← Product Code [LYZ](/productcode/LYZ) · K212436

# NBR Synthetic Nitrile Examination Gloves (K212436)

_Shandong Langtai International Trade Co., Ltd. · LYZ · Dec 23, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K212436

## Device Facts

- **Applicant:** Shandong Langtai International Trade Co., Ltd.
- **Product Code:** [LYZ](/productcode/LYZ.md)
- **Decision Date:** Dec 23, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

## Device Story

NBR Synthetic Examination Gloves are disposable, non-sterile, ambidextrous gloves composed of polyvinyl chloride, polyurethane, and DOTP, with 1%-3% nitrile added to enhance tensile strength and elongation. Available in sizes S, M, L, and XL, the blue-colored gloves are intended for use by healthcare professionals or in clinical settings to provide a barrier against contamination between the examiner and the patient. The device is worn on the hands during medical examinations. It functions as a physical barrier; no electronic or automated processing is involved. The device is intended for over-the-counter use.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM D5250 (dimensions), ASTM D5151 (watertightness/holes, AQL 2.5), ASTM D6124 (residual powder < 2.0mg), and ASTM D412 (tensile strength/elongation before and after aging). Biocompatibility confirmed via ISO 10993-10 (irritation/sensitization) and ISO 10993-11 (acute systemic toxicity).

## Technological Characteristics

Materials: Polyvinyl chloride, polyurethane, DOTP. Ambidextrous, blue, non-sterile. Sizes: S, M, L, XL. Thickness: ≥0.08mm (finger/palm). Tensile strength: ≥11MPa. Ultimate elongation: ≥300%. Standards: ASTM D5250, ASTM D5151, ASTM D6124, ASTM D412, ISO 10993-10, ISO 10993-11.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Synmax Synthetic Patient Examination Vinyl Gloves, Powder Free, Blue ([K153028](/device/K153028.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 23, 2021

Shandong Langtai International Trade Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801,No.161,East Lujiazui Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K212436

Trade/Device Name: NBR Synthetic Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYZ. Dated: October 18, 2021 Received: October 29, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212436

Device Name NBR Synthetic Examination Gloves

#### Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

| Type of Use (Select one or both, as applicable)                                                         |
|---------------------------------------------------------------------------------------------------------|
| <input type="checkbox"/> Residential Use (Per 441.030 MAI Statute): Primary Residence                   |
| <input checked="" type="checkbox"/> Rent-Free Occupation (Per 441.030 MAI Statute): Secondary Residence |

__ Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary (K212436)

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

## 1.0 Submitter's Information

SHANDONG LANGTAI INTERNATIONAL TRADE CO.,LTD. Name: Address: 8/F,office Building,Weigiao Aluminum Deep,Processing Industrial Park ,ChanshanTown,Zouping,Binzhou City, Shandong, China Tel: +86-15550323002 Contact: Ping Wang

# Designated Submission Correspondent

Name: Shanghai Truthful Information Technology Co., Ltd. Address: Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Contact: Mr. Boyle Wang Email: Info@truthful.com.cn

Date of Preparation: Dec.21st,2021

## 2.0 Device Information

| Trade name:          | NBR Synthetic Examination Gloves       |
|----------------------|----------------------------------------|
| Common name:         | Patient Examination Gloves             |
| Classification name: | Non-powdered patient examination glove |
| Model(s):            | S, M, L, XL                            |
| Product code:        | LYZ                                    |
| Regulation number:   | 21CFR880.6250                          |
| Classification:      | Class I                                |
| Panel:               | General Hospital                       |

## 3.0 Predicate Device Information

Manufacturer: Zibo Huiying Medical Products, Co. Ltd. Device: Synmax Synthetic Patient Examination Vinyl Gloves,Powder Free,Blue 510(k) number: K153028

## 4.0 Indication for Use

{4}------------------------------------------------

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner.

# 5.0 Device Description

The subject device is powder free vinyl examination gloves, adding 1%-3% nitrile to improve the tensile strength and ultimate elongation. The subject device is blue color. lt can be available in four specifications: S,M,L and XL. The subject device is non-sterile.

# 6.0 Technological Characteristic Comparison Table

| Table1-General Comparison   |                                                                                                                                                                                |                                                                                                                                                                                |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item                        | Subject Device<br>(K212436)                                                                                                                                                    | Predicate Device<br>(K153028)                                                                                                                                                  |
| Product Code                | LYZ                                                                                                                                                                            | LYZ                                                                                                                                                                            |
| Regulation No.              | 21CFR880.6250                                                                                                                                                                  | 21CFR880.6250                                                                                                                                                                  |
| Class                       | I                                                                                                                                                                              | I                                                                                                                                                                              |
| Intended Use                | A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner. | A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands to prevent contamination between patient and examiner. |
| Material                    | Poly Vinyl Chloride<br>Polyurethane<br>Di-(2-ethylhexyl)<br>Terephthalate(DOTP)                                                                                                | Poly Vinyl Chloride<br>Polyurethane<br>Diisononyl Phthalate (DINP)                                                                                                             |
| Powdered or<br>Powered free | Powdered free                                                                                                                                                                  | Powdered free                                                                                                                                                                  |
| Design Feature              | Ambidextrous                                                                                                                                                                   | Ambidextrous                                                                                                                                                                   |
| Colorant                    | Blue                                                                                                                                                                           | Blue                                                                                                                                                                           |
| Labeling Information        | Single-use indication, powder free, device color, device name, glove size and quantity, Non-Sterile                                                                            | Single-use indication, powder free, device color, device name, glove size and quantity, Non-Sterile                                                                            |
| Dimensions(mm)              | Length:<br>>=230;<br>Width:                                                                                                                                                    | Length:<br>Average over 234 on M size;<br>Width:                                                                                                                               |

