Checkpoint Guardian Intraoperative Lead, Medium, Checkpoint Guardian Intraoperative Lead, Large
Applicant
Checkpoint Surgical
Product Code
ETN · Ear, Nose, Throat
Decision Date
Aug 26, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.1820
Device Class
Class 2
Indications for Use
The Checkpoint Guardian Intraoperative Lead is a single-use, sterile medical device accessory intended to be used with Checkpoint Stimulators to provide electrical stimulation of exposed motor nerves or muscle tissue to identify nerves and to test nerve and muscle excitability.
Device Story
Single-use, sterile accessory for Checkpoint Stimulator/Locator family; wraps around surgically exposed nerves to provide hands-free electrical stimulation. Input: electrical current from Checkpoint Stimulator. Transformation: delivers charge-balanced, cathodic-first, biphasic current-controlled waveform to nerve/muscle tissue via button contact electrode. Output: localized nerve/muscle stimulation. Used in surgical settings by surgeons to identify nerves and assess excitability. Benefits: enables hands-free nerve stimulation during surgery, facilitating nerve identification and functional testing.
Clinical Evidence
Bench testing only. Includes design verification testing (mechanical/electrical specs), EMI/EMC testing, and biocompatibility testing per ISO 10993-1:2018. No clinical data.
Technological Characteristics
304 Stainless Steel electrode; Pellethane 2363-80A TPU insulation; button contact form factor; single-use; ETO sterilized. Compliant with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-10. Connectivity: wired accessory to Checkpoint Stimulator.
Indications for Use
Indicated for use with Checkpoint Stimulators to provide electrical stimulation of exposed motor nerves or muscle tissue to locate and identify nerves and to test nerve and muscle excitability in patients undergoing surgical procedures.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
August 26, 2021
Checkpoint Surgical Ben Cottrill Product Development Manager 6050 Oak Tree Blvd., Suite 360 Independence. Ohio 44131
Re: K212355
Trade/Device Name: Checkpoint Guardian Intraoperative Lead, Medium, Checkpoint Guardian Intraoperative Lead, Large Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: PDQ, ETN Dated: July 28, 2021 Received: July 29, 2021
Dear Ben Cottrill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (0S) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shu-Chen Peng, Ph.D. Assistant Director DHT1C: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212355
Device Name Checkpoint Guardian Intraoperative Lead
Indications for Use (Describe)
The Checkpoint Guardian Intraoperative Lead is a single-use, sterile medical device accessory intended to be used with Checkpoint Stimulators to provide electrical stimulation of exposed motor nerves or muscle tissue to identify nerves and to test nerve and muscle excitability.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Special 510(k) Summary
## Checkpoint Guardian Intraoperative Lead
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
### 1 Applicant Checkpoint Surgical 6050 Oak Tree Blvd., Suite 360 Independence, OH 44131 Contact Person: Ben Cottrill New Product Development Manager 2 Date Prepared 28/JUL/2021 చి Name of Device
| Device Trade Name: | Checkpoint Guardian Intraoperative Lead |
|-----------------------------------|----------------------------------------------------------|
| Device Common Name: | Intraoperative Lead |
| Class:<br>Regulation Description: | Class II<br>Surgical nerve stimulator/locator (874.1820) |
| Classification Name: | Neurosurgical Nerve Locator<br>Stimulator, Nerve |
| Product Code: | PDQ<br>ETN |
| Classification Panel: | Neurology<br>Ear Nose & Throat |
| Predicate Device | |
| Manufacturer | Trade Name | Product Code | Submission # |
|---------------------|------------|--------------|--------------|
| Checkpoint Surgical | Checkpoint | ETN | K092292 |
#### 5 Description of Device
4
The Checkpoint Guardian Intraoperative Leads are single-patient disposable accessories for providing stimulus to a targeted nerve. The accessories are intended for use with any device from the Checkpoint Stimulator/Locator Family. The lead electrode is intended to wrap around nerves that have been surgically exposed, allowing the surgeons to provide hands free stimulus to a targeted nerve.
Model Numbers include:
{4}------------------------------------------------
- REF 9525 Checkpoint Guardian Intraoperative Lead, Medium (7mm nominal diameter) .
- REF 9526 Checkpoint Guardian Intraoperative Lead, Large (10mm nominal diameter) .
#### 6 Indications for Use
The Checkpoint Guardian Intraoperative Lead is a single-use, sterile medical device accessory intended for use in conjunction with a Checkpoint Stimulator to provide electrical stimulation of exposed motor nerves or muscle tissue to locate and identify nerves and to test nerve and muscle excitability.
