3M(TM) Comply(TM) Hydrogen Peroxide Indicator Tape 1228

K212081 · 3M Company · JOJ · Sep 5, 2021 · General Hospital

Device Facts

Record IDK212081
Device Name3M(TM) Comply(TM) Hydrogen Peroxide Indicator Tape 1228
Applicant3M Company
Product CodeJOJ · General Hospital
Decision DateSep 5, 2021
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.2800
Device ClassClass 2

Intended Use

Use the 3M™ Comply™ Hydrogen Peroxide Indicator Tape 1228 to secure packs and as an external pack process indicator to differentiate unprocessed items from items processed in the following sterilizers: STERRAD 100® Sterilization System STERRAD 100S® Sterilization System STERRAD NX® Sterilization System (Standard and Advanced cycles) STERRAD 100NX® Sterilization System (Standard, Flex, Express and Duo cycles) STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles) STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express and Duo cycles) V-PRO® 1 Low Temperature Sterilization System (Lumen cycle) V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles) V-PRO® maX Low Temperature Sterilization System (Lumen, and Flexible cycles) V-PRO® 60 Low Temperature Sterilization System (Lumen, Non Lumen and Flexible cycles) V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Flexible, and Fast Non Lumen cycles) V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles) STERIZONE® VP4 Sterilizer (Cycle 1) The tape is suitable for use on non-woven disposable wraps and peel pouches. The chemical indicator stripes turn from blue toward pink after exposure to vaporized hydrogen peroxide.

Device Story

External chemical process indicator tape; non-cellulosic plastic backing with pressure-sensitive adhesive; chemical indicator stripes. Used in healthcare facilities to secure sterilization packs; differentiates processed from unprocessed items. Upon exposure to vaporized hydrogen peroxide, indicator stripes transition from blue to pink. Provides visual confirmation of exposure to sterilization process; aids in inventory management and sterility assurance workflows.

Clinical Evidence

Bench testing only. Performance verified via color change testing and minimum exposure parameter determination in the STERIZONE® VP4 Sterilizer (Cycle 1) per FDA guidance for chemical indicators.

Technological Characteristics

Non-cellulosic plastic backing; pressure-sensitive adhesive; chemical indicator stripes (blue to pink transition). 18-month shelf life; 6-month endpoint reaction stability. No electronic components or software.

