Atellica CH Enzymatic Creatinine_3 (ECre3)

K212223 · Siemens Healthcare Diagnostics, Inc. · JFY · Nov 24, 2021 · Clinical Chemistry

Device Facts

Record IDK212223
Device NameAtellica CH Enzymatic Creatinine_3 (ECre3)
ApplicantSiemens Healthcare Diagnostics, Inc.
Product CodeJFY · Clinical Chemistry
Decision DateNov 24, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1225
Device ClassClass 2

Indications for Use

The Atellica® CH Enzymatic Creatinine_3 (ECre3) assay is for in vitro diagnostic use in the quantitative determine of creatinine in human serum, plasma (lithium heparin and dipotassium EDTA), and urine using the Atellica® CH Analyzer. Such measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis.

Device Story

Atellica® CH Enzymatic Creatinine_3 (ECre3) assay; in vitro diagnostic reagent for Atellica® CH Analyzer. Input: human serum, plasma (lithium heparin/K2-EDTA), or urine samples. Principle: coupled enzymatic reaction; creatininase converts creatinine to creatine; creatinase converts creatine to sarcosine; sarcosine oxidase produces hydrogen peroxide; peroxidase catalyzes oxidative condensation of 4-aminoantipyrine and EMSE to form reddish-purple quinone pigment. Output: absorbance measured at 545/694 nm as endpoint reaction; concentration calculated. Used in clinical laboratories by technicians/pathologists. Output aids clinicians in diagnosing/treating renal disease and monitoring dialysis. Benefits: provides quantitative renal function assessment.

Clinical Evidence

Bench testing only. Precision/reproducibility evaluated per CLSI EP05-A3 (N=80-225 replicates). Linearity evaluated per CLSI EP6-A (deviation <5%). Interference testing per CLSI EP07-3rd edition. Method comparison against predicate (N=105 serum, N=102 urine) showed high correlation (R=1.000 and 0.999). Traceability to NIST SRM 967 and IDMS alignment (N=47, R=0.991) confirmed. No clinical trials performed.

Technological Characteristics

Enzymatic creatinine assay; reagents include creatinase, sarcosine oxidase, peroxidase, 4-aminoantipyrine, and EMSE. Endpoint spectrophotometric measurement at 545/694 nm. Traceable to NIST SRM 967. Automated dilution capability. Compatible with Atellica® CH Analyzer. Software-controlled automated analysis.

Indications for Use

Indicated for quantitative determination of creatinine in human serum, plasma (lithium heparin, dipotassium EDTA), and urine for diagnosis/treatment of renal diseases and monitoring renal dialysis.

