T2 is digital X-Ray imaging equipment for dental professionals that convert X-Ray signals into digital signals to acquire 2D images and reconstruct them into 3D images, using Panoramic (Pano), Cephalometric (Ceph), and Computed Tomography (CT) technology for the diagnosis of the anatomical structure of the oral and maxillofacial area.
Device Story
T2 is a digital X-ray imaging system (CT, Panoramic, Cephalometric) for dental clinics. System components include X-ray generator, controller, rotating supporter, image sensor, PC, and software. The rotating apparatus (360-degree rotation) carries the generator and sensor; X-rays penetrate the patient, are converted to electrical signals by the sensor, and transmitted to a PC. Software reconstructs 2D/3D images for display. Operated by dentists, physicians, or X-ray technicians. Output assists in diagnosing oral/maxillofacial anatomical structures. Benefits include precise diagnostic imaging for treatment planning.
Clinical Evidence
No clinical studies were submitted. Substantial equivalence is supported by bench testing, including software verification and validation, electrical safety, and electromagnetic compatibility testing.
Indicated for dental professionals to acquire 2D and 3D X-ray images of the oral and maxillofacial area for diagnostic purposes in adult and pediatric patients.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
September 21, 2021
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is an emblem representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The word "FDA" is in a larger, bolder font, and the words "U.S. FOOD & DRUG" are stacked above the word "ADMINISTRATION".
Osstem Implant Co., Ltd. % Peter Lee QA/RA Manager Hiossen Inc. 85 Ben Fairless Dr. FAIRLESS HILLS PA 19030
Re: K212303
Trade/Device Name: T2 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: OAS, Dated: July 20, 2021 Received: July 23, 2021
Dear Peter Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel M. Digitally signed by
Laurel M. Burk -S
Burk -S
Date: 2021.09.21
09:01:35 -04'00'
, for
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212303
Device Name T2
Indications for Use (Describe)
T2 is digital X-Ray imaging equipment for dental professionals that convert X-Ray signals into digital signals to acquire 2D images and reconstruct them into 3D images, using Panoramic (Pano), Cephalometric (Ceph), and Computed Tomography (CT) technology for the diagnosis of the anatomical structure of the oral and maxillofacial area.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
m Implant Co., Ltd ong, 51, Mayu-ro 238beon-gil, Siheung-si, Gyeonggi-do, 15079, Republic of Korea 82 70 4626 0701 Fax: +82 70 4394 0404 www.osstem.com
## K212303 - 510(k) Summary
### Date: July 20, 2021
### 1. Company and Correspondent making the submission
| - Submitter's Name | : Osstem Implant Co., Ltd. |
|------------------------|-------------------------------------------------------------------------------------------------|
| - Address | : 2 Floor, B-dong, 51, Mayu-ro 238beon-gil, Siheung-si<br>Gyeonggi-do, 15079, Republic of Korea |
| - Contact | : Mr. Jinwoo Bae |
| - Phone | : +82-70-4626-0701 |
| - Correspondent's Name | : Hiossen Inc. |
| - Address | : 85 Ben Fairless Dr. Fairless Hills, PA 19030 |
| - Contact | : Mr. Peter Lee |
| - Phone | : +1-267-759-7031 |
### 2. Proposed Device
| Trade or (Proprietary) Name | : T2 |
|-----------------------------|---------------------------------------|
| Classification Name | : X-Ray, Tomography, Computed, Dental |
| Regulation Number | : 21CFR892.1750 |
| Device Classification | : Class II |
| Classification Product Code | : OAS |
### 3. Predicated Device
Primary Predicate -T1-CS, T1-C, Osstem Implant Co., Ltd. (K183475)
### 4. Description
- T2 is a digital X-ray CT, panoramic, and Cephalo imaging system device composed of X-ray generator, X-ray controller, X-ray supporter, image processing unit (sensor), PC, and software. The apparatus attached to the equipment column is a structure that can be rotated 360 by the system control unit. This system control unit actuates the motor control, X-ray generator, and image processing unit (sensor). The height controlling unit controls the column and adjusts the height of the equipment. The X-ray generator and image processing unit (sensor) are attached to the rotating apparatus. When the rotating apparatus starts the rotation, X-ray is irradiated from the X-ray generator (generating unit). This X-ray irradiation penetrates the subject and reaches the image processing unit (sensor), and then is converted into electric signals to secure imagery information. Inside the imaging section of the image processing unit (sensor), real time X-ray input is
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Osstem Implant. The logo is in orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, lighter font. There is a small circle above the right side of the word "OSSTEM".
