K212505 · The O R Company Pty, Ltd. · LKF · Jan 10, 2022 · Obstetrics/Gynecology
Device Facts
Record ID
K212505
Device Name
DUMI ManipulatOR
Applicant
The O R Company Pty, Ltd.
Product Code
LKF · Obstetrics/Gynecology
Decision Date
Jan 10, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The use of the DUMI ManipulatOR is indicated in diagnostic laparoscopy, mini laparotomy, fertility exams, and salpingoplastic procedures where manipulation of the uterus is required.
Device Story
DUMI ManipulatOR is a sterile, single-use, disposable uterine manipulator. Device consists of a catheter, handle, and distal apple-shaped balloon. Physician inserts device into uterus; balloon is inflated with air via syringe through a luer connector on an inflation valve. Three-way hub accommodates fluid for dye studies. Device enables manipulation of anteverted and retroverted uteri during gynecological procedures. Used in OR or clinical settings by physicians. Provides mechanical leverage for tissue positioning; facilitates diagnostic visualization and dye studies. Benefits include effective uterine control and simplified procedural workflow.
Clinical Evidence
Bench testing only. No clinical data. Preclinical testing included physical, mechanical, and biocompatibility assessments (cytotoxicity, sensitization, acute systemic toxicity). Performance testing verified balloon capacity, fatigue, burst pressure, air leakage, spring load locking force, component joint strength, tip deflection, and luer connector integrity. All tests met predetermined acceptance criteria.
Technological Characteristics
Sterile, single-use, disposable uterine manipulator. Features include a catheter, handle, and 10cc apple-shaped balloon. Materials are well-characterized and commonly used in medical devices. Connectivity: manual luer connector. Sterilization: ETO (ISO 11135:2014). Standards: ISO 13485, ISO 14971, ISO 10993, ISO 80369-7, ASTM F88, ASTM F1929, ASTM F2096, ASTM F1980, ASTM D4169, ASTM F3172.
Indications for Use
Indicated for patients undergoing diagnostic laparoscopy, mini laparotomy, fertility exams, and salpingoplastic procedures requiring uterine manipulation.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
Predicate Devices
Panpac Uterine Manipulator Injector, Model UMI 4.5 (K092980)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 10, 2022
The O.R. Company Pty Ltd Amesha Silva Quality Assurance & Regulatory Affairs Manager 1/32 Silkwood Place Carrum Downs, Victoria 3201 Australia
Re: K212505
> Trade/Device Name: DUMI ManipulatOR (DUMI-350A) Regulation Number: 21 CFR§ 884.4530 Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II Product Code: LKF Dated: November 30, 2021 Received: December 7, 2021
Dear Amesha Silva:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212505
Device Name DUMI ManipulatOR (DUMI-350A)
Indications for Use (Describe)
The use of the DUMI ManipulatOR is indicated in diagnostic laparoscopy, mini laparotomy, fertility exams, and salpingoplastic procedures where manipulation of the uterus is required.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for "The O.R. Company". The logo consists of the text "THE O.R. COMPANY" in a blue sans-serif font, with the "O.R." in white inside of a blue circle. Below the company name is the tagline "The smart way to operate" in a smaller, lighter blue font.
