>use only. | Different:<br>Indications for Use<br>were updated to<br>include clarity of<br>systems OxySoftN<br>is compatible with.<br>Additionally<br>environments for<br>use were expanded<br>to include home<br>use environments<br>and mobile<br>emergency medical<br>applications,<br>including both<br>ground and air<br>transport,<br>environments<br>which is supported<br>by compliance to<br>IEC 60601-1-<br>11:2015 and IEC<br>60601-1-12:2014. |
| Characteristic | Subject Device | Predicate Device | Comments |
| | Nellcor™ OxySoft™<br>Neonatal-Adult SpO2<br>Sensor, OxySoftN and<br>OxySoftNHC | Nellcor™ Neonatal-Adult<br>SpO2 Sensor, MAXN<br>(K012891) | |
| Purpose and<br>Function | Non-invasive continuous or<br>spot-check monitoring of<br>functional oxygen saturation<br>of arterial oxygen saturation<br>(SpO2) and pulse rate | Non-invasive continuous<br>monitoring of functional<br>oxygen saturation of arterial<br>oxygen saturation (SpO2)<br>and pulse rate | Same |
| Target<br>Population | Neonatal, infant, pediatric,<br>and adult patients | Neonatal, infant, pediatric,<br>and adult patients | Same |
| Intended User | Professionally trained<br>health care providers,<br>patients or a lay operator<br>assisting a patient in a<br>home-care environment | Professionally trained<br>health care providers,<br>patients or a lay operator<br>assisting a patient in a<br>home-care environment | Same |
| Where Used | In hospitals, hospital-type<br>facilities, intra-hospital<br>transport, mobile<br>emergency medical<br>applications including both<br>ground and air transport<br>and home environments.<br>Transport environments<br>include intra-hospital<br>transport and both ground<br>and air emergency<br>transport: road ambulances<br>and fixed-wing aircraft and<br>helicopters. | In hospitals, hospital-type<br>facilities, intra-hospital<br>transport, and home<br>environments. | Different:<br>Environments for<br>use were expanded<br>to include home<br>use environments<br>and mobile<br>emergency medical<br>applications,<br>including both<br>ground and air<br>transport,<br>environments<br>which is supported<br>by compliance to<br>IEC 60601-1-<br>11:2015 and IEC<br>60601-1-12:2014. |
| Energy source | Sensor itself has no energy<br>source. When used with a<br>compatible monitor<br>(powered by battery or AC<br>power source), the monitor<br>powers the sensor. | Sensor itself has no energy<br>source. When used with a<br>compatible monitor<br>(powered by battery or AC<br>power source), the monitor<br>powers the sensor. | Same |
| Fundamental<br>Technology | Transmittance sensor that<br>measures functional oxygen<br>saturation and pulse rate<br>non-invasively: utilizing<br>spectrophotometry, a pair of<br>light emitting diodes (LEDs) | Transmittance sensor that<br>measures functional oxygen<br>saturation and pulse rate<br>non-invasively: utilizing<br>spectrophotometry, a pair of<br>light emitting diodes (LEDs) | Same |
| Characteristic | Subject Device | Predicate Device | Comments |
| | Nellcor™ OxySoft™<br>Neonatal-Adult SpO2<br>Sensor, OxySoftN and<br>OxySoftNHC<br><br>wavelengths, and using<br>plethysmography, the time<br>varying absorbance of the<br>tissue due to the pulsatile<br>blood signal is obtained,<br>amplified and digitized from<br>a photodiode to determine<br>functional arterial oxygen<br>saturation. | Nellcor™ Neonatal-Adult<br>SpO2 Sensor, MAXN<br>(K012891)<br><br>wavelengths, and using<br>plethysmography, the time<br>varying absorbance of the<br>tissue due to the pulsatile<br>blood signal is obtained,<br>amplified and digitized from<br>a photodiode to determine<br>functional arterial oxygen<br>saturation. | |
| Performance<br>Standards | EN ISO 80601-2-61<br>ISO 80601-2-61 | EN ISO 80601-2-61 | Same |
| Key<br>Performance<br>Specifications | Pulse rate accuracy: +/- 3<br>BPM RMS<br>SpO2 Accuracy (Adult and<br>Neonate with Motion):<br>70% to 100% (+/- 3% RMS)<br>SpO2 Accuracy (Adult and<br>Neonate): 70% to 100%<br>(+/- 2% RMS)<br>SpO2 Accuracy (Low<br>Saturation): 60% to 80%<br>(+/- 3% RMS)<br>SpO2 Accuracy (Low<br>Perfusion): 70% to 100%<br>(+/- 2% RMS) | Pulse rate accuracy: +/- 3<br>BPM RMS<br>SpO2 Accuracy (Adult and<br>Neonate with Motion):<br>70% to 100% (+/- 3% RMS)<br>SpO2 Accuracy (Adult and<br>Neonate): 70% to 100%<br>(+/- 2% RMS)<br>SpO2 Accuracy (Low<br>Saturation): 60% to 80%<br>(+/- 3% RMS)<br>SpO2 Accuracy (Low<br>Perfusion): 70% to 100%<br>(+/- 2% RMS) | Same |
| External<br>Materials | Polyurethane<br>Silicone<br>Synthetic fabric<br>Aluminum polypropylene<br>Synthetic rubber<br>PVC | Polyethylene<br>Acrylic<br>Woven cotton<br>Polypropylene<br>Rayon elastic<br>PVC | Different: The<br>subject and<br>predicate device<br>have equivalent<br>materials, with the<br>exception of the<br>bandage and<br>adhesive.<br>The subject device<br>was assessed for<br>form, fit and<br>function through a<br>clinical study,<br>MDT20002VALNIC |
| Main Safety<br>Standards | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-2<br>IEC 60601-1-2 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-2<br>IEC 60601-1-2 | Different:<br>Expanded to<br>include IEC 60601-<br>1-11 and IEC |
