← Product Code [GEX](/productcode/GEX) · K212082

# Picore (K212082)

_Bio-Med USA, Inc. · GEX · Oct 6, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K212082

## Device Facts

- **Applicant:** Bio-Med USA, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Oct 6, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

PICORE Q-switched Nd : Y AG Laser System is indicated for the incision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis. 532nm Wavelength Tattoo removal light ink (red, tan, purple, orange, sky blue) Removal of Benign Epidermal Pigmented Lesions, Minor Benign Vascular Lesions. Talangiectasias Treatment of Lentigines, Cafe-Au-Lait, Sebrorrheic Keratoses, Becker's Nevi, Freckles Treatment of Post Inflammatory Hyperpigmentation (PIH) 1064nm Wavelength: Tattoo removal: dark ink (black, blue, green) Removal of Nevus of Ota, Removal or lightening of unwanted hair with or without adjuvant preparation Treatment of Common Nevi, Melasma, Skin resurfacing procedures for the treatment of acne scars. wrinkle

## Device Story

Picore is a picosecond Nd:YAG laser system; emits 1064nm and 532nm wavelengths. Physician-operated in clinical settings; uses four non-fractional handpieces (Zoom, collimated, 3-5mm, 6-8mm). System parameters (energy, spot size) controlled via front-panel touch screen; laser emission triggered by foot switch. Device transforms electrical power into pulsed laser energy for soft tissue interaction (incision, ablation, vaporization, coagulation). Output allows physicians to perform dermatologic and surgical procedures; clinical benefit includes precise tissue treatment and lesion removal. Cooling via closed-circuit water-to-air system.

## Clinical Evidence

No clinical data. Bench testing only, including electrical safety (IEC 60601-1), laser safety (IEC 60825-1), electromagnetic compatibility (IEC 60601-1-2), and software verification/validation.

## Technological Characteristics

Nd:YAG laser; 1064nm/532nm wavelengths; Q-switched pulsed mode. 10.4" touch screen interface. Closed-circuit water-to-air cooling. 3kVA power consumption. Four non-fractional handpieces. Standards: IEC 60601-1, IEC 60825-1, IEC 60601-1-2.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- BM.IRIS ([K171648](/device/K171648.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 6, 2021

Bio-Med USA Inc Young Chi CEO 27 New England Dr Ramsey, New Jersey 07446

Re: K212082

Trade/Device Name: Picore Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 1, 2021 Received: September 8, 2021

Dear Young Chi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212082

Device Name PICORE

Indications for Use (Describe)

PICORE Q-switched Nd : Y AG Laser System is indicated for the incision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.

#### 532nm Wavelength

Tattoo removal light ink (red, tan, purple, orange, sky blue) Removal of Benign Epidermal Pigmented Lesions, Minor Benign Vascular Lesions. Talangiectasias Treatment of Lentigines, Cafe-Au-Lait, Sebrorrheic Keratoses, Becker's Nevi, Freckles Treatment of Post Inflammatory Hyperpigmentation (PIH)

1064nm Wavelength:

Tattoo removal: dark ink (black, blue, green) Removal of Nevus of Ota, Removal or lightening of unwanted hair with or without adjuvant preparation Treatment of Common Nevi, Melasma, Skin resurfacing procedures for the treatment of acne scars. wrinkle

| Type of Use (Select one or both, as applicable)                                                                                                    |  |
|----------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <span style="font-size:16px">❌</span>                     Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> |  |
| <div> <span>                     □ Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div>                                      |  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of 21CFR Section 807.92.

The assigned 510(k) Number: K212082

- Date of Preparation 1. 10/05/2021

| 2. | Applicant<br>Name:<br>Contact Name:<br>Address:<br>Telephone:<br>Fax:<br>Email: | Bio-Med USA Inc.<br>Young Chi, C.E.O.<br>27 New England Drive, Ramsey, NJ 07446. U.S.A.<br>1 973 278 5222<br>1 201 934 6030<br>biomedusa@msn.com |
|----|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| 3. | Manufacturer<br>Name:<br>Address:<br>Telephone:<br>Fax:<br>Registration #       | BlueCore Co., Ltd.<br>Acehigh 21, 48 Centum-Jungang ro 12, Haeundae Gu, Busan, Rep of Korea.<br>82 51 747 4318<br>82 51 747 4319<br>3014144061   |

- 4. Identification of the Proposed Device Trade/Device Name: PICORE Common Name: Nd:YAG O-switched Surgical Laser Classification Name: Powered Laser Surgical Instrument Classification: II Product Code: GEX Regulation Number: 21 CFR 878.4810
General& Plastic Surgery

- 5. Identification of Predicate Device 510(k) Number: K171648 Trade/Device Name: BM.IRIS Manufacturer: BlueCore Co. Ltd.
- Device Description 6.

Review Panel:

The subject device PICORE is a picosecond Nd:Y AG laser instrument that uses Nd:YAG as a medium to emit a laser beam of 1064nm and 532nm wavelength. It has four non-fractional handpieces (Zoom handpiece, 7mm collimated handpiece, 3-5 mm handpiece, 6-8 mm handpiece). A physician can optimize the treatment effect for different applications by controlling the energy of the laser pulse and the spot size of the treatment beam. Laser emission can be controlled by using a Foot Switch. Laser parameters and other system features are controlled from a display panel located on the front of the power supply unit.

