VITEK 2 AST- Gram Positive Telavancin (<=0.015 - >=1 µg/mL)

K212243 · bioMerieux, Inc. · LON · Aug 8, 2022 · Microbiology

Device Facts

Record IDK212243
Device NameVITEK 2 AST- Gram Positive Telavancin (<=0.015 - >=1 µg/mL)
ApplicantbioMerieux, Inc.
Product CodeLON · Microbiology
Decision DateAug 8, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Intended Use

VITEK® 2 AST-Gram Positive Telavancin is designed for antimicrobial susceptibility testing of Gram positive microorganisms and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Positive Telavancin is a quantitative test. Telavancin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active both in vitro and in clinical infections Staphylococcus aureus (including methicillin-resistant isolates) Enterococcus faecalis (vancomycin-susceptible isolates only) The VITEK® 2 Gram-positive Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed.

Device Story

VITEK 2 AST-GP Telavancin is an automated, miniaturized, growth-based antimicrobial susceptibility test (AST) system. It uses 64-well cards containing premeasured antimicrobial concentrations and nutrient media. Input: bacterial isolate suspension (standardized via DensiCHEK Plus to 0.5 McFarland). Operation: system fills, seals, and incubates cards; optical scanner measures light attenuation every 15 minutes to monitor growth. Output: MIC values and interpretive category results. Used in clinical laboratories by technicians/microbiologists. Results assist clinicians in determining in vitro susceptibility of specific Gram-positive cocci to telavancin, aiding therapeutic decision-making. Limitations exist for specific organism/antibiotic combinations requiring alternative testing methods due to major errors or insufficient non-susceptible strain data.

Clinical Evidence

External evaluation using contemporary clinical isolates and challenge strains. Compared VITEK 2 AST-GP Telavancin performance against CLSI broth microdilution reference method. Results: S. aureus (n=520) showed 94.2% EA and 98.8% CA; E. faecalis (n=368) showed 89.9% EA and 92.9% CA. Major error rate for E. faecalis adjusted to 5.7% due to lack of intermediate breakpoint. Reproducibility was 100%.

Technological Characteristics

Automated growth-based detection using light transmittance optics. 64-well AST cards. Concentrations: 0.015, 0.125, 0.25, 0.5 µg/mL. Inoculum: 0.45-0.50% saline. Instrument: VITEK 2/VITEK 2 Compact (software v9.04). Standards: CLSI M07-A11, M100-S30. Sterilization: Not specified.

Indications for Use

Indicated for antimicrobial susceptibility testing of Gram-positive microorganisms (Staphylococcus spp., Enterococcus spp., and S. agalactiae) in clinical laboratories. Specifically for Staphylococcus aureus (including MRSA) and Enterococcus faecalis (vancomycin-susceptible).

