K212383 · Therenva Sas · OWB · Aug 31, 2021 · Radiology
Device Facts
Record ID
K212383
Device Name
EndoNaut
Applicant
Therenva Sas
Product Code
OWB · Radiology
Decision Date
Aug 31, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
AI/ML
Indications for Use
EndoNaut provides image guidance by overlaying pre-operative 3D vessel anatomy onto live fluoroscopic images in order to assist in the positioning of the guidewires, catheters and other endovascular devices.
Device Story
EndoNaut is an intraoperative navigation and guidance system for endovascular procedures. It inputs 3D preoperative CT scans (processed via EndoSize software) and live 2D intraoperative fluoroscopy/angiography. The system performs 2D-3D or 2D-2D image registration to overlay preoperative vascular anatomy onto live fluoroscopic images. It assists physicians in positioning guidewires, catheters, and endovascular devices. Used in clinical settings (OR/interventional suite) by physicians. The system includes an 'AI module' for 2D-3D fusion and a 'PAD module' for lower limb panorama creation. Output is displayed on a workstation monitor to guide navigation, potentially improving procedural accuracy and efficiency.
Clinical Evidence
Bench testing only. No clinical data was required or provided. Verification and validation included risk assessment, usability reviews, module verification, integration testing, interoperability testing with EndoSize, and performance/safety testing (electrical, EMC, functional, mechanical) conducted by a COFRAC accredited laboratory.
Technological Characteristics
System consists of EndoNaut Software and an EndoNaut Workstation (TS1CA2DS1-2). Workstation features 27" LCD main display, 15.6" touch screen secondary display, and mobile cart with isolating transformer. Connectivity via Ethernet and USB 3.0. Software is Class II, Moderate level of concern, compliant with IEC 62304, IEC 62366, and ISO 14971. Workstation compliant with IEC 60601-1, ANSI AAMI ES60601-1, IEC 60601-2, and IEC 60601-1-6.
Indications for Use
Indicated for the treatment of patients with endovascular diseases requiring procedures such as endovascular aortic aneurysm repair (AAA and TAA), angioplasty, stenting, and embolization in iliac arteries and corresponding veins. Indicated for endovascular procedures in the thorax, abdomen, pelvis, and lower limbs.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
August 31, 2021
Image /page/0/Picture/1 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
Therenva SAS % Cemil Göksu CEO 74F rue de Paris Rennes, 35000 FRANCE
Re: K212383
Trade/Device Name: EndoNaut Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB Dated: July 12, 2021 Received: August 2, 2021
Dear Cemil Göksu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows a digital signature. The signature is from Laurel M. Burk -S. The document was signed on August 31, 2021 at 10:53:59 -04'00'.
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212383
Device Name EndoNaut
#### Indications for Use (Describe)
EndoNaut is indicated for the treatment of patients with endovascular diseases and who needs, such as but not limited to the following examples:
- · endovascular aortic aneurysm repair (AAA and TAA),
- · angioplasty,
- · stenting.
- · embolization in iliac arteries and corresponding veins.
EndoNaut is indicated for endovascular procedures in the thorax, abdomen, pelvis and lower limbs.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for a company called "Therenva". The logo consists of a stylized symbol on the left and the company name on the right. The symbol is a teal-colored shape that resembles a sail or a stylized leaf, partially enclosed by a gray circle. The company name "Therenva" is written in gray, with the tagline "Share medical innovation" in a smaller font below the name.
# Section 5 - EndoNaut Special 510(K) Summary
This Special 510(k) summary is submitted in accordance with the requirements of 21 CFR Part 807.92.
Therenva SAS hereby submits this Special 510(k) to provide a notification submission for a modified device EndoNaut and evidence of substantial equivalence to to the Predicate Medical Device EndoNaut (K171829).
The accessories are also subject to this Special 510(k).
