← Product Code [FAJ](/productcode/FAJ) · K212476

# KARL STORZ Flexible HD Cysto-Urethroscope System (K212476)

_Karl Storz Endoscopy America, Inc. · FAJ · Sep 3, 2021 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K212476

## Device Facts

- **Applicant:** Karl Storz Endoscopy America, Inc.
- **Product Code:** [FAJ](/productcode/FAJ.md)
- **Decision Date:** Sep 3, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Flexible HD Cysto-Urethroscope System is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of urinary tract including the urethra, bladder, ureters, and kidneys.

## Device Story

Flexible HD Cysto-Urethroscope System provides visualization and operative access for urinary tract procedures. Device captures video signals via distal CMOS imager; transmits data to external IMAGE1 S CCU (Connect/Connect II/X-Link modules) for processing and display. Operated by clinicians in clinical settings. System requires external CCU for operation; cannot function independently. Modifications from predicate include removal of suction channel (replaced with stainless steel housing plug) and transition to permanent T-Luer. Output allows clinicians to perform diagnostic/therapeutic interventions under direct visualization. Benefits include improved access and visualization for urological procedures.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via comparative bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-18), EMC (IEC 60601-1-2), optical/mechanical performance (ISO 8600, IEC 62471), and biocompatibility (ISO 10993).

## Technological Characteristics

Flexible endoscope with CMOS imager, 5.5 mm shaft diameter, 2.3 mm working channel. Deflection: 210° up, 140° down. Internal LED light source. Connectivity: External CCU (IMAGE1 S). Sterilization: STERRAD 100NX, STERRAD NX, V-PRO maX, V-PRO 60. Biocompatibility per ISO 10993. Reprocessing per AAMI TIR 12/30, ISO 15883-5, ASTM E1837.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Flexible HD Cysto-Urethroscope System ([K191357](/device/K191357.md))
- Flexible Video Cysto-Urethroscope (C-View) ([K202957](/device/K202957.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

September 3, 2021

KARL STORZ Endoscopy America, Inc. Thomas Ostrowski Senior Regulatory Affairs Specialist 2151 E. Grand Ave El Segundo, CA 90245

Re: K212476

> Trade/Device Name: Flexible HD Cysto-Urethroscope System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FAJ, FBO Dated: August 5, 2021 Received: August 6, 2021

Dear Thomas Ostrowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K212476

Device Name Flexible HD Cysto-Urethroscope System

Indications for Use (Describe)

The Flexible HD Cysto-Urethroscope System is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of urinary tract including the urethra, bladder, ureters, and kidneys.

Type of Use (*Select one or both, as applicable*)

| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)                 |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for KARL STORZ–ENDOSKOPE. The word "STORZ" is in large, bold, blue letters. Below that, the words "KARL STORZ–ENDOSKOPE" are in smaller, blue letters.

K212476

# 8. 510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.

| Applicant:                    | KARL STORZ Endoscopy-America, Inc<br>2151 E. Grand Avenue<br>EI Segundo, CA 90245                                                                                                                    |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                      | Thomas Ostrowski<br>Senior Regulatory Affairs Specialist<br>781-910-8508 (phone)                                                                                                                     |
| Date of Preparation:          | August 27th, 2021                                                                                                                                                                                    |
| Type of 510(k)<br>Submission: | Special                                                                                                                                                                                              |
| Device Identification:        | Trade Name:  Flexible HD Cysto-Urethroscope System                                                                                                                                                   |
| Common Name:                  | Cystoscope and accessories, Flexible/Rigid (FAJ)<br>Cystourethroscope (FBO)                                                                                                                          |
| Product Code:                 | FAJ, FBO                                                                                                                                                                                             |
| Regulation Name /<br>Number:  | Endoscope and Accessories (21 CFR 876.1500)                                                                                                                                                          |
| Predicate Device(s):          | Flexible HD Cysto-Urethroscope System (K191357) – Primary<br>Flexible Video Cysto-Urethroscope (C-View) (K202957) - Secondary<br>**The above predicate devices have not been subject to any recall** |