- 7. Indications for Use - The indications for use statement of the subject device PICORE is given below. The indications for use of the PICORE is comparable to the indications for use of the predicate device BM.IRIS (K171648). Both devices are for prescription use only.
PICORE Q-switched Nd:YAG Laser System is indicated for the incision, excision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.

532nm Wavelength

{4}------------------------------------------------

Tattoo removal light ink (red, tan, purple, orange, sky blue) Removal of Benign Epidermal Pigmented Lesions, Minor Benign Vascular Lesions, Talangiectasias Treatment of Lentigines, Cafe-Au-Lait, Sebrorrheic Keratoses, Becker's Nevi, Freckles Treatment of Post Inflammatory Hyperpigmentation (PIH)

#### 1064nm Wavelength:

Tattoo removal: dark ink (black, blue, green) Removal of Nevus of Ota, Removal or lightening of unwanted hair with or without adjuvant preparation Treatment of Common Nevi, Melasma, Skin resurfacing procedures for the treatment of acne scars, wrinkle

- 8. Substantially Equivalent (SE) Comparison

| Tab 1 Technical Comparison               |                                                                                 |                                                                                                                                                                                             |                       |
|------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Technical Specification                  | Subject Device<br>(K212082)                                                     | Predicate Device<br>(K171648)                                                                                                                                                               | Comment               |
| Laser Type                               | Nd:YAG                                                                          | Nd:YAG                                                                                                                                                                                      | same                  |
| Wavelength (nm)                          | 1064/532                                                                        | 1064/532                                                                                                                                                                                    | same                  |
| Operating Mode                           | Q-switched, Pulsed                                                              | Q-switched, Pulsed                                                                                                                                                                          | same                  |
| Mode Names                               | 1064, 532, Blue Fx, Blue<br>T                                                   | 1064, 532, SLP, Blue<br>Toning                                                                                                                                                              |                       |
| Spot size (mm)                           | 2-10 @ 1064nm<br>2-10 @ 532nm                                                   | 2-10 @ 1064nm<br>1.5-8 @ 532nm                                                                                                                                                              | comparable            |
| Maximum Pulse Energy (mJ)                | 50-800 @ 1064nm (Q)<br>5-300 @ 532nm (Q)<br>50-800 @ Blue Fx<br>50-800 @ Blue T | 1200 @1064nm (Q)<br>400 @ 532nm (Q)<br>1500 @ 1064 nm (SLP, Non Q)<br>1000 @ 1064nm (Blue<br>Toning, PTP)                                                                                   | different &<br>subset |
| Maximum Pulse Power (GW)                 | 800 mJ/450ps = 1.8 GW                                                           | 1200 mJ/5ns = 0.24 GW                                                                                                                                                                       | different             |
| Pulse Duration                           | 450 ps @ 1064nm<br>380 ps @ 532nm<br>450 ps @ Blue Fx<br>450 ps @ Blue T        | 5-10ns @1064nm (Q)<br>5-10ns @ 532nm (Q)<br>5-10ns @ 1064 nm (SLP, Non Q)<br>5-10ns @ 1064nm (Blue<br>Toning, PTP)                                                                          | different             |
| Repetition Rate (Hz) @ spot<br>size (mm) | 1-10                                                                            | 10-20 @ 2-10mm<br>@1064nm (Q)<br>10-20 @ 2-10mm @<br>532nm (Q)<br>10-20@ 1-10mm @ 1064<br>nm (SLP, Non Q)<br>10-20 @ 2-7mm @<br>1064nm<br>10-18 @ 8-10mm, @<br>1064nm (Blue Toning,<br>PTP) | different &<br>subset |
| Aiming Beam                              | <3 mW @ 655 nm                                                                  | <3 mW @ 655 nm                                                                                                                                                                              | same                  |
| LCD Display                              | 10.4" touch screen                                                              | 10.4" touch screen                                                                                                                                                                          | same                  |
| Cooling                                  | Closed circuit water to air                                                     | Closed circuit water to air                                                                                                                                                                 | same                  |
| Power Consumption                        | 3kVA                                                                            | 3kVA                                                                                                                                                                                        | same                  |
| Dimensions                               | 290mmX794mmX934mm                                                               | 295mmX740mmX1680mm                                                                                                                                                                          | comparable            |
| Weight                                   | 60 kg                                                                           | 70 kg                                                                                                                                                                                       | comparable            |

{5}------------------------------------------------

Majority of the technical parameters of the subject device are comparable to and/or fall within the range of the corresponding technical parameters of the primary predicate device. The difference in technological parameters is not expected to significantly alter the safety and effectiveness of the subject device compared to the predicate device.

- 9. Non-Clinical Testing
Non clinical tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device. Biocompatibility testing was not performed since the handpieces of the subject device do not touch patients and are exactly the same as in the predicate. The following tests were conducted:

- ▲ IEC 60601-1:2005/A1:2012 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance;
- A IEC 60825-1: 2014, Safety of laser products - Part 1: Equipment classification and requirements.
- > IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests.
- A Software Validation & Verification Test
- > Bench Testing to verify the performance
- 10. Clinical Testing

No clinical study is included in this submission.

- 11. Conclusion
Based on the comparison and analysis above, the proposed subject device is determined to be Substantially Equivalent (SE) to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K212082](https://fda.innolitics.com/device/K212082)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