#### Table1-General Comparison

{5}------------------------------------------------

|                       | S: 85±5;<br>M: 95±5;<br>L: 105±5;<br>XL: 115±5 | Average over 96 on M size                    |                                                                                                                                                                                                                                                                                                                                  |                     |                                                                                           |                     |
|-----------------------|------------------------------------------------|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|-------------------------------------------------------------------------------------------|---------------------|
| Thickness(mm)         | Finger: ≥0.08;<br>Palm: ≥0.08                  | Finger: Average 0.98;<br>Palm: Average 0.096 |                                                                                                                                                                                                                                                                                                                                  |                     |                                                                                           |                     |
|                       | Physical<br>Properties                         |                                              | Before<br>Aging                                                                                                                                                                                                                                                                                                                  | Tensile<br>Strength | 11MPa, min                                                                                | Tensile<br>Strength |
|                       |                                                |                                              | Ultimate<br>Elongation                                                                                                                                                                                                                                                                                                           | 300% min            | Ultimate<br>Elongation                                                                    | Average 550%        |
|                       |                                                | After<br>Aging                               | Tensile<br>Strength                                                                                                                                                                                                                                                                                                              | 11MPa, min          | Tensile<br>Strength                                                                       | Average 14.4MPa     |
|                       |                                                |                                              | Ultimate<br>Elongation                                                                                                                                                                                                                                                                                                           | 300%min             | Ultimate<br>Elongation                                                                    | Average 500%        |
| Freedom from<br>Holes |                                                |                                              | Be free from holes when tested in accordance with<br>ASTMD5151 AQL=2.5                                                                                                                                                                                                                                                           |                     | Be free from holes when tested in accordance with<br>ASTMD5151 AQL=2.5                    |                     |
| Powder Content        |                                                |                                              | Meet the requirements of<br>ASTM D6124                                                                                                                                                                                                                                                                                           |                     | Meet the requirements of<br>ASTM D6124                                                    |                     |
| Biocompatibility      |                                                |                                              | ISO 10993-10;<br>Under the conditions of the<br>study, not an irritant or a<br>sensitizer<br>ISO 10993-11;<br>Under the condition of acute<br>systemic toxicity test, the test /<br>article did not show acute<br>systemic toxicity in vivo.<br>ISO 10993-5;<br>Under conditions of the<br>study, device extract is<br>cytotoxic |                     | ISO 10993-10;<br>Under the conditions of the<br>study, not an irritant or a<br>sensitizer |                     |

## 7.0 Summary of Non-clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests

{6}------------------------------------------------

for systemic toxicity.

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.

| Test<br>Method  | Purpose                        | Acceptance Criteria                                          | Results                                                                            |                                                                                             |               |
|-----------------|--------------------------------|--------------------------------------------------------------|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|---------------|
| ASTM<br>D5250   | Physical<br>Dimensions<br>Test | Length(mm):<br>$\ge$ 230;                                    | Length:<br>> 240/Pass;                                                             |                                                                                             |               |
|                 |                                | Width(mm):<br>S: 85±5;<br>M: 95±5;<br>L: 105±5;<br>XL: 115±5 | Width:<br>S: 86-89 /Pass<br>M: 95-97/ Pass<br>L: 106-108/ Pass<br>XL:115-117/ Pass |                                                                                             |               |
|                 |                                | Thickness (mm):<br>Finger: $\ge$ 0.08<br>Palm: $\ge$ 0.08    | Finger: 0.08-0.16/Pass<br>Palm: 0.08-0.13/Pass                                     |                                                                                             |               |
| ASTM<br>D5151   |                                | Watertightness<br>Test for<br>Detection of<br>Holes          | Meet the requirements of ASTM D5151<br>AQL 2.5                                     | 0/125/Pass                                                                                  |               |
| ASTM<br>D6124   |                                | Powder<br>Content                                            | Meet the requirements of ASTM D6124 < 2.0mg                                        | 0.12mg/Pass;                                                                                |               |
| ASTM<br>D412    | Physical<br>properties         | Before<br>Aging                                              | Tensile<br>Strength                                                                | $\ge$ 11MPa                                                                                 | 11 -18/Pass;  |
|                 |                                |                                                              | Ultimate<br>Elongation                                                             | $\ge$ 300%                                                                                  | 340-518/Pass; |
|                 |                                | After<br>Aging                                               | Tensile<br>Strength                                                                | $\ge$ 11MPa                                                                                 | 11-16/Pass;   |
|                 |                                |                                                              | Ultimate<br>Elongation                                                             | $\ge$ 300%                                                                                  | 315-438/Pass; |
| ISO<br>10993-11 | Acute Systemic<br>Toxicity     | Non- acute systemic<br>toxicity                              |                                                                                    | Under conditions of<br>the study, did not<br>show acute systemic<br>toxicity in vivo / Pass |               |

Table 2 - Summary of non-clinical performance testing

{7}------------------------------------------------

| ISO<br>10993-10 | Irritation    | Non-irritating  | Under the conditions<br>of the study, not an<br>irritant/ Pass |
|-----------------|---------------|-----------------|----------------------------------------------------------------|
| ISO<br>10993-10 | Sensitization | Non-sensitizing | Under conditions of<br>the study, not a<br>sensitizer./ Pass   |

## 8.0 _Summary of Clinical Testing

Clinical testing is not needed for this device.

# 9.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device NBR Synthetic Examination Gloves, is as safe, as effective, and performs as well as or better than the legally marketed predicated device.

---

**Source:** [https://fda.innolitics.com/device/K212436](https://fda.innolitics.com/device/K212436)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).