TheraFace Microcurrent

K212238 · Theragun, Inc. · NFO · Sep 29, 2021 · Neurology

Device Facts

Record IDK212238
Device NameTheraFace Microcurrent
ApplicantTheragun, Inc.
Product CodeNFO · Neurology
Decision DateSep 29, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Intended Use

The TheraFace microcurrent is a hand-held device for over-the-counter aesthetic purposes . The TheraFace microcurrent is indicated for facial stimulation.

Device Story

Hand-held device for over-the-counter aesthetic facial stimulation; utilizes internal lithium rechargeable battery. Device features electrode contactors applied to facial skin; delivers regulated biphasic pulsed electrical microcurrent. User adjusts output intensity via device buttons or mobile app via Bluetooth. Device includes automatic overload/no-load trip and automatic shut-off. Intended for home use by consumers; provides non-invasive electrical stimulation to facial muscles. Output parameters (voltage, current, frequency) are designed to be comparable to predicate devices to ensure safety and effectiveness. Device performance verified through bench testing, biocompatibility, and usability studies.

Clinical Evidence

Bench testing only. Evidence includes biocompatibility (ISO 10993-5, ISO 10993-10), electrical safety (IEC 60601-1, IEC 60601-1-11, IEC 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), wireless coexistence (AAMI TIR69, ANSI C63.27), and a usability engineering study (TG-Theraface microcurrent-013). No clinical trials were performed.

Technological Characteristics

Hand-held stimulator; PC plastic and stainless steel construction. Sensing/actuation: Biphasic pulsed electrical microcurrent (8.3 Hz, 60ms pulse width). Energy source: Internal lithium rechargeable battery. Connectivity: Bluetooth for mobile app control. Sterilization: Not specified. Software: Microprocessor-controlled with automatic overload/no-load trip and shut-off. Complies with IEC 60601-1, IEC 60601-1-11, and IEC 60601-2-10.

