← Product Code [PCV](/productcode/PCV) · K212201

# PENTAX Medical Auto Leakage Tester SHA-P6 (K212201)

_Pentax of America, Inc. · PCV · Oct 1, 2021 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K212201

## Device Facts

- **Applicant:** Pentax of America, Inc.
- **Product Code:** [PCV](/productcode/PCV.md)
- **Decision Date:** Oct 1, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The PENTAX Medical Auto Leakage Tester SHA-P6 is intended to perform leakage testing on PENTAX Medical flexible endoscopes.

## Device Story

Device performs leakage testing on PENTAX Medical flexible endoscopes; detects leaks by measuring air pressure changes over time. Operates via Automated Leak Test (ALT) mode: feeds air into endoscope to reach set pressure; monitors pressure stability (dry test); followed by immersion in water to confirm watertight integrity (wet test). Used in clinical settings by healthcare staff during endoscope reprocessing. Output consists of automated leak detection status; informs clinical decision-making regarding endoscope safety for patient use. Benefits include prevention of fluid ingress into complex mechanical/optical/video components of endoscopes, ensuring effective cleaning/disinfection.

## Clinical Evidence

Bench testing only. Verification testing confirmed leak detection capabilities are equivalent to the predicate device (K123704) and reference device (K093718). Device successfully tested for all functions, performance, and safety per FDA guidance and consensus standards (IEC 60601-1, IEC 60601-1-2).

## Technological Characteristics

Portable unit with built-in 2,600 mAh lithium battery. Dimensions: 154mm x 63.2mm x 95.6mm; Weight: 380g. Operates via air pressure monitoring. Electrical safety/EMC compliance: IEC 60601-1:2005+A1:2012 and IEC 60601-1-2:2014. Connectivity: air feeding tube interface only; no external data interfaces.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Automated Endoscope Leak Tester ALT-Y0003 ([K123704](/device/K123704.md))

## Reference Devices

- ZUTRON Medical Endoscope Leak Tester, Model ZUTR-10003 ([K093718](/device/K093718.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 1, 2021

PENTAX of America, Inc. William Goeller Vice President, Ouality Assurance and Regulatory Affairs 3 Paragon Drive Montvale, NJ 07645-1782

Re: K212201

Trade/Device Name: PENTAX Medical Auto Leakage Tester SHA-P6 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: PCV Dated: July 12, 2021 Received: July 14, 2021

#### Dear William Goeller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K212201

Device Name

PENTAX Medical Auto Leakage Tester SHA-P6

Indications for Use (Describe)

The PENTAX Medical Auto Leakage Tester SHA-P6 is intended to perform leakage testing on PENTAX Medical flexible endoscopes.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary

#### a. Applicant

PENTAX of America, Inc. 3 Paragon Drive Montvale, New Jersey 07645-1782

#### b. Contact/Application Correspondent

William Goeller Vice President, Quality and Regulatory Affairs PENTAX of America, Inc. 3 Paragon Drive Montvale, New Jersey 07645-1782

Phone: 201-571-2300 ext. 2318 Fax: 201-571-2340 Email: william.goeller@pentaxmedical.com

- Date Prepared: November 20, 2020 ﻦ
- d. Common Name: Leak Tester
- e. Name of the System: PENTAX Medical Auto Leakage Tester SHA-P6
- Regulation Number: 21 CFR Part 876.1500 f.
- Regulation Name: Endoscope and accessories ള.
- h. Regulatory Class: Class II
- i. Product Code: PCV
- j. Predicate Device: Automated Endoscope Leak Tester ALT-Y0003 (K123704)
- k. Reference Device: ZUTRON Medical Endoscope Leak Tester, Model ZUTR-10003 (K093718)
- Device Description: l.

