Servator M SALF Solution is intended to be used for flushing and continuous hypothermic machine perfusion of kidneys at the time of their removal from the donor in preparation for storage, transportation, and eventual transplantation into a recipient.
Device Story
Servator M SALF Solution is a sterile, non-pyrogenic, non-toxic organ perfusion and preservation solution. It is used for in-vitro flushing and temporary continuous hypothermic machine perfusion of explanted kidneys. The solution is supplied in 1000ml PVC-free bags. It is intended for use by clinical professionals during the organ procurement and transplantation process. The solution maintains organ viability during storage and transport prior to transplantation. It does not require point-of-use filtration. The device is a chemical solution and does not involve electronic or mechanical operation.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including chemical composition analysis, biocompatibility testing per ISO 10993 series, and sterility validation per ISO 17665-1.
Indicated for flushing and continuous hypothermic machine perfusion of kidneys at the time of donor removal in preparation for storage, transport, and transplantation into a recipient.
Regulatory Classification
Identification
An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 20, 2021
S.A.L.F. S.p.A. % Joyce St. Germain Owner The 510k Consulting, LLC 1449 Springleaf Drive Ormond Beach, Florida 32174
Re: K211842
> Trade/Device Name: Servator M SALF Solution Regulation Number: 21 CFR 876.5880 Regulation Name: Isolated Kidney Perfusion And Transport System And Accessories Regulatory Class: II Product Code: KDL Dated: December 10, 2021 Received: December 16, 2021
Dear Joyce St. Germain:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Glenn B. Bell. Ph.D. Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211842
Device Name Servator M SALF Solution
Indications for Use (Describe)
Servator M SALF Solution is intended to be used for flushing and continuous hypothermic machine perfusion of kidneys at the time of their removal from the donor in preparation, and eventual transportation, and eventual transplantation into a recipient.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a stylized key design. The key is composed of two intertwined lines, one in gold and the other in black. The lines form the shape of a key with a looped head and a straight shaft. The design is simple and modern, with a focus on the interlocking lines.
# The 510k Consulting LLC
to success in obtaining your medical device clearance ringleaf Dr., Ormond Beach, FL 32174 da@gmail.com
# 510(k) Summary K211842
## Submitter/Applicant
S.A.L.F. S.p.A Via Marconi, 2, Cenate Sotto, BG, Italy Phone: +39-035-940097 Contact: Dr. Carmelo Gagliano (Quality Manager) carmelo.gagliano@salfspa.it
Date Prepared: May 28,2021
## Preparer/Consultant
The 510k Consulting, LLC 1449 Springleaf Drive Ormond Beach, FL 32174
| Phone: | 904-477-3203 |
|----------|------------------------------------------------------------------|
| Contact: | Joyce St. Germain, Regulatory Consultant, joyce510kfda@gmail.com |
## Device Classification
| Trade/Model Names: | Servator M SALF Solution |
|----------------------|----------------------------------------------------------------|
| Submitter: | SALF, S.p.A., Italy |
| Common Name: | Cold Storage Solution |
| Classification Name: | System, Perfusion, Kidney, Disposable |
| Regulation Name: | Isolated kidney perfusion and transport system and accessories |
| Regulation Number: | 21 CFR 876.5880 |
| Product Code: | KDL |
| Regulatory Class: | II |
| 510k Review Panel: | Gastroenterology/Urology Panel |
## Predicate Device
The subject device claims equivalence to the following legally marketed predicate:
| 510(k) Number: | K121736 |
|----------------|--------------------------------------------|
| Date Cleared | August 16, 2013 |
| Submitter: | Waters Medical Systems, LLC, Rochester, MN |
| Trade Name: | PERF-GENQ® Pulsatile Perfusion Solution |
{4}------------------------------------------------
| Common Name: | Cold Storage Solution |
|----------------------|-------------------------------------------------------------------|
| Classification Name: | Set, Perfusion, Kidney, Disposable |
| Regulation Name: | Isolated kidney perfusion and transport system and<br>accessories |
| Regulation Number: | 21 CFR 876.5880 |
| Product Code: | KDL |
| Regulatory Class: | II |
| Medical Specialty: | Gastroenterology/Urology Panel |
## Indications for Use
Servator M SALF Solution is intended to be used for flushing and continuous hypothermic machine perfusion of kidneys at the time of their removal from the donor in preparation for storage, transportation, and eventual transplantation into a recipient.
#### Intended Use
Organ perfusion and preservation solution.