Indications for Use

Indicated for facial stimulation in adults for over-the-counter aesthetic purposes.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 29, 2021 Theragun, Inc. % Thomas Padula Vice President Regulatory Compliance Schiff & Company, Inc. 583 Mountain Avenue North Caldwell, New Jersey 07006 Re: K212238 Trade/Device Name: TheraFace microcurrent Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NFO Dated: July 15, 2021 Received: July 19, 2021 Dear Thomas Padula: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212238 Device Name TheraFace microcurrent Indications for Use (Describe) The TheraFace microcurrent is a hand-held device for over-the-counter aesthetic purposes . The TheraFace microcurrent is indicated for facial stimulation. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(K) PREMARKET NOTIFICATION FOR THERAFACE MICROCURRENT THERAGUN, INC. ### 510(k) Summary (as required by 807.92) ### (1) SUBMITTER: THERAGUN, Inc. 6100 Wilshire Blvd Suite 200 Los Angeles, CA 90048 Registration Number: 3012386142 FEI Number: 3012386142 Contact person: CJ Frederick - Director, Regulatory Compliance Telephone: 310-570-8341 Email: jaime@therabodycorp.com Date prepared: September 20, 2021 ### Application Correspondent: Contact Person: Thomas Padula Company: Schiff & Company, Inc. Address: 583 Mountain Avenue, North Caldwell, NJ 07006 Tel: 201-317-8810 Email: thomaspadula(@schiffandcompany.com ### (2) DEVICE NAME: Trade Name: TheraFace microcurrent Common Name: Stimulator, Transcutaneous Electrical, Aesthetic Purposes Classification Name: Stimulator, Transcutaneous Electrical, Aesthetic Purposes Device Classification: Class II Review Panel: Neurology; General & Plastic Surgery Regulation Number: 21 CFR 882.5890 Product Code: NFO | Sponsor | Heat In A Click | |-----------------------|-------------------------| | Device Name and Model | 2 Face / Face Evolution | | 510(k) Number | K171821 | | Product Code | NFO | | Regulation Number | 882.5890 | | Regulation Class | II | (3) PREDICATE DEVICE(S): Substantial equivalence is based on following legally marketed devices. ### (4) DESCRIPTION OF THEDEVICE: The TheraFace microcurrent device consist of main control unit and its attachment applicators for microcurrent output which can be control by the device control button or APP in cell phone via Bluetooth {4}------------------------------------------------ connection. The device powered by one internal lithium rechargeable battery which can be charged by external battery charger. The device has the electrode contractors for facial stimulation by applying an electrical micro current to face skin. The output waveform is formed of regulated Voltage of Biphasic pulse and provided with 3 levels of output intensity, which can be adjusted by user. The low-level electrical current pulse goes through the face muscle and cause face muscle contraction for facial stimulation. ### (5) INDICATIONS FOR USE: The TheraFace microcurrent is a hand-held device for over-the-counter aesthetic purposes. The TheraFace microcurrent is indicated for facial stimulation. ### (6) COMPARISON WITH PREDICATEDEVICES: Following table is a comparison of TheraFace microcurrent and predicate devices. TheraFace microcurrent is substantially equivalent in terms of the technological characteristics, features, specifications, materials, mode of operation and indications for use, to 2 Face Evolution - K 171821, cleared for marketing under510(K). | Elements of<br>Comparison | Subject Device | Primary Predicate Device | Remark | Intended Use | The TheraFace microcurrent is a<br>hand-held device for over-the-counter<br>aesthetic purpose. The TheraFace<br>microcurrent is indicated for facial<br>stimulation. | 2 Face / Face<br>Evolution is a hand-held device for over the<br>counter aesthetic purposes.<br>(1) The EMS mode is indicated for facial<br>stimulation.<br>(2) The Photon mode: The red light is<br>intended for the treatment of periorbital<br>wrinkles and the blue light is intended<br>for the treatment of the mild to moderate<br>inflammatory acne. | SE<br>Note 1<br>(Only compared with<br>its EMS function of<br>K171821) | |-------------------------------------------------|-------------------------------|--------------------------|--------|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------| | Basic specification | | | | Apply parts | Face | Face | SE | | Device Name and<br>Model | <i>TheraFace microcurrent</i> | 2 Face / Face Evolution | -- | Power Sources | Lithium battery :2x3.7V | DC 3.7V 2200mAh | SE | | 510 (K) Number | Applying | K171821 | -- | Method of Line<br>Current Isolation | Battery Supply | Battery Supply | SE | | Product Code | NFO | NFO | -- | Number of Modes for<br>Micro current<br>stimulation | l | 1 | SE | | Regulation Number | 882.5890 | 882.5890 | -- | Number of Channels<br>for Micro current<br>stimulation | l | l | SE | | Synchronous or<br>Alternating | N/A | N/A | SE | | | | | | Regulated Current or<br>Regulated Voltage | Both | Both | SE | | | | | | Software/Firmware/<br>Microprocessor<br>control | Yes | Yes | ਟਜ | | | | | | Automatic Overload<br>Trip | Yes | Yes | ਟਜ | | | | | {5}------------------------------------------------ {6}------------------------------------------------ | Automatic No-load<br>Trip | | Yes. | Yes | SE | |-------------------------------------|------------------------------|-------------------------------------------------|---------------------------------------------------|--------------| | Automatic Shut Off | | Yes. | Yes | SE | | Patient Override<br>Control | | Yes | Yes | SE | | Indicator<br>Display | On/Off<br>Status | Yes | Yes | SE | | | Low<br>Battery | Yes | Yes | SE | | | Voltage/<br>Current<br>Level | Yes | Yes | SE | | Time Range | | Micro current 5 minutes | EMS Mode (5 minutes)<br>Photon Mode (5~7 minutes) | SE | | Console weight | | 230g | 200g | SE<br>Note 2 | | Waveform | | Pulsed Biphasic | Pulsed Biphasic, | SE | | Shape | | Rectangular | Rectangular | SE | | Maximum Output<br>Voltage (+/- 10%) | | 0.24V @ $500Ω$<br>1.0V @ $2kΩ$<br>5.0V @ $10kΩ$ | 0.31V @ $500Ω$<br>1.16V @ $2kΩ$<br>5.56V @ $10kΩ$ | SE<br>Note 3 | | | | | | | | Maximum output<br>Current | 500 μA @ 500 Ohm | 620μΑ @ 500Ω | SE | | | | 500 μA @ 2k Ohm | 580μA @ 2kΩ | | | | | 500 μA @ 10k Ohm | 556μΑ @ 10kΩ | Note 3 | | | Frequency range | 8.3 Hz | 8.33 Hz | SE | | | Pulse width range | 60ms | 60ms | SE | | | Net Charge | 0 μC @ 500Ω | 0 μC @500 Ω | SE | | | Max Phase charge | 29.4µC@ 