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ August 8, 2022 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and name on the right. The seal features an eagle with its wings spread, and the FDA part includes the acronym in a blue square and the full name "U.S. Food & Drug Administration" in blue text. bioMérieux, Inc Cherece Jones Staff Regulatory Affairs Specialist 595 Anglum Rd. Hazelwood, Missouri 63042 Re: K212243 Trade/Device Name: VITEK 2 AST- Gram Positive Telavancin (≤0.015 - ≥1 µg/mL) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System Regulatory Class: Class II Product Code: LON, LTT, LTW Dated: July 14, 2021 Received: July 19, 2021 Dear Cherece Jones: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR {1}------------------------------------------------ 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212243 Device Name VITEK® 2 AST-GP Telavancin (<0.015 - >1 ug/mL) #### Indications for Use (Describe) VITEK® 2 AST-Gram Positive Telavancin is designed for antimicrobial susceptibility testing of Gram positive microorganisms and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Positive Telavancin is a quantitative test. Telavancin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active both in vitro and in clinical infections Staphylococcus aureus (including methicillin-resistant isolates) Enterococcus faecalis (vancomycin-susceptible isolates only) The VITEK® 2 Gram-positive Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------| | <div> <span> <b> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the burden time for this collection of information is estimated to average 79 hours per time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle with a blue top half and a green and yellow bottom half. The word "BIOMÉRIEUX" is written in white letters in the center of the blue half of the circle. ### 510(k) SUMMARY B. C. ### VITEK® 2 AST Gram-Positive Telavancin (≤0.015 - ≥1 µg/mL) #### A. 510(k) Submission Information: | Submitter's Name: | bioMérieux, Inc. | |----------------------|-------------------------------------------------------------------------------------------------------| | Address: | 595 Anglum Road<br>Hazelwood, MO 63042 | | Contact Person: | Cherece L. Jones<br>Staff Regulatory Affairs Specialist | | Phone Number: | 314-731-8684 | | Fax Number: | 314-731-8689 | | Date of Preparation: | July 14, 2021 | | Device Name: | | | Formal/Trade Name: | VITEK® 2 AST Gram-Positive Telavancin (≤0.015 - ≥1<br>µg/mL) | | Classification Name: | 21 CFR 866.1645<br>Fully Automated Short-Term Incubation Cycle<br>Antimicrobial Susceptibility System | | Product Code(s): | LON, LTT, LTW | | Common Name: | VITEK® 2 AST-GP Telavancin (≤0.015 - ≥1 µg/mL) | | Predicate Device: | VITEK® 2 AST- GP Dalbavancin (≤0.015 - ≥1 µg/mL)<br>(K190616) | #### D. Device Description: The principle of the VITEK®2 AST cards is based on the microdilution minimum inhibitory concentration (MIC) technique reported by MacLowry and Marsh(1) and Gerlach(0). The VITEK @ 2 AST card is essentially a miniaturized, abbreviated and automated version of the doubling dilution technique(3). {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle that is split into two halves. The top half is blue and contains the word "BIOMÉRIEUX" in white, sans-serif font. The bottom half is yellow and green. Each VITEK® 2 AST card contains 64 wells. A control well which only contains microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The isolate to be tested is diluted to a standardized concentration with 0.45 – 0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling, sealing and loading operation. The VITEK® 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card. ### E. Substantial Equivalence Information: The similarities and differences of the VITEK® 2 AST-GP Telavancin (≤0.015 - ≥1 µg/mL) when compared to the predicate device, VITEK® 2 AST-GP Dalbavancin (≤0.015 - ≥1 µg/mL) (K190616), are described in the Table 1 below. | Item | Device:<br>VITEK® 2 AST-GP Telavancin<br>(≤0.015 - ≥1 µg/mL) | Predicate:<br>VITEK® 2 AST-GP Dalbavancin<br>(≤0.015 - ≥1 µg/mL) (K190616) | |--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Similarities | | | | Intended Use | VITEK® 2 AST-Gram Positive<br>Telavancin is designed for<br>antimicrobial susceptibility testing of<br>Gram positive microorganisms and is<br>intended for use with the VITEK® 2<br>and VITEK® 2 Compact Systems as<br>a laboratory aid in the determination<br>of <i>in vitro</i> susceptibility to<br>antimicrobial agents. VITEK® 2<br>AST-Gram Positive Telavancin is a<br>quantitative test. Telavancin has been<br>shown to be active against most<br>strains of the microorganisms listed<br>below, according to the FDA label<br>for this antimicrobial.<br><br>Active both <i>in vitro</i> and in clinical<br>infections<br><i>Staphylococcus aureus</i> (including<br>methicillin-resistant isolates)<br><i>Enterococcus faecalis</i> (vancomycin- | VITEK® 2 AST-Gram Positive<br>Dalbavancin is designed for<br>antimicrobial susceptibility testing of<br>Gram positive microorganisms and is<br>intended for use with the VITEK® 2<br>and VITEK® 2 Compact Systems as a<br>laboratory aid in the determination of<br><i>in vitro</i> susceptibility to antimicrobial<br>agents. VITEK® 2 AST-Gram Positive<br>Dalbavancin is a quantitative test.<br>Dalbavancin has been shown to be<br>active against most strains of the<br>microorganisms listed below,<br>according to the FDA label for this<br>antimicrobial.