← Product Code [LNH](/productcode/LNH) · K212367

# IMRIS iMRI 3T V (K212367)

_Deerfield Imaging, Inc. · LNH · Aug 23, 2021 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K212367

## Device Facts

- **Applicant:** Deerfield Imaging, Inc.
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Aug 23, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The IMRIS iMRI 3T V is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the inages and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The IMRIS iMRI 3T V system may also be used for imaging during intra-operative and interventional procedures when performed with MR safe devices or MR conditional devices approved for use with the MR scanner. The IMRIS iMRI 3T V MRI systems may also be used for imaging in a multi-room suite.

## Device Story

IMRIS iMRI 3T V is a traditional 3T MRI system mounted on an overhead rail system for intra-operative and multi-room use. Magnet resides in a Diagnostic Room (DR) and moves into an Operating Room (OR) via sliding doors. System components include Siemens MAGNETOM Vida MRI, Magnet Mover Assembly, RF coils, application platform, OR table, head fixation device, IMRISeye, and horseshoe headrest. Operated by radiologists and surgeons to acquire high-resolution images pre-incision, intraoperatively, and post-closure. Output consists of cross-sectional images, spectroscopic images, and derived physical parameters. Physician interprets images to assist in clinical diagnosis and surgical decision-making. Benefits include real-time imaging capability within the surgical environment.

## Clinical Evidence

Bench testing only. Verification and validation included compliance with IEC 60601-1, IEC 60601-2, and IEC 60601-2-33. Board-certified radiologist confirmed image quality is sufficient for diagnostic use. No clinical trial data presented.

## Technological Characteristics

3T MRI system; overhead rail-mounted magnet mover assembly; Siemens MAGNETOM Vida core; RF coils; OR table assembly; head fixation device; IMRISeye; horseshoe headrest. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33, ISO 14971, IEC 62366-1, IEC 62304, NEMA MS 4, MS 9, MS 6, ISO 10993-1. Connectivity: Networked/multi-room suite. Software: Siemens syngo MR.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- iMRI 3T S ([K133692](/device/K133692.md))

## Reference Devices

- MAGNETOM Vida (with syngo MR XA20A) ([K192924](/device/K192924.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 23, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name in a sans-serif font.

Deerfield Imaging, Inc. dba IMRIS % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive. Suite 510k SAINT PAUL MN 55114

## Re: K212367

Trade/Device Name: IMRIS iMRI 3T V Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, LNI, MOS Dated: July 29, 2021 Received: July 30, 2021

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

Device Name IMRIS iMRI 3T V System

#### Indications for Use (Describe)

The IMRIS iMRI 3T V is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body or extremities.

Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the inages and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

The IMRIS iMRI 3T V system may also be used for imaging during intra-operative and interventional procedures when performed with MR safe devices or MR conditional devices approved for use with the MR scanner.

The IMRIS iMRI 3T V MRI systems may also be used for imaging in a multi-room suite.

Type of Use (Select one or both, as applicable)

| <div> <span> <b>                          X Prescription Use (Part 21 CFR 801 Subpart D)                     </b> </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <b>                           Over-The-Counter Use (21 CFR 801 Subpart C)                     </b> </span> </div>   |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for IMRIS. The word "IMRIS" is in bold, black, sans-serif font. To the right of the word is a blue circle with two horizontal lines inside of it. The circle is slightly offset from the word.

# 510(k) Summary

This 510(k) summary is submitted in accordance with the requirements of 21 C.F.R. Part §807.92.

# 1. Submitter's name and address

| Submitter:                | Deerfield Imaging, Inc. dba IMRIS             |
|---------------------------|-----------------------------------------------|
|                           | 5101 Shady Oak Rd                             |
|                           | Minnetonka, MN 55343                          |
|                           | USA                                           |
|                           | Establishment Registration Number: 3010326005 |
| Date Prepared:            | May 3, 2021                                   |
| Contact Person:           | Tracy Brinkmeyer                              |
|                           | VP Quality and Regulatory                     |
| Telephone:                | (612) 412-6330                                |
| Fax:                      | (866) 992-3224 Email: tbrinkmeyer@imris.com   |
| Alternate Contact Person: | Sanjay Shah, P. Eng.                          |
|                           | Regulatory Affairs Consultant                 |
| Telephone:                | (612) 812-6462                                |
| Email:                    | sshah@imris.com                               |

| 2. Device Name and Classification |                                                               |
|-----------------------------------|---------------------------------------------------------------|
| Trade Name:                       | IMRIS iMRI 3T V                                               |
| Common Name:                      | MRDD (Magnetic Resonance Diagnostic Device)                   |
| Classification Name:              | System, Nuclear Magnetic Resonance Imaging (21 CFR §892.1000) |
| Product Code:                     | Primary: LNH<br>Secondary: LNI, MOS                           |

