← Product Code [FMI](/productcode/FMI) · K212015

# BD Nano 2nd Gen Pen Needle (K212015)

_Becton, Dickinson and Company · FMI · Nov 19, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K212015

## Device Facts

- **Applicant:** Becton, Dickinson and Company
- **Product Code:** [FMI](/productcode/FMI.md)
- **Decision Date:** Nov 19, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

Becton Dickinson Pen Needle is intended for use with pen injector devices for the subcutaneous injection of drugs.

## Device Story

BD Nano 2nd Gen Pen Needle is a single-use, sterile, disposable hypodermic needle assembly; consists of needle, hub, inner shield, and wide outer cover. Used with pen injectors for subcutaneous drug delivery. Operated manually by patients or clinicians in home or clinical settings. Device facilitates drug administration by connecting to pen injector; needle penetrates skin to deliver medication. Benefits include reliable subcutaneous access for drug therapy. Subject device introduces alternate polypropylene resin materials for hub and shield components compared to predicate.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including functional performance (flow rate, bond strength, torque, dose accuracy) per ISO 11608-2 and biocompatibility testing (cytotoxicity, reactivity, sensitization, toxicity, genotoxicity, pyrogenicity, implantation, endotoxin) per ISO 10993-1.

## Technological Characteristics

32G x 4mm stainless steel cannula with 5-bevel tip. Components (hub, inner shield, outer cover) made of polypropylene. Single-use, sterile (gamma irradiation). Manual insertion. Complies with ISO 11608-2 (functional performance) and ISO 10993-1 (biocompatibility).

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- BD Contoured Base Pen Needle ([K182320](/device/K182320.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 19, 2021

Becton Dickinson and Company Charlton Foo Staff Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417

### Re: K212015

Trade/Device Name: BD Nano 2nd Gen Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: Class II Product Code: FMI Dated: October 25, 2021 Received: October 26, 2021

### Dear Charlton Foo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801) medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

Submission Number (if known)

Device Name

BD Nano™ 2nd Gen Pen Needle

Indications for Use (Describe)

Becton Dickinson Pen Needle is intended for use with pen injector devices for the subcutaneous injection of drugs.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Submitted By:  | Charlton Foo<br>Staff Regulatory Affairs Specialist<br>Becton Dickinson and Company<br>1 Becton Drive<br>Franklin Lakes, NJ 07417<br>Tel: 201 847 6869<br>Fax: 201 847 5307 |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | June 23, 2021                                                                                                                                                               |
| Device Name:   |                                                                                                                                                                             |
|                | Trade Name: BD Nano™ 2nd Gen Pen Needle                                                                                                                                     |
|                | Common Name: Pen Needle                                                                                                                                                     |
|                | Classification: Class II device; 21 CFR 880.5570,<br>(hypodermic single lumen needle)                                                                                       |
|                | Product Code: FMI (hypodermic single lumen needle)                                                                                                                          |

Legally marketed predicate device to which substantial equivalence is being claimed: K182320: BD Contoured Base Pen Needle

#### Device Description:

The BD Nano™ 2nd Gen Pen Needle is designed for use with pen injectors for subcutaneous injection of a desired dose of drugs. It consists of a needle, base, and shield assembly. The BD Nano™ 2nd Gen Pen Needle is offered in a 32 gauge size and 4mm length. It is a singleuse disposable device that is provided sterile. The BD Nano™ 2nd Gen Pen Needle is nontoxic and non-pyrogenic.

#### Intended Use:

Becton Dickinson Pen Needle is intended for use with pen injector devices for the subcutaneous injection of drugs.

The intended use of the subject device remains the same as the predicate device.

{4}------------------------------------------------

### Comparison with Predicate Devices:

The subject device has the same fundamental scientific technology and device performance as the predicated device (K182320). The purpose of this submission is to market the BD Nano™ 2nd Gen Pen Needle device with a proposed alternate resin material. The table below provides a side-by-side comparison of the subject device compared to its predicate.

