K212393 · Sis Medical AG · LOX · Mar 14, 2022 · Cardiovascular
Device Facts
Record ID
K212393
Device Name
OPN NC PTCA Dilatation Catheter
Applicant
Sis Medical AG
Product Code
LOX · Cardiovascular
Decision Date
Mar 14, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OPN NC PTCA Dilatation Catheter is intended for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The balloon dilatation catheter is also indicated for post deployment expansion of balloon expandable coronary stents.
Device Story
The OPN NC PTCA Dilatation Catheter is a sterile, single-use, rapid-exchange catheter used by interventional cardiologists in clinical settings. It features a distal non-compliant double-layer polyamide balloon and a coaxial polymer shaft. The device is positioned over a 0.014" guidewire to the target stenosis. An inflation device connects to the proximal luer hub to inflate the balloon, which exerts pressure to dilate the vessel or expand a stent. Radiopaque marker bands facilitate fluoroscopic positioning. The device operates at a nominal pressure of 10 atm with a rated burst pressure of 35 atm. By restoring vessel patency or ensuring proper stent apposition, the catheter improves myocardial perfusion and patient outcomes.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including dimensional verification, simulated use, compliance, coating integrity, particulate testing, inflation/deflation, corrosion, abrasion/puncture resistance, lubricity, flex/kink, torque, tensile strength, rated burst pressure, balloon fatigue, and radiopacity. Biocompatibility testing included cytotoxicity, intracutaneous reactivity, complement activation, pyrogenicity, sensitization, hemolysis, thrombogenicity, and acute systemic toxicity.
Indicated for patients requiring balloon dilatation of stenotic coronary arteries or bypass grafts to improve myocardial perfusion, and for post-deployment expansion of balloon-expandable coronary stents.
Regulatory Classification
Identification
Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.
Special Controls
*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 14, 2022
SIS Medical AG c/o H. Semih Oktay President CardioMed Device Consultants, LLC 1783 Forest Drive Suite 254 Annapolis, Maryland 21401
Re: K212393
Trade/Device Name: OPN NC PTCA Dilatation Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Regulatory Class: Class II Product Code: LOX Dated: February 7, 2022 Received: February 8, 2022
Dear H. Semih Oktay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K212393
Device Name
OPN NC PTCA Dilatation Catheter
#### Indications for Use (Describe)
The OPN NC PTCA Dilatation Catheter is intended for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The balloon dilatation catheter is also indicated for post deployment expansion of balloon expandable coronary stents.
| Type of Use (Select one or both, as applicable) | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <table><tr><td><div> <span>☑</span><span> Prescription Use (Part 21 CFR 801 Subpart D)</span> </div></td></tr><tr><td><div> <span>☐</span><span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div></td></tr></table> | <div> <span>☑</span><span> Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>☐</span><span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
| <div> <span>☑</span><span> Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | | |
| <div> <span>☐</span><span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary (as required by 21 CFR 807.92)
Image /page/3/Picture/3 description: The image shows the logo for SIS Medical. The logo consists of the letters "SIS" in black, followed by a pink and gray heart-shaped symbol. To the right of the heart is the word "MEDICAL" in black. The logo appears to be for a medical company.
Swiss | Interventional
# OPN NC PTCA Dilatation Catheter
#### 510(k) K212393
| Date Prepared: | March 4, 2022 |
|------------------------|------------------------------------------------------------------------------|
| Applicant: | SIS Medical AG<br>Hungerbüelstrasse 12a<br>CH-8500 Frauenfeld<br>Switzerland |
| Contact Name: | H. Semih Oktay, PhD |
| Title: | President, CardioMed Device Consultants |
| Email: | soktay@cardiomedllc.com |
| Telephone: | (410) 271-2088 |
| Fax: | (410) 674-2133 |
| Trade Name: | OPN NC PTCA Dilatation Catheter |
| Device Classification: | Class II per 21 CFR §870.5100 |
| Classification Name: | Catheters, Transluminal Coronary Angioplasty, Percutaneous |
| Product Code: | LOX |
| Predicate Devices: | NC Euphora Rapid Exchange Balloon Dilatation<br>Catheter, K141090 |
### INTENDED USE/INDICATIONS FOR USE:
The OPN NC PTCA Dilatation Catheter is intended for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The balloon dilatation catheter is also indicated for post deployment expansion of balloon expandable coronary stents.
## DEVICE DESCRIPTION:
The OPN NC PTCA Dilatation Catheter is a sterile, single use, rapid exchange catheter with a distal non-compliant double layer balloon attached to a flexible distal polymer shaft. The balloon is made from Polyamide and is available in
{5}------------------------------------------------
diameters of 1.5, 2.0, 2.5, 3.0, 3.5, 4.0 and 4.5mm and in balloon lengths of 10, 15 and 20mm.
The distal polymer shaft consists of a two coaxial lumens, with one lumen for inflation and the other to enable the use of a quide wire for positioning the catheter. The catheter is compatible with a 0.014" guidewire, which enters at the distal tip of the catheter and exists at the RX port.