| Characteristic | Subject Device | Predicate Device | Comments |
| | Nellcor™ OxySoft™<br>Neonatal-Adult SpO2<br>Sensor, OxySoftN and<br>OxySoftNHC<br>IEC 60601-1-11<br>IEC 60601-1-12 | Nellcor™ Neonatal-Adult<br>SpO2 Sensor, MAXN<br>(K012891) | 60601-1-12 to<br>address emergency<br>and home use<br>environments. |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
The subject device, the Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, models OxySoftN and OxysoftNHC. has the same principle of operation. intended use, patient population and fundamental technology as the predicate device, Nellcor™ Neonatal-Adult SpO2 Sensor, MAXN.
The differences in technological characteristics do not alter the fundamental performance or the clinical use of the device, these differences do not raise different questions of safety and effectiveness based on the performance data and clinical studies, summarized below.
### PERFORMANCE DATA SUMMARY
The following performance data were provided to support the substantial equivalence determination for the modification being made as part of this submission.
### Biocompatibility Testing
The Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, models OxySoftNand OxysoftNHC are classified as surface devices with permanent (>30 days) skin contact. The biocompatibility evaluation was conducted in accordance with International Standard (ISO) 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process" and FDA Biocompatibility Guidance "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.""
The biocompatibility assessment included the following:
- Cytotoxicity (MEM elution method) ●
- Sensitization (Guinea pig maximization) ●
- Intracutaneous reactivity
- . Chemical characterization and toxicological risk assessment
### Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, models OxySoftN and OxysoftNHC. The sensor complies with the IEC 60601-1 standard for safety and the IEC 60601-1-2 standard for EMC.
### Non-clinical Performance Testing
{10}------------------------------------------------
Non-clinical testing for the Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, models OxySoftN and OxysoftNHC included the following:
- Mechanical
- Use Life
- Type 1 Light Shunting ●
- Storage Altitude .
- Storage Temperature and Humidity ●
- Ambient Light
- Ingress Protection
- Shock and Vibration ●
- Time to Post
- . Adhesive Liner Card and Initial Peel Force
- Nellcor Functional Compatibility ●
- Surface Temperature Rise Measurement System
- . Operational Altitude
- Operational Humidity and Temperature
- Transient Temperature and Humidity ●
- Pulse Rate Motion
- Low Perfusion ●
- Human Factors
Non-clinical testing was performed to demonstrate that the User Requirements and Product Requirements have been fulfilled. Specifically, testing was completed to demonstrate that the subject device is substantially equivalent to the predicate device and as safe and effective with respect to design characteristics and performance. A Human Factors Validation study was used to evaluate shortening of the subject device cable and confirmed that this design attribute did not contribute to any or existing usability issues. Ingress Protection testing was performed to ensure subject device functions, including the new bandage material, after being subjected to particle IP3X and liquid ingress IPX3 conditioning per IEC 60529. To demonstrate that the new adhesive material of the subject device adds benefit but does not impact clinical use or performance, biocompatibility testing was performed.
### Animal Performance Testing
Not applicable. No animal performance testing was required to demonstrate device safety and effectiveness.
### Clinical Studies
To support the substantial equivalence of the subject device with the expanded indications for use Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, OxySoftNHC, clinical studies were conducted.
### MDT20002VALNIC: Performance evaluation of the OxySoftN neonatal sensor in the NICU compared to SaO2
A clinical validation study, Performance Evaluation of the Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, OxySoftN (subject device) in the NICU Compared to SaO2 was performed. Arterial samples were collected on neonates to demonstrate form, fit, and function (clinical performance) of the Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, OxySoftN(subject
{11}------------------------------------------------
device). The standard reference for the SpO2 accuracy as read by pulse oximetry equipment was compared to SaO2 values obtained from CO-oximeter analysis of simultaneously drawn arterial blood. The observed SpO2 accuracy was 2.68% in a study of 27 patients with ages of 1 to 9 days, weight from 640 to 4710 grams, and 64 observations made spanning a range of 82.1% to 99.3% SaO2. There was acceptable evidence to demonstrate safe form and fit of the OxySoftN sensor when applied to 31 patients in the intended neonate population.
