K212479 · Cotton High Tech S.L. · HEB · Oct 8, 2021 · Obstetrics/Gynecology
Device Facts
Record ID
K212479
Device Name
Cohitech Reusable Tampon Applicator
Applicant
Cotton High Tech S.L.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Oct 8, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
COHITECH Reusable Tampon Applicator is intended to be used to insert a digital menstrual tampon into the vagina.
Device Story
Reusable, non-sterile medical device designed for manual insertion of digital menstrual tampons into the vagina. Device consists of polyethylene outer housing and inner pusher, plus polypropylene storage/carrying cases. User loads digital tampon into applicator; manually operates pusher to eject tampon into vaginal canal. Intended for personal, over-the-counter use by individuals during menstruation. Requires cleaning with unscented soap after each use and disinfection with 70% isopropyl alcohol after each menstrual cycle. Use-life is 4 years. Benefits include reusable alternative to single-use applicators, reducing waste.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and biocompatibility evaluation. Biocompatibility testing (cytotoxicity, vaginal irritation, delayed hypersensitivity) confirmed device is non-cytotoxic, non-irritating, and non-sensitizing. Bench testing confirmed physical dimensions, injection/ejection forces, tampon compatibility, and 4-year use-life met acceptance criteria.
Technological Characteristics
Materials: Polyethylene (outer housing/pusher), Polypropylene (storage/carrying cases). Dimensions: 73.3mm length; 16mm or 19mm diameter. Weight: 5.4g-5.7g. Non-sterile; user-reprocessed via soap cleaning and 70% isopropyl alcohol disinfection. Mechanical operation; no energy source or software.
Indications for Use
Indicated for individuals requiring assistance with the insertion of digital menstrual tampons into the vagina.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The FDA acronym and name are in blue, with the acronym in a larger font size than the name. The name is written as "U.S. FOOD & DRUG ADMINISTRATION".
October 8, 2021
Cotton High Tech S.L. Anna Garcia Regulatory Affairs and Product Certification Manager Colònia La Rabeia S/N Balsareny, Barcelona 08660 Spain
Re: K212479
> Trade/Device Name: COHITECH Reusable Tampon Applicator Regulation Number: 21 CFR§ 884.5470 Regulation Name: Unscented Menstrual Tampon Regulatory Class: II Product Code: HEB, HIL Dated: August 5, 2021 Received: August 9, 2021
Dear Anna Garcia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212479
Device Name COHITECH Reusable Tampon Applicator
#### Indications for Use (Describe)
COHITECH Reusable Tampon Applicator is intended to be used to insert a digital menstrual tampon into the vagina.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "COHITECH Cotton High Tech S.L.". The word "COHITECH" is in large, bold, black letters at the top of the image. Below that, in smaller gray letters, is the phrase "Cotton High Tech S.L.". Below the text is a large blue circle with a white circle in the center, creating a ring shape.
K212479 Page 1 of 4
# 510(k) Summary K212479
| SUBMITTER NAME: | Cotton High Tech S.L. |
|---------------------------|-----------------------------------------------------------------------------------|
| SUBMITTER ADDRESS: | Colònia La Rabeia, s/n<br>08660 Balsareny<br>BARCELONA<br>SPAIN |
| CONTACT: | Anna Garcia<br>Regulatory Affairs and Product Certification Manager |
| PHONE: | + 34 93 839 16 28 |
| FAX: | + 34 93 839 19 44 |
| e-mail: | agarcia@cohitech.net |
| Summary Preparation Date: | 10/04/2021 |
| DEVICE TRADE NAME: | COHITECH Reusable Tampon Applicator |
| COMMON NAME: | Reusable Tampon Applicator |
| REGULATION NUMBER: | 21 CFR 884.5470 |
| REGULATION NAME: | Unscented Menstrual Tampon |
| REGULATORY CLASS: | II |
| PRODUCT CODE: | HEB (unscented menstrual tampon) and<br>HIL (scented-deodorized menstrual tampon) |
# PREDICATE DEVICE:
| Company | Product | 510(k)# |
|------------|------------------------------------|---------|
| THINX INC. | re.t.a™ Reusable Tampon Applicator | K180850 |
The predicate device has not been subject to a design-related recall.
# DEVICE DESCRIPTION:
The COHITECH Reusable Tampon Applicator is a non-sterile, single user, reusable medical device. It is intended to be sold on its own, without a pre-loaded tampon. The applicator requires the user to load the applicator with a legally marketed digital menstrual tampon. The use-life of the subject applicator is 4 years. The applicator requires the user to clean and disinfect the applicator before initial use, before longterm storage and at the end of the menstrual cycle, as well as to clean the device after each use.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "COHITECH Cotton High Tech S.L.". The word "COHITECH" is in large, bold, black letters at the top. Below that, in smaller gray letters, is the phrase "Cotton High Tech S.L." Underneath the text is a large blue circle with a white circle in the center, creating a ring shape.