#### 1. Submitter information
| Manufacturer Name: | Therenva SAS<br>74F, rue de Paris<br>35000 Rennes<br>France |
|--------------------------------|-------------------------------------------------------------------------------------------------------|
| Contact Person: | Mrs Audrey Gallois, OA & RA Leader<br>Phone: +33 6 86 95 44 39<br>E-mail: audrey.gallois@therenva.com |
| Establishment Registration Nº: | 3011240766 |
| Date prepared: | 12-Jul-21 |
### 2. Device Identification
| Trade Name: | EndoNaut |
|----------------------------|----------------------------------------------|
| Regulation Name: | Interventional Fluoroscopic X-ray System |
| Regulatory Class: | Class II |
| Product Code: | OWB |
| Classification Regulation: | 21 CFR 892.1650 |
| Classification Panel: | Radiology |
| Accessories: | 1. Separate interventional tools workstation |
| | 2. EndoSize Software |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Therenva. The logo consists of a stylized teal-colored shape resembling a leaf or wing, connected to a gray circle. To the right of the circle is the word "Therenva" in gray, block letters. Below "Therenva" is the tagline "Share medical innovation" in a smaller, lighter gray font.
## 3. Predicate device and accessories
- Predicate Medical Device Software: EndoNaut (K171829) ●
- Predicate Accessories:
- Separate interventional tools workstation, TS1CA2DS1-1 (referred to in K171829 о 510(k) submission) and Cydar Medical's Cydar EV (K160088).
- EndoSize v3.1 (K160376) o
### 4. Description of the device
EndoNaut system is an imaging solution for intraoperative navigation and guidance tool for endovascular procedures (aorto-iliac and peripheral).
EndoNaut provides localization assistance by combining 3D preoperative scans and 2D intra-operative fluoroscopy imagery to help position guides, catheters and other vascular devices.
EndoNaut Software is interoperable with EndoSize which is a standalone Software designed and developed by Therenva to enable case planning strategy and device (endoprosthesis) selection before endovascular procedure. EndoSize is used by practitioners (in the preparation phase of the operating procedure) or by endoprosthesis manufacturers to visualize vascular structures and/or carry out an extract of the vascular structure from the preoperative CT scan. EndoSize is intended to be used for pre-operational planning and sizing. EndoSize is medical device software which obtained a substantial equivalence determination and FDA clearance through the CDRH premarket notification process (510(K)) (NºK160376).
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image is a logo for Therenva, a company that shares medical innovation. The logo features a stylized symbol on the left, consisting of a teal crescent shape intersecting with a gray circle. To the right of the symbol is the company name, "Therenva," in gray, block letters. Below the name is the tagline, "Share medical innovation," in a smaller, lighter gray font.
# 5. Comparison to the cleared (legacy) device and substantial equivalent discussion
| Medical Device Software Name | EndoNaut<br>(Predicate Device)<br>K171829 | EndoNaut<br>(Subject Device) | Comparable Properties an<br>Substantial<br>Equivalence Discussion |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Medical Device Software<br>Trade Name | EndoNaut | EndoNaut | Identical |
| Accessory 1 | Separate interventional tools<br>workstation | EndoNaut Workstation | Different.<br>Designation adjustment<br>only.<br>We have reclarified the qual-<br>ification of hardware versus<br>software. By definition,<br>here, the hardware (Endo-<br>Naut Workstation) is an ac-<br>cessory to the software med-<br>ical device. To better differ-<br>entiate what is applicable to<br>one and the other, we have<br>identified the software by<br>"EndoNaut Software" and<br>the hardware by "EndoNaut<br>Workstation". The Endo-<br>Naut Workstation alone is<br>not subject to any regulatory<br>submission. It is integrated<br>with the System submitted<br>for FDA approval under the<br>name, EndoNaut. |