<br><br>Active both <i>in vitro</i> and in clinical<br>infections<br><i>Staphylococcus aureus</i> (including<br>methicillin-resistant isolates)<br><i>Enterococcus faecalis</i> (vancomycin- | #### Table 1: Substantial Equivalence {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle with a blue top half and a yellow-green gradient bottom half. The word "BIOMÉRIEUX" is written in white, sans-serif font in the center of the blue portion of the circle. | Item | Device:<br>VITEK® 2 AST-GP Telavancin<br>(≤0.015 - ≥1 µg/mL) | Predicate:<br>VITEK® 2 AST-GP Dalbavancin<br>(≤0.015 - ≥1 µg/mL) (K190616) | |---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Similarities | | | | | susceptible isolates only) | susceptible isolates only)<br><i>Streptococcus agalactiae</i> | | | The VITEK® 2 Gram-positive<br>Susceptibility Card is intended for use<br>with the VITEK® 2 Systems in clinical<br>laboratories as an <i>in vitro</i> test<br>to determine the susceptibility of<br><i>Staphylococcus</i> spp., <i>Enterococcus</i><br>spp., and <i>S. agalactiae</i> to<br>antimicrobial agents when used as<br>instructed. | The VITEK® 2 Gram-positive<br>Susceptibility Card is intended for use<br>with the VITEK® 2 Systems in clinical<br>laboratories as an <i>in vitro</i> test<br>to determine the susceptibility of<br><i>Staphylococcus</i> spp., <i>Enterococcus</i><br>spp., and <i>S. agalactiae</i> to<br>antimicrobial agents when used as<br>instructed. | | Test Methodology | Automated quantitative antimicrobial<br>susceptibility test for use with the<br>VITEK® 2 and VITEK® 2 Compact<br>Systems to determine the <i>in vitro</i><br>susceptibility of microorganisms | Same | | Inoculum | Saline suspension of organism | Same | | Test Card | Gram-Positive (AST-GP)<br>Susceptibility Card | Same | | Analysis Algorithms | Growth Pattern Analysis | Same | | Instrument | VITEK® 2 and VITEK® 2 Compact<br>Systems | Same | | Differences | | | | Antimicrobial Agent | Telavancin | Dalbavancin | | Concentrations | 0.015, 0.125, 0.25, 0.5 | 0.0625, 0.125, 0.25, 0.5 | #### F. Indications for Use: VITEK® 2 AST-Gram Positive Telavancin is designed for antimicrobial susceptibility testing of Gram positive microorganisms and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Positive Telavancin is a quantitative test. Telavancin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active both in vitro and in clinical infections Staphylococcus aureus (including methicillin-resistant isolates) Enterococcus faecalis (vancomycin-susceptible isolates only) {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle with a blue upper half and a yellow-green lower half. The word "BIOMÉRIEUX" is written in white, sans-serif font in the center of the blue portion of the circle. The VITEK® 2 Gram-positive Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed. #### G. Performance Overview and Conclusion: VITEK® 2 AST-GP Telavancin (≤0.015 - ≥1 ug/mL) demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009). The Premarket Notification (510|k]) presents data in support of VITEK® 2 AST-GP Telavancin. An external evaluation was conducted with contemporary and stock clinical isolates, as well as a set of challenge strains. The external evaluations were designed to confirm the acceptability of AST-GP Telavancin by comparing its performance with the CLSI broth microdilution reference method incubated at 16-20 hours for Staphylococci and Enterococci. 20-24 hours for Streptococci. The data is representative of performance on both the VITEK® 2 and VITEK® 2 Compact instrument platforms. The VITEK® 2 AST-GP Telavancin (≤0.015 - ≥1 µg/mL) demonstrated acceptable performance as presented in Table 2 below: | Antimicrobial | Comment | Essential Agreement Category | | | | Category Agreement | | | | % Reproducibility | | |---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|-----|-----|-----|--------------------|--------------|------------------|-----|-------------------|--| | | | % Error | | | | % Error | | | | | | | | | %EA | VME | ME | mE | %CA | VME | ME | mE | | | | | #, E<br>Enterococcus<br>faecalis | 89.9<br>(331/368) | N/A | N/A | N/A | 92.9<br>(342/368) | 0.0<br>(0/1) | 7.1<br>(26/367)* | N/A | | | | Telavancin | #,E<br>Staphylococcus<br>aureus | 94.2<br>(490/520) | N/A | N/A | N/A | 98.8<br>(514/520) | 0.0<br>(0/0) | 1.2<br>(6/520) | N/A | 100.0 | | | | * Five major errors for Enterococcus faecalis were within essential agreement with the reference method; due to<br>the lack of an intermediate breakpoint these five major errors were considered acceptable and the major error rate<br>is adjusted to 5.7% which will still require retesting of non-susceptible strains with an alternate method.<br><br>VITEK 2 AST-Gram Positive Telavancin MIC values tended to be in exact agreement or at least one doubling<br>dilution higher when testing E. faecalis, and S. aureus compared to the CLSI reference broth microdilution<br>method. | | | | | | | | | | | Table 2: VITEK® 2 AST-GP Telavancin Performance Kev: <sup># =</sup> US Food and Drug Administration 510(k) cleared {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle with two different colors. The top half of the circle is dark blue, and the bottom half is yellow-green. The company name, "BIOMÉRIEUX", is written in white letters in the center of the blue portion of the circle. E = External performance data Quality Control demonstrated acceptable results. #### H. References: - 1. MacLowry, J.D. and Marsh, H.H., Semi-automatic Microtechnique for Serial Dilution Antibiotic Sensitivity Testing in the Clinical laboratory, Journal of Laboratory Clinical Medicine, 72:685-687, 1968. - 2. Gerlach, E.H., Microdilution 1: A Comparative Study, p. 63-76. Current Techniques for Antibiotic Susceptibility Testing. A. Balows (ed.), Charles C. Thomas, Springfield, IL, 1974. - 3. Barry, A.L., The Antimicrobic Susceptibility Test, Principles and Practices, Lea and Febiger, Philadelphia, PA, 1976.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...