Indications for Use

Indicated for use as an external pack process indicator to differentiate processed from unprocessed items in specified vaporized hydrogen peroxide sterilization systems. Suitable for use on non-woven disposable wraps and peel pouches.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. September 5, 2021 3M Company Mary Fretland Regulatory Affairs Specialist 3M Center, Building 275-5W-06 St. Paul, Minnesota 55144 Re: K212081 Trade/Device Name: 3M1M Comply 100 Hydrogen Peroxide Indicator Tape 1228 Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: August 6, 2021 Received: August 9, 2021 Dear Mary Fretland: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K212081 Device Name 3M™ ComplyTM Hydrogen Peroxide Indicator Tape 1228 Indications for Use (Describe) Use the 3M™ Comply™ Hydrogen Peroxide Indicator Tape 1228 to secure packs and as an external pack process indicator to differentiate unprocessed items from items processed in the following sterilizers: STERRAD 100® Sterilization System STERRAD 100S® Sterilization System STERRAD NX® Sterilization System (Standard and Advanced cycles) STERRAD 100NX® Sterilization System (Standard, Flex, Express and Duo cycles) STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles) STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express and Duo cycles) V-PRO® 1 Low Temperature Sterilization System (Lumen cycle) V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles) V-PRO® maX Low Temperature Sterilization System (Lumen, and Flexible cycles) V-PRO® 60 Low Temperature Sterilization System (Lumen, Non Lumen and Flexible cycles) V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Flexible, and Fast Non Lumen cycles) V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles) STERIZONE® VP4 Sterilizer (Cycle 1) The tape is suitable for use on non-woven disposable wraps and peel pouches. The chemical indicator stripes turn from blue toward pink after exposure to vaporized hydrogen peroxide. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | Prescription Use (Part 21 CFR 801 Subpart D) | | |----------------------------------------------|--------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | <div> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the 3M logo. The logo is in red and consists of the number "3" followed by the letter "M". The logo is simple and recognizable. # 510(k) Summary for 3M™ Comply™ Hydrogen Peroxide Indicator Tape 1228 # K212081 #### Sponsor Information: 3M Company 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000 Contact: Mary Fretland Regulatory Affairs Specialist Phone Number: (651) 737-2296 Fax Number: (651) 737-5320 Date of Summary: August 02, 2021 {4}------------------------------------------------ #### PREMARKET NOTIFICATION [510(k)] 3M™ ComplyTM Hydrogen Peroxide Indicator Tape 1228 ### 1. Device Name and Classification: | Common or Usual Name: | Chemical Indicator | |------------------------|----------------------------------------------------| | Proprietary Name: | 3M™ Comply™ Hydrogen Peroxide Indicator Tape 1228 | | Classification Name: | Indicator, physical/chemical sterilization process | | Device Classification: | Class II, 21 CFR § 880.2800 | | Product Code: | JOJ | #### 2. Predicate Device: K203285 3M™ Comply™ Hydrogen Peroxide Indicator Tape 1228 ### 3. Description of Device: The 3MTM Comply™ Hydrogen Peroxide Indicator Tape 1228 consists of a noncellulosic plastic backing material with a pressure-sensitive adhesive on one side and chemical indicator stripes on the other side. The chemical indicator stripes turn from blue toward pink after exposure to vaporized hydrogen peroxide. #### 4. Indications for Use Use the 3M™ Comply™ Hydrogen Peroxide Indicator Tape 1228 to secure packs and as an external pack process indicator to differentiate unprocessed items from items processed in the following sterilizers: | STERRAD 100® Sterilization System | |-----------------------------------------------------------------------------------------------------------| | STERRAD 100S® Sterilization System | | STERRAD NX® Sterilization System (Standard and Advanced cycles) | | STERRAD 100NX® Sterilization System (Standard, Flex, Express and Duo cycles) | | STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles) | | STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express and Duo cycles) | | V-PRO® 1 Low Temperature Sterilization System (Lumen cycle) | | V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles) | | V-PRO® maX Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles) | | V-PRO® 60 Low Temperature Sterilization System (Lumen, Non Lumen and Flexible cycles) | | V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Non Lumen, Flexible, and Fast Non Lumen cycles) | | V-PRO® s2 Low Temperature Sterilization System (Lumen, Non Lumen, Flexible, and Fast cycles) | | STERIZONE® VP4 Sterilizer (Cycle 1) | The tape is suitable for use on non-woven disposable wraps and peel pouches. The chemical indicator stripes turn from blue toward pink after exposure to vaporized hydrogen peroxide. {5}------------------------------------------------ ## 5. Technological Characteristic Comparison Table | Feature | Submission Device:<br>3MTM Comply™ Hydrogen<br>Peroxide Indicator Tape 1228 | Predicate Device (K203285):<br>3MTM ComplyTM Hydrogen<br>Peroxide Indicator Tape 1228 | Comparison | |---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------| | Indications for<br>Use | Use the 3M™ ComplyTM<br>Hydrogen Peroxide Indicator Tape<br>1228 to secure packs and as an<br>external pack process