{4}------------------------------------------------

## 3. Predicate Devices:

| Predicate Device Name | 510(K)  | Decision<br>Date | Product<br>Code | Manufacturer |
|-----------------------|---------|------------------|-----------------|--------------|
| iMRI 3T S             | K133692 | Feb 10, 2014     | LNH             | IMRIS        |

| Reference Device Name                  | 510(k)  | Decision<br>Date  | Product<br>Code | Manufacturer               |
|----------------------------------------|---------|-------------------|-----------------|----------------------------|
| MAGNETOM Vida<br>(with syngo MR XA20A) | K192924 | March 11,<br>2020 | LNH,<br>LNI,MOS | Siemens Healthcare<br>GmbH |

#### Device Description: 4.

The IMRIS iMRI 3T V is a traditional MRI unit that has been suspended on an overhead rail system, and is designed to operate inside an RF shielded room to facilitate intra-operative and multi-room use. The magnet is normally situated in a Diagnostic Room (DR) until imaging is requested. The system retains all standard diagnostic features of an MRI system in the DR. The DR is separated from the intra-operative Operating Room (OR) by sliding doors that are part of the facility structure.

The IMRIS iMRI 3T V is a tool for radiologists and surgeons, used to acquire images for diagnostic, intraoperative, or interventional procedures. For OR purposes, high-resolution images can be obtained immediately prior to surgical incision, intraoperatively, and after wound closure. The IMRIS iMRI 3T V is based on the IMRIS iMRI 3T S cleared under 510(k) K133692 and the Siemens MAGNETOM Vida MRI system cleared under 510(k) K192924. The major components of the IMRIS iMRI 3T V system are: the Siemens MAGNETOM Vida MRI system with minor modifications, IMRIS Magnet Mover System, RF coils, Application platform, OR Table Assembly, Head fixation device, IMRISeye, and Horseshoe headrest.

#### 5. Indications for Use:

The IMRIS iMRI 3T V is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body or extremities.

Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for IMRIS. The text "IMRIS" is in bold, black font. To the right of the text is a blue circle with two horizontal white lines inside.

#### IMRIS iMRI 3T V

The IMRIS iMRI 3T V system may also be used for imaging during intra-operative and interventional procedures when performed with MR safe devices or MR conditional devices approved for use with the MR scanner.

The IMRIS iMRI 3T V MRI systems may also be used for imaging in a multi-room suite.

- Comparison with Predicate Devices: 6.
The IMRIS iMRI 3T V is based on the IMRIS iMRI 3T S cleared under 510(k) K133692 and the Siemens MAGNETOM Vida MRI system cleared under 510(k) K 192924. There are no changes made to the Siemens syngo® MR software in the IMRIS iMRI 3T V system. The IMRIS iMRI 3T V intra-operative features, including the Magnet Mover Assembly, OR Patient Table, Intra-operative Coils, Application platform, IMRISeye, and Head Fixation Device are substantially equivalent to the same intra-operative features of the predicate iMRI 3T S system. The IMRIS iMRI 3T V does not raise any new safety or effectiveness issues related to the use of a moving MRI system in an intra-operative setting.

- 7. Standards:

