← Product Code [EZL](/productcode/EZL) · K212077

# Teleflex Rusch SoftSimplastic Foley Catheters (K212077)

_Teleflexmedical, Inc. · EZL · Jul 27, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K212077

## Device Facts

- **Applicant:** Teleflexmedical, Inc.
- **Product Code:** [EZL](/productcode/EZL.md)
- **Decision Date:** Jul 27, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5130
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

2 Way SoftSimplastic Catheters: Indicated where routine transurethral drainage of the bladder is required either postoperatively or for patients with conditions requiring transurethral urine drainage. 3 Way SoftSimplastic Catheters: Indicated where routine transurethral drainage of the bladder is required either postoperatively, for patients with conditions requiring transurethral urine drainage and for patients requiring bladder irrigation.

## Device Story

Rusch SoftSimplastic Foley Catheter is a single-use, sterile, balloon-retention catheter for transurethral bladder drainage or irrigation. Device consists of a transparent PVC shaft (containing DOTP plasticizer) with 2 or 3 lumens, a proximal funnel with inflation valve, and a distal latex retaining balloon. Used in clinical settings by healthcare professionals. Input is manual insertion into the urethra; device functions by anchoring in the bladder via balloon inflation to facilitate urine drainage or irrigation fluid delivery. Output is the physical drainage of urine or delivery of irrigation fluid. Benefits include effective bladder management for postoperative or symptomatic patients. Device is radiopaque for visualization.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing and biocompatibility evaluation per ISO 10993-1, including cytotoxicity, sensitization, reactivity, mutagenicity, systemic toxicity, and implantation tests. Performance verified via ISO 20696 and ASTM F623 standards for mechanical integrity, flow rates, and balloon safety.

## Technological Characteristics

Materials: PVC shaft with DOTP plasticizer, latex balloon. Dimensions: 14-24Fr, 30/75ml balloon capacity. Features: 2 or 3 lumen, luer taper/funnel, radiopaque. Sterilization: Ethylene Oxide (ISO 11135-1). Standards: ISO 20696:2018, ASTM F623-99, ASTM F640-12, ISO 10993 series.

## Regulatory Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

## Predicate Devices

- Rusch Simplastic Foley Catheters ([K162989](/device/K162989.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

July 27, 2023

Teleflex Medical, Inc. Kelley Breheim Sr. MDR Regulatory Project Manager 3015 Carrington Mill Blvd. Morrisville, NC 27560

Re: K212077

> Trade/Device Name: Rusch SoftSimplastic Foley Catheters Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: EZL Dated: June 23, 2023 Received: June 23, 2023

Dear Kelley Breheim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica K. Nguyen -S

Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K212077

Device Name Rusch SoftSimplastic Foley Catheters

#### Indications for Use (Describe)

2 Way SoftSimplastic Catheters:

Indicated where routine transurethral drainage of the bladder is required either postoperatively or for patients with conditions requiring transurethral urine drainage.

3 Way SoftSimplastic Catheters:

Indicated where routine transurethral drainage of the bladder is required either postoperatively, for patients with conditions requiring transurethral urine drainage and for patients requiring bladder irrigation.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary [As required by 21 CFR 807.92] Date Prepared: July 26, 2023 510(k) Number: K212077

# Submitter's Name / Contact Person

Manufacturer Teleflex Medical, Incorporated 3015 Carrington Mill Blvd Morrisville, NC 27560 USA Phone: 612-500-1344

Contact Person Kelley Breheim Sr. MDR Regulatory Project Manager

#### General Information

| Trade Name:          | Rusch SoftSimplastic Foley Catheters                                                                              |
|----------------------|-------------------------------------------------------------------------------------------------------------------|
| Common / Usual Name: | Disposable Balloon-Retention Catheter                                                                             |
| Classification Name: | Urological Catheter and Accessories                                                                               |
| Regulation Number:   | 21 CFR 876.5130                                                                                                   |
| Product Code:        | EZL                                                                                                               |
| Device Class:        | Class II                                                                                                          |
| Predicate Device     | K162989, Rusch Simplastic Foley Catheters. This predicate device has not been subject to a design-related recall. |

#### Device Description

The Rusch SoftSimplastic Foley Catheter is a balloon retention type catheter and is single use, disposable and sterile. The catheters are made of transparent PVC. They have a 2 lumen or 3 lumen shaft with proximal funnel, inflation valve and a distal retaining balloon made of latex. Balloon capacity is in ml and the shaft size in French gauge (Fr.) as indicated on the funnel of each individual catheter.

