Brainlab Elements, Brainlab Elements Contouring, Brainlab Elements Fibertracking, Brainlab Elements Image Fusion, Brainlab Elements Image Fusion Angio
K212420 · Brainlab AG · LLZ · Dec 16, 2021 · Radiology
Device Facts
Record ID
K212420
Device Name
Brainlab Elements, Brainlab Elements Contouring, Brainlab Elements Fibertracking, Brainlab Elements Image Fusion, Brainlab Elements Image Fusion Angio
Applicant
Brainlab AG
Product Code
LLZ · Radiology
Decision Date
Dec 16, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K212420 · Dec 16, 2021
Brainlab Elements, Brainlab Elements Contouring, Brainlab Elements Fibertracking, Brainlab Elements Image Fusion, Brainlab Elements Image Fusion Angio
Brainlab AG
Retrospective patient data
Retrospective patient data was used in validation testing to compare the performance of modified algorithms in the subject device against the algorithms present in the predicate devices.
Brainlab Elements Contouring provides an interface with tools and views to outline, refine, combine and manipulate structures in patient image data. The generated 3D structures are not intended to create physical replicas used for diagnostic purposes. The device itself does not have clinical indications. Brainlab Elements Fibertracking is an application for the processing and visualization of cranial white matter tracts based on Diffusion Tensor Imaging (DTT) data for use in treatment planning procedures. The device itself does not have clinical indications. Brainlab Elements Image Fusion is an application for the co-registration of image data within medical procedures by using rigid and deformable registration methods. It is intended to align anatomical structures between data sets. The device itself does not have clinical indications. Brainlab Elements Image Fusion Angio is a software application that is intended to be used for the co-registration of cerebrovascular image data. The device itself does not have clinical indications.
Device Story
Brainlab Elements is a software suite for medical image processing used in hospital/surgical environments by clinicians for treatment planning. Inputs include DICOM medical images (e.g., DTI, DSA). The device performs image co-registration (rigid/deformable), automatic/manual segmentation of anatomical structures, 3D visualization, and fiber tract processing. Outputs are processed images, 3D models, and measurements displayed on a workstation. Clinicians use these outputs to delineate structures and plan surgical or radiotherapy interventions. The device aids decision-making by providing enhanced visualization and spatial alignment of patient data, potentially improving treatment accuracy.
Clinical Evidence
No clinical trials performed. Evidence consists of bench testing, including verification and validation of software against design specifications (IEC 62304, ISO 14971, IEC 62366). Validation included accuracy testing of co-registration, segmentation, and fiber tract generation using retrospective patient data and phantom data, with acceptance criteria derived from scientific literature.
Technological Characteristics
Software-based medical image management and processing system. Features include DICOM data handling, rigid/deformable co-registration, 3D visualization, and segmentation algorithms. Operates on standard hospital computing hardware. Software developed under major level of concern. Connectivity via PACS/storage media. No specific hardware materials or energy sources; purely software-based.
Indications for Use
Indicated for use by medical professionals (neurosurgeons, ENT/CMF surgeons, radiation oncologists, medical physicists) for processing, visualization, and analysis of medical images to support surgical or radiotherapy treatment planning. No specific patient population or contraindications defined.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.
Brainlab AG % Chiara Cunico Manager Regulatory Affairs Olof-Palme-Str. 9 Munich, Bayern 80809 GERMANY
December 16, 2021
# Re: K212420
Trade/Device Name: Brainlab Elements, Brainlab Elements Contouring, Brainlab Elements Fibertracking, Brainlab Elements Image Fusion, Brainlab Elements Image Fusion Angio Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ, JAK Dated: November 10, 2021 Received: November 15, 2021
Dear Chiara Cunico:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212420
#### Device Name
Brainlab Elements, Brainlab Elements Contouring (4.0), Brainlab Elements Fibertracking (2.0), Brainlab Elements Image Fusion (4.0), Brainlab Elements Image Fusion Angio (1.0)
#### Indications for Use (Describe)
Brainlab Elements Contouring provides an interface with tools and views to outline, refine, combine and manipulate structures in patient image data. The generated 3D structures are not intended to create physical replicas used for diagnostic purposes. The device itself does not have clinical indications.
Brainlab Elements Fibertracking is an application for the processing and visualization of cranial white matter tracts based on Diffusion Tensor Imaging (DTT) data for use in treatment planning procedures. The device itself does not have clinical indications.
