Exactech® Equinoxe® Laser Cage Glenoid

K212356 · Exactech, Inc. · KWS · Mar 4, 2022 · Orthopedic

Device Facts

Record IDK212356
Device NameExactech® Equinoxe® Laser Cage Glenoid
ApplicantExactech, Inc.
Product CodeKWS · Orthopedic
Decision DateMar 4, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3660
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemi-arthroplasty is determined by the surgeon to be the preferred method of treatment. - The cemented primary humeral stem, long/revision stem, fracture stems, and all Equinoxe glenoids are intended for cemented fixation. - The press-fit humeral stems are intended for press-fit applications but may be used with bone cement at the discretion the surgeon. - The reverse humeral components are intended to be used in cemented applications or in revision cases when the humeral component is well-fixed/stable, as deemed by the orthopaedic surgeon. - Humeral Heads are intended for use in cemented and press-fit applications.

Device Story

The Exactech Equinoxe Laser Cage Glenoid is a shoulder prosthesis component used in total shoulder arthroplasty. It consists of an Ultra High Molecular Weight Polyethylene (UHMWPE) body molded onto Ti-6Al-4V peripheral pegs and a central cage. The metal components are additively manufactured using direct metal printing (DMP) technology. The device is implanted by an orthopaedic surgeon in a clinical setting to replace the glenoid surface of the shoulder joint. It is intended for cemented fixation. The device functions as a mechanical bearing surface to restore joint mobility and reduce pain in patients with degenerative joint disease or fractures. It is provided sterile for single use.

Clinical Evidence

Bench testing only. Testing included porous structure characterization, glenoid fixation, peg shear resistance, peg pull-off, and peg bending fatigue. Bacterial endotoxin testing was performed in accordance with USP <161>, USP <85>, and ANSI/AAMI ST72.

Technological Characteristics

Materials: UHMWPE body, Ti-6Al-4V peripheral pegs and central cage. Manufacturing: Additive manufacturing (direct metal printing). Form factor: Four sizes, four augment angle options. Fixation: Cemented. Sterilization: Provided sterile for single use. Conforms to recognized performance standards for shoulder joint metal/polymer semi-constrained cemented prostheses.

Indications for Use

Indicated for skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint requiring total or hemi-arthroplasty. Specific indications include rheumatoid arthritis, osteoarthritis, osteonecrosis, post-traumatic degenerative problems, congenital abnormalities, humeral head fracture with displacement, and revision of failed prostheses. Reverse shoulder system indicated for degenerative diseases with grossly deficient, irreparable rotator cuff or failed joint replacement with superior migration of the humeral head.

