The Relivion® transcutaneous electrical nerve stimulator is indicated for the acute treatment of migraine with or without aura in patients 18 years of age or older. It is a prescription device to be self-used at home.
Device Story
Relivion is a non-invasive, wearable neurostimulator headset for migraine treatment. Device features three pairs of electrodes contacting forehead and occiput to stimulate trigeminal (supraorbital/supratrochlear) and greater occipital nerve branches. Powered by rechargeable 3.7V Li-Po battery; delivers electrical pulses to induce neuromodulation. User operates device via onboard interface to activate/deactivate and adjust intensity; treatment runs for 60 minutes. Optional mobile app connects via Bluetooth for status display, alerts, and patient migraine status reporting. New modification adds Physician Interface for remote follow-up on patient migraine attacks and treatment history. Healthcare providers use data from Physician Interface to monitor patient progress. Device benefits patients by providing non-pharmacological acute migraine relief.
Clinical Evidence
No clinical data presented. Substantial equivalence supported by bench testing and software verification/validation of the modified Physician Interface and mobile application. Testing confirmed output parameters meet specifications and cybersecurity risks were addressed.
Technological Characteristics
Transcutaneous electrical nerve stimulator; plastic PA and silicone housing; 90g weight; 209mm x 128mm x 39mm. Rechargeable 3.7V Li-Po battery. Bluetooth connectivity. Single-use replaceable electrode pads with water-releasing covers. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, IEC 60601-2-10, IEC 62366. Software-controlled with one fixed program.
Indications for Use
Indicated for acute treatment of migraine with or without aura in patients 18 years of age or older. Prescription device for home self-use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator to treat headache is a device used to apply an electrical current to a patient's cranium through electrodes placed on the skin.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Appropriate analysis/testing must validate electromagnetic compatibility and electrical, mechanical, and thermal safety.
(3) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm
2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system must be fully characterized.(4) Electrical performance, adhesive integrity, shelf life, reusability, and current distribution testing of the electrodes must be conducted.
(5) Appropriate software verification, validation, and hazard analysis must be performed.
(6) Clinical performance data must demonstrate that the device is safe and effective as a treatment for headache in the indicated patient population.
(7) Labeling must include the following:
(i) Appropriate contraindications such as not for use in subjects with an implanted metallic or electronic device in the head, a cardiac pacemaker, or an implanted or wearable defibrillator.
(ii) Appropriate warnings such as not to apply the device on the neck or chest, not to use the device in the presence of electronic monitoring equipment, not to use in the bath or shower, not to use while sleeping, not to use while driving, not to use while operating machinery.
(iii) Appropriate precautions such as the long-term effects of chronic use of the device are unknown.
(iv) A summary of the expected risks and benefits of using the device.
(v) A summary of the clinical performance data, including information on the patient population for which the device has and has not been demonstrated to be effective, and any adverse events and complications.
(vi) Information on how the device operates and the typical sensations experienced during treatment.
(vii) A detailed summary of the device technical parameters.
(viii) An expiration date/shelf life for the electrodes and the number of times they can be reused.
(ix) Disposal instructions.
In combination with the general controls of the Food Drug & Cosmetic Act, the Transcutaneous Electrical Nerve Stimulator to Treat Headache is subject to the following special controls:
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 2, 2021
Neurolief Ltd. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street, Suite 2320 Philadelphia, Pennsylvania 19103
Re: K212106
Trade/Device Name: Relivion Regulation Number: 21 CFR 882.5891 Regulation Name: Transcutaneous Electrical Nerve Stimulator To Treat Headache Regulatory Class: Class II Product Code: PCC Dated: July 6, 2021 Received: July 6, 2021
Dear Janice Hogan:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Patrick Antkowiak Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
Device Name Relivion®
Indications for Use (Describe)
The Relivion® transcutaneous electrical nerve stimulator is indicated for the acute treatment of migraine with or without aura in patients 18 years of age or older. It is a prescription device to be self-used at home.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D)
- Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unlessit displays a currently valid OMB number. "
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
### SPECIAL 510(k) SUMMARY
Image /page/3/Picture/1 description: The image shows the logo for Neuroleaf Brain Innovation. The logo consists of a stylized brain shape on the left, colored in a light blue. To the right of the brain shape, the word "Neuroleaf" is written in a larger, light blue font. Below "Neuroleaf", the words "Brain Innovation" are written in a smaller, light blue font.