Brainlab Elements Image Fusion is an application for the co-registration of image data within medical procedures by using rigid and deformable registration methods. It is intended to align anatomical structures between data sets. The device itself does not have clinical indications.
Brainlab Elements Image Fusion Angio is a software application that is intended to be used for the co-registration of cerebrovascular image data. The device itself does not have clinical indications.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image is an abstract design featuring a pattern of triangles in varying shades of blue. The triangles are arranged in a way that creates a sense of depth and movement. The color palette transitions from a darker blue on the left to a lighter, almost white, blue on the right. The overall effect is modern and geometric.
# RAINLAB
# 510(k) Summary
K212420
### November 10, 2021
| General Information | |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Brainlab AG; Olof-Palme-Str.9, 81829, Munich, Germany |
| Establishment Registration | 8043933 |
| Trade Name | Brainlab Elements, Brainlab Elements Contouring, Brainlab Elements<br>Fibertracking, Brainlab Elements Image Fusion, Brainlab Elements<br>Image Fusion Angio |
| Classification Name | System, image processing, radiological |
| Product Code | LLZ, JAK |
| Regulation Number | 892.2050 |
| Regulatory Class | Class II |
| Panel | Radiology |
| Predicate Device(s) | Mimics Medical (K183105) – Primary Predicate<br>iPlan (K113732) – Second predicate<br>Image Fusion (K170816) – Third predicate |
| Reference Device | Brainlab Elements Image Fusion Angio (K190042) |
| Contact Information | |
| Primary Contact | Alternate Contact |
| Chiara Cunico | Regulatory Affairs Brainlab |
| Manager RA | Phone: +49 89 99 15 68 0 |
| Phone: +49 89 99 15 68 0 | Fax: +49 89 99 15 68 5033 |
| Email: chiara.cunico@brainlab.com | Email: regulatory.affairs@brainlab.com |
# 1. Indication for Use
Brainlab Elements Contouring provides an interface with tools and views to outline, refine, combine and manipulate structures in patient image data. The generated 3D structures are not intended to create physical replicas used for diagnostic purposes. The device itself does not have clinical indications.
Brainlab Elements Fibertracking is an application for the processing and visualization of cranial white matter tracts based on Diffusion Tensor Imaging (DTI) data for use in treatment planning procedures. The device itself does not have clinical indications.
Brainlab Elements Image Fusion is an application for the co-registration of image data within medical procedures by using rigid and deformable registration methods. It is intended to align anatomical structures between data sets. The device itself does not have clinical indications.
Brainlab Elements Image Fusion Angio is a software application that is intended to be used for the co-registration of cerebrovascular image data. The device itself does not have clinical indications.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image is an abstract design featuring a pattern of interconnected triangles. The triangles vary in shades of blue, ranging from a deep, saturated blue on the left to a lighter, almost white blue on the right. White lines connect the triangles, creating a geometric, fragmented appearance.
### 2. Device Description
Brainlab Elements is a medical device for processing of medical images, that is used to support treatment planning of surgical or radio-therapeutical procedures.
The Brainlab Elements applications transfer DICOM data to and from picture archiving and communication systems (PACS) and other storage media devices. They include modules for 2D & 3D image viewing, image processing, image co-registration, image segmentation and 3D visualization of medical image data for treatment planning procedures.
Brainlab Elements main software functionalities include:
- Visualization of medical image data in DICOM format -
- Co-registration of different imaging modalities by using both rigid and deformable registration methods
- Processing of co-registered data to highlight differences between distinct scanning sequences or to assess the response to a treatment
- Contouring and delineation of objects and anatomical structures -
- -Automatic segmentation of anatomical structures
- Manipulation of objects and seqmented structures (e.g. splitting, mirroring, etc.) -
- -Measuring tools
- Co-reqistration of cerebrovascular image data -
- Visualization of Diffusion Tensor Imaging (DTI) based data and processing of such data to visualize e.g. cranial white matter tracts
The intended users are medical professionals. Typical users as defined in the use case descriptions are:
- -Image Guided Surgery: Neurosurgeons, Ear-Nose-Throat (ENT) surgeons and Cranio-Maxillofacial (CMF) surgeons including their assistants
- Radiotherapy: medical professionals who perform radiation treatment planning (medical physicists, radiation oncologists, dosimetrists, physicians, etc.)
The device shall be used in a hospital office environment or rooms appropriate for surgical interventions or radiotherapy planning.