| Recognition<br>Number | Product Area             | Reference<br>Number and<br>Date                                                 | Title of Standard                                                                                                                                                                                          | Standards<br>Development<br>Organization |
|-----------------------|--------------------------|---------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| 19-4                  | General II (ES/<br>EMC)  | ES606011:2005/(R)2012<br>and A1:2012,<br>C1:2009/(R)2012 and<br>A2:2010/(R)2012 | Medical electrical equipment<br>- Part 1: general<br>requirements for basic safety<br>and essential performance                                                                                            | ANSI /AAMI                               |
| 19-8                  | General II (ES/<br>EMC)  | 60601-1-2, Ed.<br>4.0:2014-02                                                   | Medical electrical equipment<br>- Part 1-2: General<br>requirements for basic safety<br>and essential performance -<br>Collateral Standard:<br>Electromagnetic<br>disturbances - Requirements<br>and tests | IEC                                      |
| 12-295                | Radiology                | 60601-2-33, Ed.<br>B:2015                                                       | Medical electrical equipment<br>- Part 2-33: Particular<br>requirements for the basic<br>safety and essential<br>performance of magnetic<br>resonance equipment for<br>medical diagnosis                   | IEC                                      |
| TIATKI'S IIVIRI 21  V |                          |                                                                                 |                                                                                                                                                                                                            |                                          |
| Recognition<br>Number | Product Area             | Reference<br>Number and<br>Date                                                 | Title of Standard                                                                                                                                                                                          | Standards<br>Development<br>Organization |
| 5-125                 | General I (QS/<br>RM)    | 14971:2019                                                                      | Medical devices -<br>Application of risk<br>management to medical<br>devices                                                                                                                               | ISO                                      |
| 5-129                 | General I (QS/<br>RM)    | 62366-1:2015+<br>AMD1:2020                                                      | Medical devices -<br>Application of usability<br>engineering to medical<br>devices, including<br>Amendment 1                                                                                               | ANSI<br>AAMI<br>IEC                      |
| 13-79                 | Software/<br>Informatics | 62304: Edition 1.1 2015-<br>06                                                  | Medical device software -<br>Software life cycle processes                                                                                                                                                 | IEC                                      |
| 12-232                | Radiology                | MS 4:2010                                                                       | Acoustic Noise<br>Measurement Procedure for<br>Diagnosing<br>Magnetic Resonance<br>Imaging<br>Devices                                                                                                      | NEMA                                     |
| 12-288                | Radiology                | MS 9:2008<br>(R2014)                                                            | Standards Publication<br>Characterization of Phased<br>Array<br>Coils for Diagnostic<br>Magnetic<br>Resonance Images                                                                                       | NEMA                                     |
| 12-195                | Radiology                | MS 6-2008 (R2014)                                                               | Determination of Signal-to-<br>Noise<br>Ratio and Image Uniformity<br>for Single-Channel Non-<br>Volume Coils in Diagnostic<br>MR Imaging                                                                  | NEMA                                     |
| 2-220                 | Biocompatibility         | 10993- 1:2009<br>/(R)2013                                                       | Biological evaluation of<br>medical devices - Part 1:<br>evaluation and testing within<br>a risk management process                                                                                        | AAMI<br>ANSI<br>ISO                      |

The IMRIS iMRI 3T V conforms to the following FDA recognized consensus standards:

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for IMRIS. The text "IMRIS" is in bold, black, sans-serif font. To the right of the text is a blue circle with two horizontal lines through it. The circle is slightly offset to the right of the text.

510(k) Premarket Notification IMBIS IMBI

# 8. Summary of Studies:

The IMRIS iMRI 3T V has been designed to provide MRI imaging in an intra-operative setting in the same manner as the predicate iMRI 3T S System. The IMRIS iMRI 3T V intra-operative features, including the Magnet Mover Assembly, RF Coils, Application platform, OR Patient Table, Head fixation device, IMRISeye, and Horseshoe headrest are substantially equivalent to the same intra-operative features of the 

{7}------------------------------------------------

predicate iMRI 3T S. The IMRIS iMRI 3T V does not raise any new safety or effectiveness issues related to the use of a moving MRI system in an intra-operative setting.

The IMRIS iMRI 3T V does not raise any new safety issues related to static magnetic field effects, changing magnetic field effects, RF heating, acoustic noise, effectiveness issues related to specification volume, signal to noise, image uniformity, and geometric distortion, slice profile, thickness, and gap, or high contrast spatial resolution.

The IMRIS iMRI 3T V verification and validation support a determination of substantial equivalence.

#### 9. Summary of non-clinical data

Design Verification and Validation Test (Bench Testing)

The IMRIS iMRI 3T V system passed the following tests and meets product specifications. IMRIS has performed a number of V&V tests as that includes the following:

- IEC 60601-1 compliance .
- IEC 60601-2 compliance
- . IEC 60601-2-33 compliance
- . Board certified Radiologist confirmation that images produced by the device are of sufficient quality for diagnostic use.

#### 10. Conclusion:

The IMRIS iMRI 3T V has the same intended use and same basic technological characteristics as the predicate device, iMRI 3T S.. While there are some differences in technological characteristics/features compared to the predicate device, the differences have been tested and the conclusions from all verification and validation data suggest that the features bear an equivalent safety and performance profile to that of the predicate device.

The IMRIS iMRI 3T V is substantially equivalent to the predicate device iMRI 3T S (K133692) based on the included studies and analysis.

---

**Source:** [https://fda.innolitics.com/device/K212367](https://fda.innolitics.com/device/K212367)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