PENTAX Medical Auto Leakage Tester SHA-P6 (Hereinafter "the Device") is intended to be used to perform leakage testing on PENTAX Medical flexible endoscopes are complex devices that consist of various mechanical, optical, and video elements that might be affected by exposure to fluid. Therefore, endoscope leak testing is a crucial procedure for the proper endoscope cleaning and disinfecting process. The Device is designed to detect leaks by measuring the difference in air pressure over time using the Automated Leak Test mode (ALT

{4}------------------------------------------------

mode). The Device conducts two types of tests: a dry leakage test and a wet leakage test. The dry leakage test is performed when the device feeds air into the endoscope until a set pressure is reached, and then monitors the pressure over a specific period of time. The wet leakage test is conducted after confirming the absence of a leak during the dry leakage test, by immersing the endoscope in potable water to test for loss of integrity in its watertight construction. The Device is compatible with PENTAX Medical flexible endoscopes.

## m. Statement of Indications for Use

The PENTAX Medical Auto Leakage Tester SHA-P6 is intended to be used to perform leakage testing on PENTAX Medical flexible endoscopes.

# n. Statement of Substantial Equivalence

The Device is substantially equivalent to the predicate Automated Endoscope Leak Tester ALT-Y0003 (K123704) (Hereinafter "the Predicate Device"). Both devices are used for leak testing of flexible endoscopes in combination with a connecting tube feeding air into the endoscope.

The two devices are substantially equivalent and have:

- similar Intended Use
- use the same Operating Principle ●
- incorporate the same Basic System Design

The main differences are:

- The Device only has an Automated Leak Test mode (ALT) for determining the ● watertight integrity of the endoscope.
- The Predicate Device has and Automated Leak Test mode (ALT) and A Manual Leak Test mode (MLT).
- The predicate device has additional features such as RFID tracking and internal . memory to store test results. The Device does not have these additional features.

{5}------------------------------------------------

# K212201

Page 2 of 6

# o. Comparison Table

| Item                                         | Predicate Device             | Reference Device                                                                                                     | The Device                                                                                                                   | Remark                                                                                                         |              |
|----------------------------------------------|------------------------------|----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|--------------|
| Product Name                                 |                              | Automated Endoscope Leak<br>Tester                                                                                   | ZUTRON Medical Endoscope<br>Leak Tester                                                                                      | PENTAX Medical Auto<br>Leakage Tester                                                                          | -            |
| Model Number                                 |                              | ALT-Y0003                                                                                                            | ZUTR-10003                                                                                                                   | SHA-P6                                                                                                         | -            |
| 510(k) Number                                |                              | K123704                                                                                                              | K093718                                                                                                                      | N/A                                                                                                            | -            |
| Intended Use                                 |                              | This equipment is intended to<br>be used to perform and record<br>leakage testing on Olympus<br>flexible endoscopes. | The Zutron Medical ZUTR-10003<br>Endoscope Leak Tester is<br>designed to detect interior and<br>exterior leaks in endoscopes | This product is intended to be<br>used to perform leakage<br>testing on PENTAX Medical<br>flexible endoscopes. | Similar *1   |
| AC adapter                                   | DC output<br>voltage/current | 45 VA                                                                                                                | 12V, 3.3A                                                                                                                    | 7.5 V, 1.6 A                                                                                                   | Similar      |
|                                              | AC input<br>frequency        | 50/60 Hz                                                                                                             | 50/60 Hz                                                                                                                     | 50/60 Hz                                                                                                       | Identical    |
|                                              | AC input<br>voltage          | 100 - 120 V                                                                                                          | 100 - 240 V                                                                                                                  | 100 - 240 V                                                                                                    | Similar      |
| Built - in<br>packaged<br>lithium<br>battery | Battery<br>capacity          | N/A                                                                                                                  | N/A                                                                                                                          | 2,600 mAh                                                                                                      | Different *2 |
| Size                                         |                              | 280 mm (W) x 201 mm (H) x<br>184 mm (D)<br>11.0 in (W) x 7.9 in (H) x 7.2<br>in (D)                                  | 127 mm (W) x 88.9 mm (H) x<br>196.85 mm (D)<br>5.00 in (W) x 3.5 in (H) x 7.75 in<br>(D                                      | 154 mm (W) x 63.2 mm (H) x<br>95.6 mm (D)<br>6.06 in (W) x2.49 in (H) x<br>3.76 in (D)                         | Similar      |
| Weight                                       |                              | 5.5 kg (12.1 lbs.)                                                                                                   | 1.4 kg (3.10 lbs.) (main unit only)                                                                                          | 380 g (0.84 lbs.) (main unit<br>only)                                                                          | Different *3 |
|                                              |                              |                                                                                                                      |                                                                                                                              | Page 4 of 6                                                                                                    |              |
| Item                                         | Predicate Device             | Reference Device                                                                                                     | The Device                                                                                                                   | Remark                                                                                                         |              |
| Automated Leakage Testing<br>(ALT)           | Yes                          | Yes                                                                                                                  | Yes                                                                                                                          | Identical                                                                                                      |              |
| Manual Leakage Testing<br>(MLT)              | Yes                          | No                                                                                                                   | No                                                                                                                           | Different *4                                                                                                   |              |