#### 3. Substantial Equivalence
| ltem | Information | 510 (k) | Product Code;<br>Subsequent<br>Product Code(s) | Subject Device |
|--------------------------------|----------------|---------|------------------------------------------------|----------------------------------|
| Primary<br>Predicate<br>device | Mimics Medical | K183105 | LLZ | Brainlab Elements Contouring 4.0 |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image is an abstract design featuring a pattern of interconnected triangles in varying shades of blue and white. The triangles are arranged in a seemingly random fashion, creating a sense of depth and complexity. The color gradient transitions from a darker blue on the left to a lighter, almost white shade on the right, adding a sense of movement and visual interest. The overall effect is modern and geometric, with a clean and crisp aesthetic.
# BRAINLAB
| Item | Information | 510 (k) | Product Code;<br>Subsequent<br>Product Code(s) | Subject Device |
|-------------------------------|-------------------------------------------------------------------------------------------------|---------|------------------------------------------------|------------------------------------------------------------------------------|
| Second<br>Predicate<br>device | iPlan (iPlan Cranial, iPlan<br>Stereotaxy, iPlan ENT,<br>iPlan CMF, iPlan View,<br>iPlan Spine) | K113732 | JAK; LLZ | Brainlab Elements Fibertracking<br>2.0; Brainlab Elements<br>Countouring 4.0 |
| Third<br>Predicate<br>device | Image Fusion 3.0 | K170816 | LLZ; JAK | Brainlab Elements Image Fusion<br>4.0 |
| Reference<br>Device | Image Fusion Angio 1.0 | K190042 | LLZ; JAK | Brainlab Elements Image Fusion<br>Angio 1.0.1 |
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
At a high level, the subject and predicate devices are based on the following same technological elements:
#### Image Fusion and Image Fusion Angio:
- -Principle of image co-registration (rigid and deformable)
- -Verification tools
- Viewing of images and DICOM data -
#### Contouring:
- Object segmentation using manual, semi-automatic or fully automatic algorithms -
- 3D model calculation
- -Measuring functionalities
- Verification tools -
- Viewing of images and DICOM data -
#### Fibertracking:
- -Automatic DTI data processing
- Region-of-interest (ROI) definition -
- 3D visualization -
In addition, the subject device indications for use are equivalent to that of the predicate device.
The following technological differences exist between the subject and predicate device(s):
- New/modified user interface -
- Fibertracking algorithm improvements (compared to predicate device iPlan) -
- -Contouring: support of 2D-DSA images (compared to predicate device iPlan)
- -Functionality: The device shares the main functionalities of the predicate devices, but is organized in a different way and tools and views are slightly different.
Validation of differences compared to predicate devices:
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image is an abstract design featuring a series of interconnected triangles. The triangles are filled with varying shades of blue, creating a gradient effect from darker to lighter hues. White lines separate the triangles, adding contrast and definition to the geometric shapes.
- INLAB
- The new/modified user interface and views as well as the modified interactions with tools were subject of several formative and summative usability tests.
- -Validation tests using retrospective patient data and phantom data were performed to compare the results of the modified algorithm to the results of the algorithm which was present in the predicate device.
#### 4. Performance Data
Product specifications and the implementation of risk control measures have been tested in verification tests for the device according to IEC 62304 and ISO 14971.
Validation tests were performed to demonstrate that the products fulfill critical state of the art requirements:
- -Accuracy of co-registrations, generated by the devices Elements Image Fusion and Elements Image Fusion Angio
- -Accuracy of automatically segmented objects, generated by the device Elements Contouring
- Accuracy of fiber tracts, generated by the device Elements Fibertracking -
In all cases, acceptance criteria for the validation tests were derived from scientific literature.
Usability tests were performed to demonstrate the devices meet usability requirements as defined in IEC 62366.
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a major level of concern, since:
- Prior to mitigation of hazards, a failure of the Software could result in death or serious injury of the patient, because the software provides information that directly drives a decision regarding treatment or therapy
- A malfunction of, or a latent design flaw in, the software could lead to an erroneous diagnosis or delay in delivery of appropriate medical care that would likely lead to Minor Injury.
#### 5. Conclusion
The comparison of Brainlab Elements with the predicate devices shows that the Brainlab Elements have similar functionality, intended use, technological characteristics and users like the predicate device(s). Verification and validation activities ensured that the design specifications are met and that Brainlab Elements does not introduce new issues concerning safety and effectiveness. Hence, Brainlab Elements is substantially equivalent to the predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.