← Product Code [NFO](/productcode/NFO) · K212342

# ZIIP+ Device (K212342)

_Ziip, Inc. · NFO · Sep 23, 2021 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K212342

## Device Facts

- **Applicant:** Ziip, Inc.
- **Product Code:** [NFO](/productcode/NFO.md)
- **Decision Date:** Sep 23, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The ZIIP+ Device is intended for facial and neck stimulation and is indicated for over-the-counter cosmetic use.

## Device Story

The ZIIP+ is a handheld, battery-powered facial toning device used with conductive gel to deliver transcutaneous electrical microcurrents to the face and neck. Operated by the user in a home setting, the device applies electrical stimulation to the skin to promote cellular response. It features Bluetooth Low Energy (BLE) connectivity. The device is controlled by internal firmware that manages output levels, which are indicated to the user via LED illumination. The user applies the device to the skin, and the electrical output affects facial and neck tissue for cosmetic purposes. The device includes safety features such as automatic shut-off, automatic no-load trip, and patient override control.

## Clinical Evidence

Bench testing only. No clinical data was provided. Testing included electrical safety (IEC 60601-1, 60601-1-2, 60601-1-11), electromagnetic compatibility (EMC), wireless coexistence, and software verification and validation. Biocompatibility was established by demonstrating that skin-contacting materials are identical to the predicate ZIIP device.

## Technological Characteristics

Handheld, battery-powered (rechargeable Lithium-Ion) transcutaneous electrical nerve stimulator. Materials are biocompatible. Output: Pulsed biphasic, modulated square wave. Connectivity: Bluetooth Low Energy (BLE). Safety standards: IEC 60601-1, 60601-1-2, 60601-1-11. Software-controlled with LED indicators for status and current levels. Not sold sterile.

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- ZIIP Skincare Device ([K161484](/device/K161484.md))
- NūFACE® Trinity Plus ([K201782](/device/K201782.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 23, 2021

ZIIP, Inc. % Heather Tanner Principle Consultant Hill Regulatory Consulting, LLC 1910 15th Ave E Seattle, Washington 98112

Re: K212342

Trade/Device Name: ZIIP+ Device Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: July 23, 2021 Received: July 28, 2021

Dear Heather Tanner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Heather Dean, PhD Assistant Director. Acute Injurv Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212342

Device Name ZIIP+ Device

Indications for Use (Describe)

The ZIIP+ Device is indicated for facial and neck stimulation and is indicated for over-the-counter cosmetic use.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) SUMMARY

510(k) Notification K212342

#### 1. GENERAL INFORMATION

Applicant: ZIIP Inc. 1425 Leimert Blvd Suite 203 Oakland, CA 94602

Contact Person: David Mason President, ZIIP, Inc. Contact information: david@ziipbeauty.com

Date Prepared: July 23, 2021

### 2. DEVICE INFORMATION

Trade Name: ZIIP+ Device

Generic/Common Name: Facial Toning Device

Classification: Transcutaneous Electrical Nerve Stimulatory - CFR 882.5890

Product Code: NFO

#### 3. PREDICATE DEVICE(S)

- K161484 ZIIP Skincare Device ●
- K201782 NūFACE® Trinity Plus ●

#### 4. INDICATIONS FOR USE

The ZIIP+ Device is intended for facial and neck stimulation and is indicated for over-the-counter cosmetic use.

#### 5. DEVICE DESCRIPTION

The ZIIP+ Device is a hand-held, battery-powered device used with conductive gel to stimulate the face superficially through application of transcutaneous electrical currents. The device is powered by a Lithium-lon rechargeable battery, and it is shipped with a portable battery charger.