indicator to<br>differentiate unprocessed items<br>from items processed in the<br>following sterilizers:<br>STERRAD 100® Sterilization<br>System<br>STERRAD 100S® Sterilization<br>System<br>STERRAD NX® Sterilization<br>System (Standard and<br>Advanced cycles)<br>STERRAD 100NX®<br>Sterilization System (Standard,<br>Flex, Express and Duo cycles)<br>STERRAD NX® with<br>ALLClear® Technology<br>Sterilization System (Standard<br>and Advanced cycles)<br>STERRAD 100NX® with<br>ALLClear® Technology<br>Sterilization System (Standard,<br>Flex, Express and Duo cycles)<br>V-PRO® 1 Low Temperature<br>Sterilization System (Lumen<br>cycle)<br>V-PRO® 1 Plus Low<br>Temperature Sterilization<br>System (Lumen and Non<br>Lumen cycles)<br>V-PRO® maX Low<br>Temperature Sterilization<br>System (Lumen, Non Lumen,<br>and Flexible cycles)<br>V-PRO® 60 Low Temperature<br>Sterilization System (Lumen,<br>Non Lumen and Flexible<br>cycles)<br>V-PRO® maX 2 Low<br>Temperature Sterilization<br>System (Lumen, Non Lumen,<br>Flexible, and Fast Non Lumen<br>cycles)<br>V-PRO® s2 Low Temperature<br>Sterilization System (Lumen,<br>Non Lumen, Flexible, and Fast<br>cycles) | Use the 3M™ ComplyTM<br>Hydrogen Peroxide Indicator Tape<br>1228 to secure packs and as an<br>external pack process indicator to<br>differentiate unprocessed items<br>from items processed in the<br>following sterilizers:<br>STERRAD 100® Sterilization<br>System<br>STERRAD 100S® Sterilization<br>System<br>STERRAD NX® Sterilization<br>System (Standard and<br>Advanced cycles)<br>STERRAD 100NX®<br>Sterilization System (Standard,<br>Flex, Express and Duo cycles)<br>STERRAD NX® with<br>ALLClear® Technology<br>Sterilization System (Standard<br>and Advanced cycles)<br>STERRAD 100NX® with<br>ALLClear® Technology<br>Sterilization System (Standard,<br>Flex, Express and Duo cycles)<br>V-PRO® 1 Low Temperature<br>Sterilization System (Lumen<br>cycle)<br>V-PRO® 1 Plus Low<br>Temperature Sterilization<br>System (Lumen and Non<br>Lumen cycles)<br>V-PRO® maX Low<br>Temperature Sterilization<br>System (Lumen, Non Lumen,<br>and Flexible cycles)<br>V-PRO® 60 Low Temperature<br>Sterilization System (Lumen,<br>Non Lumen and Flexible<br>cycles)<br>V-PRO® maX 2 Low<br>Temperature Sterilization<br>System (Lumen, Non Lumen,<br>Flexible, and Fast Non Lumen<br>cycles)<br>V-PRO® s2 Low Temperature<br>Sterilization System (Lumen,<br>Non Lumen, Flexible, and Fast<br>cycles) | Similar. The<br>submission<br>device adds the<br>STERIZONE®<br>VP4 Sterilizer<br>(Cycle 1) to the<br>Indications for<br>Use. | | Feature | Submission Device:<br>3M™ Comply™ Hydrogen<br>Peroxide Indicator Tape 1228 | Predicate Device (K203285):<br>3M™ Comply™ Hydrogen<br>Peroxide Indicator Tape 1228 | Comparison | | | STERIZONE® VP4 Sterilizer<br>(Cycle 1)<br>The tape is suitable for use on non-<br>woven disposable wraps and peel<br>pouches. The chemical indicator<br>stripes turn from blue toward pink<br>after exposure to vaporized<br>hydrogen peroxide. | The tape is suitable for use on non-<br>woven disposable wraps and peel<br>pouches. The chemical indicator<br>stripes turn from blue toward pink<br>after exposure to vaporized<br>hydrogen peroxide. | | | Substrate | Non-cellulosic plastic | Identical | | | Biocompatibility | The exposure to health care<br>professionals is minimal and well<br>below any identified toxic<br>thresholds for the compounds. | Identical | | | Color Change | Blue toward pink | Identical | | | Detection | Hydrogen Peroxide | Identical | | | Stability of the<br>endpoint reaction | Six (6) months | Twelve (12) months | Alignment of<br>claim with<br>similar 3M<br>products. | | Shelf life | Eighteen (18) months | Identical | | {6}------------------------------------------------ #### PREMARKET NOTIFICATION [510(k)] 3M™ ComplyTM Hydrogen Peroxide Indicator Tape 1228 The 3M™ Comply™ Hydrogen Peroxide Indicator Tape 1228 is the same design as the previously cleared device of the same model number (the predicate) which is sold under the tradename 3M™ Comply™ Hydrogen Peroxide Indicator Tape 1228 (K203285). No changes have been made to the device materials or fundamental technology. #### 6. Summary of Non-clinical Testing To demonstrate performance in the newly claimed sterilizer and cycles, nonclinical testing was performed in accordance with the FDA Guidance for Industry and Staff: Premarket Notification [510(k)] Submissions for Chemical Indicators, issued December 19, 2003. Reference Table 6.1 for testing completed in the STERIZONE® VP4 Sterilizer (Cycle 1). #### Table 6.1 Summary of Nonclinical Testing | Test | Purpose | Acceptance Criteria | Result | |--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|--------| | Color Change in<br>Health Care<br>Facility Cycle | To demonstrate the color change of<br>the device when used in the<br>STERIZONE® VP4 Sterilizer<br>(Cycle 1) | Color change from blue toward<br>pink | Pass | | Minimum<br>Exposure<br>Parameters | To determine the minimum time<br>required for the color change of the<br>device when used in the<br>STERIZONE® VP4 Sterilizer<br>(Cycle 1) | Determination of the minimum<br>time for color to change from<br>blue toward pink | Pass | {7}------------------------------------------------ ## 7. Conclusion Based on the non-clinical performance data, the 3M™ Comply™ Hydrogen Peroxide Indicator Tape 1228 is as safe, as effective, and performs as well as or better than the legally marketed predicate device, 3M™ Comply™ Hydrogen Peroxide Indicator Tape 1228 cleared under K203285, Class II (21 CFR 880.2800), product code JOJ.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...