Regulatory Classification

Identification

A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY ## I Background Information: A 510(k) Number K212223 B Applicant Siemens Healthcare Diagnostics Inc. C Proprietary and Established Names Atellica® CH Enzymatic Creatinine_3 (ECre3) D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JFY | Class II | 21 CFR 862.1225 - Creatinine Test System | CH - Clinical Chemistry | ## II Submission/Device Overview: A Purpose for Submission: New Device B Measurand: Creatinine C Type of Test: Quantitative, enzymatic Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} K212223 - Page 2 of 11 # III Intended Use/Indications for Use: ## A Intended Use(s): See Indications for Use below. ## B Indication(s) for Use: The Atellica® CH Enzymatic Creatinine_3 (ECre3) assay is for in vitro diagnostic use in the quantitative determine of creatinine in human serum, plasma (lithium heparin and dipotassium EDTA), and urine using the Atellica® CH Analyzer. Such measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis. ## C Special Conditions for Use Statement(s): Rx - For Prescription Use Only ## D Special Instrument Requirements: Atellica® CH Analyzer/Trinidad CH System # IV Device/System Characteristics: ## A Device Description: The Atellica® CH Enzymatic Creatinine_3 (ECre3) assay reagents are packaged in 2 packs with 2 wells per pack. The wells have a divider which keeps the two liquid chambers separate. In pack 1 (P1) and pack 2 (P2) both wells are used. Atellica® CH ECre3 Pack 1 (P1) consist of 23.5 mL of Creatinase (&lt;133,350 U/L); sarcosine oxidase (&lt;39,000 U/L); N-ethyl-N-(3-methylphenyl)-N'-succinyl-ethylenediamine (EMSE) (0.04%); detergent; preservatives, bovine serum albumin (0.1%), bovine catalase. Atellica® CH ECre3 Pack 2 (P2) consist of 11.5 mL of Peroxidase; creatininase (&lt;600,000 U/L); 4-aminoantipyrine (4-AA) (0.001%); detergent; sodium azide (&lt; 0.1%); preservatives. ## B Principle of Operation: This assay uses a creatinine enzymatic method that measures the concentration of creatinine through a series of coupled enzymatic reactions. A "pretreatment" reaction removes endogenous creatine and sarcosine using creatinase and sarcosine oxidase. First, creatinine is enzymatically converted by creatininase into creatine. Creatine is then enzymatically converted to sarcosine by creatinase. This is followed by the oxidation of sarcosine by sarcosine oxidase to produce hydrogen peroxide. In the presence of peroxidase, the hydrogen peroxide allows for the oxidative condensation of 4-aminoantipyrine and N-ethyl-N-(3-methylphenyl)-N'-succinylethylenediamine to produce a reddish purple quinone pigment. The absorbance of this quinone pigment is measured as an endpoint reaction at 545/694 nm. {2} V Substantial Equivalence Information: A Predicate Device Name(s): ADVIA Chemistry Enzymatic Creatinine_2 B Predicate 510(k) Number(s): K070727 C Comparison with Predicate(s): | Device & Predicate Device(s): | K212223 | K070727 | | --- | --- | --- | | Device Trade Name | Atellica® CH Enzymatic Creatinine_3 (ECre3) | ADVIA Chemistry Enzymatic Creatinine_2 (ECre_2) | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | For in vitro diagnostic use in the quantitative determination of creatinine in human serum, plasma (lithium heparin and potassium EDTA), and urine. Such measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis. | Same | | Assay Principle | Enzymatic | Same | | Sample Type | Serum, plasma (lithium heparin and K2 EDTA), urine | Same | | General Device Characteristic Differences | | | | Assay Range | Serum/Plasma: 0.15–30.00 mg/dL Urine: 2.00–245.00 mg/dL | Serum/Plasma: 0.10–30.00 mg/dL Urine: 1.00–245.00 mg/dL | K212223 - Page 3 of 11 {3} K212223 - Page 4 of 11 | Calibrators | Atellica CH Chemistry Calibrator (CHEM CAL) | ADVIA Chemistry Calibrator (K050374) | | --- | --- | --- | VI Standards/Guidance Documents Referenced: - Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline–Third Edition. (CLSI EP05-A3). - Interference Testing in Clinical Chemistry (CLSI EP07-ED3). - Measurement Procedure Comparison and Bias Estimation Using Patient Samples (CLSI EP09c-ED3). - Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline–Second Edition (EP17-A2). - Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline (CLSI EP25-A). - Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline – Third Edition (CLSI EP28-A3c). - Establishing and Verifying an Extended Measuring Interval Through Specimen Dilution and Spiking (CLSI EP34-ED1) VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: Eleven (11) samples comprised of 8 serum and 3 urine sample pools were aliquoted and frozen prior to the start of the study. Testing was performed with three (3) Atellica® CH ECre3 reagent lots on one (1) Atellica® CH Analyzer. For each testing day, new serum and urine pool aliquots were thawed and assayed. Each sample was assayed in two (2) replicates per run, with two (2) runs per day separated by a minimum of 2 hours between runs, for 20 days, yielding a total 80 replicates. Reproducibility was determined in accordance with CLSI EP05-A3 guideline. Samples were assayed n=5 in 1 run for 5 days using 3 instruments and 3 reagent lots. The data were analyzed to calculate the following components of precision: repeatability, between-day, between-lot, between-instrument, and reproducibility (total). The following results were obtained: Precision summary results | Specimen Type | N | Mean mg/dL | Repeatability | | Within-Lab | | | --- | --- | --- | --- | --- | --- | --- | | | | | SD mg/dL | CV (%) | SD mg/dL | CV (%) | | Serum 1 (Diluted) | 80 | 0.41 | 0.009 | 2.2 | 0.013 | 3.2 | | Serum 2 (Native) | 80 | 0.75 | 0.008 | 1.1 | 0.015 | 2.0 | | Serum 3 (Spiked) | 80 | 1.29 | 0.010 | 0.8 | 0.030 | 2.3 | | Serum 4 (QC 1) | 80 | 1.91 | 0.012 | 0.6 | 0.024 | 1.3 | | Serum 5 (QC 2) | 80 | 3.11 | 0.009 | 0.3 | 0.026 | 0.8 | {4} | Specimen Type | N | Mean mg/dL | Repeatability | | Within-Lab | | | --- | --- | --- | --- | --- | --- | --- | | | | | SD mg/dL | CV (%) | SD mg/dL | CV (%) | | Serum 6 (Spiked) | 80 | 8.89 | 0.021 | 0.2 | 0.055 | 0.6 | | Serum 7 (Spiked) | 80 | 18.52 | 0.039 | 0.2 | 0.093 | 0.5 | | Serum 8 (Spiked) | 80 | 26.49 | 0.054 | 0.2 | 0.121 | 0.5 | | Urine 1 | 80 | 42.82 | 0.064 | 0.1 | 0.322 | 0.8 | | Urine 2 | 80 | 86.41 | 0.156 | 0.2 | 0.497 | 0.6 | | Urine 3 | 80 | 185.06 | 0.320 | 0.2 | 0.831 | 0.4 | **Reproducibility Summary results** | Specimen Type | N | Mean mg/dL | Repeatability | | Between-Day | | Between-Lot | | Between Instrument | | Total Reproducibility | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD mg/dL | CV (%) | SD mg/dL | CV (%) | SD mg/dL | CV (%) | SD mg/dL | CV (%) | SD mg/dL | CV (%) | | Serum 1 | 225 | 0.44 | 0.008 | 1.9 | 0.005 | 1.1 | 0.008 | 1.9 | 0.000 | 0.0 | 0.013 | 2.9 | | Serum QC 1 | 225 | 0.79 | 0.011 | 1.3 | 0.004 | 0.5 | 0.008 | 1.0 | 0.000 | 0.0 | 0.014 | 1.7 | | Serum 2 | 225 | 0.95 | 0.014 | 1.4 | 0.034 | 3.5 | 0.000 | 0.0 | 0.007 | 0.7 | 0.037 | 3.9 | | Serum QC 2 | 225 | 1.88 | 0.011 | 0.6 | 0.008 | 0.4 | 0.009 | 0.5 | 0.000 | 0.0 | 0.016 | 0.8 | | Serum QC 3 | 225 | 6.82 | 0.016 | 0.2 | 0.022 | 0.3 | 0.000 | 