## Device Description
The Servator M SALF is a clear, sterile, non-pyrogenic, non-toxic solution for the in-vitro flushing and temporary continuous perfusion preservation of explained kidneys. This solution has an approximate calculated osmolarity of 300 mOsm/kg, a sodium concentration of 100 mEq/L, a potassium concentration of 25 rnEq/L, and a pH of approximately 7.4 at room temperature. The primary containers used for the device are PVC free bags 1000ml, therefore they are free of phthalates. The solution may be used without any point of use filtration. Servator M SALF is intended to be used for flushing and continuous hypothermic machine perfusion of kidneys at the time of their removal from the donor in preparation for storage, transportation, and eventual transplantation into a recipient.
#### Type of device.
Sterile, non-pyrogen solution for perfusion/protection and preservation of organs. NOT FOR INJECTION.
## Package
Box containing 10 bags per carton of 1000 ml PVC free.
Model Number of Device: SERVM10DMA
{5}------------------------------------------------
## Comparison of Technological Characteristics with Predicate
- The indications for use and intended use of the subject and predicate devices are ● identical.
- . The technologies are substantially equivalent as the composition of both solutions are identical.
- The subject and predicate devices are both supplied in bags with overbags for single ● use.
- The subject and predicate devices are both supplied sterile.
- Tests were performed in order to confirm the equivalence between the subject and predicate devices. The following table compares technological and other characteristics of the subject and predicate device.
## Table of Comparison
# Technological Comparison
| | Subject Device | Predicate Device | Comparison |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Device | Servator M SALF Solution | PERF-GENQ® Pulsatile<br>Perfusion Solution | NA |
| Manufacturer | SALF spa, Italy | Waters Medical Syustems,<br>LLC, Rochester, MN | NA |
| 510(k) Number | K211842 | K121736 | NA |
| Classification &<br>Product Code | 876.5880; KDL | 876.5880; KDL | Same |
| Regulation<br>Name | Isolated Kidney Perfusion<br>and Transport System and<br>Accessories | Isolated Kidney Perfusion<br>and Transport System and<br>Accessories | Same |
| Device<br>Classification<br>Name | System, Perfusion, Kidney,<br>Disposable | System, Perfusion, Kidney,<br>Disposable | Same |
| Common Name | Cold Storage Solution | Cold Storage Solution | Same |
| Device<br>Description | The Servator M SALF<br>Solution is a clear, sterile,<br>non-pyrogenic, non-toxic<br>solution for the in-vitro<br>flushing and temporary<br>continuous perfusion<br>preservation of explained<br>kidneys. This solution has<br>an approximate calculated<br>osmolarity of 300 | The PERF-GENO Pulsatile<br>Perfusion Solution (PERF-<br>GEN Solution) is a clear,<br>sterile, non-pyrogenic, non-<br>toxic solution for the in-<br>vitro flushing and<br>temporary continuous<br>perfusion preservation of<br>explained kidneys. This<br>solution has an | Same |
| | mOsm/kg, a sodium<br>concentration of 100<br>mEq/L, a potassium<br>concentration of 25<br>mEq/L, and a pH of<br>approximately 7.4 at room<br>temperature. | approximate calculated<br>osmolarity of 300<br>mOsm/kg, a sodium<br>concentration of 100<br>mEq/L, a potassium<br>concentration of 25 mEq/L,<br>and a pH of approximately<br>7.4 at room temperature. | |
| Indication for<br>Use | Servator M SALF<br>Solution is intended to be<br>used for flushing and<br>continuous hypothermic<br>machine perfusion of<br>kidneys at the time of their<br>removal from the donor in<br>preparation for storage,<br>transportation, and<br>eventual transplantation<br>into a recipient. | The PERF-GENQ®<br>Pulsatile Perfusion Solution<br>is intended to be used for<br>flushing and continuous<br>hypothermic machine<br>perfusion of kidneys at the<br>time of their removal from<br>the donor in preparation for<br>storage, transportation, and<br>eventual transplantation into<br>a recipient. | Same |
| Intended Use | Organ storage and<br>preservation for<br>transplantation | Organ storage and<br>preservation for<br>transplantation | Same |
| Mode of<br>operation | Cold storage | Cold storage | Same |
| Container/Bag | PVC free bags | PVC free bags | Same |
| Meets UNOS<br>Policy | Yes | Yes | Same |
| pH | 6.90-7.50 at 20°C | 6.90-7.50 at 20°C | Same |
| Container | PVC free bags | PVC free bags | Same |
| Particulate<br>Matter and<br>Biocompatibility | Particle Counts less than<br>limits for Large Volume<br>Injections per USP <788>;<br>Biocompatible per ISO | Particle Counts less than<br>limits for Large Volume<br>Injections per USP <788>;<br>Biocompatible per ISO | Same |
| | 10993-1 battery of tests for<br>Externally Communicating<br>Blood Path Indirect<br>Contact for prolonged<br>periods >24 hours. | 10993-1 battery of tests for<br>Externally Communicating<br>Blood Path Indirect<br>Contact for prolonged<br>periods >24 hours. | |