{6}------------------------------------------------

# K212201

{7}------------------------------------------------

Page 5 of 6

The following are explanations regarding the differences between the subject and predicate devices:

*1 Intended use

Intended use of the Device is similar to the Predicate Device regarding leakage testing for the flexible endoscopes.

*2 Built - in packaged lithium batterv

The availability of the built - in packaged lithium battery is a minor addition and does not raise and safety and effectiveness concerns.

*3 Weight

This is due to the difference of mechanical constructions, which does not raise any safety and effectiveness concerns.

*4 Manual Leakage Testing (MLT)

The Device only has an Automated Leakage Testing (ALT) mode. The availability of the Manual Leakage Testing (MLT) in the Predicate Device is a minor functional difference and does not raise and safety and effectiveness concerns.

Comparison to the Reference device:

ZUTRON Medical Endoscope Leak Tester, ZUTR-10003 (K093718) was selected as a Reference device due to the fact that it was used as one of the predicate devices for the premarket notification application of the Automated Endoscope Leak Tester ALT-Y0003 (K123704), which is the Predicate Device under this premarket notification of the Device.

The Reference device, ZUTRON Medical Endoscope Leak Tester, ZUTR-10003 (K093718) has been used for leakage detection testing, where both subject and predicate devices were compared in their ability to find leakage within compatible endoscopes. Subject device has been found to be substantially equivalent to the Reference device in its ability to detect leaks.

# p. Software

Software verification and validation tests were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. See Section 19 Software for details.

Cybersecurity risks have been assessed according to the FDA Guidance for Industry and Staff "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" issued October 2, 2014. No cyber security risks have been detected since the Device has no other interface than an air feeding tube to an endoscope tested.

{8}------------------------------------------------

### q. EMC and Electrical Safety

The acceptable level of electromagnetic compatibility (EMC) and electrical safety (ES) for the Device were confirmed by the following standards:

IEC 60601-1-2:2014; IEC 60601-1:2005+CORR 1:2006+CORR 2:2007+A1:2012

### r. Performance Testing

The verification test was conducted to confirm that the leak detection capabilities of the Device were similar to the Predicate Device. The Endoscope Leak Tester ZUTR-10003 (K093718) was used the comparison device, which was the device used for leak detection for the Predicate Device. The Device was successfully tested for all functions, performance, and safety as per FDA Guidance and recognized consensus standards.

### s. Conclusion

After analyzing the intended use, technological characteristics (including operating principle, performance characteristics and constituent materials), and labeling, the Device did not raise any issue regarding safety and effectiveness. Accordingly, the testing concluded that the Device is substantially equivalent to the Predicate Device and the Reference Device.

---

**Source:** [https://fda.innolitics.com/device/K212201](https://fda.innolitics.com/device/K212201)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