0.0 | 0.017 | 0.2 | 0.032 | 0.5 | | Serum 3 | 225 | 7.96 | 0.021 | 0.3 | 0.034 | 0.4 | 0.000 | 0.0 | 0.019 | 0.2 | 0.045 | 0.6 | | Serum 4 | 225 | 26.86 | 0.051 | 0.2 | 0.105 | 0.4 | 0.039 | 0.1 | 0.082 | 0.3 | 0.148 | 0.6 | | Urine 1 | 225 | 42.30 | 0.106 | 0.2 | 0.102 | 0.2 | 0.000 | 0.0 | 0.171 | 0.4 | 0.225 | 0.5 | | Urine 2 | 225 | 189.56 | 0.379 | 0.2 | 0.401 | 0.2 | 0.503 | 0.3 | 1.024 | 0.5 | 1.267 | 0.7 | **2. Linearity:** A linearity study was performed following the recommendations in the CLSI EP6-A guideline. Native human serum and native human urine samples were spiked to create the high samples. Native human serum and native human urine were diluted with Atellica® CH Diluent to create the low samples. A set of 8 equally spaced creatinine sample levels were prepared by mixing a low serum/urine sample with a high serum/urine sample. The expected creatinine concentration levels used in the study for serum: 0.01, 3.93, 7.84, 11.75, 15.66, 19.58, 23.49, 27.40, and 31.31 mg/dL and for urine: 0.28, 32.34, 64.48, 96.58, 128.68, 160.78, 192.88, 224.98, and 257.07 mg/dL. Measurements were made with 4 replicates per level. The results showed that the deviation from linearity did not exceed 5%. The results of the regression analysis are summarized below: | Analyte | Sample Type | Linear Regression | | --- | --- | --- | | Creatinine | Serum | y=0.99x+ 0.02 mg/dL | | | Urine | y=0.98x + 0.05 mg/dL | The results from the linearity study support an analytical measuring range of 0.15 to 30.00 mg/dL for serum/plasma and 2.00 to 245.00 mg/dL for urine. K212223 - Page 5 of 11 {5} Dilution: The sponsor provided data to support that automatic dilution can be used to assay samples containing creatinine concentrations above the claimed analytical measuring interval. The results of the dilution study showed that 5-fold dilution could be used to measure samples within an extended measuring interval up to 150.00 mg/dL for the serum/plasma sample type and up to 1225.00 mg/dL for the urine sample type using the automated dilution routine on the Atellica® Chemistry Analyzer. 3. Analytical Specificity/Interference: The interference study was performed following the recommendations in the CLSI EP07-3rd edition guideline. Various concentrations of interferents were spiked into pooled human serum or pooled human urine containing creatinine at normal (1 mg/dL) and high (8 mg/dL) concentrations. Human serum pools were split into control and test pools. The test pools were spiked with the interferents and the control pools were spiked with an equivalent volume of diluent. Low level analyte serum pool was native serum and the high-level analyte serum pool was prepared using a concentrated creatinine spiking solution. Five (5) replicates were tested per sample. Nonsignificant interference was defined by the sponsor as the highest interferent level tested with ≤10% bias of the control for both serum and urine samples. Serum Creatinine: | Test Interferent | Highest concentration tested that showed no interference | | --- | --- | | Hemoglobin | 200 mg/dL | | Lipema | 2000 mg/dL | | Conjugated Bilirubin | 25 mg/dL | | Unconjugated Bilirubin | 25 mg/dL | | Acetaminophen | 200 μg/mL | | N-Acetylcysteine (NAC) | 37.5 mg/dL | | N-acetyl-p-benzoquinine imine (NAPQI) | 0.4 mg/dL | | Methyldopa | 22.5 μg/mL | | L-Dopa | 15 μg/mL | | Calcium Dobesilate | 0.375 mg/dL | | Cefoxitin | 6600 μg/mL | | Cephalexin | 200 μg/mL | | Dicynone (Etamsylate) | 0.59 mg/dL | | DL-proline | 11.5 mg/dL | | Dobutamine | 5 μg/mL | | Dopamine | 10 μg/mL | | Ethylglycine (N-ethylglycine) | 6 μg/mL | | Fluorocytosine | 200 μg/mL | | Metamizole (Sulpyrine) | 25 mg/dL | | Phenindione | 4 mg/dL | K212223 - Page 6 of 11 {6} K212223 - Page 7 of 11 | Test Interferent | Highest concentration tested that showed no interference | | --- | --- | | Phenylbutazone | 321 μg/mL | | Rifampicin | 2.4 mg/dL | | Salicylate | 200 μg/mL | The sponsor noted that published literature has shown that in very rare cases, gammopathy, in particular type IgM (Waldenström’s macroglobulinemia), may cause unreliable results. The package insert contains the following limitation: “Blood samples from some patients with monoclonal gammopathies may produce falsely elevated results.” Urine Creatinine: | Test Interferent | Highest concentration tested that showed no interference | | --- | --- | | 6N HCl | 0.01% HCL | | pH 4 | 4.0 pH | | pH 9 | 9.0 pH | | Acetaminophen | 200 mg/dL | | Acetic Acid | 25 mL/24 hr collection | | Albumin | 0.5 g/dL (5 g/L) | | Ascorbate | 3.0 mg/dL (199.9 μmol/L) | | Boric Acid | 1% w/v | | Conjugated Bilirubin | 50 mg./dL (855 μmol/L) | | Ethanol | 1 g/dL (216.9 mmol/L) | | Gamma Globulin | 0.5 g/dL (5 g/L) | | Glucose | 2000 mg/dL (111.1 mmol/L) | | Hemoglobin | 100 mg/dL (0.1 g/L) | | Ibuprofen | 500 mg/dL (24.3 mmol/L) | | N-Acetyl cysteine | 2 mg/dL (122/6 | | Nitric Acid | 0.6% | | Oxalic Acid | 0.1 g/dL (11.1 mmol/L) | | Sodium carbonate | 5 g/24 hr collection | 4. Assay Reportable Range: Serum/Plasma: 0.15 – 30.00 mg/dL Urine: 2.00 – 245 mg/dL 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): The assay is traceable to the National Institute of Standards and Technology (NIST) Standard Reference Material SRM 967. {7} Additionally, 47 serum samples with creatinine concentrations across the assay range were tested on Isotope Dilution Mass Spectrometry (IDMS) to demonstrate alignment of the assay with this reference procedure. | Specimen Type | Comparison Assay | N | R | Regression Equation | Sample Range (mg/dL) | | --- | --- | --- | --- | --- | --- | | Serum | IDMS | 47 | 0.991 | y=1.01x + 0.01 | 0.35-26.70 | # 6. Detection Limit: The limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) were determined in accordance with CLSI EP17-A2 guideline. For the LoB determination of serum and urine, 4 blank samples were tested with $n = 5$ replicates per sample using 3 Atellica® CH ECre3 reagent lots for 3 days on 1 Atellica CH Analyzer for a total of 60 measurements per reagent lot. The LoB was calculated non-parametrically as the 95th percentile ranked position of all blank samples. The LoD was determined parametrically using the pooled standard deviation of the highest observed LoB and results from 4 low analyte serum and urine samples prepared by diluting native serum and urine samples with Atellica® CH Diluent and tested in $n = 5$ replicates per sample using 3 Atellica® CH ECre3 reagent lots for 3 days on 1 Atellica® CH Analyzer for a total of 60 measurements per reagent lot. A LoQ study was performed using 4 low concentration serum pools and 2 reagent lots. Each sample was assayed in $n = 5$ replicates per sample on 3 Atellica® CH ECre3 reagent lots for 3 days, on 1 Atellica® CH Analyzer for a total of 60 measurements per lot for serum and urine, respectively. The LoQ was defined as the lowest concentration of creatinine at which the total analytical error is $\leq 0.10 \, \mathrm{mg/dL}$ for serum and plasma and $\leq 1.50 \, \mathrm{mg/dL}$ for urine. The results are summarized in the table below: | Matrix | LoB (mg/dL) | LoD (mg/dL) | LoQ (mg/dL) | | --- | --- | --- | --- | | Serum | 0.02 | 0.04 | 0.13 | | Urine | 0.07 | 0.19 | 1.93 | The studies support the lower limits of assay range claims for serum (0.15 mg/dL) and urine (2.00 mg/dL) and the following detection limit claims in the labeling: | Matrix | Detection Capability | Result (mg/dL) | | --- | --- | --- | | Serum/Plasma | LoB | 0.05 | | | LoD | 0.10 | | | LoQ | 0.15 | | Urine | LoB | 0.15 | | | LoD | 0.50 | | | LoQ | 2.00 | K212223 - Page 8 of 11 {8} 7. Assay Cut-Off: Not applicable. B Comparison Studies: 1. Method Comparison with Predicate Device: Method comparison studies were conducted consistent with CLSI EP09C-ED3 guideline. Native patient samples (serum N= 105, urine N=102) were used in these studies. Each sample was assayed in duplicate using the candidate method Atellica® CH ECre3 reagent tested on one Atellica® CH Analyzer and the predicate method ADVIA Chemistry ECRE_2 tested on one ADVIA 1800 Chemistry system. Only the first replicate was used for analysis. Deming regression results are as follows: | Matrix | N | Claimed measuring range (mg/dL) | Ranges of samples as determined by Predicate (mg/dL) | Regression Equation (mg/dL) | R | | --- | --- | --- | --- | --- | --- | | Serum | 105 | 0.15-30 | 0.18-28.41 | y=0.99x + 0.02 | 1.000 | | Urine | 102 | 2.00-245 | 4.71-240.47 | y=0.98x + 0.05 | 0.999 | 2. Matrix Comparison: Fifty-five (55) matched samples (serum, lithium heparin plasma, and K2-EDTA plasma) were evaluated in duplicate with 1 lot of the Atellica® CH ECre3 reagent. Five (5) of the samples were spiked to obtain higher creatinine concentrations. Only the first replicate was used in analysis. Deming regression results are as follows: | Matrix Y | Matrix X | N | Claimed measuring range (mg/dL) | Ranges of samples (mg/dL) | Regression Equation (mg/dL) | R | | --- | --- | --- | --- | --- | --- | --- | | Lithium Heparin Plasma | Serum | 55 | 0.15-30 | 0.50-26.91 | y=0.99x + 0.00 | 1.000 | | K2-EDTA Plasma | Serum | 55 | 2.00-245 | 0.50-26.91 | y=0.97x + 0.02 | 0.999 | The results of the study support the sponsor's claim that lithium heparin and K2-EDTA plasma are acceptable sample types for the candidate device. K212223 - Page 9 of 11 {9} K212223 - Page 10 of 11 C Clinical Studies: 1. Clinical Sensitivity: Not applicable 2. Clinical Specificity: Not applicable 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable D Clinical Cut-Off: Not applicable E Expected Values/Reference Range: Expected Values Transference as defined in CLSI EP28-A3c was used to verify established reference ranges for serum and plasma creatinine. Healthy male and female serum, lithium heparin plasma, and dipotassium EDTA plasma samples were obtained (23 men and 21 women) and processed with N=2 replicates, only the first replicate was used in analysis. Results of the verification study support the serum reference ranges which were established through literature. | Group | Sample Type | Reference Interval | | --- | --- | --- | | Males | Serum/Plasma^{a} | 0.73-1.18 mg/dL | | Females | Serum/Plasma^{a} | 0.55-1.02 mg/dL | | Males | Urine^{b} | 800-2000 mg/day | | Females | Urine^{b} | 600-1800 mg/day | a Junge W, Wilke B, Halabi A, Klein G. Determination of reference intervals for serum creatinine, creatinine excretion and creatinine clearance with an enzymatic and a modified Jaffé method. Clin Chim Acta. 2004;344(1-2):137-48. b Wu, AHB. Tietz Clinical Guide to Laboratory Tests. 4th ed. Philadelphia, PA: WB Saunders Company; 2006:316.. VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. {10} IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K212223 - Page 11 of 11
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...