The proximal portion of the catheter is formed from a hypotube containing the inflation lumen. At the proximal end, a hub with luer connector is attached to the hypotube, allowing connection to an inflation device. Radiopaque balloon marker bands enable accurate positioning of the device. Shaft markers for brachial and femoral techniques are also in place.
The OPN NC comes in a single useable length of 147cm. The device nominal pressure is 10 atm and the rated burst pressure is 35 atm.
## COMPARISON WITH PREDICATE DEVICES:
Comparison of the OPN NC Dilatation Catheter and predicate device show that the technological characteristics of the subject device such as the components, design, materials, sterilization method, shelf life and operating principle are similar to the currently marketed predicate and reference devices. The intended use of the subject device and the predicate are the same.
| | | Subject | Predicate |
|-----------------------------------|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name of Device | | OPN NC PTCA Dilatation<br>Catheter | NC Euphora Rapid<br>Exchange Balloon Dilatation<br>Catheter |
| Manufacturer | | SIS Medical AG | Medtronic Inc. |
| 510(k) | | K212393 | K141090 |
| Catheter Shaft<br>Characteristics | Indications for Use | Indicated for balloon dilatation of<br>the stenotic portion of the<br>coronary artery or bypass graft<br>for the purpose of improving<br>myocardial perfusion<br>The balloon dilatation catheter is<br>also indicated for post-<br>deployment expansion of balloon<br>expandable coronary stents | Indicated for balloon dilatation<br>of the stenotic portion of a<br>coronary artery or bypass graft<br>for the purpose of improving<br>myocardial perfusion<br>The balloon dilatation catheter is<br>also indicated for post-<br>deployment expansion of<br>balloon expandable stents |
| | Device Design | Rapid Exchange PTCA catheter | Rapid Exchange PTCA<br>catheter |
| | Radiopaque<br>Marker(s) | 2 Platinum/Iridium | 2 Platinum/Iridium |
| | Useable Length | 147cm | 142cm |
{6}------------------------------------------------
| | Subject | | Predicate |
|----------------------------|------------------------------|-------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Balloon<br>Characteristics | Name of Device | OPN NC PTCA Dilatation<br>Catheter | NC Euphora Rapid<br>Exchange Balloon Dilatation<br>Catheter |
| | Manufacturer | SIS Medical AG | Medtronic Inc. |
| | 510(k) | K212393 | K141090 |
| | Balloon Diameters | 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, and<br>4.5mm | 2.00, 2.25, 2.50, 2.75, 3.00,<br>3.25, 3.50, 3.75, 4.00, 4.50<br>and 5.00mm |
| | Balloon Lengths | 10, 15 and 20mm | 6, 8, 12, 15, 20 and 27mm |
| | Rated Burst<br>Pressure | 35atm | 20atm |
| | Hydrophilic coating | No | Coating present, type<br>unavailable |
| | Hydrophobic coating | Yes | unavailable |
| | Guide catheter compatibility | 1.5 – 3.5mm diameter:<br>6F (0.071" or 1.8mm)<br>4.0 - 4.5mm diameter:<br>7F (0.081" or 2.06mm) | Not known |
| | Guidewire compatibility | 0.014" (0.36mm) | 0.014" (0.36mm) |
| | Sterilization Method | EO | E-beam |
## NON-CLINICAL TESTING/PERFORMANCE DATA:
Non-clinical laboratory testing was performed on the OPN NC PTCA Dilatation Catheter to determine substantial equivalence. The following testing/assessments were performed:
- Dimensional Verification ●
- Simulated Use
- Compliance ●
- Coating Integrity ●
- Particulate ●
- Inflation/deflation ●
- Corrosion
- Abrasion/Puncture resistance ●
- Lubricity/pinch testing .
- Flex/Kink ●
- Torque ●
- Tensile Strength
- Rated burst pressure ●
- In-Stent rated burst pressure ●
- Balloon fatigue ●
- In-Stent balloon fatigue
- Radiopacity ●
The in vitro bench tests demonstrated that the OPN NC PTCA Dilation Catheter met all acceptance criteria and performed similarly to the predicate device. Performance data demonstrate that the device functions as intended and has substantially equivalent safety and effectiveness/performance outcomes to the predicate device.
### BIOCOMPATIBILITY:
Testing was performed to assess biocompatibility of the OPN NC PTCA Dilatation Catheter. The following tests were performed:
{7}------------------------------------------------
- Cytotoxicity ●
- Intracutaneous Reactivity ●
- Complement Activation (SC5b-9) ●
- Material Mediated Pyrogenicity
- Sensitization ●
- . Hemolysis (Direct and Extract)
- . Thrombogenicity
- . Acute Systemic Toxicity
The results from the testing performed showed the OPN NC PTCA Dilatation Catheter to be biocompatible.
# CONCLUSION:
The OPN NC PTCA Dilatation Catheter has the same intended use and the same or similar technological characteristics such as components, design, materials, and operating principles as the predicate device. Performance data demonstrates that the device functions as intended. The conclusions drawn from the nonclinical tests demonstrate that the OPN NC has substantially equivalent safety and effectiveness/performance outcomes to the legally marketed predicate device.
Therefore, the OPN NC PTCA Dilatation Catheter is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.