| Applicant Name: | Neurolief Ltd. |
|-----------------|-------------------------|
| | 12 Giborei Israel St. |
| | Netanya, Israel 4250412 |
| | Tel: +972-9-3730288 |
- Contact Person: Michal Kedar-Datel Clinical & Regulatory Affairs Director Neurolief Ltd. Tel: +972-9-3730288
- Date Prepared: July 6, 2021
- Relivion® Trade Name:
Classification Name: 21 CFR 882.5891 Transcutaneous electrical nerve stimulator to treat headache
- Product Code: PCC
- Classification: Class II
- Classification Panel: Neurology
- Neurolief Ltd's Relivion® Device (K203419) Predicate Device:
### Intended Use/Indication for Use:
The Relivion® transcutaneous electrical nerve stimulator is indicated for the acute treatment of migraine with or without aura in patients 18 years of age or older. It is a prescription device to be self-used at home.
{4}------------------------------------------------
## Device Description:
The modified Relivion® is an external non-invasive neurostimulator designed for transcutaneous electrical nerve stimulation. The modified Relivion® headset integrates three pairs of output electrodes which come in contact with the subject's scalp at the forehead and occiput. The electrodes deliver the stimulation pulses produced by the headset's stimulation unit to the subject's scalp. The frontal electrodes stimulate the trigeminal (supraorbital and supratrochlear) nerve branches and the posterior electrodes stimulate the greater occipital nerve branches.
The modified Relivion® includes single-use replaceable electrode pads that are positioned on-top of the electrodes prior to treatment and are wetted by the user before use, in order to provide proper conductivity between the electrodes and the scalp.
The modified Relivion® is powered by a rechargeable battery and the headset incorporates an onboard user interface that enables the user to activate/deactivate the device and to adjust the stimulation intensity. Upon treatment activation, the treatment automatically runs and ends after 60 minutes or alternatively, the user can stop the treatment when desired.
The modified Relivion® can communicate via a low energy Bluetooth link with the modified Relivion®'s dedicated mobile application on the patient's smartphone. The modified Relivion®'s mobile application is optional, and it is used by the patient to display the device status and provide indications and alerts. Additionally, the modified Relivion® mobile application also enables the patient to report their migraine headache status.
The main modification introduced in the Relivion® Device compared to the cleared predicate device (K203419) is an added optional Physician Interface that enables physicians to remotely follow-up on the patient's migraine attacks status and the treatments they performed using the Relivion device, based on the data transferred to it via the modified mobile application.
Both the modified Relivion®'s mobile application (for patients) and added physician interface (for physicians) are referred together as the optional "Patient Management Interface" (PMI) of the modified Relivion® Device.
## Technological Characteristics:
The modified Relivion® treats migraines by stimulating the trigeminal and occipital nerve branches by a transcutaneous electrical nerve stimulation. Trigeminal and occipital electrical nerve stimulation induces neuromodulation of these nerve pathways and by that reduces pain associated with migraine and associated symptoms.
The modified Relivion® includes single-use, replaceable electrode pads that are positioned on-top of the headset electrodes and are wetted by the user before each use. Water releasing covers are located on the outer side of each back occipital electrode and are used to release moisture from the
{5}------------------------------------------------
electrode pads to the scalp in order to provide proper electrical conductivity between the electrodes and the scalp.
The modified Relivion® headset adjusts to various head sizes and contours and can be worn comfortably. The headset includes two flexible arms that penetrate under the hair layers while the headset is worn.
The modified Relivion® headset incorporates an on-board user interface and can communicate via a low energy Bluetooth link with the modified Relivion®'s dedicated mobile application on the patient's smartphone. The user interface and mobile app display the device status and provides indications and alerts, in addition to an optional Physician Interface that enables physicians to remotely follow-up on the patient's migraine attacks status and treatments performed with the Relivion device.
## Performance Data:
Neurolief conducted several performance tests to demonstrate that the modified Relivion® device complies with performance standards and that it functions as intended.
Performance - Bench Testing: The modified Relivion® device underwent performance testing for the modifications introduced, including software validation and device verification tests. It was successfully verified that the modified Relivion® output parameters meet the product's specifications.
Software Testing: The PMI software was also subject to verification and validation testing, and results demonstrated that the modified Mobile Application and added Physician Interface perform as intended. Cybersecurity risks were also identified and addressed.
Because the modification to the device impacted only the software, no further testing was required.
## Substantial Equivalence:
The following table compares the modified Relivion® device to its predicate device with respect to intended use, technological characteristics and principles of operation, providing detailed information regarding the basis for the determination of substantial equivalence.