| Accessory 1 Ref. | TS1CA2DS1-1 | TS1CA2DS1-2 | Different<br>Change of the Panel PC, the<br>Control Panel model and Ca-<br>blings.<br>Due to these changes, the<br>Product number<br>"TS1CA2DS1-2" has been<br>changed. New testings have<br>been performed. |
| Accessory 2 | EndoSize Software<br>(K160376) | EndoSize Software<br>(K160376) | Identical<br>EndoSize Software is an ac-<br>cessory that is intended to<br>supplement the performance<br>of EndoNaut. |
| Identification and traceability | Name: EndoNaut<br>No UDI | Trade name: EndoNaut<br>Common name: EndoNaut<br>System<br>UDI-DI: 3760262480046 | Different<br>No risk related to idenfica-<br>tion between older and<br>newer EndoNaut device.<br>EndoNaut is now referring to<br>the EndoNaut System while<br>the EndoNaut Software is<br>named EndoNaut Software |
| Medical Device Software | EndoNaut | EndoNaut | Comparable Properties an |
| Name | (Predicate Device) | (Subject Device) | Substantial |
| | K171829 | | Equivalence Discussion |
| | | | No unit of the predicate device was sold in the USA. |
| Manufacturer | Therenva SAS | Therenva SAS | Identical |
| Product Code | OWB | OWB | Identical |
| Regulation Number | 892.1650 | 892.1650 | Identical |
| Regulation Name | Interventional Fluoroscopic<br>X-Ray System | Interventional Fluoroscopic<br>X-Ray System | Identical |
| Software Safety class (62304) | B | B | Identical |
| Level of concern | Moderate | Moderate | Identical |
| Intended use | Intended use and indications<br>for use were mixed.<br>See below Indications for<br>use. | EndoNaut provides image<br>guidance by overlaying pre-<br>operative 3D vessel anatomy<br>onto live fluoroscopic im-<br>ages in order to assist in the<br>positioning of the guide-<br>wires, catheters and other<br>endovascular devices. | Different<br>A distinction between "In-<br>tended use" and "indications<br>for use" has been made based<br>on the recommendations of<br>the FDA Guidance " The<br>510(k) Program: Evaluating<br>Substantial Equivalence in<br>Premarket<br>Notifications<br>[510(k)]" (section IV, D, 1). |
| Indications for use | EndoNaut provides image<br>guidance by overlaying pre-<br>operative 3D vessel anatomy<br>onto live fluoroscopic im-<br>ages in order to assist in the<br>positioning of the guide-<br>wires, catheters and other<br>endovascular devices.<br>EndoNaut is intended to as-<br>sist endovascular procedures<br>in the thorax, abdomen,<br>neck, pelvis and lower limbs.<br>Suitable procedures include<br>(but not limited to) endovas-<br>cular aortic aneurysm repair<br>(AAA and TAA), angio-<br>plasty, stenting and emboli-<br>zation in iliac arteries and<br>corresponding veins.<br>EndoNaut is not intended for<br>use in the X-ray guided pro-<br>cedures in the liver, kidneys<br>or pelvic organs. | EndoNaut is indicated for<br>the treatment of patients with<br>endovascular diseases and<br>who needs for example<br>(without this list being re-<br>strictive):<br>endovascular aortic an-<br>eurysm repair (AAA<br>and TAA),angioplasty,stenting,embolization in iliac ar-<br>teries and corresponding<br>veins.EndoNaut is indicated for<br>endovascular procedures in<br>the thorax, abdomen, pelvis<br>and lower limbs. | Different<br>A distinction between "In-<br>tended use" and "indications<br>for use" has been made based<br>on the recommendations of<br>the FDA Guidance " The<br>510(k) Program: Evaluating<br>Substantial Equivalence in<br>Premarket<br>Notifications<br>[510(k)]" (section IV, D, 1). |
| Medical Device Software<br>Name | EndoNaut<br>(Predicate Device)<br>K171829 | EndoNaut<br>(Subject Device) | Comparable Properties and<br>Substantial<br>Equivalence Discussion |