# DUMI ManipulatOR - 510(k) Summary
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
| Submitted by: | The O R Company Pty Ltd |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 1/32 Silkwood Rise, Carrum Downs, VIC, 3201, Australia |
| | Phone: +61 3 9413 5555 |
| | Fax: + 61 3 9413 5556 |
| Contact Person: | Name: Nicole Conway |
| | Title: Quality Assurance & Regulatory Affairs Manager |
| | Phone number: +61 3 9413 5555 |
| | Email: nicole@theorcompany.com |
| Date Prepared: | 06 January 2022 |
| Device Trade Name: | DUMI ManipulatOR (DUMI-350A) |
| Common name: | Uterine manipulator |
| Classification name: | Obstetric-gynecologic specialized manual instrument |
| Regulation number: | 884.4530 |
| Product Code: | LKF (Class II) - Cannula, Manipulator/Injector, Uterine |
| Panel: | Obstetrics/Gynecology |
| Predicate Devices: | Panpac Medical Corporation ---- Panpac Uterine Manipulator Injector, Model UMI 4.5 (K092980) |
| | The predicate device has not been subjected to a recall. |
| Device Description: | The DUMI ManipulatOR is a sterile, single use, disposable medical device designed for the intra- |
| | operative manipulation of both the anteverted and retroverted uterus as well as facilitating simple |
| | and effective dye studies. |
| | The device includes an anatomically designed balloon at the distal end that is injected with air by |
| | syringe through the luer connector of inflation valve. The three-way hub can accommodate fluid |
| | and can be used for dye studies. |
| Indications for use: | The use of the DUMI ManipulatOR is indicated in diagnostic laparoscopy, mini laparotomy, fertility |
| | exams, and salpingoplastic procedures where manipulation of the uterus is required. |
| Substantial Equivalence<br>Discussion: | Indications:<br>The ORC DUMI ManipulatOR and UMI 4.5 have almost identical indications for use. Both devices<br>are used for diagnostic procedures such as laparoscopy, minilaparotomy, fertility examinations and<br>salpingoplastic procedures where manipulation of the uterus is required. There is no difference in<br>intended use.<br><br>Technological characteristics:<br>Both the DUMI ManipulatOR and the UMI 4.5 share almost identical technological characteristics.<br>The designs of both devices consist of a catheter, handle and a balloon to help with manipulation<br>of tissue. The devices have the following differences:<br>There is a 20mm difference in length between DUMI ManipulatOR and UMI 4.5.The shape of the DUMI ManipulatOR balloon is apple shaped while it is elongated for the predicate. They are the same volume (10cc)There are minor differences in handle designThe differences do not raise different questions of safety and effectiveness.<br><br>Performance:<br>As noted in the summary of testing below, the DUMI ManipulatOR was evaluated for performance<br>and demonstrated to be as safe and effective as the predicate. |
| Asia Pacific | North America | Hong Kong | Contact |
|------------------------------------------------------------|--------------------------------------------------|----------------------------------------------------------------------------------|-------------------------------------------|
| 1/32 Silkwood Rise<br>Carrum Downs, VIC<br>3201, Australia | 1625 South Tacoma Way<br>Tacoma WA, 98409<br>USA | Flat 9/F Wah Kit<br>Commercial Centre<br>300 Des Voeux Road<br>Central Hong Kong | info@theorcompany.com<br>theorcompany.com |
| T +61 3 9413 5555<br>F +61 3 9413 5556 | T +1 253 441 6509<br>F +1 866 333 9524 | T +85 2 3611 0981<br>F +85 2 8148 8932 | |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for "THE O.R. COMPANY". The logo consists of the text "THE" followed by a blue circle containing the letters "O.R." in white. To the right of the circle is the text "COMPANY" and below that is the tagline "The smart way to operate".
| Summary of Testing/Data: | The design of The OR Company's DUMI ManipulatOR device is based on Panpac Medical<br>Corporation's Uterine Manipulator Injector (Model UMI 4.5) device. The DUMI ManipulatOR is<br>demonstrated to be substantial equivalent to the Panpac Medical Corporation's Uterine<br>Manipulator Injector (Model UMI 4.5) device. |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | A series of preclinical physical, mechanical and biocompatibility tests were performed to assess<br>the safety and effectiveness of The O.R Company Pty Ltd's DUMI ManipulatOR as compared to the<br>predicate. |
| | The tests were conducted in accordance with relevant standards listed below and have successfully<br>met predetermined acceptance criteria: |
| | ISO 13485:2016 Medical devices Quality management systems Requirements for regulatory |
| | purposes |
| | FDA 21 CFR 820 Quality System regulations |
| | ISO 14971: 2019 Medical devices application of risk management to medical devices |
| | ISO 11135:2014 Sterilization of health-care products-ETO Requirements for the development,<br>validation & routine control of a sterilization process for medical device |
| | ISO 11737-1: 2018 Sterilization of Medical devices-Microbiological methods-Part 1: Determination<br>of a population of microorganisms on products |
| | ISO 11737-2: 2019 Sterilization of medical devices - Microbiological methods - Part 2: Tests of<br>sterility performed in the definition, validation and maintenance of a sterilization process |
