← Product Code [FDS](/productcode/FDS) · K212296

# Distal Cap Model DH-32EN, Distal Cap Model DH-17EN2 (K212296)

_Fujifilm Corporation · FDS · Aug 19, 2021 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K212296

## Device Facts

- **Applicant:** Fujifilm Corporation
- **Product Code:** [FDS](/productcode/FDS.md)
- **Decision Date:** Aug 19, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Distal Cap Models DH-32EN and DH-17EN2 are intended to be used in combination with the dedicated endoscope to maintain the field of view during endoscopic procedures.

## Device Story

Distal Cap Models DH-32EN and DH-17EN2 are single-use, pre-sterilized endoscopic accessories; attached to the distal tip of compatible FUJIFILM double balloon endoscopes. The device functions as a physical hood to maintain a clear, unobstructed field of view during endoscopic procedures. Used by physicians in clinical settings; the device provides a mechanical barrier to keep tissue away from the endoscope lens. Benefits include improved visualization during endoscopic examination of the digestive tract.

## Clinical Evidence

No clinical data. Bench testing only. Performance evaluated via sterility testing (ISO 11135:2014, ISO 11607-1:2019, ASTM F1980-16), biocompatibility (ISO 10993 series), and mechanical testing to ensure secure attachment during use.

## Technological Characteristics

Single-use, EO-sterilized endoscopic distal caps. Dimensions: DH-32EN (9.7mm OD, 9.1mm length); DH-17EN2 (11.5mm OD, 8.0mm length). Compatible with specific FUJIFILM double balloon endoscopes (EN-580T, EI-580BT). Materials evaluated per ISO 10993-1:2018.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Distal Cap Model DH-39CZ ([K193123](/device/K193123.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

August 19, 2021

FUJIFILM Corporation % Jeffrey Wan Manager, Regulatory Affairs FUJIFILM Medical Systems U.S.A., Inc. 81 Hartwell Avenue, Suite 300 Lexington, MA 02421

Re: K212296

> Trade/Device Name: Distal Cap Model DH-32EN, Distal Cap Model DH-17EN2 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FDF, FDS Dated: July 21, 2021 Received: July 22, 2021

Dear Jeffrey Wan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general control's provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212296

Device Name

Distal Cap Models DH-32EN and DH-17EN2

Indications for Use (Describe)

Distal Cap Models DH-32EN and DH-17EN2 are intended to be used in combination with the dedicated endoscope to maintain the field of view during endoscopic procedures.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <span style="text-decoration: overline;">☒</span> Production/Use for 21 CFR 511.1 Subject Drugs |
|-------------------------------------------------------------------------------------------------|
| □ Compounding/Manufacturing for 21 CFR 511.4 Subject Drugs                                      |

X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY FUJIFILM Corporation Distal Cap Models DH-32EN and DH-17EN2

Date: July 21, 2021

# Submitter's Information:

FUJIFILM Corporation 798 Miyanodai Kaisei-Machi Ashigarakami-Gun, Kanagawa, 258-8538, Japan

# Contact Person:

Jeffrey Wan Manager, Regulatory Affairs Telephone: (201) 675-8947 E-Mail: jeffrey.wan@fujifilm.com

### Identification of the Proposed Device:

| Device Name:           | Distal Cap Models DH-32EN and DH-17EN2 |
|------------------------|----------------------------------------|
| Common Name:           | Endoscopic Accessory                   |
| Device Class:          | Class II                               |
| Classification Number: | 21 C.F.R. § 876.1500                   |
| Classification Name:   | Endoscope and accessories              |
| Device Panel:          | Gastroentereology/Urology              |
| Product Code:          | FDS, FDF                               |

### Predicate Devices:

- . Distal Cap Model DH-39CZ (K193123)
# Intended Use / Indications for Use

Distal Cap Models DH-32EN and DH-17EN2 are intended to be used in combination with the dedicated endoscope to maintain the field of view during endoscopic procedures.

