← Product Code [MQB](/productcode/MQB) · K212269

# Intelligent NR (K212269)

_Canon, Inc. · MQB · Sep 17, 2021 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K212269

## Device Facts

- **Applicant:** Canon, Inc.
- **Product Code:** [MQB](/productcode/MQB.md)
- **Decision Date:** Sep 17, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Pediatric

## Indications for Use

As a part of the Canon radiography system, the CXDI Control Software when used with a compatible Canon detector is intended to provide digital image capture, processing, and display for conventional film/screen radiographic examinations. This device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures including specialist areas like intensive care, trauma, and pediatric work. This device is not intended for fluoroscopic, angiographic, or mammography applications.

## Device Story

Intelligent NR is a software function within Canon CXDI Control Software (V3.10) designed to reduce noise in X-ray images captured by compatible Canon digital detectors. The device utilizes a machine-learning algorithm, pre-trained on a clinical image database to recognize noise characteristics, to process raw X-ray signals into noise-reduced images. It operates on a PC, displaying results on a monitor for clinician review. The device is used in clinical settings by healthcare professionals to assist in diagnostic imaging. By reducing image noise, the device aims to improve image quality, potentially aiding clinical decision-making. The function is locked post-release and does not perform further machine learning in the field.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, FMEA risk analysis, and verification of software performance. Hardware performance, biocompatibility, and safety remain consistent with the predicate device.

## Technological Characteristics

Software-based noise reduction function integrated into CXDI Control Software V3.10. Operates on standard PC hardware. Compatible with existing Canon digital detectors (hardware unchanged). Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-32, FCC Part 15 Subpart B, and ICES-003. Machine-learning algorithm is locked post-deployment.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Enhanced Feature Software Pack for CXDI Series ([K190368](/device/K190368.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

September 17, 2021

Image /page/0/Picture/1 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized caduceus. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Canon Inc. % Mr. Gregory Woodard Biomedical Engineer Ken Block Consulting 800 East Campbell Road, Suite 202 RICHARDSON TX 75081

## Re: K212269

Trade/Device Name: Intelligent NR Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: July 19, 2021 Received: July 20, 2021

## Dear Mr. Woodard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

542 of the Act); 21 CFR 1000-1050.

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

. for

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K212269

Device Name Intelligent NR

#### Indications for Use (Describe)

As a part of the Canon radiography system, the CXDI Control Software when used with a compatible Canon detector is intended to provide digital image capture, processing, and display for conventional film/screen radiographic examinations. This device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures including specialist areas like intensive care, trauma, and pediatric work. This device is not intended for fluoroscopic, angiographic, or mammography applications.

| Type of Use (Select one or both, as applicable)                                                     |                                                                                           |
|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| <div> <span> <b> </b>           Prescription Use (Part 21 CFR 801 Subpart D)         </span> </div> | <div> <span>           Over-The-Counter Use (21 CFR 801 Subpart C)         </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the word "Canon" in red font. The font is bold and slightly italicized. The word is centered in the image and is the only element present.