| | ISO 11607-1 : 2019 Packaging for terminally sterilized medical devices Part 1: Requirements for<br>materials, sterile barrier systems and packaging systems |
| | ISO 11607-2: 2019 Packaging for terminally sterilized medical devices Part 2: Validatio n<br>requirements for forming, sealing and assembly processes |
| | ASTM F88 /F88M-15 Standard test method for seal strength of flexible barrier materials. |
| | ASTM F1929-15 Standard test method for detecting seal leaks in porous medical packaging by dye<br>penetration. |
| | ASTM F2096-11 (2019) Standard Test Method For Detecting Gross Leaks In Packaging By Internal<br>Pressurization (Bubble Test) |
| | ISO 10993-1:2018 Biological evaluation of medical devices Part 1: Evaluation and testing within a<br>risk management process |
| | ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity |
| | ISO 10993-10:2010 Biological evaluation of medical devices – Part 10: Tests for irritation and skin<br>sensitization |
| | ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity<br>EN 14372:2004 REACH Regulation (EC) No 1907/2006- Non-DEHP |
| | ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices |
| | ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems |
| | ISO 2859-1:1999 Sampling procedures for inspection by attribute - Part 1: Sampling schemes<br>indexed by acceptance quality limit (AQL) for lot-by-lot inspection |
| | ASTM F3172-15 Design verification device size and sample selection for Endovascular devices |
| | ISO 80369-7: 2016 Small-bore connectors for liquids and gases in healthcare applications — Part 7:<br>Connectors for intravascular or hypodermic applications |
| | The specific tests undertaken included: |
| | EO Sterilization: |
| | EO Sterilization Validation EO Residues Bacteriostasis and Fungistatis |
| Asia Pacific | North America | Hong Kong | Contact |
|--------------------|-----------------------|--------------------|-----------------------|
| 1/32 Silkwood Rise | 1625 South Tacoma Way | Flat 9/F Wah Kit | info@theorcompany.com |
| Carrum Downs, VIC | Tacoma WA, 98409 | Commercial Centre | theorcompany.com |
| 3201, Australia | USA | 300 Des Voeux Road | |
| | | Central Hong Kong | |
| T +61 3 9413 5555 | T +1 253 441 6509 | T +85 2 3611 0981 | |
| F +61 3 9413 5556 | F +1 866 333 9524 | F +85 2 8148 8932 | |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for "The O.R. Company". The logo is in blue and features the company name in a simple, sans-serif font. Below the company name is the tagline "The smart way to operate".
#### Bioburden ●
### Shelf Life:
- Packaging integrity post 5 year accelerated aging ●
- Performance post 5 year accelerated aging
# Packaging:
- Seal Strength of Flexible Barrier Materials
- Dye Penetration
- Sterility ●
# Transportation:
- Handling
- Loose Load Vibration ●
- Low Pressure (High Alititude) Hazard
- Vehicle Vibration
- Handling
# Biocompatibility:
Both devices are used in the same anatomical site and the same components of both devices come in short term contact with the mucosal membranes of the uterus and the vagina.
All patient contacting components of both devices have been tested for biocompatibility and have been found to be non-toxic, non-sensitizing, non-cytotoxic and non-irritating. The raw materials have all been used in various combinations by not only The O.R Company and predicate devices but also by uterine manipulators currently on the market. The materials are all well-characterized, are all commonly used in these and other medical devices. The following tests were conducted on the subject device:
- . Cytotoxicity
- Sensitization
- Acute Systemic Toxicity ●
The performance tests listed below have demonstrated that The O.R Company Pty Ltd's DUMI ManipulatOR performs at least the same as the predicate device PanPac Medical Corporations UMI 4.5:
- Intrauterine balloon capacity and diameter test
- Intrauterine balloon fatigue test (repeat inflation) ●
- Intrauterine balloon burst test
- Air leakage test for prolonged time ●
- . Spring load locking mechanism force test
- Component joint strength test
- Tip deflection test
- Balloon adhesion strength test
- . Luer connector leakage test
| Conclusion: | The subject device, DUMI ManipulatOR share similar indications and technological characteristics<br>as the predicate device. All tests undertaken to demonstrate substantial equivalence of the DUMI<br>ManipulatOR meet the requirements of its predetermined acceptance criteria. The test results<br>confirm that the The O.R Company Pty Ltd's DUMI ManipulatOR is as safe and effective as the<br>predicate device. Therefore, the DUMI ManipulatOR is substantially equivalent to the<br>predicate. |
|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Asia Pacific | North America | Hong Kong | Contact |
|--------------------|-----------------------|---------------------------------------|-----------------------|
| 1/32 Silkwood Rise | 1625 South Tacoma Way | Flat 9/F Wah Kit<br>Commercial Centre | info@theorcompany.com |
| Carrum Downs, VIC | Tacoma WA, 98409 | 300 Des Voeux Road | theorcompany.com |
| 3201, Australia | USA | Central Hong Kong | |
| T +61 3 9413 5555 | T +1 253 441 6509 | T +85 2 3611 0981 | |
| F +61 3 9413 5556 | F +1 866 333 9524 | F +85 2 8148 8932 | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.