Regulatory Classification

Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's: (i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” (iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,” (iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and (v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” (2) International Organization for Standardization's (ISO): (i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,” (ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,” (iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,” (iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,” (v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,” (vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and (vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and (3) American Society for Testing and Materials': (i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,” (ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,” (iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,” (iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,” (v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,” (vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,” (vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and (viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ March 4, 2022 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Exactech, Inc. Kenneth Maxwell Sr. Regulatory Affairs Specialist 2320 NW 66th Ct. Gainesville, Florida 32653 Re: K212356 Trade/Device Name: Exactech® Equinoxe® Laser Cage Glenoid Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWS Dated: February 1, 2022 Received: February 2, 2022 Dear Kenneth Maxwell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Lixin Liu. Ph.D. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ | 510(k) Number (if known) | K212356 | |--------------------------|----------------------------------------| | Device Name | Exactech® Equinoxe® Laser Cage Glenoid | Indications for Use (Describe) The Equinoxe Shoulder System is indicated for use individuals with degenerative diseases or fractures of the glenolumeral joint where total or hemi-arthroplasty is determined by the preferred method of treatment. - The cemented primary humeral stem, longrevision stems, and all Equinoxe glenoids are intended for cemented fixation. - The press-fit humeral stems are intended for press-fit applications but may be used with bone cement at the surgeon. - . The reverse humeral components are intended to be used in cemented applications or in revision cases when the humeral component is well-fixed/stable, as deemed by the orthopaedic surgeon. - Humeral Heads are intended for use in cemented and press-fit applications. Clinical indications for the PRIMARY (P), LONG/REVISION (L), and FRACTURE (F) humeral components are as follows: | P | L | F | Indications | |---|---|---|------------------------------------------------------------------------------------------------------------------------| | √ | √ | | Rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic degenerative problems | | √ | √ | | Congenital abnormalities in the skeletally mature | | √ | | | Primary and secondary necrosis of the humeral head. | | √ | | √ | Humeral head fracture with displacement of the tuberosities | | √ | √ | | Pathologies where arthrodesis or resectional arthroplasty of the humeral head are not acceptable | | √ | √ | | Revisions of humeral prostheses when other treatments or devices have failed (where adequate fixation can be achieved) | | | | √ | Displaced three-part and four-part upper humeral fractures | | | √ | | Spiral and other fractures of the mid-humerus (in combination with glenohumeral degenerative diseases) | | | √ | | Revision of failed previous reconstructions when distal anchorage is required | | √ | √ | | To restore mobility from previous procedures (e.g. previous fusion) | The Equinoxe Reverse Shoulder System is in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head. The Equinoxe Platform Fracture Stem is in skeletally mature individuals with acute fracture of the proximal humerus and displacement of the tuberosities, displaced 3- and 4-part fractures (hemi-athroplasty), or acute fracture of the proximal humerus with failure of the glenohumeral joint (primary total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is also indicated for acute fracture of the proximation with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff resulting in superil head (reverse total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is indicated for cemented use only. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ #### This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, includingthe time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collections for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov " An agency may not conduct or sponsor, and a person is not respond to, a collection of informationunless it displays a currently valid OMB number." {4}------------------------------------------------ ## Exactech® Equinoxe® Laser Cage Glenoid Traditional 510(k) - 510(k) Summarv | Company | Exactech, Inc.