### MDT20037CYBOXY: Validation of the OxySoftN to Determine the Gentleness on Fragile Skin
A clinical study was performed on healthy elderly subjects (known for having fragile skin) to evaluate evaporative water loss (TEWL), and amount of skin protein disruption between the Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, OxySoftN (subject device) and the Nellcor™ Neonatal-Adult SpO2 Sensor, MAXN (predicate device) to evaluate gentleness of the new adhesive OxySoftN (subject device). The clinical study data demonstrated the adhesive removed significantly less cells than MAXN sensor (predicate device) and caused a minimal increase in TEWL on subjects likely to have fragile skin.
### MDT20047OXSNOF: Sensor-Off Validation of the OxySoftN sensor in Healthy Adults
A clinical study, Nellcor™ Sensor-Off Validation of the Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, OxySoftN (subject device) in healthy adults was performed to further ensure the clinical use of the sensor, and validated that the investigational OxySoftN oximetry sensor displayed a "Sensor Off" message within 1 minute after the sensor is removed at least 90% of the time. A sequence of steps for different sensor off configurations was performed per subject. This study was completed to determine whether simulated and real-world situations known to represent challenges to pulse oximetry have an effect on the ability of the "Sensor Off" feature to function according to product design specifications. The study demonstrated the sensor off feature performance with a sample proportion of 0.996 and 95% lower bound for the proportion of success of 0.983. The results demonstrate the device posts a "Sensor Off" message within 1 minute after the sensor is removed at least 90% of the time.
### MDT20038CYBADH: Multiple Reposition Peel Adhesion on Human Skin Validation, OxySoftN
The objective of this study was to demonstrate that the Nellcor™ OxySoft ™ Neonatal-Adult SpO2 Sensor, OxySoftN (subject device) remains adhered to the skin after the 18th repositioning by measuring peel adhesion as the sensor was serially removed and reapplied to alternating sites on the volar forearm of healthy adults using standardized in-vivo testing procedures. The study concluded that the OxySoftN Sensor can be repositioned and remains adhered to the skin after the 18th repositioning using the above mentioned standardized in-vivo testing procedures.
### MDT20028OXYLOV: SpO2 Accuracy Validation of the OxySoftN sensor via Reference CO-Oximetry in Healthy, Well-Perfused Subject
The purpose of this study was to directly compare the Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor. OxySoftN (subject device) saturation measurements, during normal to low saturation, to saturation measurements made by a multi-wavelength CO-oximeter, taken from
{12}------------------------------------------------
arterial blood samples from healthy adult subjects. Additionally, prototype pulse rate measurements were compared to 3-lead ECG pulse rate measurements made via a multiparameter monitor. Results of this study concluded, when using the N-600x, PM1000N and Oxicable devices, paired with the OxySoftN sensor, during a saturation range of 60-80%, the svstem demonstrates expected SpO2 accuracy of ±3% Arms. When using the N-600x. PM1000N and Oxicable devices, paired with the OxySoftN sensor during a saturation range of 70% - 100%, the system demonstrates expected SpO2 accuracy of ±2% Arms. During a saturation range of 60% - 100%, the system demonstrates expected PR accuracy of ≤3 BPM.
### MDT20006OXYVMT: SpO2 Accuracy Validation of the OxySoft sensor via Reference CO-Oximetry Motion Study
The primary objective of the study is to investigate the performance of the Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, OxySoftN (subject device) for pulse rate and saturation accuracy in a diverse subject population during motion and non-motion conditions and over a specified saturation range. When using the N-600x device, in conjunction with the OxySoftN sensor, during a saturation range of 70% - 100% and motion conditions, the system demonstrates expected SpO2 accuracy of ±3% Arms. When using the N-600x device, in conjunction with the OxySoftN sensor, during a saturation range of 70% - 100% and motion conditions, the system demonstrates expected PR accuracy of ≤5 BPM. When using the N-395 device, in conjunction with the OxySoftN sensor, during a saturation range of 70% - 100% and non-motion conditions, the system demonstrates expected SpO2 accuracy of ±3% Arms. The N-395 device uses a leqacy calibration scheme where the calibration curve is stored in the monitor and the appropriate curve to be used is identified by a resistor value in the sensor connector. The RCAL calibration scheme was replaced by the sensor EPROM digital calibration scheme but to ensure efficacy of the legacy N-395 device the RCAL calibration scheme was included in testing.
### CONCLUSIONS
The comprehensive verification and validating testing of non-clinical performance testing and clinical validation studies, demonstrate that the Nellcor™ OxySoft ™ Neonatal-Adult SpO2 Sensor, OxySoftN/ OxySoftNHC is substantially equivalent to the predicate device with respect to clinical use case, characteristics and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.