{5}------------------------------------------------
| | Predicate Devices (K092292) | Guardian Intraoperative Lead | Remark |
|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Indications<br>for Use | The Checkpoint® is a single-use<br>device intended to provide electrical<br>stimulation of exposed motor nerves<br>or muscle tissue to locate and<br>identify nerves and to test nerve and<br>muscle excitability. | The Checkpoint Guardian<br>Intraoperative Lead is a single-use,<br>sterile medical device accessory<br>intended to be used with<br>Checkpoint Stimulators to provide<br>electrical stimulation of exposed<br>motor nerves or muscle tissue to<br>locate and identify nerves and to<br>test nerve and muscle excitability. | Same |
| Stimulation<br>Electrical<br>Parameters<br>and<br>Waveform | Charge balanced cathodic-first<br>biphasic current-controlled waveform<br>compliant with IEC 60601 2-<br>10:2012+A1:2016. | Charge balanced cathodic-first<br>biphasic current-controlled<br>waveform compliant with IEC 60601<br>2-10:2012+A1:2016. | Same |
| Stimulation<br>Electrode<br>Material | 304 Stainless Steel | 304 Stainless Steel | Same |
| Stimulation<br>Electrode | Probe | Button contact | SE |
| Charge<br>Density<br>(µC/cm² ) | Maximum of 12.7 µC/cm² through<br>minimum contact surface area. | Maximum of 10.5 µC/cm² through<br>minimum contact surface area. | SE |
| Insulating<br>Material<br>at/around<br>electrode<br>contact | Medical grade polyolefin, MT2000 | Pellethane 2363-80A TPU (a<br>thermoplastic polyurethane<br>elastomer), used extensively for<br>both cardiac and deep brain<br>stimulation leads. | SE |
| Sterilization | Supplied Sterile via validated and<br>monitored ETO cycle | Supplied Sterile via SAME validated<br>and monitored ETO cycle | Same |
| Shelf Life | 2 Years | 2 Years | Same |
| Packaging<br>Config. &<br>Materials | Stimulator with protective probe<br>sleeve in sterile single use Tyvek<br>pouch and paperboard shelf carton. | Lead on HDPE backer card, in<br>same packaging parts with same<br>processes as stimulator. | SE |
#### Summary Comparison of Technological Characteristics to Predicate 7
{6}------------------------------------------------
#### 8 Nonclinical Testing
The Guardian Leads underwent a battery of bench and user tests in accordance with in-house procedures. Design Verification Testing (DVT) conducted on the lead to support this submission demonstrated that all requirements were met including Checkpoint's mechanical and electrical specifications, as well as functionality / performance requirements.
Electrical Safety, general functional testing, and Electromagnetic Immunity / Electromagnetic Compatibility (EMI / EMC) testing for the Checkpoint Stimulator-Locator family was configured and executed in a worst-case configuration to ensure that the stimulation system meets system electrical design specifications and statutory requlations when used in conjunction with associated accessories, including the proposed lead.
Biocompatibility testing was performed on the lead, with all endpoints conforming to limits of BS EN ISO 10993-1:2018 and 2016 FDA guidance titled 'Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" for its intended use as a limited exposure (<24 hours) externally communicating tissue contacting device Incremental testing on the wire and connector of the subject device was also provided by its manufacturer and certified to have met all relevant biocompatibility endpoints.
Design Analysis and Evaluation Testing evaluated device robustness and correct implementation of the design intent.
A Packaging Assessment was performed to support a determination of substantial equivalence. The same sterile barrier system, sales, and shipping packaging as well as the same labeling stock and inks as the Checkpoint Stimulator Family are used for the proposed lead. The lead was adopted into the same sterilization cycle as the stimulator as per
Electrical Safety Standards applicable to the proposed lead:
- IEC 60601-1:2005 +A12013, Medical electrical equipment Part 1: General requirements for ● basic safety and essential performance
- IEC 60601-1-2:2014, Medical electrical equipment Part 1-2: General requirements for . safety - Collateral standard: Electromagnetic compatibility - requirements and tests
- . IEC 60601-2-10:2012+A1:2016, Medical Electrical Equipment- Part 2.10: Particular Requirements for Safety and Essential Performance of Nerve & Muscle Stimulators
Standards used to perform Biocompatibility Testing:
- . EN ISO 10993-1:2018, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process
- EN ISO 10993-5:2009, Biological evaluation of medical devices Part 5: Tests for in vitro . cvtotoxicity
- . BS EN ISO 10993-7:2008, Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals
- . EN ISO 10993-10:2013, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- BS EN ISO 10993-11:2018, Biological evaluation of medical devices Tests for systemic . toxicitv
Results of this testing provide reasonable assurance that the proposed device design meets the
{7}------------------------------------------------
performance requirements and performs as intended.
#### 9 Conclusions
The Checkpoint Guardian Intraoperative Leads have the same indication for use and share similar technological characteristics with their predicate devices, for which they are accessories. Any technological differences between the predicate device (K092292) and the Checkpoint Guardian Intraoperative Leads have been analyzed in terms of risks and addressed through performance testing which demonstrates that the Intraoperative Lead meets its intended use. There are no new concerns around safety or effectiveness. Thus, the leads have been shown to be substantially equivalent to legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.