The device consists of a polyethylene outer tampon housing and inner pusher, a polypropylene storage case, and a carrying case (optional).
This device is available in two sizes, one size for reqular and super digital tampons and another for super and super plus digital tampons. The total applicator length is 73.3 ± 0.2 mm for both sizes. The overall applicator diameter (outer housing and inner pusher) is 16 ± 0.1 mm for the regular/super digital tampon applicator and 19 ± 0.1 mm for the super/super plus digital tampon applicator. The applicator weight is 5.4 ± 0.2 for the regular/super tampon applicator and 5.7 ± 0.2 g for the super/super plus tampon applicator. The storage case weight is 7.9 ± 0.1 g and carrying case weight is 26.9 ± 0.1 g
# INDICATIONS FOR USE STATEMENT:
COHITECH Reusable Tampon Applicator is intended to be used to insert a digital menstrual tampon into the vagina.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS AND INDICATIONS FOR USE:
The following table compares the technological characteristics and indications for use statement of the subject and the predicate device:
| Description | Subject Device | Predicate Device |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | COHITECH Reusable<br>Tampon Applicator | re.t.a™ Reusable<br>Tampon Applicator |
| Manufacturer<br>Applicant | /<br>Cotton High Tech, S.L. | THINX INC |
| 510(k) Number | K212479 | K180850 |
| Product Code | HEB, HIL | HEB, HIL |
| Regulation Number | 884.5470 | 884.5470 |
| Indication for use | COHITECH Reusable<br>Tampon Applicator is<br>intended to be used to<br>insert a digital<br>menstrual tampon into<br>the vagina. | The re.t.a.™ reusable<br>tampon applicator is<br>intended to be used to<br>insert a digital<br>menstrual tampon into<br>the vagina. |
| Usability | Reusable | Reusable |
| Use-life | 4 years | 2 years |
| Tampon compatibility | Digital tampons of<br>various sizes (regular,<br>super, and super plus). | Digital tampons of<br>various sizes |
| Device Design | Outer tampon housing<br>and inner pusher | Sleeve with slit, pusher<br>and outer cover |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "COHITECH Cotton High Tech S.L.". The word "COHITECH" is in large, bold, black letters at the top. Below that, in smaller gray letters, is the phrase "Cotton High Tech S.L." Underneath the text is a large blue circle with a smaller white circle inside, creating a ring shape.
The subject and predicate device have the same intended use - to deliver digital menstrual tampons into the vagina. The subject device differs from the predicate device in technological characteristics. The subject device has a different device design, use-life, and material composition.
The differences in technological characteristics between the subject and predicate device do not raise different questions of safety and effectiveness.
# SUMMARY DISCUSSION OF NON-CLINICAL PERFORMANCE DATA
### Biocompatibility
The following biocompatibility studies were performed in accordance with the 2020 FDA quidance Use of International Standard ISO 10993-1 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and standard ISO 10993-1:
- Cytotoxicity (ISO 10993-5:2009) O
- Vaginal Irritation (ISO 10993-10:2010) O
- Delayed Hypersensitivity (ISO 10993-10:2010) O
The results showed that the subject device was non-cytotoxic, non-irritating, and a non-sensitizer.
#### Bench Testing
The following performance characteristics were assessed to demonstrate that the proposed subject device met applicable design and performance requirements:
- Device weight o
- o Physical dimensions
- Functional evaluation, including: o
- Injection force of the inner pusher into the outer tampon housing с component
- Ejection force of the inner pusher out of the outer tampon housing o component
- Tampon compatibility and ejection force testing o
- o Use-life testing
The results of the non-clinical performance data were acceptable and met the established acceptance criteria.
#### Reprocessing, Cleaning, and Disinfecting
The cleaning instructions provided for the COHITECH Reusable Tampon Applicator were developed per the recommendations in the 2015 FDA quidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "COHITECH Cotton High Tech S.L.". The word "COHITECH" is in large, bold, black letters at the top. Below that, in smaller gray letters, is the phrase "Cotton High Tech S.L.". Below the text is a large blue circle with a white circle in the center.
Labeling." The COHITECH Reusable Tampon Applicator should be washed with unscented soap after every use and disinfected in 70% isopropyl alcohol after every menstrual cycle.
The COHITECH Reusable Tampon Applicator is not provided sterile and is not intended to be sterilized by users.
# CONCLUSION:
The subject and predicate device have the same intended use and the technological differences do not raise different questions of safety or effectiveness. The results of the non-clinical testing described above demonstrate that COHITECH Reusable Tampon Applicator is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.