{4}------------------------------------------------

| Device Description:                   | The Flexible HD Cysto-Urethroscope System (Part Number: 11272VH(U)-TL) is intended to be used with the IMAGE1 S CCU consisting of Connect / Connect II and X-Link modules (cleared in K201135). Identical to the predicates, the scope cannot be operated on its own because it produces image data output in the form of video signals which require external processing and display. When the scope is used with the compatible CCU, it becomes the Flexible HD Cysto-Urethroscope System, which can then be operated as intended and indicated. |                                                                                                                             |                                                                 |                                      |  |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|--------------------------------------|--|
|                                       | The Flexible HD Cysto-Urethroscope System consists of the following key components:<br>11272VH-TL / 11272VHU-TL – Cysto-Urethroscopes with positive (VH-TL) / contra-positive (VHU-TL) deflection control and permanent T-Luer affixed to working channel TC200US / TC201US / TC301US – Connect/Connect II/Link camera control unit (CCU) modules for image processing                                                                                                                                                                             |                                                                                                                             |                                                                 |                                      |  |
|                                       | The subject device is a lower cost alternative to the primary predicate and incorporates the following differences. <b>Housing Plug:</b> Due to lack of clinical need, the subject device replaces the suction channel of the primary predicate device with a permanent stainless steel housing plug. <b>Fixed LUER:</b> The subject device uses a removable T-Luer whereas the primary predicate supported a removable T-Luer or Stopcock.                                                                                                        |                                                                                                                             |                                                                 |                                      |  |
| Intended Use and Indications for use: | The Flexible HD Cysto-Urethroscope System is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of urinary tract including the urethra, bladder, ureters, and kidneys. The minor modifications to the Flexible HD Cysto-Urethroscope System do not change the intended use and indications for use. The intended use / indications for use are identical between the proposed device and the predicate devices.                                                                            |                                                                                                                             |                                                                 |                                      |  |
| Technological Characteristics:        | The subject device is a modification of the primary predicate, 11272VH/VHU, cleared via K191357 and also incorporates design features of the secondary predicate device, 11272VE/VUE, cleared via K202957. A comparison between the proposed device and primary predicate device is included in the table below:                                                                                                                                                                                                                                   |                                                                                                                             |                                                                 |                                      |  |
|                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Proposed Device                                                                                                             | Primary Predicate<br>K191357                                    | Secondary Predicate<br>K202957       |  |
|                                       | Physical Characteristics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                             |                                                                 |                                      |  |
|                                       | Type of scope                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Flexible                                                                                                                    | Same                                                            | Same                                 |  |
|                                       | Insertion Shaft Diameter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 5.5 mm                                                                                                                      | Same                                                            | 5.2 mm                               |  |
|                                       | Insertion Shaft<br>Length                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 37 cm                                                                                                                       | Same                                                            | Same                                 |  |
|                                       | Working Channel<br>Diameter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 2.3 mm                                                                                                                      | Same                                                            | Same                                 |  |
|                                       | Deflection (°)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Up: 210°<br>Down: 140°                                                                                                      | Same                                                            | Same                                 |  |
|                                       | Optical Characteristics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                             |                                                                 |                                      |  |
|                                       | Type of Imager                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | CMOS                                                                                                                        | Same                                                            | Same                                 |  |
|                                       | Field of View                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 100°                                                                                                                        | Same                                                            | Same                                 |  |
|                                       | Direction of View                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 0°                                                                                                                          | Same                                                            | Same                                 |  |
|                                       | Depth of Field                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 3-50 mm                                                                                                                     | Same                                                            | 5-50 mm                              |  |
|                                       | On-axis Resolution<br>(minimal)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 40 lp/mm @ 3 mm<br>2.5 lp/mm @ 50 mm                                                                                        | Same                                                            | 16 lp/mm @ 5 mm<br>1.8 lp/mm @ 50 mm |  |
|                                       | Light Source                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Internal LED                                                                                                                | Same                                                            | Same                                 |  |
|                                       | Material / Design Characteristics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                             |                                                                 |                                      |  |
|                                       | Suction Port /<br>Channel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | No                                                                                                                          | Yes                                                             | No                                   |  |
|                                       | Luer Type                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Permanent T-Luer                                                                                                            | Bayonet with<br>removable T-Luer<br>or Stopcock<br>assembly     | Same as proposed<br>device           |  |
|                                       | Cleaning, Disinfection and Sterilization Methods                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                             |                                                                 |                                      |  |
|                                       | Cleaning                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Manual                                                                                                                      | Same                                                            | Same                                 |  |
|                                       | Sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | STERRAD 100NX<br>(FLEX and DUO<br>cycles),<br>STERRAD NX<br>(Advanced Cycle),<br>V-PRO maX<br>(Flexible Cycle),<br>V-PRO 60 | Same with<br>additional support<br>for SSIE (Standard<br>Cycle) | Same as proposed<br>device           |  |
|                                       | High Level<br>Disinfection                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Revital-Ox<br>RESERT                                                                                                        | Same                                                            | Same                                 |  |
| Non-Clinical<br>Performance Data:     | There are no performance standards or special controls developed under Section<br>514 of the FD&C Act for endoscopes, however, the subject device follows FDA<br>recognized consensus standards identified in the primary and secondary<br>predicate device submissions:<br><br>• Electrical Safety and EMC<br>o IEC 60601-1:2005/(R)2012, Ed. 3.1<br>o IEC 60601-1-2:2014, 4th Ed.<br>o IEC 60601-2-18:2009, Ed. 3.0<br><br>• Optical/Mechanical Performance<br>o IEC 62471:2006<br>o ISO 8600                                                    |                                                                                                                             |                                                                 |                                      |  |