| Feature                       | Subject Device: BD Nano™<br>2nd Gen Pen Needle                                                                              | Predicate Device: BD<br>Contoured Base Pen Needle<br>(K182320)                                                              | Comparison                                                           |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| 510(k) Number                 | Pending                                                                                                                     | K182320                                                                                                                     | N/A                                                                  |
| Product Code                  | FMI                                                                                                                         | FMI                                                                                                                         | Unchanged                                                            |
| Regulation<br>Number          | 21 CFR 880.5570                                                                                                             | 21 CFR 880.5570                                                                                                             | Unchanged                                                            |
| Class                         | II                                                                                                                          | II                                                                                                                          | Unchanged                                                            |
| Intended Use                  | Becton Dickinson Pen Needle is<br>intended for use with pen injector<br>devices for the subcutaneous<br>injection of drugs. | Becton Dickinson Pen Needle is<br>intended for use with pen injector<br>devices for the subcutaneous<br>injection of drugs. | Unchanged                                                            |
| Needle Gauge                  | 32G                                                                                                                         | 32G                                                                                                                         | Unchanged                                                            |
| Needle Length                 | 4mm                                                                                                                         | 4mm                                                                                                                         | Unchanged                                                            |
| Pen Needle<br>Components      | Needle Hub<br>Inner Needle Shield<br>Wide Outer Cover<br>Cannula (32G x 4mm 5 bevel)<br>Tear Drop Label                     | Needle Hub<br>Inner Needle Shield<br>Wide Outer Cover<br>Cannula (32G x 4mm 5 bevel)<br>Tear Drop Label                     | Unchanged                                                            |
| Needle<br>insertion<br>method | Manual                                                                                                                      | Manual                                                                                                                      | Unchanged                                                            |
| Provided<br>Sterile           | YES (Gamma Irradiation)                                                                                                     | YES (Gamma Irradiation)                                                                                                     | Unchanged                                                            |
| Hub                           | Polypropylene                                                                                                               | Polypropylene                                                                                                               | Introduction of<br>alternate polymer<br>from a secondary<br>supplier |
| Needle                        | Stainless Steel                                                                                                             | Stainless Steel                                                                                                             | Unchanged                                                            |
| Lubricant                     | Medical Grade Lubricant                                                                                                     | Medical Grade Lubricant                                                                                                     | Unchanged                                                            |
| Feature                       | Subject Device: BD Nano™<br>2nd Gen Pen Needle                                                                              | Predicate Device: BD<br>Contoured Base Pen Needle<br>(K182320)                                                              | Comparison                                                           |
| Inner Needle<br>Shield        | Polypropylene                                                                                                               | Polypropylene                                                                                                               | Introduction of<br>alternate polymer<br>from a secondary<br>supplier |
| Wide Outer<br>Cover           | Polypropylene                                                                                                               | Polypropylene                                                                                                               | Introduction of<br>alternate polymer<br>from a secondary<br>supplier |
| Tear Drop<br>Label            | Paper                                                                                                                       | Paper                                                                                                                       | Unchanged                                                            |

{5}------------------------------------------------

The modifications to the materials were assessed through performance testing per ISO 11608-2:2012 and ISO 10993-1:2018. The differences between the predicate and subject device do not raise any new questions of safety or effectiveness.

# Testing:

### Non-Clinical Test Summary

The subject device has the same technological characteristics as the predicate device cleared in K182320. BD has validated the design of the subject device as part of its design control process in accordance with the Quality System Regulation. This testing included functional performance per ISO 11608-2: Needle-based injection systems for medical use – Requirements and test methods- Part 2: Needles and Biocompatibility testing per ISO 10993-1: Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing.

Functional Performance Testing include the following tests per ISO 11608-2:

- Determination of Flow Rate through Needle .
- Bond between Hub and Cannula .
- . Pen Installation and Removal Torque
- Dose Accuracy ●

### Biocompatibility Testing included the following tests per ISO 10993-1:

{6}------------------------------------------------

- Cytotoxicity .
- Intracutaneous Reactivity .
- Skin Sensitization .
- Acute Systemic Toxicity .
- Subacute / Subchronic Toxicity .
- Genotoxicity Bacterial and Mammalian .
- Material-Mediated Pyrogenicity .
- Implantation .
- . Bacterial Endotoxin

### Clinical Test Summary

No clinical study was included in this submission

# Conclusion:

The results of functional performance and biocompatibility testing passed and successfully demonstrated the subject device met requirements. The non-clinical testing has demonstrated the subject device is substantially equivalent to its predicate device.

---

**Source:** [https://fda.innolitics.com/device/K212015](https://fda.innolitics.com/device/K212015)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