{6}------------------------------------------------
| Parameter | Neurolief Ltd.'s<br>Relivion®<br>(Subject Device) | Neurolief Ltd.'s<br>Relivion®<br>(Predicate) | Comparison |
|---------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | General Characteristics | | |
| 510(k) number | Pending | K203419 | N/A |
| Classification | 21 CFR § 882.5891 | 21 CFR § 882.5891 | Same |
| Product Code | PCC | PCC | Same |
| Product Class | Class II | Class II | Same |
| Regulation Name | Transcutaneous Electrical Nerve<br>Stimulator to Treat Headache | Transcutaneous Electrical Nerve<br>Stimulator to Treat Headache | Same |
| Indications for<br>Use | The Relivion® transcutaneous electrical<br>nerve stimulator is indicated for the acute<br>treatment of migraine with or without aura<br>in patients 18 years of age or older. It is a<br>prescription device to be self-usedat<br>home. | The Relivion® transcutaneous electrical<br>nerve stimulator is indicated for the acute<br>treatment of migraine with or without aura<br>in patients 18 years of age or older. It is a<br>prescription device to be self-used at<br>home. | Same |
| Technology | Transcutaneous Electrical<br>Nerve Stimulator | Transcutaneous Electrical<br>Nerve Stimulator | Same |
| Invasiveness | Non-Surgical | Non-Surgical | Same |
| Electrode<br>Locations | Forehead and Occiput | Forehead and Occiput | Same |
| Nerves over<br>which electrodes<br>are placed | Supratrochlear and supraorbital branches<br>of the trigeminal nerve bilaterally and<br>the occipital nerves | Supratrochlear and supraorbital branches<br>of the trigeminal nerve bilaterally and<br>the occipital nerves | Same |
| Energy | Electric | Electric | Same |
| Power Source | Rechargeable Li-Po 3.7 V Battery | Rechargeable Li-Po 3.7 V Battery | Same |
| Software-<br>controlled | Yes, 1 fixed program | Yes, 1 fixed program | Same |
| Constant Current | Yes | Yes | Same |
| Constant Voltage | No | No | Same |
| Software<br>Function | Controls the output of the device<br>and device indicators | Controls the output of the device<br>and device indicators | Same |
| Timer Settings | Yes | Yes | Same |
| Parameter | Neurolief Ltd.'s<br>Relivion®<br>(Subject Device) | Neurolief Ltd.'s<br>Relivion®<br>(Predicate) | Comparison |
| Patient override<br>control method | On/Off button | On/Off button | Same |
| Button Types | On/Off Button and buttons to adjust<br>intensity of electrical stimulus | On/Off Button and buttons to adjust<br>intensity of electrical stimulus | Same |
| Functional<br>features | Visual and auditory indicators inform the<br>user when the device is on vs. off and help<br>them troubleshoot if it is not working<br>properly (e.g., indicates if device is<br>active/non-active, low battery indication<br>and if electrical connection between device<br>and skin is unacceptable) | Visual and auditory indicators inform the<br>user when the device is on vs. off and help<br>them troubleshoot if it is not working<br>properly (e.g., indicates if device is<br>active/non-active , low battery indication<br>and if electrical connection between device<br>and skin is unacceptable) | Same |
| Bluetooth<br>Capable | Yes | Yes | Same |
| Associated<br>Mobile<br>Application<br>(for patients) | Yes | Yes | Similar<br>Both devices include an optional<br>mobile app to display device status.<br>treatment duration, and battery<br>status, with the modified device<br>application also enabling patients to<br>report their migraine headache<br>status, this feature does not alter the<br>therapeutic effect. |
| Associated<br>Physician<br>Interface<br>(for physicians) | Yes | No | Different<br>The modified device has an optional<br>a ssocia ted Physician Interface that<br>enables physicians to remotely<br>follow-up on the patient's migraine<br>attacks status and treatments<br>performed with the Relivion device.<br>this feature does not alter the<br>therapeutic effect. |
| Max leakage<br>current | None (battery operated) | None (battery operated) | Same |
| Parameter | Neurolief Ltd.'s<br>Relivion®<br>(Subject Device) | Neurolief Ltd.'s<br>Relivion®<br>(Predicate) | Comparison |
| Electrodes | Relivion® electrode | Relivion® electrode | Same |