{4}------------------------------------------------

### 6. SUBSTANTIAL EQUIVALENCE

The ZIIP+ is substantially equivalent to the legally marketed ZIIP device (K161484) and the NuFACE Trinity Plus Device (K201782). The ZIIP+ Device has the identical output characteristics, principles of operation, and treatment method, and it has a similar indication for use as the predicate ZIIP Device. In addition, the ZIIP+ Device has the same intended use and indication for use [facial and neck toning] as the NuFACE Trinity Plus Device, as well as substantially equivalent technological characteristics and principles as operation, including Bluetooth Low Energy (BLE) capability.

### 7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

| General Device Characteristics       |                                                                                                                  |                                                                                                            |                                                                                                                                 |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
|                                      | ZIIP+ Subject Device                                                                                             | ZIIP Predicate<br>Device                                                                                   | NuFace Trinity Plus Predicate<br>Device                                                                                         |
| Manufacturer                         | ZIIP Inc.                                                                                                        | ZIIP Inc.                                                                                                  | NuFace                                                                                                                          |
| K Number                             | K212342                                                                                                          | K161484                                                                                                    | K201782                                                                                                                         |
| Device Name                          | ZIIP+                                                                                                            | ZIIP                                                                                                       | NuFace Trinity Plus                                                                                                             |
| OTC/Rx                               | OTC                                                                                                              | OTC                                                                                                        | OTC                                                                                                                             |
| Med Dev Class                        | Class II                                                                                                         | Class II                                                                                                   | Class II                                                                                                                        |
| Product Code                         | NFO                                                                                                              | NFO                                                                                                        | NFO                                                                                                                             |
| Common Name                          | Transcutaneous Electrical<br>Nerve Stimulator                                                                    | Transcutaneous<br>Electrical Nerve<br>Stimulator                                                           | Transcutaneous Electrical Nerve<br>Stimulator                                                                                   |
| Mechanism of Action                  | Microcurrent electrical<br>therapy to deliver skin<br>stimulation through cellular<br>response                   | Microcurrent<br>electrical therapy to<br>deliver skin<br>stimulation through<br>cellular response          | Microcurrent electrical therapy to<br>deliver skin stimulation through<br>cellular response                                     |
| Device Type                          | 21 CFR, Part 882.5890                                                                                            | 21 CFR, Part<br>882.5890                                                                                   | 21 CFR, Part 882.5890                                                                                                           |
| Indications for Use<br>/Intended Use | The ZIIP+ Device is<br>intended for facial and neck<br>stimulation and is indicated<br>for over-the-counter use. | The ZIIP Device is<br>intended for facial<br>stimulation and is<br>indicated for over-<br>the-counter use. | The NuFace Trinity Plus Device is<br>intended for facial and neck<br>stimulation and is indicated for over-<br>the-counter use. |
| Material/ Biocompatibility           | Biocompatible materials<br>typically used in medical<br>devices and identical to<br>predicate ZIIP device        | Biocompatible<br>materials typically<br>used in medical<br>devices.                                        | Biocompatible materials typically<br>used in medical devices.                                                                   |
| Power source                         | Internal Rechargeable<br>Lithium Ion Battery                                                                     | Internal<br>Rechargeable Lithium<br>Ion Battery                                                            | Internal Rechargeable Lithium Ion<br>Battery                                                                                    |
| Performance standard                 | No known required<br>performance standards                                                                       | No known required<br>performance<br>standards                                                              | No known required performance<br>standards                                                                                      |
| Sterility                            | Not applicable - this device<br>is not sold sterile                                                              | Not applicable - this<br>device is not sold<br>sterile                                                     | Not applicable - this device is not<br>sold sterile                                                                             |