### Device Description

Distal Cap Models DH-32EN and DH-17EN2 are pre-sterilized, single-use accessories that are attached to the tip of an applicable FUJIFILM double balloon endoscope to maintain a clear, unobstructed field of view during endoscopic procedures.

{4}------------------------------------------------

# Comparison of Technological Characteristics

A comparison of technological characteristics between DH-32EN and DH-17EN2 and the predicate DH-39CZ is provided below:

|                                                   | Subject Device #1<br>DH-32EN                                                                                                                                                                 | Subject Device #2<br>DH-17EN2                                                                                                                                                                | Predicate Device<br>DH-39CZ                                                                                                                                          |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number                                     | To be assigned                                                                                                                                                                               | To be assigned                                                                                                                                                                               | K193123                                                                                                                                                              |
| Product code                                      | FDF, FDS                                                                                                                                                                                     | FDF, FDS                                                                                                                                                                                     | FDF, FDS                                                                                                                                                             |
| Manufacturer                                      | FUJIFILM Corporation                                                                                                                                                                         | FUJIFILM Corporation                                                                                                                                                                         | FUJIFILM Corporation                                                                                                                                                 |
| Intended Use                                      | Distal Cap Models DH-<br>32EN and DH-17EN2<br>are intended to be used<br>in combination with the<br>dedicated endoscope to<br>maintain the field of view<br>during endoscopic<br>procedures. | Distal Cap Models DH-<br>32EN and DH-17EN2<br>are intended to be used<br>in combination with the<br>dedicated endoscope to<br>maintain the field of view<br>during endoscopic<br>procedures. | This hood is intended to<br>be used in combination<br>with compatible<br>endoscopes to maintain<br>the field of view during<br>observation of the<br>digestive tract |
| Outer<br>diameter                                 | 9.7±0.2mm                                                                                                                                                                                    | 11.5±0.2mm                                                                                                                                                                                   | 14.2±0.5mm                                                                                                                                                           |
| Maximum<br>diameter of<br>compatible<br>endoscope | 13.3mm                                                                                                                                                                                       | 17.8mm                                                                                                                                                                                       | 18.2mm                                                                                                                                                               |
| Total length                                      | 9.1mm                                                                                                                                                                                        | 8.0mm                                                                                                                                                                                        | 12.0mm                                                                                                                                                               |
| Distance from<br>the tip                          | 1.3mm                                                                                                                                                                                        | 1.5mm                                                                                                                                                                                        | 2.0mm                                                                                                                                                                |
| Compatible<br>endoscopes                          | EN-580T<br>El-580BT                                                                                                                                                                          | EN-580T<br>EI-580BT                                                                                                                                                                          | EC-600HL                                                                                                                                                             |
| Sterile method                                    | EO                                                                                                                                                                                           | EO                                                                                                                                                                                           | EO                                                                                                                                                                   |
| Single use                                        | Yes                                                                                                                                                                                          | Yes                                                                                                                                                                                          | Yes                                                                                                                                                                  |

# Performance Data

Sterility of the subject devices was evaluated using the following consensus standards: ISO 11135:2014, ISO 11607-1:2019, ASTM F1980-16.

Biocompatibility of the subject device was evaluated using the following consensus standards: ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-7:2008, ISO 10993-10:2010. Biocompatibility testing was performed in accordance with FDA's guidance, "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," published September 4, 2020.

{5}------------------------------------------------

Bench testing was conducted to demonstrate that the subject devices are attached securely such that they will not detach during use.

Additional performance specifications were evaluated against pre-defined acceptance criteria.

### Conclusions

The subject devices Distal Cap Models DH-32EN and DH-17EN2 share the same intended use and similar indications to the predicate device. Bench testing demonstrates that the subject devices are as safe and effective as the predicate device. Thus, Distal Cap Models DH-32EN and DH-17EN2 are substantially equivalent to the listed predicate device.

---

**Source:** [https://fda.innolitics.com/device/K212296](https://fda.innolitics.com/device/K212296)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