# 510(k) SUMMARY

| Submitter:                         | Canon Inc.<br>30-2 Shimomaruko, 3-chrome<br>Ohta-ku, Tokyo 146-8501 Japan                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Contact Person:                    | Mr. Akira Hirai<br>General Manager<br>TEL: 81-3-3758-2111;<br>FAX: 044-739-6695<br>hirai.akira@mail.canon                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |
| Date Prepared:                     | July 19, 2021                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |
| Submission Type:                   | Traditional 510(k) Submission                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |
| Proposed Device:                   | Manufacturer: Canon Inc.<br>Trade Name: Intelligent NR<br>Common Name: Solid State X-Ray Imager (Flat Panel/Digital Imager)<br>Classification Name: Stationary X-ray System<br>Classification: 892.1680, Stationary X-ray System<br>Product Code: MQB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |
| Predicate Device:                  | 510k Number: K190368<br>Manufacturer: Canon Inc.<br>Trade Name: Enhanced Feature Software Pack for CXDI Series<br>Common Name: Solid State X-Ray Imager (Flat Panel/Digital Imager)<br>Classification Name: Stationary X-ray System<br>Classification: 892.1680, Stationary X-ray System<br>Product Code: MQB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |
| Device Description:                | The Intelligent NR (Intelligent Noise Reduction) function is a part of the CXDI<br>Control software for the Canon detectors (with version 3.10 of the CXDI Control<br>Software) that makes use of the Intelligent NR function to reduce noise for X-ray<br>images taken using the Canon detectors. The Intelligent NR function was developed<br>using machine-learning and trained by using an existing clinical image database to<br>learn the characteristics of noises and create noise reduced images. The Intelligent<br>NR function does not perform machine learning after release to users. The CXDI<br>control software, which Intelligent NR is a part of, provides system control, controls<br>GUI on the monitor, and processes images. The Intelligent NR function works on a<br>PC and displays to a monitor. The Intelligent NR function is used in conjunction<br>with the cleared Canon detectors compatible with CXDI Control Software V3.10.<br>The firmware within compatible Canon detectors to be used with this device is<br>unchanged, and no firmware update is necessary for compatibility with the<br>Intelligent NR. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |
| Indication for Use:                | As a part of the Canon radiography system, the CXDI Control Software when used<br>with a compatible Canon detector is intended to provide digital image capture,                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |
|                                    | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |  |
| Trade Name                         | Intelligent NR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Enhanced Feature Software Pack for CXDI Series                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |  |
| 510(k)<br>Submitter<br>[Number]    | Canon Inc.<br>[TBD]                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Canon Inc.<br>[K190368]                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |  |
| Indication for<br>Use              | As a part of the Canon radiography system, the CXDI<br>Control Software when used with a compatible Canon<br>detector is intended to provide digital image capture,<br>processing, and display for conventional film/screen<br>radiographic examinations. This device is intended to<br>replace radiographic film/screen systems in all general<br>purpose diagnostic procedures including specialist<br>areas like intensive care, trauma, and pediatric work.<br>This device is not intended for fluoroscopic,<br>angiographic, or mammography applications.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | As a part of the CXDI series radiography system, the<br>CXDI Control Software when used with a compatible<br>CXDI detector is intended to provide digital image<br>capture, processing, and display for conventional<br>film/screen radiographic examinations. This device is<br>intended to replace radiographic film/screen systems<br>in all general purpose diagnostic procedures including<br>specialist areas like intensive care, trauma, and<br>pediatric work. This device is not intended for<br>fluoroscopic, angiographic, or mammography<br>applications. |  |
| Software /<br>Version              | CXDI Control Software V3.10                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | CXDI Control Software V2.17                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |
| Scatter<br>Correction              | An image can be created (with high contrasts) by<br>using software algorithm, in clinical field, without a<br>grid                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | An image can be created (with high contrasts) by<br>using software algorithm, in clinical field, without a<br>grid                                                                                                                                                                                                                                                                                                                                                                                                                                                     |  |
| One Shot<br>Long Length<br>Imaging | One exposure to obtain images across multiple<br>detectors, automatically stitched together with ability<br>to make manual adjustments after the automatic<br>stitching.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | One exposure to obtain images across multiple<br>detectors, automatically stitched together with ability<br>to make manual adjustments after the automatic<br>stitching.                                                                                                                                                                                                                                                                                                                                                                                               |  |
| Intelligent NR<br>Function         | Process to reduce noise from the taken images by<br>using function machine trained on characteristics of<br>noises included in X-ray image signals using an<br>existing clinical image database                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |  |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Canon" in a bold, red font. The letters are closely spaced together, creating a solid block of text. The font is sans-serif, and the overall impression is one of a strong, recognizable brand logo. The background is plain white.

processing, and display for conventional film/screen radiographic examinations. This device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures including specialist areas like intensive care, trauma, and pediatric work. This device is not intended for fluoroscopic, angiographic, or mammography applications.

Summary of Comparison with the predicate devices shows the characteristics of the proposed Technological device to be substantially equivalent to the predicate device. Characteristics:

The Operation Manual provides detailed instructions and information for safe and effective use of the device, and users are expected to adhere to the instructions and other information. The Operation Manual explains how to use the software and other equipment.

Performance: The hardware within the Canon Digital Radiography Canon Detectors used with the Intelligent NR has not been modified, the detectors have the same performance, biocompatibility, effectiveness, and safety and is substantially equivalent to the predicate device. The risks and hazardous impacts of the device modification were analyzed by FMEA methodology. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented as part of product design. The overall assessment concluded that all identified risks and hazardous conditions were successfully mitigated and accepted.

> As reported in prior submissions to FDA, the compatible detectors comply with the U.S. Performance Standard for radiographic equipment and with relevant voluntary

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Canon" in a bold, red font. The letters are slightly blurred, giving the logo a soft appearance. The logo is simple and recognizable, representing the well-known electronics company.

safety standards for Electrical Safety and Electromagnetic Compatibility testing, specifically IEC standards 60601-1, 60601-1-2, 60601-1-3, and 60601-2-32. The wireless detectors also comply with the FCC test standard for SAR, specifically 47CFR 2.1093 and for EMI test regulations FCC Part 15 Subpart B:2012 Class A and ICES-003 Issue 5:2012 Class A.

- Conclusion: Canon Inc. considers the Intelligent NR to be substantially equivalent to the predicate device listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, and established medical use.

---

**Source:** [https://fda.innolitics.com/device/K212269](https://fda.innolitics.com/device/K212269)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