Osstem Implant Co., Ltd.
2 Floor, B-dong, 51, Mayu-ro 238beon-gil, Siheung-si, Gyeonggi-do, 15079, Republic of Korea Tel: +82 70 4626 0701 Fax: +82 70 4394 0404 www.osstem.com
converted into electric signals and consecutively combined, resulting in imagery information. The combined panoramic imagery information is then sent to the PC and saved in patient management software
| | Proposed Devices | Predicated Devices | Remark | |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Device Name | T2 | T1-CS, T1-C | Different | |
| 510(k) No. | Proposed | K183475 | Different | |
| Manufacturer | Osstem Implant Co., Ltd. | Osstem Implant Co., Ltd. | Same | |
| Indications for Use | T2 is digital X-Ray imaging<br>equipment for dental<br>professionals that convert X-<br>Ray signals into digital<br>signals to acquire 2D images<br>and reconstruct them into 3D<br>images, using Panoramic<br>(Pano), Cephalometric<br>(Ceph), and Computed<br>Tomography (CT)<br>technology for the diagnosis<br>of the anatomical structure of<br>the oral and maxillofacial<br>area. | T1-C, T1-CS are are a<br>computed tomography x-ray<br>system intended to produce<br>panoramic, cephalometric or<br>crossectional images of the<br>oral anatomy on a real time<br>basis by computer<br>reconstruction of X-ray<br>image data from the same<br>axial plane taken at different<br>angleds. It provides<br>diagnostic details of the<br>anatomic structures by<br>acquiring 360° rotational<br>image sequences of oral and<br>maxillofacial area from a<br>precise treatment planningin<br>adult and pediatric dentistry.<br>The device is operated and<br>used by physicians, dentists,<br>and X-ray technicians. | Same | |
| Perfomance Specification | Panoramic, cephalometric<br>and computed tomography | Panoramic, cephalometric<br>and computed tomography | Same | |
| Input Voltage | 100-240VAC | 100-240 VAC | Same | |
| Tube Voltage | 60-95KV | 60-100 kV | Similar<br>(included<br>within<br>range of<br>predicated<br>product) | |
| Tube Current | 5-11mA | 5-16 mA | Similar<br>(included<br>within<br>range of<br>predicated<br>product) | |
| Focal Spot Size | 0.5mm | 0.5 mm | Same | |
| Exposure Time | Max. 22s | Max. 22 s | Same | |
| 2D Image Viewing Program | Model Name: 2D02 | Model Name: 2D01, 2D02 | Similar | |
| | | | | (included<br>within<br>predicated<br>product) |
| | 3D Image Viewing Program | Model Name: 3D02 | Model Name: 3D01, 3D02 | Similar<br>(included<br>within<br>predicated<br>product) |
| Anatomical Sites | | Maxillofacial | Maxillofacial | Same |
| Image<br>Sensor | Computed<br>Tomography | 1515DXT<br>SEN1 | 1515DXT<br>SEN1 | Same |
| | Panoramic | 1515DXT<br>SEN1 | 1515DXT<br>SEN1 | Same |
| | Cephalometric | XID-C24DC | XID-C24DC | Same |
| | | 1515DXT : Max. 15x9<br>SEN1 : Max 15x10<br>(Stiching Mode : Max 15 X 15) | 1515DXT : Max. 15x9<br>SEN1 : Max 15x10 | Similar<br>(included<br>within<br>range of<br>predicated<br>product) |
| | Size of Imaging Volume<br>(cm) | | | |
| Pixel<br>Resolution | Computed<br>Tomography | 1515DXT :<br>- 3.94 lp/mm<br>SEN1 :<br>- 4.0 lp/mm | 1515DXT :<br>- 3.94 lp/mm<br>- 1.97 lp/mm<br>SEN1 :<br>- 4.0 lp/mm<br>- 2.0 lp/mm | Same |
| | Panoramic | 1515DXT : 3.94 lp/mm<br>SEN1 : 4.0 lp/mm | 1515DXT : 3.94 lp/mm<br>SEN1 : 4.0 lp/mm | Same |
| | Cephalometric | XID-C24DC : 5 lp/mm | XID-C24DC : 5 lp/mm | Same |
| Pixel Size | Computed<br>Tomography | 1515DXT :<br>- 127 μm * 127 μm<br>SEN1 :<br>- 125 μm * 125 μm | 1515DXT :<br>- 127 μm * 127 μm<br>- 254 μm * 254 μm<br>SEN1 :<br>- 125 μm * 125 μm<br>- 250 μm * 250 μm | Similar<br>(included<br>within<br>range of<br>predicated<br>product) |
| | Panoramic | 1515DXT :<br>- 127 μm * 127 μm<br>SEN1 :<br>- 125 μm * 125 μm | 1515DXT :<br>- 127 μm * 127 μm<br>SEN1 :<br>- 125 μm * 125 μm | Same |