{5}------------------------------------------------

{6}------------------------------------------------

|                               | Biocompatibility  ISO 10993-1:2009/(R) 2013 ISO 10993-5:2009/(R) 2014 ISO 10993-10:2010 ISO 10993-11:2006/(R) 2010   Reprocessing  AAMI TIR 12:2010 ISO 15883-5:2005 AAMI TIR 30:2011 AAMI/ANSI/ISO 11737-1:2006/ (R)2011 ASTM E1837-96:2014                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial<br>Equivalence:   | The intended use, operating principles, technological characteristics and features<br>are similar, if not identical, between that subject device and the predicate<br>devices, Flexible HD Cysto-Ureteroscope System (K191357) and Flexible<br>Video Cysto-Uretheroscope (C-View) (K202957). The minor difference<br>between the subject and predicate devices that does not raise new or different<br>questions or safety and effectiveness are listed above in Technological<br>Characteristics section.<br><br>As proven by the comparisons, the differences do not raise different questions of<br>safety and effectiveness because the intended use, operating principles,<br>technological characteristics, and features are similar, if not identical. Both<br>systems also comply with identical standards and safety testing, where<br>applicable.<br><br>Substantial equivalence to the effectiveness of the subject device is supported by<br>the comparison of the images and standard image quality characteristics<br>including, but not limited to the performance testing listed above. |
| Clinical Performance<br>Data: | Clinical performance is not required to demonstrate substantial equivalence to<br>the predicate devices. Comparative analysis is sufficient to establish substantial<br>equivalence.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Conclusion:                   | The Flexible HD Cysto-Urethroscope System is substantially equivalent to its<br>predicate devices. The comparative analysis demonstrates that the device is as<br>safe and effective as the legally marketed devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |

---

**Source:** [https://fda.innolitics.com/device/K212476](https://fda.innolitics.com/device/K212476)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com