da Vinci SP Firefly Imaging System

K212101 · Intuitive Surgical, Inc. · NAY · Nov 23, 2021 · Gastroenterology, Urology

Device Facts

Record IDK212101
Device Nameda Vinci SP Firefly Imaging System
ApplicantIntuitive Surgical, Inc.
Product CodeNAY · Gastroenterology, Urology
Decision DateNov 23, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The da Vinci SP Firefly Imaging System is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. The da Vinci SP Firefly Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, using near-infrared imaging.

Device Story

The da Vinci SP Firefly Imaging System is an adjunct imaging system for the da Vinci SP Surgical System. It utilizes an endoscope controller to provide visible light or NIR excitation laser light. The endoscope illuminates the surgical site; NIR light fluoresces intravenously-injected indocyanine green (ICG). The system captures these signals and displays them at the Surgeon Console. Used in clinical settings by surgeons to perform minimally invasive surgery, the device provides real-time visual assessment of vessels, blood flow, and tissue perfusion. This visual feedback assists surgeons in intraoperative decision-making, potentially improving surgical outcomes by highlighting tissue perfusion and vascular structures.

Clinical Evidence

Bench testing and animal (porcine) model validation were performed. Bench testing included endoscope design verification, illumination reliability, and human factors evaluation. Animal testing evaluated safety and effectiveness for surgical use. No clinical data on human subjects provided.

Technological Characteristics

Fully-integrated adjunct imaging system for da Vinci SP Surgical System. Components: endoscope controller (light source/NIR excitation laser), endoscope, and software. Connectivity: integrated with da Vinci SP Surgical System. Sterilization: not specified. Software: integrated system software functions.

Indications for Use

Indicated for surgeons performing minimally invasive surgery requiring real-time endoscopic visible and near-infrared fluorescence imaging for visual assessment of vessels, blood flow, and tissue perfusion.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The FDA logo is in blue. Intuitive Surgical, Inc. Connor Mccarty Sr. Regulatory Engineer 1266 Kifer Road Sunnyvale, California 94086 ### Re: K212101 Trade/Device Name: da Vinci SP Firefly Imaging System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: NAY, IZI Dated: October 22, 2021 Received: October 25, 2021 #### Dear Connor Mccarty: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mark W. Trumbore Ph.D., GWCPM Acting Assistant Director THT4A1: Robotically-Assisted Surgical Device Team DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use Submission Number (if known) K212101 Device Name da Vinci SP Firefly Imaging System #### Indications for Use (Describe) The da Vinci SP Firefly Imaging System is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. The da Vinci SP Firefly Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, using near-infrared imaging. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | Contact Details | 21 CFR 807.92(a)(1) | |-----------------------------|--------------------------------------------------| | Applicant Name | Intuitive Surgical, Inc. | | Applicant Address | 1266 Kifer Road Sunnyvale CA 94086 United States | | Applicant Contact Telephone | 805-798-4205 | | Applicant Contact | Mr. Connor McCarty | | Applicant Contact Email | connor.mccarty@intusurg.com | | Device Name | 21 CFR 807.92(a)(2) | |---------------------|--------------------------------------------------| | Device Trade Name | da Vinci SP Firefly Imaging System | | Common Name | Endoscope and accessories | | Classification Name | System, Surgical, Computer Controlled Instrument | | Regulation Number | 876.1500 | | Product Code | NAY | | Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) | |------------------------------------|---------------------| |------------------------------------|---------------------| | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | |-------------|----------------------------------------------------------|--------------| | K210918 | da Vinci Firefly Imaging System | NAY | ## Device Description Summary The da Vinci SP Firefly Imaging System is a fully-integrated, adjunct imaging system for the da Vinci SP Surgical System. The da Vinci SP Firefly Imaging System consists of enhanced existing components of the da Vinci SP Surgical System: the endoscope controller, the endoscope, and supporting software functions. The endoscope controller provides a light source, either visible light or a near-infrared (NIR) excitation laser. The endoscope transmits visible light from the endoscope controller to illuminate the surgical site. The near-infrared light is used to fluoresce an intravenously-injected imaging agent, indocyanine green (ICG), which is imaged by the endoscope and displayed at the Surgeon Console. ## Intended Use/Indications for Use The da Vinci SP Firefly Imaging System is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. The da Vinci SP Firefly Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, using near-infrared imaging. ## Indications for Use Comparison The indications for use of the subject device are a subset of the predicate device. The indications for use differ between the subject and predicate device based on the parent device indications for use. The parent da Vinci SP Surgical System indications for use do not include surgical procedures utilizing bile duct imaging, and so this indication for the subject device is not included. The intended use is identical between the predicate and subject devices are intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. ## 21 CFR 807.92(a)(5) 21 CFR 807.92(a)(5) 21 CFR 807.92(a)(4) {4}------------------------------------------------ ## Technological Comparison The technological characteristics of the endoscope controller component of the predicate and subject devices are identical. The parent system software and endoscope components differ between the predicate and subject devices. #### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) The da Vinci SP Firefly Imaging System underwent a series of bench tests to verify requirements and risk mittigations. Based on the subject device impact on the parent da Vinci SP Surgical System, bench testing was limited to endoscope design verification, illumination reliability testing, and a human factors evaluation. The da Vinci SP Firefly Imaging System underwent validation testing using an animal model (porcine) to evaluate its safety and effectiveness for use in surgery. Bench testing and clinical validation testing on the subject device confirmed that no issues of safety or effectiveness, analogous to the results of the predicate device verification testing. Therefore, the test results demonstrate that the subject device is substantially equivalent to its predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...