| Labelling | 1 label in the "about" section<br>for the Medical Device Soft-<br>ware with the Trade Name<br>"EndoNaut"<br>+<br>1 label for the Separate inter-<br>ventional tools workstation<br>with the Trade Name "Endo-<br>Naut" and Product Number<br>= TS1CA2DS1-1 + unique<br>Serial (production) Number | 1 label in the "about" section<br>for the Medical Device Soft-<br>ware with the Trade Name<br>"EndoNaut Software".<br>+<br>1 label on the workstation<br>with the Trade Name "Endo-<br>Naut" (for the whole Sys-<br>tem) and Workstation model<br>Number = TS1CA2DS1-2 +<br>unique Serial (production)<br>Number (for the whole Sys-<br>tem) | Different<br>Each component of the En-<br>doNaut System has its<br>proper labelling.<br><br>Complete review of Label-<br>ling design. Use of symbols<br>of ISO 15223-1.<br>Addition of UDI.<br><br>Identical<br>The system has a unique<br>identifier (S/N). |
| Directions for use (User<br>Guide(s)) | 1 User guide for EndoNaut | 1 User Guide for EndoNaut<br>Software<br>1 User Guide for EndoNaut<br>Workstation +<br>1 addendum for informing<br>about the conditions govern-<br>ing the marketing of Endo-<br>Naut. | Different<br>The directions for use have<br>been enhanced to clearly dis-<br>tinguish the instructions spe-<br>cific to the medical device<br>Software, "EndoNaut SW"<br>from the instructions specific<br>to the EndoNaut Work-<br>station, accessory to Endo-<br>Naut SW. |
| Hardware compatibility | Software only product; runs<br>on a separate interventional<br>tools (imaging) workstation. | EndoNaut Software is the<br>class II medical device Soft-<br>ware which runs on a sepa-<br>rate interventional tools (im-<br>aging) workstation, the so-<br>named EndoNaut Work-<br>station which is the acces-<br>sory of the medical device<br>software. | Different<br>EndoNaut (legacy device)<br>was intended to be sold with<br>and installed on the separate<br>interventional tools work-<br>station (ref: TS1CA2DS1-1)<br>or installed on any hardware<br>meeting the minimum re-<br>quirements.<br>The conditions for installing<br>the software on the work-<br>station are identical.<br>Only the model number of<br>the workstation has changed<br>(ref: TS1CA2DS1-2). |
| Software Operating System | Win Pro 7 SP1x64<br>Win Pro 10 64bits | Win Pro 10 64bits | Different<br>Support for Windows 7<br>ended on January 14, 2020. |
| Medical Device Software<br>Name | EndoNaut<br>(Predicate Device)<br>K171829 | EndoNaut<br>(Subject Device) | Comparable Properties an<br>Substantial<br>Equivalence Discussion |
| Software interoperability | EndoNaut software requires<br>the use of EndoSize software<br>(K160376) to prepare patient<br>data and perform preopera-<br>tive sizing.<br>Data imported from En-<br>doSize include 3D volume,<br>preoperative images, sizing<br>report (comments and meas-<br>urements), and snapshots<br>taken during sizing. | EndoNaut requires the use of<br>EndoSize<br>software<br>(K160376) to prepare patient<br>data and perform preopera-<br>tive sizing.<br>Data imported from En-<br>doSize include 3D volume,<br>preoperative images, sizing<br>report (comments and meas-<br>urements), and snapshots<br>taken during sizing. | Similar<br>Some clarifications are<br>made.<br>Preoperative data is a man-<br>datory input for the use of AI<br>module, but it is not for the<br>PAD module.<br>Preoperative data include<br>pre-op CT images and sizing<br>report (in case of AI module<br>use).<br>The stent placement strategy<br>and sizing are usually per-<br>formed pre-operatively via<br>the use of software devices<br>such as EndoSize.<br>Alternatives to EndoSize ex-<br>ist: other sizing or visualiza-<br>tion software. In such cases,<br>the data is then printed on<br>paper and used in the operat-<br>ing room as is. |