| | Cephalometric | XID-C24DC<br>- 100 μm * 100 μm | XID-C24DC<br>- 100 μm * 100 μm | Same |
| | | | | |
| Active Area | | [CT]<br>- 1515DXT : 146 x 146 mm<br>- SEN1 : 160 x 128 mm<br>[Pano]<br>- 1515DXT : 146 x 8.13 mm<br>- SEN1 : 160 x 8 mm | [CT]<br>- 1515DXT : 146 x 146 mm<br>- SEN1 : 160 x 128 mm<br>[Pano]<br>- 1515DXT : 146 x 8.13 mm<br>- SEN1 : 160 x 8 mm | Same |
### 5. Substantial Equivalence Matrix
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a bold, sans-serif font above the word "IMPLANT" in a smaller, lighter font. The logo is simple and modern.
**Osstem Implant Co., Ltd.**
OSSTERM IMPLANT CO., Ltd.
2 Floor, B-dong, 51, Mayu-ro 238beon-gil, Gibentg-si, Gyeonegi-do, 15079, Republic of Korea
Tel: +82 70 4626 0701 Fax: +82 70 4394 0404 www.osstem
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, lighter font. The logo is simple and modern.
n Implant Co., Ltd
on-gil. Siheung-si. Gyeonggi-do. 15079. Republic of Korea Fax: +82 70 4394 0404 www.osstem.com
| [Ceph]<br>- 1.8 x 150 mm | [Ceph]<br>- 1.8 x 150 mm | |
|--------------------------|--------------------------|--|
|--------------------------|--------------------------|--|
### 6. Indications for Use
- T2 is digital X-Ray imaging equipment for dental professionals that converts X-Ray signals into digital signals to acquire 2D images and reconstruct them into 3D images, using Panoramic (Pano), Cephalometric (Ceph), and Computed Tomography (CT) technology for the diagnosis of the anatomical structure of the oral and maxillofacial area.
### 7. Summary of Non-clinical Performance Testing
Software validation and verification test, EMC/Electrical Safety tests, and performance tests were conducted and the test results support that the subject device is substantially equivalent to the predicate device
### Non-Clinical Performance Data
We performed the applicable non-clinical tests in the FDA guidance document for solid state x-ray detectors.
### Software Verification and Validation Testing
Software was designed and developed according to a software development process. Also, Software verification and validation test were conducted and documented in accordance with FDA guidance: "The content of premarket submissions for software contained in medical devices, on May 11, 2005". The software for this device contains "Moderate" level of concern software.
#### Safety, EMC and Performance Data
T2 complies with the electrical safety and electromagnetic compatibility requirements established by the standards IEC60601-1, IEC60601-1-2.
### 8. Summary of Clnical Testing
No clinical studies are submitted.
### 9. Conclusion
The indication for use and the technical characteristics of the subject device is substantially equivalent to the predicate device. It is substantially equivalent to predicate device in output characteristics and operation mode (CT, Cephalo, Panorama). The major difference is size of image volume however, the non-clinical and clinical study results show that the subject device is substantially equivalent to the predicate device. In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and Based
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a stylized font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, sans-serif font. The logo is simple and modern, and it is likely used to represent the company's brand.
Osstem Implant Co., Ltd.
2 Floor, B-dong, 51, Mayu-ro 238beon-gil, Siheung-si, Gyeonggi-do, 15079, Republic of Korea
Tel: +82 70 4626 0701 Fax: +82 70 4394 0404 www.osstem.com
on the information provided in this submission, we conclude that the subject device is safe and effective and substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.