#### General Device Characteristics

{5}------------------------------------------------

| Human Factors        | Ergonomic handheld design                         | Ergonomic handheld design             | Ergonomic handheld design          |
|----------------------|---------------------------------------------------|---------------------------------------|------------------------------------|
| Electrical Safety    | Compliant with IEC 60601-1, 60601-1-2, 60601-1-11 | Compliant with IEC 60601-1, 60601-1-2 | IEC 60601-1, 60601-1-2, 60601-1-11 |
| Type of Energy       | Electrical current                                | Electrical current                    | Electrical current                 |
| Charging Method      | External wall adaptor                             | External wall adaptor                 | Wireless charging                  |
| Charging Circuitry   | Internal to device                                | Internal to device                    | Internal to device                 |
| Special Requirements | Conductive gel                                    | Conductive gel                        | Conductive gel primer              |
| Wireless Capability  | Bluetooth Low Energy (BLE)                        | N/A                                   | Bluetooth Low Energy (BLE)         |

## Detailed Output Characteristics

| Detailed Output<br>Specifications                      | ZIIP Subject Device                                                                                          | ZIIP Predicate Device                                                                                        | NuFace Trinity Plus                                                    |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Power Source(s)                                        |                                                                                                              |                                                                                                              |                                                                        |
| a) Method of Line<br>Current Isolation                 | One rechargeable Lithium-Ion Battery and External<br>Charger Isolation                                       | One rechargeable Lithium-Ion Battery and External<br>Charger Isolation                                       | One rechargeable Lithium-Ion Battery and External<br>Charger Isolation |
| b) Patient Leakage<br>Current                          | External Charger Included                                                                                    | External Charger Included                                                                                    | External Charger Included                                              |
| 1. Normal<br>Condition                                 | $46 \mu A$                                                                                                   | $46 \mu A$                                                                                                   | N/A - Battery Operated                                                 |
| 2. Fault Condition                                     | $46 \mu A$                                                                                                   | $46 \mu A$                                                                                                   | N/A - Battery Operated                                                 |
| Number of Output<br>Modules                            | 1                                                                                                            | 1                                                                                                            | 1                                                                      |
| Number of Output<br>Channels                           | 1                                                                                                            | 1                                                                                                            | 1                                                                      |
| a) Synchronous or<br>Alternating                       | N/A - 1 Output Channel                                                                                       | N/A - 1 Output Channel                                                                                       | N/A - 1 Output Channel                                                 |
| b) Method of<br>Channel<br>Isolation                   | N/A - 1 Output Channel                                                                                       | N/A - 1 Output Channel                                                                                       | N/A - 1 Output Channel                                                 |
| Regulated Current or<br>Regulated Voltage              | Both                                                                                                         | Both                                                                                                         | Both                                                                   |
| Software/<br>Firmware/<br>Microprocessor<br>Controlled | Yes                                                                                                          | Yes                                                                                                          | Yes                                                                    |
| Automatic Overload Trip                                | Not required because of<br>circuit design (Current and<br>Voltage Limited by Circuit<br>Design and Firmware) | Not required because of<br>circuit design (Current and<br>Voltage Limited by Circuit<br>Design and Firmware) | Not required because of<br>circuit design                              |
| Automatic No-Load Trip                                 | Yes (Reversion to Fixed<br>Voltage Output)                                                                   | Yes (Reversion to Fixed<br>Voltage Output)                                                                   | Yes                                                                    |
| Automatic Shut Off                                     | Yes                                                                                                          | Yes                                                                                                          | Yes                                                                    |
| Patient Override Control                               | Yes                                                                                                          | Yes                                                                                                          | Yes                                                                    |
| Indicator Display                                      |                                                                                                              |                                                                                                              |                                                                        |
| a) On/Off Status                                       | Yes (LED Illumination on<br>Conduction)                                                                      | Yes (LED Illumination on<br>Conduction)                                                                      | Yes                                                                    |
| b) Low Battery                                         | Yes                                                                                                          | Yes                                                                                                          | Yes                                                                    |
| c) Voltage/Current<br>Level                            | Yes (LED Illumination on<br>Target Current Levels)                                                           | Yes (LED Illumination on<br>Target Current Levels)                                                           | Yes                                                                    |