<br>2320 NW 66th Court<br>Gainesville, FL 32653 | |----------------------|---------------------------------------------------------------------------------------------------------------| | Contact Person | Kenneth C. Maxwell II<br>Senior Regulatory Affairs Specialist<br>Phone: (352) 377-1140<br>Fax: (352) 378-2617 | | Date | 4 March 2022 | | Proprietary Name | Exactech® Equinoxe® Laser Cage Glenoid | | Common Name | Shoulder prosthesis, glenoid component | | Classification Name | Shoulder joint metal/polymer semi-constrained cemented<br>prosthesis (Class II per 21 CFR §888.3660) | | Product Code | KWS | | Classification Panel | Orthopedic | ## DEVICE DESCRIPTION The Equinoxe Laser Cage Glenoids are intended to be used with Exactech Equinoxe Humeral Head components in Total Shoulder Arthroplasty. The Laser Cage Glenoids are composed of an Ultra High Molecular Weight Polyethylene (UHMWPE) body molded onto Ti-6Al-4V peripheral pegs and a central cage. The central cage and peripheral pegs of the implant are additively manufactured using direct metal printing (DMP) technology. The Laser Cage Glenoids are available in four sizes and four augment angle options. ## INDICATIONS FOR USE The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemi-arthroplasty is determined by the surgeon to be the preferred method of treatment. - The cemented primary humeral stem, long/revision stem, fracture stems, and all Equinoxe glenoids are intended for cemented fixation. - . The press-fit humeral stems are intended for press-fit applications but may be used with bone cement at the discretion the surgeon. - . The reverse humeral components are intended to be used in cemented applications or in revision cases when the humeral component is well-fixed/stable, as deemed by the orthopaedic surgeon. - Humeral Heads are intended for use in cemented and press-fit applications. ● Clinical indications for the PRIMARY (P), LONG/REVISION (L), and FRACTURE (F) humeral components are as follows: | P | L | F | Indications | |---|---|---|---------------------------------------------------------------------------------------------| | √ | √ | | Rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic degenerative problems | | √ | √ | | Congenital abnormalities in the skeletally mature | | √ | | | Primary and secondary necrosis of the humeral head. | | √ | | √ | Humeral head fracture with displacement of the tuberosities | {5}------------------------------------------------ ## Exactech® Equinoxe® Laser Cage Glenoid Traditional 510(k) - 510(k) Summarv | √ | √ | Pathologies where arthrodesis or resectional arthroplasty of the humeral head are not acceptable | | |---|---|------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------| | √ | √ | Revisions of humeral prostheses when other treatments or devices have failed (where adequate fixation can be achieved) | | | | | √ | Displaced three-part and four-part upper humeral fractures | | | √ | | Spiral and other fractures of the mid-humerus (in combination with glenohumeral degenerative diseases) | | √ | √ | | Revision of failed previous reconstructions when distal anchorage is required | | √ | √ | | To restore mobility from previous procedures (e.g. previous fusion) | The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head. The Equinoxe Platform Fracture Stem is indicated for use in skeletally mature individuals with acute fracture of the proximal humerus and displacement of the tuberosities, displaced 3- and 4part fractures of the proximal humerus (hemi-arthroplasty), or acute fracture of the proximal humerus with failure of the glenohumeral joint (primary total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is also indicated for acute fracture of the proximal humerus in combination with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff resulting in superior migration of the humeral head (reverse total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is indicated for cemented use only. # TECHNOLOGICAL CHARACTERISTICS The rationale for substantial equivalence is based on consideration of the following device use and characteristics: - Indications for Use. The proposed Laser Cage Glenoids and the predicate devices have ● identical indications for use. - Materials. The proposed Laser Cage Glenoids and the predicate devices are composed of identical biocompatible materials. - Design Features. The proposed Laser Cage Glenoids and the predicate devices have ● the same design features. - Dimensions. The proposed Laser Cage Glenoids and the predicate devices are ● dimensionally comparable. - . Sterilization. The proposed Laser Cage Glenoids and the predicate devices are provided sterile for single use only. - Performance Requirements. The proposed Laser Cage Glenoids and the predicate ● devices conform to the same recognized performance standards. {6}------------------------------------------------ # LEGALLY MARKETED DEVICES TO WHICH SUBSTANTIAL EQUIVALENCE IS CLAIMED Primary Predicate: | 510k Number | Trade or Proprietary or Model Name | Manufacturer | |-------------|------------------------------------|----------------| | K113309 | Equinoxe® Cage Glenoid | Exactech, Inc. | Additional Predicates: | 510k Number | Trade or Proprietary or Model Name | Manufacturer | |-------------|-----------------------------------------------|----------------| | K121220 | Equinoxe 12° Posterior Augment Pegged Glenoid | Exactech, Inc. | | K111379 | Equinoxe 16° Posterior Augment Pegged Glenoid | Exactech, Inc. | #### REFERENCE DEVICE | 510k Number | Trade or Proprietary or Model Name | Manufacturer | |-------------|---------------------------------------|----------------| | K192097 | Equinoxe® Stemless Humeral Components | Exactech, Inc. | ## NON-CLINICAL TESTING The Equinoxe Laser Cage Glenoid has been tested in the following test modes: - · Porous Structure Characterization - · Glenoid Fixation - · Peg Shear Resistance - · Peg Pull-Off - · Peg Bending Fatigue Bacterial endotoxin testing was conducted in accordance with USP <161>, USP <85>, and ANSI/AAMI ST72 to ensure the proposed Laser Cage Glenoid meets recommended limits per FDA's Guidance Document Submission and Review of Sterility Information in Premarket (510(k)) Submission for Devices Labeled as Sterile. ## SUBSTANTIAL EQUIVALENCE CONCLUSION Based on consideration of indications for use, technological characteristics, biocompatibility of the proposed devices, and results of non-clinical testing, it was concluded the Equinoxe Laser Cage Glenoid demonstrates substantially equivalence to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...