| Device<br>Standards of<br>Conformity | ISO 10993-1<br>ISO 10993-4<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-11<br>ISO 14971<br>ISO 15223-1<br>ISO 17665-1<br>USP 39 <71><br>USP 39 <85> | ISO 10993 Series<br>Exact test series of<br>predicate device are<br>unknown | Subject<br>device<br>passed<br>according<br>to ISO<br>Standards |
| Protecting<br>Overwrap bag | Yes | Yes | Same |
| Bag connections | 1 flip off, 1 needle point | 1 flip off, 1 needle point | Same |
| Single use only | Yes | Yes | Same |
| Sterilization<br>method | Steam | Steam | Same |
| Nominal value | 1000 mL bags | 1000 mL bags | Same |
| Shelf Life | 24 months | 24 months | Same |
| Storage<br>Temperature | Indoors with temperature at<br>2°-<br>25°C, without freezing. | Indoors with temperature at<br>2° - 25°C, without freezing. | Same |
| Interaction with<br>other medical<br>technology | Check the expiration date<br>stated on the container. The<br>expiration date refers to the<br>product in unopened<br>packaging.<br>CAUTION: do not use the<br>device after the expiry date<br>stated on the container.<br>Even though within its | Check the expiration date<br>stated on the container. The<br>expiration date refers to the<br>product in unopened<br>packaging.<br>CAUTION: do not use the<br>device after the expiry date<br>stated on the container.<br>Even though within its shelf | Same |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
| shelf life (device not<br>expired), the solution must<br>not be used, if the<br>container is damaged or if<br>there are visible particles,<br>precipitates or<br>contaminations. | life (device not expired), the<br>solution must not be used, if<br>the container is damaged or<br>if there are visible particles,<br>precipitates or<br>contaminations. |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
## Solution qualitative and quantitative composition / Same for both Subject and Predicate
| CONSTITUENT | Amount 1000 ml | Concentration (mM) |
|---------------------------------|-------------------|--------------------|
| Adenine (free base) | 0.68 g | 5 |
| Calcium Chloride (dihydrate) | 0.068 g | 0.5 |
| Dextrose (+) | 1.80 g | 10 |
| Glutathione (reduced) | 0.92 g | 3 |
| HEPES (free acid) | 2.38 g | 10 |
| Hydroxyethyl Starch | 50.0 g | N/A |
| Magnesium Gluconate (anhydrous) | 1.13 g | 5 |
| Mannitol | 5.4 g | 30 |
| Potassium Phosphate (monobasic) | 3.4 g | 25 |
| Ribose, D(-) | 0.75 g | 5 |
| Sodium Gluconate | 17.45 g | 80 |
| Sodium Hydroxide | 0.70 g | N/A |
| Sterile Water for injection | to 1000 ml Volume | N/A |
The above comparison shows the subject and predicate devices are identical in each attribute/characteristic.
NOTE. . . The composition list is identical for the subject and predicate device composition and therefore, the subject and predicate devices are identical in chemical composition.
The above comparison shows the subject and predicate devices are substantially equivalent in technology characteristics.
The Servator M SALF Solution has the same indication for use, intended use, design, materials, packaging and other technological characteristics to the predicate device.
## Non-Clinical Performance Data
The following performance data is provided in support of the substantial equivalence determination. All tests performed are included in this submission.
{9}------------------------------------------------
Biocompatibility ... is required for this device. The tests were all performed according to the ISO 10993 series that are listed in the Table of Comparison above. The subject device passed all biocompatibility test standards.
Sterilization and Shelf Life . . . is required for the subject device. The Validation of Sterility was performed, and the results passed according to ISO 17655-1. Steam sterilization and storage conditions are the subject and predicate device. Shelf life for the subject is 24 months and the predicate device is 12 months.
Electrical Safety and EMC . . . testing was not applicable for this device.
Software ... was not applicable for this device.
Performance Testing ... was completed as a direct comparison between the subject and predicate device. The chemical comparisons and leachables performance testing demonstrated the substantial equivalence of this device to the predicate.
## Conclusion
The subject and predicate devices have the same indications for use and the same intended use.
Both devices are substantially equivalent in design, materials, packaging and other technological characteristics and performance (since they have, in fact, the same chemical composition).
The Servator M SALF Solution does not raise any questions regarding safety and effectiveness and is equivalent to the predicate device. The non-clinical data supports and demonstrates the safety of the device.
The conclusion is that Servator M SALF Solution warrants a finding of substantial equivalence to the legally marketed PERF-GENQ® Pulsatile Perfusion Solution, and therefore, should have clearance for premarket activities in the United States.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.