500 Ohm | 26.3μC@ 500 Ohm | SE Note 4 | | | Maximum Current<br>Density | 0.2mA/cm² @500Ω | 0.33 mA/cm²@ 500Ω | SE Note 5 | | | Maximum Power<br>Density | 23.1μW/cm² @500Ω | 4.34 μW/cm² @ 500Ω | SE Note 5 | | | ON time | 60 ms | 60 ms | SE | | | OFF time | 60 ms | 60 ms | SE | | | Contraction and<br>Relaxation time | 60 ms ON/OFF | 60 ms ON/OFF | SE | | {7}------------------------------------------------ {8}------------------------------------------------ | Material of device and<br>construction | PC Plastic & Stainless Steel | ABS Plastic & Stainless Steel | SE Note 6 | |----------------------------------------|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------| | Bio-compatibility | All user directly contacting materials<br>are compliance with ISO10993-5,<br>ISO10993-10 | All user directly contacting materials are<br>compliance with ISO10993-5 and<br>ISO10993-10 requirements. | SE | | Electrical Safety | Comply with IEC 60601-1<br>IEC 60601-1-11<br>IEC 60601-2-10 | Comply with: IEC 60601-1<br>IEC 60601-1-11<br>IEC 60601-2-10 | SE | | EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | SE | ## Comparison in Detail(s): Note 1: There is multiple function operation mode in the predicate device, the subject device TheraFace microcurrent only has one micro-current operation mode. To this microcurrent mode, the subject device can be substantial equivalent to predicate device K171821. Note 2: Even there is console weight difference between subject device and predicate devices; but they all are portable medical device and comply with IEC60601-1 testing, such minor difference would not affect safety and effectiveness issue. Note 3:(Maximum Output Voltage and Maximum Output Current): The effect of micro current stimulation is determined by micro current output waveform and output current. There is only minor difference about the output voltage and current between the subject device and the predicated devices, it can still obtain the same effect because our output voltage and output current are in the range which is between the value of K171821. Also, the subject device complies with IEC 60601-1, and IEC60601-2-10 which means we have proved its safety as well as the effectiveness comparing with the predicate devices. Therefore, the subject device and predicate devices are substantially {9}------------------------------------------------ equivalence on output voltage and current. Note 4: (Max phase charge) Even there is minor difference on the max phase charge between subject device and predicate device, but the pulse waveform is Pulsed Biphasic, and its net charge is 0 µC; And the micro current waveform parameters comply with IEC60601-2-10, so the difference on max phase charge would not raise safety issue; and it would not affect effectiveness for facial stimulation. Therefore, such minor difference would not affect safety and effectiveness issue. Note 5 (Maximum current density and Maximum power density): The effect of micro current stimulation on facial skin are determined by micro-current output waveform and output current. The maximum current density and power density is the parameters related to safety of medical device. For maximum current density, there is only little difference on it between the subject device and the predicated device K171821, it is little smaller than the value of K171821, so it would not raise safety issue. And viewing from max current density value in the predicate device K171821, the value of subject device is in the range of max current density between predicate device K171821. Therefore, such minor difference would not affect safety and effectiveness issue. For maximum power density, the value of maximum power density of subject device is larger than the predicate device. But the maximum power density meets with the maximum allowed value 0.25 (W/cm²) required in FDA guidance. Also, the subject device complies with IEC 60601-1, and IEC60601-2-10 which means we have proved its safety as well as the effectiveness comparing with the predicate devices. Therefore, such minor difference would not affect safety and effectiveness issue. ### Note 6 Even there is different for the material between subject device and predicate device, but its material complies with ISO10993-5, and ISO10993-10. ### Finial Conclusion: The TheraFace microcurrent device has the same indications use and technology characteristics as the predicate devices. TheraFace microcurrent device is as safe, as effective, and performs as well as the predicate devices. {10}------------------------------------------------ ## 510(K) PREMARKET NOTIFICATION FOR THERAFACE MICROCURRENT THERAGUN, INC. ### (7) PERFORMANCE STSTANDARDS APPLIED: A series of studies were completed to demonstrate the substantial equivalence of TheraFace microcurrent to the predicate devices. All testing was conducted in accordance with and in conformance to applicable device regulations and guidance. Results of all testing demonstrate the device is non-toxic, is comparable to other currently marketed devices and is substantially equivalent to legally marketed predicates and included: ### Biocompatibility - ISO 10993-5:2009, biological evaluation of medical devices -- part 5: tests for in vitro cytotoxicity. (Biocompatibility). - ISO 10993-10 :2010, biological evaluation of medical devices part 10: tests for irritation and skin sensitization. (Biocompatibility). Electrical Safety and Electromagnetic Compatibility - IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012 (or IEC 60601-1:2012 reprint) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - IEC / EN 60601-1-2: 2014-02, Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - collateral standard: Electromagnetic Compatibility. - IEC 60601-1-11 :2015. Medical electrical equipment -- part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home health care environment. - IEC 60601-2-10: 2012, Medical electrical equipment -- Part 2: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators. TheraFace Micro-current Test Report, Theragun study TG-TheraFace-1010 AAMI TIR69:2017, ANSI C63.27-2017 Wireless Coexistence Test Usability Engineering Usability Study Report, Theragun study TG-Theraface microcurrent-013 ### (8) CONCLUSION: TheraFace microcurrent has the same indications for use and technology characteristics as the predicate devices. TheraFace microcurrent is as safe, as effective, and performs as well as the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...