← Product Code [NKG](/productcode/NKG) · K211957

# QUARTEX Occipito-Cervico-Thoracic Spinal System, Globus Navigation Instruments (K211957)

_Globus Medical, Inc. · NKG · Aug 18, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K211957

## Device Facts

- **Applicant:** Globus Medical, Inc.
- **Product Code:** [NKG](/productcode/NKG.md)
- **Decision Date:** Aug 18, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3075
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The QUARTEX™ Occipito-Cervico-Thoracic Spinal System implants are intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1-C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facebility. These implants are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage turnors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, rods may be connected to occipital cervical thoracic or thoracolumbar stabilization systems ranging in diameter from 3.2mm, using corresponding connectors.

## Device Story

System comprises spinal implants (rods, screws, hooks, connectors, plates) and manual navigation instruments. Implants provide spinal immobilization/stabilization. Navigation instruments assist surgeons in locating anatomical structures during open or minimally invasive spinal surgery. Instruments are designed for use with Medtronic StealthStation® System; navigation relies on CT/MR models, fluoroscopy, or digitized landmarks to guide screw placement. Surgeon uses output to navigate anatomy during instrumentation. Benefits include precise implant placement and spinal stabilization.

## Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F1717 (static/dynamic compression bending) and ASTM F1798 (static interconnection testing). Bacterial endotoxin testing (BET) performed per ANSI/AAMI ST-72:2011. No clinical data provided.

## Technological Characteristics

Implants composed of titanium alloy, stainless steel, or cobalt chromium molybdenum (CoCr) alloy. Manual surgical instruments. Navigation instruments interface with Medtronic StealthStation® System. Mechanical testing per ASTM F1717, ASTM F1798. Sterilization via standard methods (implied).

## Regulatory Identification

Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

## Special Controls

*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
*e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.

## Predicate Devices

- QUARTEX™ OCT Spinal System ([K161591](/device/K161591.md))
- ELLIPSE® OCT Spinal System ([K090565](/device/K090565.md), [K110963](/device/K110963.md))
- PROTEX® CT OCT Spinal System ([K050391](/device/K050391.md))

## Reference Devices

- CREO® Stabilization System ([K180210](/device/K180210.md), [K124058](/device/K124058.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 18. 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image, there is the Department of Health & Human Services logo. To the right of that, there is a blue square with the letters "FDA" in white. To the right of the blue square, there is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Globus Medical Inc. Jennifer Antonacci Group Manager, Regulatory Affairs 2560 General Armistead Ave. Audubon, Pennsylvania 19403

### Re: K211957

Trade/Device Name: QUARTEX™ Occipito-Cervico-Thoracic Spinal System, Globus Navigation Instruments Regulation Number: 21 CFR 888.3075 Regulation Name: Posterior Cervical Screw System Regulatory Class: Class II Product Code: NKG, KWP, OLO Dated: June 23, 2021 Received: June 24, 2021

### Dear Jennifer Antonacci:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K211957

#### Device Name

QUARTEXTM Occipito-Cervico-Thoracic Spinal System

### Indications for Use (Describe)

The QUARTEX™ Occipito-Cervico-Thoracic Spinal System implants are intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1-C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facebility. These implants are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage turnors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, rods may be connected to occipital cervical thoracic or thoracolumbar stabilization systems ranging in diameter from 3.2mm, using corresponding connectors.

| Type of Use (Select one or both, as applicable)                                                                                                                                                     |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <div style="margin-right:5px;"> <span style="font-size: 20px;">☑</span> </div> <div>Research Use (Part 21 CFR 201.321, Subpart B)</div> </div>      |
| <div style="display:flex; align-items:center;"> <div style="margin-right:5px;"> <span style="font-size: 20px;">☐</span> </div> <div>Routine Production Use (21 CFR 201.321, Subpart C)</div> </div> |

> Prescription Use (Part 21 CFR 801 Subpart D)

_ | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K211957

### Device Name Globus Navigation Instruments

### Indications for Use (Describe)

Globus Navigation Instruments are intended to be used during the preparation and placement of Globus screws (QUARTEX™, CREO®, REVERE®, REVOLVE®, ELLIPSE®, PROTEX® CT, and SI-LOK®) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, qather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary: QUARTEX™ Additional Implants

| Company:        | Globus Medical Inc.<br>2560 General Armistead Ave.<br>Audubon, PA 19403<br>610-930-1800                                                                                                                                                                                                                                                   |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:        | Jennifer Antonacci, Ph.D.<br>Group Manager, Regulatory Affairs                                                                                                                                                                                                                                                                            |
| Date Prepared:  | August 17, 2021                                                                                                                                                                                                                                                                                                                           |
| Device Name:    | QUARTEX™ Occipito-Cervico-Thoracic Spinal System,<br>Globus Navigation Instruments                                                                                                                                                                                                                                                        |
| Classification: | QUARTEX™ Occipito-Cervico-Thoracic Spinal System<br>Per 21 CFR §888.3075 Posterior Cervical Screw System<br>Product Code: NKG<br>Regulatory Class: II, Panel Code: 87<br><br>Per 21 CFR §888.3050 Spinal Interlaminal Fixation Orthosis<br>Product Code: KWP<br>Regulatory Class: II, Panel Code: 87<br><br>Globus Navigation Instruments |

Per 21 CFR §882.4560 Stereotaxic Instrument Product Code: OLO Regulatory Class: II; Panel Code: 87

Primary Predicate: QUARTEX™ OCT Spinal System (K161591)

Addtl. Predicates: ELLIPSE®OCT Spinal System (K090565, K110963) PROTEX® CT OCT Spinal System (K050391)

Reference Device: CREO® Stabilization System (K180210, K124058)

### Purpose:

The purpose of this submission is to request clearance for additional QUARTEX™ implants and instruments.

### Device Description:

The QUARTEX™ Occipito-Cervico-Thoracic Spinal System includes 3.5mm-4.0mm jointed, straight and pre-bent rods, tapered rods, polyaxial screws, hooks, locking caps, t-connectors, lateral connectors, parallel connectors, in-line connectors, rod-to-rod connectors, rod extension clamps, QUARTEX™ H-LINK™ 

{5}------------------------------------------------

integrated rod, and occipital plates and screws. The implants are composed of titanium alloy, stainless steel, or cobalt chromium molybdenum (CoCr) alloy.

QUARTEX™ constructs may be connected to stabilization systems including ELLIPSE®, PROTEX® CT, PROTEX®, CREO®, REVERE®, or BEACON® Systems using corresponding connectors. The QUARTEX™ System includes manual surgical instruments.

### Indications for Use:

The QUARTEXTM Occipito-Cervico-Thoracic Spinal System implants are intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1-C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.q. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. These implants are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, rods may be connected to occipital cervical thoracic or thoracolumbar stabilization systems ranging in diameter from 3.2mm to 6.5mm, using corresponding connectors.

Globus Navigation Instruments are intended to be used during the preparation and placement of Globus screws (QUARTEXTM, CREO®, REVERE®, REVOLVE®, ELLIPSE®, PROTEX® CT, and SI-LOK®) during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

# Performance Data:

Mechanical testing (static and dynamic compression bending and static interconnection testing) was conducted in accordance with ASTM F1717, ASTM F1798, and the "Guidance for Industry and FDA Staff, Guidance for Spinal System 510(k)s," May 3, 2004. Performance data demonstrate substantial equivalence to the predicate devices. Bacterial endotoxin testing (BET) was also conducted in accordance with ANSI/AAMI ST-72:2011.

{6}------------------------------------------------

### Technological Characteristics:

Subject implants have the same technological characteristics as the predicate devices including design, intended use, material composition, function, and range of sizes.

## Basis of Substantial Equivalence:

Subject QUARTEX™ implants have been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence of the subject devices to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K211957](https://fda.innolitics.com/device/K211957)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