| Data management | The user can import and<br>manage patient data within<br>the software. Patient data in-<br>clude pre-op CT images and<br>sizing report. | The user can import and<br>manage patient data within<br>the software. Patient data in-<br>clude pre-op CT images and<br>sizing report. | Similar<br>EndoNaut Software now en-<br>ables import and export of<br>data with a PACS. |
| Visualization | Intra-operative fluoroscopy<br>or angiography, pre-opera-<br>tive CT scan image, pre-op-<br>erative 3D scanner volume<br>reconstruction (if any in case<br>of PAD module).<br>AI module only:<br>Before and during the inter-<br>vention, the user can access<br>information from pre-opera-<br>tive sizing report such as pre-<br>op CT images, measure-<br>ments, comments, snap-<br>shots and strategy. | Intra-operative fluoroscopy<br>or angiography, pre-opera-<br>tive CT scan image, pre-op-<br>erative 3D scanner volume<br>reconstruction (if any in case<br>of PAD module).<br>AI module only:<br>Before and during the inter-<br>vention, the user can access<br>information from pre-opera-<br>tive sizing report such as pre-<br>op CT images, measurements,<br>comments, snap-<br>shots and strategy. | Identical |
| Export | Take and export snapshots.<br>Export panoramas in case of<br>PAD module. | Take and export snapshots.<br>Export panoramas in case of<br>PAD module. | Identical |
| Medical Device Software<br>Name | EndoNaut<br>(Predicate Device)<br>K171829 | EndoNaut<br>(Subject Device) | Comparable Properties an<br>Substantial<br>Equivalence Discussion |
| 3D-2D / 2D-2D Registration | AI module:<br>Display 2D-3D fusion: 3D<br>volume pre-op overlay on<br>per-op 2D fluoroscopy.<br>Semi-automatic registration<br>(automatic or manual initial-<br>ization, automatic computa-<br>tion and manual validation). | AI module:<br>Display 2D-3D fusion: 3D<br>volume pre-op overlay on<br>per-op 2D fluoroscopy.<br>Semi-automatic registration<br>(automatic or manual initial-<br>ization, automatic computa-<br>tion and manual validation). | Identical |
| | Lower limbs module:<br>Panorama creation: Acquisi-<br>tion and save of fluoroscopy<br>and angiography stage by<br>stage keeping the same C-<br>Arm orientation.<br>Display 2D-2D fusion: 2D<br>pre-op angiographic overlay<br>on per-op 2D fluoroscopy.<br>Synchronization between<br>current per-op 2D fluoros-<br>copy and 2D fluoroscopy<br>from recorded panorama. | PAD (lower limbs) module:<br>Panorama creation: Acquisi-<br>tion and save of fluoroscopy<br>and angiography stage by<br>stage keeping the same C-<br>Arm orientation.<br>Display 2D-2D fusion: 2D<br>pre-op angiographic overlay<br>on per-op 2D fluoroscopy.<br>Synchronization between<br>current per-op 2D fluoros-<br>copy and 2D fluoroscopy<br>from recorded panorama. | |
| Dynamic update on C-arm /<br>table / patient motion | Automatic motion<br>detection; registration is<br>updated manually. | Automatic motion detection<br>Registration: auto-<br>matic/manual initialization<br>and manual user validation. | Different<br>Improvement of semi-auto-<br>matic registration (already<br>existing requirement). The<br>change does not significantly<br>affect the use of the device.<br>No new risks or possible er-<br>rors were detected or identi-<br>fied. New clinical data were<br>not necessary. V&V activi-<br>ties were performed and suc-<br>cessful. No additional ques-<br>tions raised for safety and ef-<br>fectiveness. |
| Tools | AI module:<br>Draw markers on intra-oper-<br>ative images, locate/track<br>points between per-op and<br>pre-op images, and take<br>measurements on pre-op CT<br>scan images and fusion<br>view.<br>Lower limbs module:<br>Draw lesions markers on<br>panorama (stenosis and | AI module:<br>Draw markers on intra-oper-<br>ative images, locate/track<br>points between per-op and<br>pre-op images, and take<br>measurements on pre-op CT<br>scan images and fusion<br>view.<br>PAD (lower limbs) module:<br>Draw lesions markers on<br>panorama (stenosis and | Identical…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.