| Indicator display:<br>Unit functioning | Yes | Yes | Same |
| Low battery<br>indicator | Yes | Yes | Same |
| Standards:<br>IEC 60601-1 | Yes | Yes | Same |
| IEC 60601-1-2 | Yes | Yes | Same |
| IEC 60601-1-6 | Yes | Yes | Same |
| IEC 60601-1-11 | Yes | Yes | Same |
| IEC 60601-2-10 | Yes | Yes | Same |
| IEC 62366 | Yes | Yes | Same |
| Weight | 90 gr | 90 gr | Same |
| Dimensions | 209mm x 128mm x 39mm | 209mm x 128mm x 39mm | Same |
| Housing<br>materials | Plastic PA + Silicone | Plastic PA + Silicone | Same |
| Maximum Time<br>Device Used | 60 minutes | 60 minutes | Same |
| Net Charge (µC)<br>per pulse | 0 | 0 | Same |
| Pulse Duration<br>(µsec) | 850 | 850 | Same |
| Frequency (Hz) | 80 | 80 | Same |
| Maximum output<br>voltage (V): @500<br>ohms @2000<br>ohms @10000<br>ohms | 3 front electrodes / 6 back electrodes<br>12 front electrodes / 24 back electrodes<br>60 front electrodes / 100 back electrodes | 3 front electrodes / 6 back electrodes<br>12 front electrodes / 24 back electrodes<br>60 front electrodes / 100 back electrodes | Same |
| Maximum output<br>current (mA):<br>@500 ohms<br>@2000 ohms<br>@10000 ohms | 6 front electrodes /12 back electrodes<br>6 front electrodes /12 back electrodes<br>6 front electrodes /10 back electrodes | 6 front electrodes /12 back electrodes<br>6 front electrodes /12 back electrodes<br>6 front electrodes /10 back electrodes | Same |
| Parameter | Neurolief Ltd.'s<br>Relivion®<br>(Subject Device) | Neurolief Ltd.'s<br>Relivion®<br>(Predicate) | Comparison |
| Max Phase<br>Amplitude | 6 mA front electrodes/ 12 mA back<br>electrodes; with a load of a 4.7 uF<br>capacitor parallel with 2.2K ohms<br>resistance | 6 mA front electrodes/ 12 mA back<br>electrodes; with a load of a 4.7 uF capacitor<br>parallel with 2.2K ohms resistance | Same |
| Maximum phase<br>charge (µC)<br>@500Ω | 4.8 | 4.8 | Same |
| Maximum<br>Current Density,<br>(mA/cm², r.m.s.)<br>@500Ω | 1.93 front electrodes/<br>2.78 back electrodes | 1.93 front electrodes/<br>2.78 back electrodes | Same |
| Maximum<br>Average Power<br>Density,<br>(W/cm²) @500Ω | 0.0000116 front electrodes /<br>0.000034 back electrodes | 0.0000116 front electrodes /<br>0.000034 back electrodes | Same |
| Maximum<br>Average Current<br>(average absolute<br>value, mA)<br>@500Ω | 0.38 front electrodes /<br>0.76 back electrodes | 0.38 front electrodes /<br>0.76 back electrodes | Same |
| Phase rise time | 5 µS | 5 µS | Same |
| Modulation<br>Options<br>Amplitude<br>Frequency<br>Duration | 0- 12 mA<br>Fixed @ 80 Hz<br>330- 400 µS | 0- 12 mA<br>Fixed @ 80 Hz<br>330- 400 µS | Same |
| Phase decay time | 2 µS | 2 µS | Same |
| Ramp<br>Modulations<br>Ramp Up<br>Ramp Down | Manually<br>Manually | Manually<br>Manually | Same |
## Table 1: Neurolief, Ltd.'s Modified Relivion® Substantial Equivalence
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
As described in the comparison table above, the subject modified Relivion® and the predicate Relivion® device (K203419) share the same intended use and indications, technological characteristics, and same principles of operation. The minor differences in the technological characteristics do not alter the overall therapeutic effect of the device. Any differences between the modified Relivion® and its predicate (K203419) were carefully evaluated through performance testing. The modified Relivion® device underwent performance testing, including bench testing and software validation testing. These performance tests confirmed that the modified Relivion® complies with the same special controls and the same consensus and performance standards, on which FDA based its clearance of the company's Relivion® device (K203419). These tests demonstrated that the differences in the technological characteristics between the subject and predicate device do not adversely impact performance and that the subject modified Relivion® is substantially equivalent to its predicate device (K203419).
### Conclusions:
Therefore, based on the same intended use, indications, technological characteristics, and same principles of operation, the modified Relivion® device is substantially equivalent to its predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.