{6}------------------------------------------------

| Output Specifications                                      | ZIIP Subject Device                   | ZIIP Predicate Device                 | NuFace Trinity Plus                         |
|------------------------------------------------------------|---------------------------------------|---------------------------------------|---------------------------------------------|
| Waveform                                                   | Pulsed Biphasic                       | Pulsed Biphasic                       | Pulsed Monophasic                           |
| Shape                                                      | Modulated Square Wave                 | Modulated Square Wave                 | Modulated Square Wave                       |
| Maximum Output Voltage                                     | 154mV@500Ω<br>465mV@52KΩ<br>2.2V@10KΩ | 154mV@500Ω<br>465mV@52KΩ<br>2.2V@10KΩ | Not publicly available                      |
| Maximum Output Current                                     | 309μA@500Ω<br>232μΑ@2ΚΩ<br>202μΑ@10ΚΩ | 309μA@500Ω<br>232μΑ@2ΚΩ<br>202μΑ@10ΚΩ | Not publicly available                      |
| Burst Mode (i.e. pulse<br>trains)                          | N/A - no burst mode                   | N/A - no burst mode                   | 20 (10 positive and 10 negative)            |
| Pulses per second                                          |                                       |                                       | 8.3 (@ 8.3 Hz)                              |
| Burst duration (seconds)                                   |                                       |                                       | 2.4s (@ 8.3 Hz)                             |
| Duty Cycle [Line (b) x<br>Line (c)] (on time per<br>burst) |                                       |                                       | Duty Cycle: 50%<br>On time per burst: 20.2s |
| On time (seconds)                                          | Constant                              | Constant                              | 60 ms (@8.3 Hz)                             |
| Off time (seconds)                                         | None                                  | None                                  | 60 ms (@8.3 Hz)                             |
| Maximum Phase Charge<br>(μC)                               | 6.16 µC@500Ω                          | 6.16 µC@500Ω                          | 23.06 μC@500Ω                               |
| Maximum Current<br>Density (mA/cm2)                        | 0.34 mA/cm2 @500Ω                     | 0.34 mA/cm2 @500Ω                     | 0.78 mA/cm2 @500Ω                           |
| Maximum Power Density<br>(µW/cm2)                          | 3.44 W/cm2@500Ω                       | 3.44 W/cm2@500Ω                       | 3.22 µW/cm2@500Ω                            |

### 8. NONCLINICAL TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION

Comprehensive performance testing was performed to support the substantial equivalence determination for the ZIIP+ Device, including the following:

- . Electrical Safety Testing, including EMC. Specifically, the ZIIP+ Device was tested and found to be in compliance with:
	- o IEC 60601- 1 -2 (Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – requirements and tests) for radiated and power line conducted emissions
	- IEC 60601-1 (Medical electrical equipment: Part 1: General requirements for basic o safety and essential performance) for Electrical Safety
	- o IEC 60601-1-11 (Medical Electrical Equipment: Part 1-11: Requirements for medical electrical systems used in the home healthcare environment) for home healthcare devices
- Software Verification and Validation Testing
	- The results of software verification and validation confirmed that the ZIIP+ Device meets and complies with the applicable software requirements specifications.
- Biocompatibility Evaluation
	- The skin-contacting materials for ZIIP+ are identical to the skin-contacting materials for the predicate ZIIP device, which was evaluated and determined to be biocompatible; therefore, the ZIIP+ Device is biocompatible and substantially equivalent to the predicate device.

{7}------------------------------------------------

In summary, the safety and performance testing results for EMC, Electrical Safety, Wireless Coexistence, and Verification & Validation concluded that the ZIIP+ Device complies with the applicable standards and meets bench testing performance requirements.

## 9. CONCLUSION

The submitted documentation and performance testing demonstrates that the ZIIP+ device is substantially equivalent to the cited predicate devices in its intended use, principles of operation, design, and overall technological characteristics.

---

**Source:** [https://fda.innolitics.com/device/K212342](https://fda.innolitics.com/device/K212342)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