{4}------------------------------------------------

| Product Codes | Size (Fr) | Product Description                                  |
|---------------|-----------|------------------------------------------------------|
| 4426/662430   | 14 – 24   | SoftSimplastic 2 Way 30ml Catheter Coude/Mercier Tip |
| 5706/664130   | 16 – 24   | SoftSimplastic 3 Way 30ml Catheter Couvelaire Tip    |
| 5707/662530   | 16 – 24   | SoftSimplastic 2 Way 30ml Catheter Couvelaire Tip    |
| 5708/664176   | 18 – 24   | SoftSimplastic 3 Way 75ml Catheter Couvelaire Tip    |

## Indications for Use

## 2 Way SoftSimplastic Catheters:

Indicated where routine transurethral drainage of the bladder is required either postoperatively or for patients with conditions requiring transurethral urine drainage.

## 3 Way SoftSimplastic Catheters:

Indicated where routine transurethral drainage of the bladder is required either postoperatively, for patients with conditions requiring transurethral urine drainage and for patients requiring bladder irrigation.

## Comparison of Technological Characteristics with the Predicate Device

A comparison of the technological characteristics between the Rusch SoftSimplastic Foley Catheters and the predicate device are provided in the following table.

| Comparison of Technological Characteristics |                                         |                                      |
|---------------------------------------------|-----------------------------------------|--------------------------------------|
|                                             | Subject Device                          | Primary Predicate Device             |
| Characteristic                              | Rusch SoftSimplastic<br>Foley Catheters | Rusch Simplastic Foley<br>Catheters  |
| Classification Name                         | Catheter, Retention Type,<br>Balloon    | Catheter, Retention Type,<br>Balloon |
| Product Code                                | EZL                                     | EZL                                  |
| Regulation Number                           | 876.5130                                | 876.5130                             |
| Class                                       | II                                      | II                                   |

{5}------------------------------------------------

| Indications for Use | 2 Way SoftSimplastic<br>Catheters:<br>Indicated where routine<br>transurethral drainage of the<br>bladder is required either<br>postoperatively or for<br>patients with conditions<br>requiring transurethral urine<br>drainage.<br><br>3 Way SoftSimplastic<br>Catheters:<br>Indicated where routine<br>transurethral drainage of the<br>bladder is required either<br>postoperatively, for patients<br>with conditions requiring<br>transurethral urine drainage<br>and for patients requiring<br>bladder irrigation. | These catheters are indicated<br>for routine transurethral<br>drainage of the bladder or for<br>routine post-operative<br>transurethral drainage and<br>irrigation of the bladder. |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

| Comparison of Technological Characteristics |                                                           |                                                                 |
|---------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------------|
| Characteristic                              | Subject Device<br>Rusch SoftSimplastic<br>Foley Catheters | Primary Predicate Device<br>Rusch Simplastic Foley<br>Catheters |
| Shelf Life                                  | 5 Year                                                    | 1 Year                                                          |
| Size Range                                  | 14-24Fr                                                   | 12-28Fr                                                         |
| Balloon Sizes                               | 30 and 75 ml                                              | 30 and 75 ml                                                    |
| Balloon Material                            | Latex                                                     | Latex                                                           |
| Shaft Material                              | PVC with dioctyl<br>terephthalate (DOTP)<br>plasticizer   | PVC with bis(2-ethylhexyl)<br>phthalate (DEHP) plasticizer      |
| Connections                                 | Luer Taper/Funnel                                         | Luer Taper/Funnel                                               |
| Lumen                                       | Two way and three way                                     | Two way and three way                                           |
| Population                                  | Adult                                                     | Adult and Pediatric                                             |

{6}------------------------------------------------

| Biocompatibility        | Materials have been tested<br>per ISO 10993 | Materials have been tested<br>per ISO 10993 |
|-------------------------|---------------------------------------------|---------------------------------------------|
| Sterile                 | Yes                                         | Yes                                         |
| Method of Sterilization | Ethylene Oxide                              | Ethylene Oxide                              |
| Single Use or Reusable  | Single Use                                  | Single Use                                  |
| Coated or Uncoated      | Uncoated                                    | Uncoated                                    |
| Radiopacity             | Yes                                         | Yes                                         |

In this 510(k), the sponsor proposed modifications to their previously cleared device, including a material change, an extension of shelf-life and minor labeling changes. Differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness.

## Summary of Non-Clinical Performance Testing

## Biocompatibility

The biocompatibility evaluation for the Rusch SoftSimplastic Foley Catheters was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.

| Test                                                                         | Standard     |
|------------------------------------------------------------------------------|--------------|
| Cytotoxicity                                                                 | ISO 10993-5  |
| Sensitization                                                                | ISO 10993-10 |
| Intracutaneous Reactivity                                                    | ISO 10993-23 |
| Bacterial Reverse Mutation Assay (ISO)                                       | ISO 10993-3  |
| Acute Systemic Toxicity Test (ISO)                                           | ISO 10993-11 |
| Material Mediated Pyrogenicity Test (ISO/USP)                                | ISO 10993-11 |
| Mouse Lymphoma Assay (ISO)                                                   | ISO 10993-3  |
| Implantation Test (ISO)                                                      | ISO 10993-6  |
| 28-Day Systemic Toxicity (Dual Route Repeated<br>Exposure Method) Test (ISO) | ISO 10993-11 |

# Performance - Bench

The bench testing performed verifies that the performance of the subject device is substantially equivalent in terms of critical performance characteristics to the predicate device. The following table of test methods and standards (including packaging and sterility) were relied upon for a determination of substantial equivalence.

{7}------------------------------------------------

| Test                                      | Reference to Standard<br>(if applicable)                                                                 |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Visual inspection - Shaft                 | ISO 20696:2018                                                                                           |
| Visual inspection - Balloon               | ISO 20696:2018                                                                                           |
| Visual inspection - Funnel                | ISO 20696:2018                                                                                           |
| Visual inspection - Tip                   | ISO 20696:2018                                                                                           |
| Effective Shaft Length                    | ISO 20696:2018                                                                                           |
| Strength of Catheter                      | ISO 20696:2018                                                                                           |
| Connector Security                        | ISO 20696:2018                                                                                           |
| Balloon Safety                            | ISO 20696:2018                                                                                           |
| Inflation and Deflation Test              | ISO 20696:2018                                                                                           |
| Flow Rate Through Catheter                | ISO 20696:2018                                                                                           |
| Kink Stability                            | ISO 20696:2018                                                                                           |
| Peak Tensile Force - Shaft                | ISO 20696:2018                                                                                           |
| Peak Tensile Force - Funnel               | ISO 20696:2018                                                                                           |
| Peak Tensile Force - Balloon              | ISO 20696:2018                                                                                           |
| Peak Tensile Force - Tip                  | ISO 20696:2018                                                                                           |
| Inflated Balloon Response to Traction     | ISO 20696:2018                                                                                           |
| Flow Rate Through Drainage Lumen          | ASTM F623-99                                                                                             |
| Balloon Integrity (Resistance to Rupture) | ASTM F623-99                                                                                             |
| Inflated Balloon Response to Traction     | ASTM F623-99                                                                                             |
| Balloon Volume Maintenance                | ASTM F623-99                                                                                             |
| Balloon and Shaft Size                    | ASTM F623-99                                                                                             |
| Deflation Reliability                     | ASTM F623-99                                                                                             |
| Radiopacity                               | ASTM F640-12                                                                                             |
| Ethylene Oxide Sterilization              | ISO 11135-1                                                                                              |
| EO Residuals                              | ISO 10993-7                                                                                              |
| Packaging                                 | ISO 11607-1<br>ISO 11607-2<br>ASTM D4169<br>ASTM F1886/F1886M<br>ASTM F88/F88M<br>EN 868-5<br>ASTM F2096 |

{8}------------------------------------------------

# Conclusion

The Rusch SoftSimplastic Foley Catheter has the same intended use, technological characteristics, and construction as its predicate, with the exception of different PVC plasticizer material. Performance test results demonstrate that the subject device is as safe and effective as the predicate device. The Rusch SoftSimplastic Foley Catheter is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K212077](https://fda.innolitics.com/device/K212077)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
